K250687 · Dexcowin Global, Inc. · MUH · Oct 24, 2025 · Dental
Device Facts
Record ID
K250687
Device Name
Cocoon Solo (DX-7020s)
Applicant
Dexcowin Global, Inc.
Product Code
MUH · Dental
Decision Date
Oct 24, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.1800
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Cocoon Solo (DX-7020s) is a portable x-ray system indicated for taking diagnostic dental x-rays for both pediatric and adult patients.
Device Story
Handheld, portable dental X-ray system; used by dentists or dental technicians for radiographic examination. Device generates ionizing radiation to expose external image receptors (film, PSP, or digital sensors). Operates via 16.8 Vdc Li-Polymer battery. Clinician positions device manually to capture dental images; output is an exposed receptor for subsequent processing and diagnostic review. Benefits include portability for point-of-care dental imaging. No image detectors are included with the device.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC 60601-1 series), EMC (IEC 60601-1-2), and performance testing per 21 CFR 1020.30/31.
Technological Characteristics
Handheld portable X-ray system; 70kV fixed voltage; 1.7/2.0mA fixed current; 0.4mm focal spot; 180mm source-to-skin distance (SSD); 2.3mmAl total filtration. Powered by 16.8 Vdc Li-Polymer rechargeable battery. Compatible with film, PSP, or digital sensors. Complies with IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, and 21 CFR 1020.30/31.
Indications for Use
Indicated for taking diagnostic dental x-rays for pediatric and adult patients.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 24, 2025
Dexcowin Global Inc
% Dave Kim
Regulatory Affairs Consultant
Mtech Group LLC
7505 Fannin St. Suite 610
HOUSTON, TX 77054
Re: K250687
Trade/Device Name: Cocoon Solo (DX-7020s)
Regulation Number: 21 CFR 872.1800
Regulation Name: Extraoral source x-ray system
Regulatory Class: Class II
Product Code: MUH
Dated: September 26, 2025
Received: September 26, 2025
Dear Dave Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250687 – Dave Kim
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}
K250687 – Dave Kim
Page 3
See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed
by Gabriela M. for
Rodal -S
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-Ray Systems Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250687 | |
| Device Name Cocoon Solo (DX-7020s) | |
| Indications for Use (Describe) | |
| Cocoon Solo (DX-7020s) is a portable x-ray system indicated for taking diagnostic dental x-rays for both pediatric and adult patients. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary K250687
The following 510(k) summary is being submitted as required by 21 CFR 868.5120;
1. Submitter: Dexcowin Global Inc.
Office # 878 Regus Center 155
Lake Ave., Suite 800
Pasadena, CA 91101
Manufacturer: Dexcowin Co., Ltd.
905, 2, Gasan digital 1-ro, GeumCheon-Gu
Seoul, Republic of Korea 08591
82-2-2027-2880
Ryu, Seung-Bum (sbryu@dexcowin.com)
Contact Person: Mtech Group LLC
(Official Correspondent) Dave Kim, MBA
7505 Fannin St. Suite 610, Houston, TX 77054, US
Tel: 713-467-2607
Email: davekim@mtechgroupllc.com
Date Prepared: September 25, 2025
2. Device Identification
Device Trade Name: Cocoon Solo (DX-7020s)
Common Name: Portable X-ray System
Classification Number: 21 CFR 872.1800
Device Classification: Extraoral source x-ray system
Regulation Name: II
Product Code: MUH
3. Predicate Device
Device Trade Name: Cocoon [DX-7017 / DX-7020]
510K Number: K173046
Common Name: Portable X-ray System
Classification Number: 21 CFR 872.1800
Device Classification: II
Product Code: MUH
4. Device Description
Cocoon Solo (DX-7020s) is a handheld, portable X-ray device designed for dental radiographic examination and diagnosis for pediatric and adult patients by exposing a X-ray image receptor to ionizing radiation.
The X-ray tube is located inside the device body to be used with conventional film (F-speed or greater film), PSP (Phosphor plates), or digital X-ray sensors.
The image detectors (an integral part of a fully-functional diagnostic x-ray system) are not part of the submission.
This device should only be used by trained and qualified dentists or dental technicians.
The subject device Cocoon Solo (DX-7020s) is strongly based on the predicate (from the same manufacturer). Changes to the predicate device are only minor and they don't seem to affect device safety or effectiveness.
{5}
2
5. Indications for Use
Cocoon Solo(DX-7020s) is a portable x-ray system indicated for taking diagnostic dental x-rays for both pediatric and adult patients.
6. Non-clinical Performance Test
The following performance was completed on the subject device in support of the substantial equivalence determination.
- Electrical Safety and EMC
- Specifications of Raw Material
- Software Validation
- Biocompatibility
- Risk Assessment
- All tests were performed in accordance with Iso standards and recognized by FDA.
- None of the standards was adapted for application to the device under review.
- There were no requirements of any deviation from the standards applied.
- No difference exist between the subject device and the sample test device unit.
Bench Testing for Image Performance has not been performed because the changes from the predicate system are only minor and the x-ray detectors are not part of the submission. The image quality should stay unchanged from the predicate performance.
7. Safety, EMC Data
Cocoon Solo(DX-7020s) complies with industry standards such as IEC 60601-1 Series to minimize electrical, mechanical and radiation hazards.
- Electrical, mechanical, environmental safety and performance testing which are mentioned in the standard IEC 60601-1, IEC 60601-1-3 and IEC 60601-2-65 were performed.
- EMC testing was conducted in accordance with Standard IEC 60601-1-2
- Performance testing performed according to FDA 21CFR 1020.30, 21CFR 1020.31 standards, Software Validation.
FDA Specific Guidance Documents associated with the device:
- Radiation Safety Consideration for X-ray Equipment Designed for Hand-Held Use.
- Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff"
- Pediatric Information for X-ray Imaging Device Premarket Notifications - Guidance for Industry and Food and Drug Administration Staff".
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions - Guidance for Industry and Food and Drug Administration Staff."
All test results were satisfied
8. Clinical Test Conclusion
Clinical testing was not required for this submission.
9. Technical Characteristics and Substantial Equivalence
The following table compares technological characteristics of the subject and predicate device.
{6}
Table 1. General Device Characteristics Comparison Table
| | Candidate Device | Predicate Device | | Substantial Equivalence Analysis |
| --- | --- | --- | --- | --- |
| 510(k) Number | K250687 | K173046 | | - |
| Device Name | COCOON SOLO (DX-7020s) | COCOON | | - |
| Manufacturer | Dexcowin Co., Ltd | Dexcowin Co., Ltd. | | - |
| Product Code | MUH | MUH | | - |
| Indication for Use | Cocoon Solo(DX-7020s) is a portable x-ray system indicated for taking diagnostic dental x-rays for both pediatric and adult patients | Cocoon is a portable X-ray System indicated for taking diagnostic dental x-rays for both pediatric and adult patients using intraoral film or digital sensors. | | Substantial Equivalence |
| Tube Voltage | 70kV (fixed) | 70kV (fixed) | | Substantial Equivalence |
| Tube Current | 1.7mA/2.0mA (fixed) | 2.0mA (fixed) | | Substantial Equivalence |
| Exposure Time | 0.05 – 1.00 seconds | Same | | Substantial Equivalence |
| Focal Spot | X-ray tube(D-041): 0.4 mm | X-ray tube (D-041) | 0.4mm | Substantial Equivalence |
| | | X-ray tube (DXDR-070) | 0.3mm | |
| Total Filtration | 2.3 mmAl (Inherent Filtration: 1.0mmAl) | X-ray tube(D-041) | 2.2 mmAl (Inherent Filtration: 1.0mmAl) | Substantial Equivalence |
| | | X-ray tube(DXDR-070) | 2.62 mmAl (Inherent Filtration: 1.82mmAl) | |
| SSD | 180mm | Same | | Substantial Equivalence |
| Cone Diameter | 58mm | Same | | Substantial Equivalence |
| Weight | 5 lbs | 4 lbs | | Substantial Equivalence |
| Detector Compatibility | Film/Digital Sensor/PSP | Same | | Substantial Equivalence |
| Power Supply | 16.8 Vdc Li-Polymer rechargeable battery | Same | | Substantial Equivalence |
The subject device and predicate device have no significant differences in most parameters and the differences do not affect the substantial equivalence of the subject device when compared to the predicate device.
# 10. Conclusion
The Cocoon Solo (DX-7020s) shows substantial equivalence to the predicate device. All differences were evaluated according to the tests and deemed not to raise new issues of safety or effectiveness. The modifications did not affect the device's intended use or alter the device's fundamental scientific technology. This modified subject device is at least as safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.