Ligence Heart

K252105 · Ligence Uab · QIH · Sep 26, 2025 · Radiology

Device Facts

Record IDK252105
Device NameLigence Heart
ApplicantLigence Uab
Product CodeQIH · Radiology
Decision DateSep 26, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAcceptanceObservedDev DSTest DS
Lateral mitral-annular early-diastolic velocity (lateral e')Individual Equivalence Coefficient (IEC) 95% confidence interval upper level < 0.25Met non-inferiority criterion; ICC ≥ 0.90Not specifiedRetrospective non-interventional validation study: 524 echocardiographic studies assessed by three human readers
Left-ventricular end-diastolic volume (LVEDV MOD)Individual Equivalence Coefficient (IEC) 95% confidence interval upper level < 0.25Met non-inferiority criterion; ICC ≥ 0.90Not specifiedRetrospective non-interventional validation study: 524 echocardiographic studies assessed by three human readers
Left-ventricular ejection fraction (LVEF MOD)Individual Equivalence Coefficient (IEC) 95% confidence interval upper level < 0.25Met non-inferiority criterion; ICC ≥ 0.90Not specifiedRetrospective non-interventional validation study: 524 echocardiographic studies assessed by three human readers
Left-ventricular end-systolic volume (LVESV MOD)Individual Equivalence Coefficient (IEC) 95% confidence interval upper level < 0.25Met non-inferiority criterion; ICC ≥ 0.90Not specifiedRetrospective non-interventional validation study: 524 echocardiographic studies assessed by three human readers

Indications for Use

Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi). Limitations: • Poor image capture will lead to poor annotations and subsequent measurements. • Multiple image quality algorithms are used to filter out images of poor quality. • Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. • If no gender is present, female referenced guideline values will be used for conclusions. • If Body Surface Area (BSA) is not present, indexed values cannot be provided. • During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software. • Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy.

Device Story

Ligence Heart is an automated image post-processing software for adult transthoracic echocardiography. It ingests DICOM images; utilizes machine-learning-based view classification, cardiac cycle selection, and border detection to quantify left-ventricular function; and generates structured reports. Used in clinical settings by cardiologists or sonographers, the software provides automated measurements (LVEDV, LVESV, LVEF, lateral e') for review. Clinicians may accept, edit, or replace automated results. The device streamlines echocardiography workflows, supporting clinical decision-making by providing reproducible quantitative data.

Clinical Evidence

Retrospective non-interventional study of 524 adult echocardiographic studies. Performance compared against three expert sonographers. Primary endpoint: Individual Equivalence Coefficient (IEC) with non-inferiority margin < 0.25. All four parameters (LVEF, LVEDV, LVESV, lateral e') met non-inferiority criteria. Intraclass Correlation Coefficients (ICC) were ≥ 0.90. No device-related adverse events reported.

Technological Characteristics

Software-only platform operating on off-the-shelf PC hardware. Utilizes machine-learning-based algorithms for view classification, cardiac cycle selection, and border detection. Compliant with DICOM standards for image input/output. No specific hardware materials or energy sources; software-based processing.

Indications for Use

Indicated for adult patients undergoing transthoracic echocardiography to support cardiologists or sonographers in clinical decision-making. Not validated for congenital heart disease, valve disease, pericardial disease, intra-cardiac lesions, prosthetic/surgically altered valves, implantable intracardiac devices, or irregular (non-sinus) rhythms.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION September 26, 2025 Ligence, UAB % Raymond Kelly Consultant Arazy Group Consultants Inc. 3422 Leonardo Lane New Smyrna Beach, Florida 32168 Re: K252105 Trade/Device Name: Ligence Heart Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: September 5, 2025 Received: September 5, 2025 Dear Raymond Kelly: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K252105 - Raymond Kelly Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K252105 - Raymond Kelly Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Ligence Heart Page 10 of 47 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252105 | ? | | Please provide the device trade name(s). | | ? | | Ligence Heart | | | | Please provide your Indications for Use below. | | ? | | Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi). Limitations: • Poor image capture will lead to poor annotations and subsequent measurements. • Multiple image quality algorithms are used to filter out images of poor quality. • Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. • If no gender is present, female referenced guideline values will be used for conclusions. • If Body Surface Area (BSA) is not present, indexed values cannot be provided. • During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software. • Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} Page 1 of 5 510(k) Summary 21 CFR 807.92 K252105 I. SUBMITTER (21 CFR 807.92(a)(1)) Submitter: Ligence, UAB Taikos pr. 54 Kaunas, 51305, Lithuania Phone: +37061144345 Contact Person: Alvaro Perez-Moreno, Ph.D. Date Prepared: July 2, 2025 II. DEVICE (21 CFR 807.92(a)(2)) Name of Device: Ligence Heart Common or Usual Name: Automated Radiological Image Processing Software Classification Name: Medical Image Management and Processing System (§892.2050) Regulatory Class: II Product Code: QIH III. PREDICATE DEVICE (21 CFR 807.92(a)(3)) Us2.v1, K210791 This predicate has not been subject to a design-related recall. No reference devices were used in this submission. IV. DEVICE DESCRIPTION (21 CFR 807.92(a)(4)) Ligence Heart is an image post-processing software used for viewing and quantifying adult cardiac ultrasound DICOM studies. The device is intended to generate structured measurement reports for echocardiography analysis and aid qualified sonographers and cardiologists in their decision-making process. Ligence Heart automatically identifies standard transthoracic echo views with machine-learning-based view classification, cardiac cycle selection, and border detection, then {5} Page 2 of 5 generates reproducible quantitative left-ventricular volumetric and functional measurements. The results are inserted into a PACS-compatible report that the reviewing cardiologist or sonographer can accept, edit, supplement with additional manual measurements, or entirely replace with manual measurements. The software also organizes, displays, and compares each measurement with reference-guideline ranges. Completed reports are exported in PDF, streamlining routine echocardiography workflow while leaving final diagnostic responsibility with the clinician. ## V. INDICATIONS FOR USE (21 CFR 807.92(a)(5)) Ligence Heart is a fully automated software platform that processes, analyses and makes measurements on acquired transthoracic cardiac ultrasound images, automatically producing a full report with measurements of several key cardiac structural and functional parameters. The data produced by this software is intended to be used to support qualified cardiologists or sonographers for clinical decision making. Ligence Heart is indicated for use in adult patients. Ligence Heart has not been validated for the assessment of congenital heart disease, valve disease, pericardial disease, and/or intra-cardiac lesions (e.g., tumors, thrombi). ## Limitations: - Poor image capture will lead to poor annotations and subsequent measurements. - Multiple image quality algorithms are used to filter out images of poor quality. - Our software complements good patient care and does not exempt the user from the responsibility to provide supervision, clinically review the patient, and make appropriate clinical decisions. - If no gender is present, female referenced guideline values will be used for conclusions. - If Body Surface Area (BSA) is not present, indexed values cannot be provided. - During image acquisition, inappropriate use of the echo machine, use of non-cardiac ultrasound probes, or use of suboptimal settings (e.g., gain, contrast, depth) may lead to lower accuracy of the software. - Ligence Heart has not been validated in patients with prosthetic or surgically altered cardiac valves, implantable intracardiac devices, or irregular (non-sinus) rhythm; these conditions may affect automated measurement accuracy. {6} Page 3 of 5 VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE (21 CFR 807.92(a)(5) &amp; 21 CFR 807.92(a)(6)) | Attribute | Predicate Device: K210791 | Subject Device | | --- | --- | --- | | Product Code | QIH | QIH | | Regulation | 892.2050 | 892.2050 | | Classification | II | II | | Rx/OTC | Rx | Rx | | Physical Characteristics | Software package that operates utilizing off-the-shelf hardware | Software package that operates utilizing off-the-shelf hardware | | User interface | The software is designed for use on a personal computer that has received images from a compatible PACS | The software is designed for use on a personal computer that has received images from a compatible PACS | | Automation level | Fully automated, including clip selection | Fully automated, including clip selection | | User confirmation / rejection of result | Yes | Yes | | Manual editing of automated result by user | Yes | Yes | | Input Data | Echocardiographic images | Echocardiographic images | | Communication protocol | DICOM Standard Compliance | DICOM Standard Compliance | | Patient Populations | Adults | Adults | Both the subject and predicate devices share the following technological characteristics: Variations in automated-measurement implementation: Ligence Heart automatically generates four core left-ventricular measurements essential for efficient and accurate routine cardiac assessments: lateral mitral-annular early-diastolic velocity (lateral e', Le), left-ventricular end-diastolic volume by the biplane Method of Disks (LVEDV MOD), left-ventricular ejection fraction by the biplane Method of Disks (LVEF MOD), and left-ventricular end-systolic volume by the biplane Method of Disks (LVESV MOD). {7} Page 4 of 5 # VII. PERFORMANCE DATA (21 CFR 807.92(b)) Ligence Heart was developed and tested in accordance with Ligence Heart's life cycle development processes. The device has been subject to extensive safety and performance testing. Verification and validation test results established that the device meets its design requirements and intended use. Specifically, software verification was conducted at the unit, integration, and system levels. Risk management analysis generated multiple risk mitigation measures and verification activities. A cybersecurity analysis and testing were conducted to verify that security measures are properly implemented in Ligence Heart to protect Ligence Heart's data, including patient health information (PHI). A summative human factors study was performed to validate that Ligence Heart can be used safely and effectively by the intended user population. A formal retrospective non-interventional validation study was conducted at a U.S. based independent Echocardiography Core Laboratory using 600 previously acquired echocardiograms to evaluate the performance of the device for each claimed measurement. Test datasets were strictly segregated from algorithm training datasets, as they are from completely separate cohorts. A cumulative total of 76 studies were excluded because of missing views or corrupt metadata. Ultimately, 524 echocardiographic studies were assessed by all three human readers, satisfied the predefined quality criteria, and were included in the final analysis. Baseline demographic characteristics of the cohort are summarized in the table below. | Participants characteristics | | | --- | --- | | Age, years | 49.48 ± 13.56 | | Male, n (%) | 319 (60.88%) | | Female, n (%) | 205 (39.12%) | | Height, cm | 173.36 ± 10.52 | | Weight, kg | 88.92 ± 21.11 | | LVEF, % | 34.45% ± 15.78% | | Normal LVEF (>50%), n (%) | 120 (22.90%) | | Reduced LVEF (<50%), n (%) | 404 (77.10%) | {8} Page 5 of 5 | Participants characteristics | | | --- | --- | | Mildly reduced LVEF (40-50%), n (%) | 64 (12.21%) | | Moderately reduced LVEF (30-40%), n (%) | 85 (16.22%) | | Severely reduced LVEF (<30%), n (%) | 255 (48.66%) | Primary Clinical Performance Endpoint was Individual Equivalence Coefficient (IEC). Agreement between Ligence Heart and three independent expert sonographers was quantified using the reference-scaled IEC. Automated measurements are deemed interchangeable when the upper level of the IEC 95% confidence interval was &lt; 0.25, a non-inferiority margin derived from individual-bioequivalence principles. All four parameters met the non-inferiority criterion, confirming that Ligence Heart measurements are interchangeable with expert human measurements. Secondary agreement metrics were consistent with the IEC result. Intraclass Correlation Coefficients (ICC) were ≥ 0.90 for all four parameters. No device-related adverse events, malfunctions, or unmediated issues occurred during the study. These combined non-clinical and clinical data demonstrate that Ligence Heart satisfies its specified performance criteria and support determination of substantial equivalence. # VIII. CONCLUSIONS (21 CFR 807.92(b)) Ligence Heart is substantially equivalent to the predicate device Us2.v1 (K210791). The subject and predicate share the same intended use, core technological characteristics, and fundamental principles of operation. No distinctions between the two systems do not introduce new questions of safety or effectiveness. Comprehensive non-clinical testing, including software verification and validation, cybersecurity assessment, and summative human-factors validation, demonstrated that the software platforms as intended and can be used safely under the specified conditions. Complementary clinical validation confirmed that automated measurements of LVEF, LVEDV, LVESV, and lateral e' are interchangeable with expert manual measurements and remain within normal inter-observer variability. Taken together, the non-clinical and clinical data show that Ligence Heart is at least as safe and effective as the predicative device for post processing analysis of adult transthoracic echocardiograms.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...