K122114 · GE Healthcare · IYN · Oct 3, 2012 · Radiology
Device Facts
Record ID
K122114
Device Name
LOGIQ S7 EXPERT, LOGIQ S7 PRO
Applicant
GE Healthcare
Product Code
IYN · Radiology
Decision Date
Oct 3, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
Device Story
LOGIQ S7 Expert and LOGIQ S7 Pro are mobile, general-purpose diagnostic ultrasound systems. They acquire, process, and display ultrasound data using various transducers. The system features a console with a keyboard, specialized controls, a 7-inch LCD touch screen, and a 19-inch LCD display. Operated by qualified physicians in clinical settings, the device provides real-time imaging and fluid flow analysis. Outputs include B-mode, M-mode, PW/CW Doppler, Color Doppler, Power Doppler, and 3D/4D imaging. These outputs assist clinicians in diagnostic evaluations, such as infertility monitoring, needle guidance, and elastography. The system aids clinical decision-making by providing visual diagnostic information, potentially improving patient outcomes through accurate assessment of anatomy and pathology.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including acoustic output measurements, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, NEMA UD 3, and ISO 10993-1 standards.
Technological Characteristics
Mobile ultrasound console (62x86x175 cm) with 19-inch LCD display and 7-inch touch interface. Supports multiple transducers (e.g., C1-5-D, L8-18i-D, 11L-D, 9L-D, S4-10-D, 8C, ML6-15, P2D, 3Sp-D, IC5-9-D, 3CRF-D, RAB4-8-D). Imaging modes: B, M, PW/CW Doppler, Color/Power Doppler, 3D/4D, Harmonic, Coded Pulse. Connectivity: DICOM. Biocompatible patient-contact materials. Safety compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2/3, ISO 10993-1.
Indications for Use
Indicated for ultrasound evaluation of fetal, abdominal, pediatric, small organ (breast, testes, thyroid), neonatal/adult cephalic, cardiac (adult/pediatric), peripheral vascular, musculoskeletal (conventional/superficial), urological (prostate), transrectal, transvaginal, and intraoperative (abdominal, thoracic, vascular, neurosurgical) applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Voluson E6/E8/E8 Expert Diagnostic Ultrasound System (K112213)
Submission Summary (Full Text)
{0}------------------------------------------------
Superficial;
Urology
and
Conventional
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters are stylized and have a flowing, cursive appearance. The logo is black and white.
ОСТ Э 2012
# GE Healthcare 510(k) Premarket Notification Submission
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
Date: July 16, 2012
Submitter: GE Healthcare 9900 Innovation Dr Wauwatosa, WI 53226
| Primary Contact Person: Bryan Behn |
|------------------------------------|
| Regulatory Affairs Manager |
| GE Healthcare |
| T:(414)721-4214 |
| F:(414)918-8275 |
| Device: Trade Name: | LOGIQ S7 Expert and LOGIQ S7 Pro Ultrasound System |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | LOGIQ S7 Expert and LOGIQ S7 Pro |
| Classification Names:<br>Product Code: | Class II<br>Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Predicate Device(s): | K111582 LOGIQ S8 Diagnostic Ultrasound System<br>K092271 LOGIQ E9 Diagnostic Ultrasound System<br>K101874 LOGIQ P6 Diagnostic Ultrasound System<br>K113690 LOGIQ e Diagnostic Ultrasound System<br>K112213 Voluson E6/E8/E8 Expert Diagnostic Ultrasound<br>System |
| Device Description: | The LOGIQ S7 Expert AND LOGIQ S7 Pro is a full featured,<br>general purpose diagnostic ultrasound system which consists of a<br>mobile console approximately 62 cm wide, 86 cm deep and 175<br>cm high that provides digital acquisition, processing and display<br>capability. The user interface includes a computer keyboard,<br>specialized controls, 7-inch LCD touch screen and color 19-inch<br>LCD image display. |
| Intended Use: | The device is intended for use by a qualified physician for<br>ultrasound evaluation of Fetal; Abdominal; Pediatric; Small<br>Organ (breast, testes, thyroid); Neonatal Cephalic; Adult<br>Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; |
Musculo-skeletal
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters and the border are in black, contrasting with the white background.
(including prostate); Transrectal; Transvaginal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
The LOGIO S7 Expert and LOGIQ S7 Pro employs the same Technology: fundamental scientific technology as its predicate devices
Summary of Non-Clinical Tests: Determination of
Substantial Equivalence: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. LOGIQ S7 Expert and LOGIO S7 Pro and its applications comply with voluntary standards:
- IEC60601-1, Medical Electrical Equipment -1. Part 1: General Requirements for Safety
- 2. IEC60601-1-2,Medical Electrical Equipment -Part 1-2:General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests
- 3. IEC60601-2-37, Medical Electrical Equipment -Part 2-37:Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment
- 4. NEMA UD 3, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- 5. ISO10993-1. Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing- Third Edition
- 6. NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- 7. ISO14971, Application of risk management to medical devices
- 8. NEMA, Digital Imaging and Communications in Medicine (DICOM) Set. (Radiology)
{2}------------------------------------------------
The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- . Requirements Reviews
- . Design Reviews
- . Testing on unit level (Module verification)
- . Integration testing (System verification)
- . Final Acceptance Testing (Validation)
- Performance testing (Verification)
- . Safety testing (Verification)
Transducer materials and other patient contact materials are biocompatible.
### Summary of Clinical Tests:
.
The subject of this premarket submission, LOGIQ S7 Expert and LOGIQ S7 Pro, did not require clinical studies to support substantial equivalence.
- Conclusion: GE Healthcare considers the LOGIQ S7 Expert and LOGIQ S7 Pro to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling the perimeter, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion. In the center of the seal is a stylized representation of a caduceus, a symbol often associated with healthcare and medicine.
10903 New Hampshire Avenue Silver Spring, MD 20993
OCT 3 2012
GE Healthcare c/o Brvan Behn 9900 Innovation Dr. Wauwatosa. WI 53226
Re: K122114 Trade Name: LOGIO S7 Expert LOGIO S7 Pro Ultrasound System Regulation Number: 21 CFR 8892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Codes: IYN, IYO, ITX Dated: July 16, 2012 Received: July 17, 2012
Dear Mr. Behn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the LOGIO S7 Expert and LOGIQ S7 Pro, as described in your premarket notification:
Transducer Model Number
| C1-5-D | L8-18i-D | 11L-D |
|---------|----------|-------|
| 9L-D | S4-10-D | 8C |
| ML6-15 | P2D | 3Sp-D |
| IC5-9-D | P6D | |
| 3CRF-D | RAB4-8-D | |
{4}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Joshua Nipper at (301) 796-6524.
Sincerely Yours.
Mark D. Okerblom
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure(s)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular shape. The circle has decorative swirls or flourishes extending outward from the top and bottom, giving it a vintage or classic appearance.
510(k) Number (if known):
### K122114
Device Name:
LOGIQ S7 Expert and LOGIQ S7 Pro
Indications for Use:
The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal and Intraoperative (abdominal, thoracic, vascular and neurosurgical).
Prescription Use_X_ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NA (Part 21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mhhmf D. O
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
/2-11.
510(k) Number
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The letters and the circle are black, while the background is white.
### Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro Ultrasound System
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | N | N | N | N | N | N | N | N | N | N | 5,6 |
| Abdominal[1] | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Pediatric | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Small Organ[2] | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Neonatal Cephalic | N | N | N | N | N | N | N | N | N | N | 5,6 |
| Adult Cephalic | N | N | N | N | N | N | N | N | N | N | 5,6 |
| Cardiac Adult & Pediatric | N | N | N | N | N | N | N | N | N | N | |
| Peripheral Vascular | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Musculo-skeletal Conventional | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Musculo-skeletal Superficial | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Other[1] | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal(8) | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Transvaginal | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Transuretheral | | | | | | | | | | | |
| Intraoperative(8) | N | N | N | N | N | N | N | N | N | N | 3,5,6 |
| Intraoperative Neurological | N | N | N | N | N | N | N | N | N | N | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA;
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Muhi D. Khm
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Division of Haclological Dovice Evaluation and Safety
610K 122 114
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular border. The letters and the border are white, set against a black background, creating a high-contrast design. The logo is simple, recognizable, and represents the General Electric brand.
### Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with C1-5-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other (Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5,6 |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Pediatric | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac Adult & Pediatric | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[4] | P | P | P | P | P | P | P | P | P | P | 3,5,6 |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[3] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative[8] | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal. GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
15| 3D/4D Imaging mode
{6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
(Division Sign Off)
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Division of Radiological Devices
510K 12.2 11.4
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular frame. The letters are stylized and connected, creating a unique and recognizable design. The logo is in black and white.
# Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with 9L-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics(7) | P | P | P | | P | P | P | P | P | P | 5,6 |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | 3,5,6 |
| Pediatric | P | P | P | | P | P | P | P | P | P | 3,5,6 |
| Small Organ(2) | P | P | P | | P | P | P | P | P | P | 3,5,6 |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac Adult & Pediatric | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | 3,5,6 |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | 3,5,6 |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | 3,5,6 |
| Other(4) | | | | | | | | | | | |
| Exam Type. Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative[8] | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication: P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801,109)
(Division Sign-Off)
Devices
Division of Radiological Devices Division of Radiological Devices
Office of in Vitro Diagnostic Device Evaluation and Safety
S'OK. 112 114
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circular border. The letters and the border are black, contrasting with the white background. The logo is a well-known symbol associated with the General Electric company.
# Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with ML6-15 Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | 3,5,6 | |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | 3,5,6 | |
| Neonatal Cephalic | N | N | N | | N | N | N | N | N | N | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac Adult & Pediatric | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | 3,5,6 | |
| Musculo-skeletal Conventional | N | N | N | | N | N | N | N | N | N | 3,5,6 | |
| Musculo-skeletal Superficial | N | N | N | | N | N | N | N | N | N | 3,5,6 | |
| Other[4] | | | | | | | | | | | | |
| Exam Type. Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative[8] | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K101874)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Mhhael D. Khm
(Division signa
Division of Radiological Devices Division of Radiological Device Evaluation and Safety
510K 112.114
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are white and are set against a black circular background. The black circle has a white outline with small white accents around the perimeter.
# Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with IC5-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | | |
|-------------------------------|-------------------|---|---------|-------------------|-------|---------|---------|-------------------|---------|-------|---------|--|--|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined Harmonic | | Coded | Other | | |
| Anatomy/Region of Interest | | | Doppler | Doppler Doppler | | Doppler | Doppler | Modes | Imaging | Pulse | [Notes) | | |
| Ophthalmic | | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | b | P | P | 5,6 | | |
| Abdominal(!) | | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | | |
| Cardiac Adult & Pediatric | | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | | |
| Other (4) | P | P | P | | P | P | P | P | P | b | 3,5,6 | | |
| Exam Type, Means of Access | | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | 3,5,6 | | |
| Transvaginal | P | P | P | | P | P | P | P | P | P | 3.5,6 | | |
| Transuretheral | | | | | | | | | | | | | |
| Intraoperative[8] | | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
*(Division Sign-Off)*
Division of Radiological Devices Division of Radiological Device Evaluation and Safety
510K 172 119
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The logo is black and white.
## Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with 3CRF-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|----------------|------------------|-------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics(7) | P | P | P | | P | P | P | P | P | P | 5,6 |
| Abdominal(1) | P | P | P | | P | P | P | P | P | P | 5,6 |
| Pediatric | P | P | P | | P | P | P | P | P | P | 5,6 |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac Adult & Pediatric | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other(4) | P | P | P | | P | P | P | P | P | P | 5,6 |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal(8) | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative(8) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K111582)
[1] Abdominal includes Renal, GYN/Pelvic. Notes:
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
-
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
---
(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety
510K-
ડર્
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, creating a strong contrast. The logo is simple, recognizable, and represents the General Electric brand.
# Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with L8-18i-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|----------|----------|-------|------------------|--|
| Clinical Application | B | M | PW | CW | Color | Color M | Power | Combined | Harmonic | Coded | Other<br>(Notes) | |
| Anatomy/Region of Interest | | | Doppler | Doppler | Doppler | Doppler | Doppler | Modes | Imaging | Pulse | | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | P | P | P | | P | P | P | P | P | P | 5,6 | |
| Small Organ[2] | P | P | P | | P | P | P | P | P | P | | |
| Neonatal Cephalic | P | P | P | | P | P | P | P | P | P | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac Adult & Pediatric | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | | P | P | P | P | P | P | 3,5,6 | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | 3,5,6 | |
| Musculo-skeletal Superficial | P | P | P | | P | P | P | P | P | P | 3,5,6 | |
| Other[4] | | | | | | | | | | | | |
| Exam Type. Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative[8] | P | P | P | | P | P | P | P | P | P | 3,5,6 | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA(K111582)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
*(Division Sign-Off)*
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
810K 172114
ﺮ
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a stylized circular border. The letters and the border are solid black, contrasting with the white background. The logo has a slightly rough or textured appearance, possibly due to the image's source or reproduction method.
# Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with S4-10-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|----------------|------------------|-------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes | Harmonic Imaging | Coded Pulse | Other [Notes) | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Pediatric | P | P | P | P | P | P | P | P | P | P | 5 | |
| Small Organ[2] | P | P | P | P | P | P | P | P | P | P | 5 | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | 5 | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac Adult & Pediatric | P | P | P | P | P | P | P | P | P | P | P | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | | |
| Exam Type. Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative[8] | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K111582)
(1) Abdominal includes Renal, GYN/Pelvic. Notes:
(2) Small organ includes breast, testes and thyroid
(3) Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Archi D. Ofil
(Division Sign-Off)
Division of Radiological Devices Office of In Diagnostic Device Evaluation and Safety
510K
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are black, creating a strong contrast. The logo is simple and recognizable.
### Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with P2D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | P | P | | | | | | | |
| Cardiac Adult & Pediatric | | | P | P | | | | | | | |
| Peripheral Vascular | | | P | P | | | | | | | |
| Musculo-skeletal Conventional | | | | . | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | |
| Exam Type. Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative[8] | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA(K111582)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
{6} Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Marl D. Kirk
(Division Sign-Off) Division of Radiological Devices Office of In Vitro Division of Radiological Dovice Evaluation and Safety
510K. 12-114
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined within a circle. The letters are stylized and connected, creating a recognizable and iconic symbol for the company.
# Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with P6D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics(7) | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ(2) | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | P | P | | | | | | | | |
| Cardiac Adult & Pediatric | | | P | P | | | | | | | | |
| Peripheral Vascular | | | P | P | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other (4) | | | | | | | | | | | | |
| Exam Type. Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative(8) | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA (K111582)
Notes: [1] Abdominal includes Renal, GYN/Pelvic.
[2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity
[4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode
[6] Needle guidance imaging
[7] Includes infertility monitoring of follicle development
[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Prescription User (Per 21 CFR 801.109)
Mull D. O'K
(Division Sign-Off)
Division of Radiological Devices Division of Radiological Device Evaiuation and Safety
510K D2-114
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, contrasting with the white background.
# Diagnostic Ultrasound Indications for Use Form GE LOGIQ S7 Expert and LOGIQ S7 Pro with RAB4-8-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Mo…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.