PORTABLE X-RAY SYSTEM, MODEL PORT-X II

K063121 · Genoray Co., Ltd. · EHD · Jan 11, 2007 · Dental

Device Facts

Record IDK063121
Device NamePORTABLE X-RAY SYSTEM, MODEL PORT-X II
ApplicantGenoray Co., Ltd.
Product CodeEHD · Dental
Decision DateJan 11, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1800
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The PORT-X II Portable X-Ray system is intended to be used by trained dentists and dental technicians as an extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects.

Device Story

PORT-X II is a portable, hand-held dental X-ray system; powered by a 22.2VDC rechargeable Li-Polymer battery. Device integrates X-ray tubehead, controls, and power source into a single enclosure. Operated by trained dentists or dental technicians in clinical settings. Principle of operation: high voltage electricity generates X-rays in the tube; radiation penetrates patient anatomy (hand, tooth, jaw) to create images on external receptors (chemical film or digital sensors). Features include a 'Dead Man' type shooting controller to prevent accidental or continuous exposure and an audible buzzer during radiation emission. Output is diagnostic X-ray images used by clinicians to inform dental and diagnostic decision-making. Benefits include portability for point-of-care imaging.

Clinical Evidence

Bench testing only. Compliance with UL/IEC 60601-1, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, and EN/IEC 60601-1-2 (EMC) standards. No clinical data provided.

Technological Characteristics

Portable hand-held X-ray system; 22.2VDC Li-Polymer battery power; X-ray tubehead, controller, and beam limiting device integrated in single enclosure. 'Dead Man' switch control. Complies with IEC 60601-1, 60601-2-7, 60601-2-28, 60601-2-32, and EN/IEC 60601-1-2. No software-based image processing or algorithmic analysis described.

Indications for Use

Indicated for use by trained dentists and dental technicians as an extraoral X-ray source for diagnostic imaging of hand, tooth, and jaw in adult and pediatric patients using intraoral image receptors.

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ GENORAY Co., Ltd. ### Kola3121 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92. Date: September 4, 2006 ### JAN 1 1 2007 - 1. Company and Correspondent making the submission: Name - GENORAY Co., Ltd. Address - #512 Byucksan Technopia 434-6, Sangdaewon1-Dong, Jungwon-Gu, Seongnam-City, Gyeonggi-Do, 462-716, Korea Telephone - +82-31-737-8020 Fax – +82-31-737-8025 Contact - Yong Jin Park / Engineer E-mail - yjpark@genoray.com 2. Device : | Trade/proprietary name | : Portable X-Ray System/PORT-X II | |------------------------|-----------------------------------| | Common Name | : Portable X-Ray System | | Classification Name | : Extraoral source x-ray system | - 3. Predicate Devices : | Manufacturer | : Aribex, Inc. | |---------------|------------------------------------------| | Device | : NOMAD Dental X-ray System | | 510(k) Number | : K051795(Decision Date - Jul. 14. 2005) | - 4. Classifications Names & Citations : 21CFR 872.1800, EHD - Extraoral source x-ray system, Class 2 #### 5. Description : - 5.1 General PORT-X II is a portable dental X-ray system that operates on 22.2VDC supplied by a rechargeable Li-Polymer battery pack. The X-ray tubehead, X-ray controls, and power #512 Byucksan Technopia 434-6, Sangdaewon1-Dong, Jungwon-Gu, Seongnam-City, Gyeonggi-Do, 462-716, Korea {1}------------------------------------------------ # GENORAY Co., Ltd. source are assembled into a single hand-held enclosure. The package includes spare batteries, a battery charger. #### 5.2 Outline This equipment generates and controls X-ray in order to diagnose of hand, tooth and jaw. It is composed of X-ray generator, controller and beam limiting device. 800年会になる最後のアフェアム『 Operating principle is that X-ray generated by high voltage electricity into X-ray tube, which penetrates hand, tooth and jaw, and makes X-ray images on receptor (Chemical Film or Digital Sensor). #### 5.3 Operating Principle Operating principle is that X-ray generated by high voltage electricity into X-ray tube, which penetrates hand, tooth and jaw, and makes X-ray images on receptor (Chemical Film or Digital Sensor). #### 5.4 Features - Leakage Current of External Body should be satisfied with the Leakage Current Requirement of IEC 60601-1. - When the Radiation is shooting, buzzer makes sound. And then, when the shooing is Stopped, the sound is turned off. - X-Ray Shooting Controller: < Dead Man Type> Without opening Dead Man Controller, shooting X-Ray and continuous shooting should be not operated. #### 6. Indication for use : The PORT-X II Portable X-Ray system is intended to be used by trained dentists and dental technicians as an extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects. - 7. Comparison with predicate device : GENORAY Co., Ltd. believes that the Portable X-Ray System(PORT-X 11) is #512 Byucksan Technopia 434-6, Sangdaewon1-Dong, Jungwon-Gu, Seongnam-City, Gyeonggi-Do, 462-716, Korea {2}------------------------------------------------ ## GENORAY Co., Ltd. substantially equivalent to NOMAD Dental X-ray System of Aribex, Inc .. - 8. Safety, EMC and Performance Data : The portable x-ray system, PORT-X II, will comply with applicable requirements of the Underwriters Laboratories Standard for Safety-UL/IEC 60601-1, IEC 60601-2-7, IEC 60601-2-28 and IEC 60601-2-32. All required documents and reports will be submitted to the appropriate oversight agency to establish compliance with the applicable requirements. EMC testing was conducted by EMC Compliance Co., Ltd. in accordance with Standard EN/IEC 60601-1-2(2001). All test results were satisfactory. - 9. Conclusions : In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification GENORAY Co., Ltd. concludes that the Portable X-Ray System(PORT-X II) is safe and effective and substantially equivalent to predicate devices as described herein. - 10. GENORAY Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA. END {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows a circular logo with text around the perimeter and a symbol in the center. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The symbol in the center consists of three curved lines that resemble a stylized human figure. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 GENORAY Co., Ltd. % Mr. Marc M. Mouser Senior Project Engineer, Medical Devices Underwriters Laboratories, Inc. 2600 N.W. Lake Road CAMAS WA 98607-8542 JAN 1 1 2007 Re: K063121 Trade/Device Name: Portable X-Ray System/PORT-X II Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD Dated: December 22, 2006 Received: December 26, 2006 Dear Mr. Mouser: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Corneric Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. TTo grateral controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affective your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/3/Picture/10 description: The image shows a circular logo with the letters "PA" in the center. Above the letters, the numbers "1906-2006" are displayed. Below the letters, the word "Centennial" is written in a cursive font. Three stars are located at the bottom of the logo. Protecting and Promoting Public Health {4}------------------------------------------------ Page 2 - Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours. Nancy C. Hogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number(if known): K063121 Device Name: Portable X-Ray System/ PORT-X II Indications for Use: The PORT-X II Portable X-Ray system is intended to be used by trained dentists and dental technicians as an extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. Its use is intended for both adult and pediatric subjects. AND/OR Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (Part 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) David A. Seggerman (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number 2063121 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...