K242599 · Carlsmed, Inc. · LLZ · Dec 20, 2024 · Radiology
Device Facts
Record ID
K242599
Device Name
aprevo® Digital Planning
Applicant
Carlsmed, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 20, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Intended Use
The aprevo® Digital Planning software is intended to be used by trained, medically knowledgeable design personnel to generate surgical plans for skeletally mature patients. The device processes a 3D spine model in conjunction with inputs from healthcare professionals to produce 3D models of spinal corrections and anatomical measurements.
Device Story
aprevo® Digital Planning is a rules-based, semi-automated surgical planning software. It takes 3D spine models (DICOM) as input. Trained Carlsmed design personnel operate the software, incorporating inputs from healthcare professionals. The device uses landmark determination and surgical plan generation algorithms to semi-automatically place landmarks and apply correctional operations. It produces 3D models of spinal corrections and anatomical measurements, which are exported as a surgical plan. Surgeons review these outputs to inform clinical decision-making regarding spinal alignment and implant placement. The device aims to assist in creating accurate, patient-specific surgical plans.
Clinical Evidence
Bench testing only. The software was evaluated against internal specifications and performance standards. A comparative study demonstrated that surgical plans generated by the semi-automated aprevo® Digital Planning software were substantially equivalent to plans manually created by medically knowledgeable operators, with 90% confidence and reliability.
Technological Characteristics
Rules-based, semi-automated software for 3D spine model processing. Inputs: DICOM files. Outputs: 3D spinal correction models and anatomical measurements. Features semi-automated landmark determination and surgical plan generation algorithms. Operates as a standalone software system for internal use by trained design personnel.
Indications for Use
Indicated for generating surgical plans for skeletally mature patients requiring spinal correction.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
K170172 — UNiD Spine Analyzer · Medicrea International · May 24, 2017
K180091 — UNiD Spine Analyzer · Medicrea International · Feb 8, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
Carlsmed, Inc. Karen Liu VP Quality and Regulatory 1800 Aston Ave. Suite 100 Carlsbad, CA 92008
December 20, 2024
Re: K242599
Trade/Device Name: aprevo® Digital Planning Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: LLZ Dated: November 22, 2024 Received: November 25, 2024
Dear Karen Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{2}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb, Ph.D Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
Submission Number (if known)
K242599
Device Name
aprevo® Digital Planning
#### Indications for Use (Describe)
The aprevo® Digital Planning software is intended to be used by trained, medically knowledgeable design personnel to generate surgical plans for skeletally mature patients. The device processes a 3D spine model in conjunction with inputs from healthcare professionals to produce 3D models of spinal corrections and anatomical measurements.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(K) Summary
#### Applicant Contact Information
| Company Name | Carlsmed, Inc. |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Address | 1800 Aston Ave, Ste 100<br>Carlsbad, CA 92008 |
| Company Telephone | 760-766-1926 |
| Contact Person | Karen Liu, VP Quality and Regulatory<br>Carlsmed, Inc.<br>1800 Aston Ave, Ste 100<br>Carlsbad, CA 92008<br>Phone: 760-766-1926<br>Email: regulatory@carlsmed.com |
| Date Prepared | November 22, 2024 |
#### Device Name
| Trade or proprietary<br>name | aprevo® Digital Planning |
|-------------------------------------|--------------------------|
| Device classification<br>regulation | §892.2050 |
| Common name | Image processing system |
| Classification | Class II |
| Product code | LLZ |
| Classification panel | Radiology |
### Predicate Device
| 510(k) number | Product code | Trade name | Manufacturer |
|---------------|--------------|-----------------------------|----------------|
| K222195 | LLZ | aprevo® Digital<br>Workflow | Carlsmed, Inc. |
### Description of the Device
The aprevo® Digital Planning, is a rules-based, semi-automated surgical planning software that, based on surgeon input, measures the 3D spine model and results in a spinal correction plan. The outputs consist of a surgical plan and 3D spinal correction assets, which are then reviewed by the surgeon. The device is operated by trained, medically knowledgeable Carlsmed design personnel.
### Indications for Use
The aprevo® Digital Planning software is intended to be used by trained, medically knowledgeable design personnel to generate surgical plans for skeletally mature patients. The device processes a 3D spine model in conjunction with inputs from healthcare professionals to produce 3D models of spinal corrections and anatomical measurements.
{5}------------------------------------------------
## Technological Characteristics
The aprevo® Digital Planning software offers the option to semi-automatically place landmarking points and apply a series of correctional operations. The aprevo® Digital Planning software is substantially equivalent to the predicate devices through a comparison including the intended use, design, technological characteristics, and functionality.
## Substantial Equivalence
The software functionality is equivalent to the predicate device (K222195) based on the intended use and technological characteristics and does not raise any new question of safety and effectiveness. The table below includes details on a comparison between the subject and predicate devices.
| Characteristic | Subject Device | Primary Predicate Device | Substantial<br>Equivalence? |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Name | aprevo® Digital Planning | aprevo® Digital Workflow | |
| Clearance<br>Number | K242599 | K222195 | |
| Regulation<br>Number | 892.2050 | 892.2050 | Identical |
| Device Class | II | II | Identical |
| Product Code | LLZ | LLZ | Identical |
| Intended Use<br>and Indications<br>for Use | The aprevo® Digital Planning<br>software is intended to be used by<br>trained, medically knowledgeable<br>design personnel to generate<br>surgical plans for skeletally mature<br>patients. The device processes a 3D<br>spine model in conjunction with<br>inputs from healthcare professionals<br>to produce 3D models of spinal<br>corrections and anatomical<br>measurements. | aprevo® Digital Workflow software is<br>intended to view, store, and manipulate<br>3-D models to visualize surgical plans for<br>spinal alignment. The device inputs a 3-D<br>spine model which is used by trained,<br>medically knowledgeable design<br>personnel in conjunction with inputs from<br>healthcare professionals to produce 3-D<br>models of spinal correction and<br>anatomical measurements. The device<br>outputs are used by healthcare<br>professionals for placement of surgical<br>implants. | Substantially<br>Equivalent |
| | | Technological Characteristics | |
| Algorithms | ●<br>Landmark determination<br>algorithm<br>●<br>Surgical plan generation<br>algorithm | ● Landmarking and correction plan<br>generation is manually conducted by<br>operators | Substantially<br>Equivalent |
| Compatible Input<br>File Types | DICOM | DICOM | Identical |
| Process<br>Functionality | Semi-automatic landmark and<br>surgical plan generation | Manual landmark and surgical plan<br>generation | Substantially<br>Equivalent |
| User Interaction | User interaction is that the operator<br>provides the surgical inputs, then<br>initiates the landmarking and<br>surgical plan generation algorithms | User interaction is that the operator must<br>manually place landmarking points and<br>perform a series of operations to create a<br>surgical plan | Substantially<br>Equivalent |
| Export Outputs | Surgical plan | Surgical plan | Identical |
| Intended User<br>Population | Internal Carlsmed personnel only | Internal Carlsmed personnel only | Identical |
{6}------------------------------------------------
### Non-clinical Testing
The aprevo® Digital Planning software has been evaluated in accordance with internal software specifications and applicable performance standards. The surgical plans generated by aprevo® Digital Planning were shown to be substantially equivalent, with 90% confidence and reliability, to plans manually created by medically knowledgeable operators. In addition, the software development, verification and validation procedures ensure that the device performs according to specifications, user requirements, and the FDA guidance document Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
### Conclusion
Based on the intended use, indications for use, functions, technological characteristics, performance testing, and comparison to the predicate device, the subject aprevo® Digital Planning device has been shown to be substantially equivalent to the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.