K090433 · Beijing Wandong Medical Equipment Co., Ltd. · LNH · Mar 25, 2009 · Radiology
Device Facts
Record ID
K090433
Device Name
I_OPEN 0.4T
Applicant
Beijing Wandong Medical Equipment Co., Ltd.
Product Code
LNH · Radiology
Decision Date
Mar 25, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The i Open 0.4T system is an open, whole body scanner. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the head, body, or extremities. The images produced by the i Open 0.4T system reflect the special distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (TI), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
Device Story
i Open 0.4T is an open, permanent-magnet MRI system (0.4 Tesla) using Nd-B-Fe magnet material. Device captures proton magnetic resonance signals from head, body, spine, and extremities. System utilizes gradient subsystems for magnetic field control and RF subsystems for signal transmission/reception. Software performs image reconstruction, review, and post-processing. Operated by trained clinicians in clinical settings; output consists of 2D/3D diagnostic images (T1/T2 weighted, MIP-MRA, fat/water suppression, MRCP). Physicians interpret images to assist in diagnosis. Benefits include non-invasive visualization of internal structures.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including performance metrics for signal-to-noise ratio, image uniformity, geometric distortion, slice profile/thickness/gap, and high-contrast spatial resolution, conforming to FDA-recognized NEMA standards.
Technological Characteristics
0.4 Tesla permanent MRI system; magnet material Nd-B-Fe. Employs gradient and RF subsystems for signal processing. Supports 2D/3D imaging sequences (SE, FSE, STIR, FLAIR, GRE, TOF). Connectivity includes DICOM printing. Software runs on Windows XP.
Indications for Use
Indicated for diagnostic imaging of head, body, spine, and extremities in patients requiring MRI. Produces transverse, sagittal, coronal, and oblique images based on proton density, T1, T2, and flow properties. Contraindications not specified.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 ad 21CFR 807.92.
## l. General Information
## Establishment:
| Name: | Beijing Wandong Medical Equipment Co., Ltd. |
|----------|---------------------------------------------|
| Address: | No.7, SanJianFang NanLi, ChaoYang District, |
| | Beijing, CHINA, 100024 |
| Phone: | +86 10 84575844 |
| Fax: | +86 10 84575842 |
### Registration Number: 3004859018
| Contact Person: | Mr. Wang Weimin | | |
|-----------------|-----------------|-----------------|--|
| | Manager, | | |
| | Phone: | +86 10 84575844 | |
| | Fax: | +86 10 84575842 | |
| | E_Mail: | wmw@263.net.cn | |
Date of Summary Preparation: Dec 17, 2008
traditional Type of submission:
Device Name:
●Trade Name: i Open 0.4T
### ●Classification Number:
Magnetic Resonance Diagnostic Device, CFR 892.1000 90-LNH
●Classification: Class II
## ·Performance Standards:
None established under Section 514 the Food, Drug, and Cosmetic Act.
{2}------------------------------------------------
# II. Safety and Effectiveness Information.
#### Device Description: .
See Part C <Device Description> document.
#### . Intended Use:
The i Open 0.4T system is an open, whole body scanner. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the head, body, or extremities. The images produced by the i Open 0.4T system reflect the special distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (TI), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
- Anatomical Region: Head, Body, Spine, Extremities
- Nucleus excited: Proton
- Diagnostic uses:
T1,T2 weighted Proton density weighted MIP-MRA Water suppress imaging fat suppress imaging MRCP
Imaging capabilities:
2D Spin Echo (SE) 2D,3D Fast Spin Echo(FSE) 2D Short Tau Inversion Recovery (STIR) 2D Fluid Attenuated Inversion Recovery (FLAIR) 2D,3D Rewinded Gradient Echo (2D, 3D Rewinded-GRE) 2D, 3D Spoiled Gradient Echo (2D,3D Spoiled-GRE) 2D,3D Time of Flight Angiography (TOF)
#### . Technological Characteristics (comparison with predicate device):
The i Open 0.4T is a 0.4 Tesla permanent MRI system. The magnet is mainly made of Nd-B-Fe material. The system software based on Windows XP is an interactive program integrated with scanning control, image reconstruction, reviewing, post-processing, DICOM printing.
#### Predicated Device: .
K974212: Hitachi AIRIS II K001334: AIRIS II Version 4.1 Software
{3}------------------------------------------------
K090433
Page 3 of 3
#### ● Statement of Substantial Equivalence:
The i Open 0.4T is of comparable type and substantially equivalent to Hitachi AIRIS II (K974212) and AIRIS II version 4.1 Software (K001334) in that they are similar in technology and intended uses. Both of these systems are open-permanent-magnet MRI Imaging System, use Gradient Subsystem to provide controlled and uniform gradient magnet fields in the X, Y and Z directions, and use RF Subsystem to complete the function of RF signal transmitting/receiving and processing. Image reconstruction is controlled by console that has an interactive user interface, and the system produces 2D and 3D image that can be filmed or electronically stored for future review. Both of these systems have the traditional MRI units.
#### General Safety and Effectiveness Concerns: .
Operation of the i Open 0.4T is substantially equivalent to the commercially available AIRIS II. The following are the safety parameter with action levels:
- Maximum Static Field ●
- � Rate of Change of Magnetic Ficld
- . RF Power Deposition
- . Acoustic Noise Levels
and performance levels:
- Specification Volume .
- Signal to Noise
- . Image Uniformity
- . Geometric Distortion
- . Slice Profile, Thickness and Gap
- High Contrast Spatial Resolution
specifical by the FDA guidance document for MR Diagnostic Devices that will be evaluated. The i Open 0.4T will conform to the FDA recognized NEMA Standards for the measurement of performance and safety parameters and the international IDC standard for safety issues with Magnetic Resonance Imaging Devices. This will assure that the performance of this device can be considered safe and effective with respect to currently available system.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three stripes forming its wing, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" encircles the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Beijing Wandong Medical Equipment Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K090433
Trade/Device Name: i_Open 0.4 T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 11, 2009 Received: March 12, 2009
## Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
MAR 2.6 2009
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html.
Sincerely yours,
Janine M. Morris
anine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): __ K0 90433
Device Name: i Open 0.4T
Indications for Use:
The i Open 0.4T system is an open, whole body scanner. It is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the head, body, or extremities. The images produced by the i Open 0.4T system reflect the special distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.
- Anatomical Region: Head, Body, Spine, Extremities
- Nucleus excited: Proton
- Diagnostic uses:
T1,T2 weighted Proton density weighted MIP-MRA Water suppress imaging fat suppress imaging MRCP
Imaging capabilities:
2D Spin Echo (SE) 2D,3D Fast Spin Echo(FSE) 2D Short Tau Inversion Recovery (STIR) 2D Fluid Attenuated Inversion Recovery (FLAIR) 2D,3D Rewinded Gradient Echo (2D, 3D Rewinded-GRE) 2D, 3D Spoiled Gradient Echo (2D,3D Spoiled-GRE) 2D,3D Time of Flight Angiography (TOF)
Prescription Use Yes (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
| (Division Sign-Off) | |
|---------------------|--|
|---------------------|--|
Division of Reproductive, Abdominal and
Radiological Devices
| 510(k) Number | K090433 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.