Magnifico Open (100009900); Magnifico MSK ( 100009910)
Applicant
Esaote, S.p.A.
Product Code
LNH · Radiology
Decision Date
Jun 26, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The general-purpose magnetic resonance imaging (MR) device is designed to scan any targeted area of the body, to collect, display and analyze MR images and other real-time imaging procedures. lmaqinq portions of calf, knee, ankle, foot, thigh, hand, wrist, forearm, elbow, arm, shoulder, hip, lumbar column, cervical column, thoracic spine, pelvis, temporomandibular joint (included only for "Open" configuration), head (included only for "Open" configuration).
Device Story
Magnetic resonance imaging (MRI) system; utilizes 0.4T permanent NdFeB magnet; C-shape design. Inputs: MR signals from anatomical regions (limbs, spine, head). Processing: EVO 23 software release; includes LSDWI (diffusion-weighted imaging) for brain, MR Angiography, and Maximum Intensity Projection (MIP) post-processing. Output: MR images displayed for physician review. Used in clinical settings by radiologists/technicians. Assists in diagnosing diseases/conditions by providing soft tissue contrast. Benefits: improved diagnostic visualization of fat, water, and muscle compared to CT.
Clinical Evidence
Bench testing only. Software verification and validation performed for new features (LSDWI, MRA, MIP). Performance testing included image quality assessment, electromagnetic compatibility, and surface heating tests for the new body coil. Sample clinical images were reviewed by an ACR-registered radiologist and deemed to be of good diagnostic quality.
Indicated for patients requiring MRI scanning of the calf, knee, ankle, foot, thigh, hand, wrist, forearm, elbow, arm, shoulder, hip, lumbar/sacral/cervical/thoracic spine, pelvis, TMJ, and head (Open configuration only) for diagnostic imaging purposes.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 26, 2024
Esaote S.p.A. Alberto Carcagnì Regulatory Affairs Specialist Via Melen 77 Genova. 16152 Italy
Re: K241133
Trade/Device Name: Magnifico Open (100009900); Magnifico MSK (100009910) Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: June 6, 2024 Received: June 6, 2024
Dear Alberto Carcagnì:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel M. Krainak, Ph.D Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K241133
Device Name Magnifico Open (100009900): Magnifico MSK ( 100009910)
#### Indications for Use (Describe)
The general-purpose magnetic resonance imaging (MRI) device is designed to scan any targeted area of the body, to collect, display and analyze MR images and other real-time imaging procedures.
Imaging portions of calf, knee, ankle, foot, thigh, hand, wrist, forearm, elbow, arm, shoulder, hip, lumbar column, sacral column, cervical column, thoracic spine, pelvis, temporomandibular joint (included only for "Open" configuration), head (included only for "Open" configuration).
MRI provides better soft tissue contrast than CT and can differentiate better between fat, water, muscle, and other soft tissue than CT (CT is usually better at images provide information to physicians and can be useful in diagnosing a wide variety of diseases and conditions.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) #: K241133
510(k) Summary
Prepared on: 2024-06-26
21 CFR 807.92(a)(1)
### Contact Details
| Applicant Name | Esaote S.p.A. | |
|-----------------------------|-----------------------------------------------------------|--------------|
| Applicant Address | Via Melen 77 Genova 16152 Italy | |
| Applicant Contact Telephone | +39 338 7170634 | |
| Applicant Contact | Mr. Alberto Carcagni | |
| Applicant Contact Email | alberto.carcagni@esaote.com | |
| Device Trade Name | Magnifico Open (100009900);<br>Magnifico MSK ( 100009910) | |
| Common Name | Magnetic resonance diagnostic device | |
| Classification Name | System, Nuclear Magnetic Resonance Imaging | |
| Regulation Number | 892.1000 | |
| Product Code(s) | LNH | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K212419 | Magnifico Open, Magnifico MSK | LNH |
| K032232 | Hitachi Airis Elite | LNH |
| K153736 | 3Mensio WORKSTATION | LLZ |
The new Magnifico EVO 23 employs the same fundamental scientific technology as its predicate device, the first version of Esaote Magnifico, cleared via K212419.
The modifications, do not affect the intended use or alter the fundamental scientific technology of the device.
The software, used on the proposed Magnifico system has been modified, from EVO 21 to include the
• LSDWI sequence for brain examination
- · MR Angiography
· Maximum Intensity Projection (MIP)
- and to support:
· Additional multi-channel technology coil for body
- · Enhancement of PC board with new processor
Maqnifico keeps all the other technological characteristics of the first cleared, version, as the 0.4T permanent Magnet, based on NdFeB (neodymium), C-shape.
{4}------------------------------------------------
# Intended Use/Indications for Use
The general-purpose magnetic resonance imaging (MR) device is designed to scan any targeted area of the body, to collect, display and analyze MR images and other real-time imaging procedures.
lmaqing portions of calf, knee, ankle, foot, thigh, hand, wrist, forearm, elbow, arm, shoulder, hip, lumbar column, cervical column, thoracic spine, pelvis, temporomandibular joint (included only for "Open" configuration), head (included only for "Open" configuration).
## Indications for Use Comparison
The indications for use of the new version of Esaote Magnifico MRI system are the first version of Magnifico, registered via K212419.
Only the Outcome related to diagnosis, due to the introduction of Magnifico (Magnifico Magnifico MSK), contains, for new Magnifico EVO23, a distinguish of the configurations. This is irrelevant because is referred to the more complete Version of Magnifico
## Technological Comparison
The technological characteristics of this device are similar to the primary predicate device. There has been no change to the magnet design or control system. The base elements of the operating system software are identical to the primary predicate device but the software have been improved with new functions like MR Angiography and DW. Both are consolidated standard MRI functions, cleared also with a low field MRI as Hitachi Airis Elite.
The MIP feature is exactly the same that was firstly realized by 3Mension (in Esaote Group) for a S.a.M.D. called Warkstation and, successively, included in a Esacte software tool, called 3D Viewer registered, with G-scan Brio. Both S.a.M.D, Workstation and 3D viewer, have been FDA cleared, respectively via K153736 and K180592.
The additional body coil has the same technical characteristics of the coils cleared with Primary Predicate device Magnifico.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Software verification and validation activities are provided. Bench testing outputs from new features (LSDW, MRA and MIP), image quality test, relevant electromagnetic compatibility tests. Surface heating testing and all relevant performance tests the Body coil were provided. Sample clinical images for the new features were reviewed by a ACR registered radiologist and determined to be of good diagnostic quality. The testing and software documentation demonstrate that the Magnifico MRI system with EVO 23 Software release is substantially equivalent to the predicate device, and conform to applicable medical device safety and performance requirements.
# 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
## 21 CFR 807.92(a)(5)