K170172 · Medicrea International · LLZ · May 24, 2017 · Radiology
Device Facts
Record ID
K170172
Device Name
UNiD Spine Analyzer
Applicant
Medicrea International
Product Code
LLZ · Radiology
Decision Date
May 24, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
Indications for Use
The UNiD Spine Analyzer is intended for assisting healthcare professionals in viewing and measuring images as well as planning orthopedic surgeries. The device allows surgeons or service providers to perform generic, as well as spine related measurements on images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for placement of surgical implants. Clinical judgement and experience are required to properly use the software.
Device Story
Web-based software for orthopedic surgical planning; inputs standard image formats (JPEG, PNG, GIF). Enables surgeons or service providers to perform generic and spine-specific anatomical measurements; provides tools for osteotomy planning and templating of surgical implants (screws, cages, rods). Output consists of annotated images and surgical plans; plans may be validated by surgeons for manufacturing of patient-specific rods by Medicrea. Used in clinical settings; requires clinical judgment for interpretation and manipulation of images. Facilitates pre-operative planning and implant selection, potentially improving surgical accuracy and patient outcomes.
Clinical Evidence
Bench testing only. Verification and validation performed per FDA software guidance. Accuracy testing compared UNiD Spine Analyzer against Surgimap 2.0 using random lines and angles. Distance measurements: mean error 0.23mm (SD 0.42mm). Angle measurements: mean error 0.2° (SD 0.4°). Surgical tool testing (wedges/cages) showed mean errors of 0.25° (SD 0.44°) for wedges and 0.4° (SD 0.5°) for cages. No clinical data presented.
Technological Characteristics
Web-based software; PC compatible; runs on Windows and MAC. Features include image viewing, generic/spine measurement tools, osteotomy module, and implant templating. Software is classified as moderate level of concern. Cybersecurity controls implemented per FDA guidance.
Indications for Use
Indicated for healthcare professionals, including surgeons and service providers, to assist in viewing/measuring images and planning orthopedic surgeries, specifically for spine-related procedures and surgical implant placement.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 24, 2017
Medicrea International % Mr. David Ryan VP Product Development and Marketing 5389 route de Strasbourg - Vancia 69140 Rillieux-la-Pape FRANCE
Re: K170172
Trade/Device Name: UNiD Spine Analyzer Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 25, 2017 Received: April 27, 2017
Dear Mr. Ryan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
For
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | | | |
|-----------------------------------------|--|--|--|
| Food and Drug Administration | | | |
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
#### Indications for Use
510(k) Number (if known) K170172
Device Name UNiD Spine Analyzer
#### Indications for Use (Describe)
The UNiD Spine Analyzer is intended for assisting healthcare professionals in viewing and measuring images as well as planning orthopedic surgeries. The device allows surgeons or service providers to perform generic, as well as spine related measurements on images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for placement of surgical implants. Clinical judgement and experience are required to properly use the software.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--|--|--|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | | | | |
| Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov | Department of Health and Human Services | | | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | | | |
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}------------------------------------------------
#### MEDICREA INTERNATIONAL'S UNiD SPINE ANALYZER
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the UNiD Spine Analyzer.
#### 1. Submitter: Contact Person:
MEDICREA INTERNATIONAL 5389 Route de Strasbourg - Vancia 69140 RILLIEUX LA PAPE FR
David RYAN MEDICREA INTERNATIONAL 5389 Route de Strasbourg - Vancia 69140 RILLIEUX LA PAPE FR
Date Prepared: April 25, 2017
- 2. Trade name:
## Classification Name:
Class II, Picture archiving and communications system (21 C.F.R. 892.2050)
## Classification and Regulation:
LLZ: system, image processing, radiological
- 3. Predicate or legally marketed devices which are substantially equivalent
#### Primary predicate:
- SURGIMAP 2.0 (NEMARIS INC, K141669)
#### 4. Description of the device
UNiD Spine Analyzer is a software solution developed for the medical community. It is intended to be used to view images and perform spine related measurements and plan surgical procedures. The planning of surgical procedures can be done either by MEDICREA as part of the service of designing patient specific implant (surgeons will have to validate the planning submitted by MEDICREA before the manufacturing of any implants) or by the surgeon himself. The image formats supported encompass the standard image formats (jpeg, png, gif). Measurements (generic, measuring and surgical tools) can be overlaid to each image. UNiD Spine Analyzer offers the ability to plan certain surgical procedures, such as osteotomies of the spine, and templating implants (screws, cages and rods). Patient
{5}------------------------------------------------
specific rods can be ordered to be manufactured by MEDICREA. UNiD Spine Analyzer is a web-based software.
#### 5. Indications for Use
The UNiD Spine Analyzer is intended for assisting healthcare professionals in viewing and measuring images as well as planning orthopedic surgeries. The device allows surgeons or service providers to perform generic, as well as spine related measurements on images, and to plan surgical procedures. The device also includes tools for measuring anatomical components for placement of surgical implants. Clinical judgment and experience are required to properly use the software.
## 6. Substantial equivalence claimed to predicate devices
The table below compares the features and technological characteristics of UNiD Spine Analyzer to its predicate device:
| Feature | UNiD Spine Analyzer | Surgimap 2.0 |
|---------------------------------------------------------------------------|---------------------|---------------------------|
| Computer | PC Compatible | PC Compatible |
| Operating System | Windows + MAC | Windows + MAC |
| Image Input | Local | Local + PACS connectivity |
| Runs on Server | Yes | No |
| Osteotomy Module | Yes | Yes |
| Generic<br>measurements | Yes | Yes |
| Spine measurements | Yes | Yes |
| Pre-operative planning | Yes | Yes |
| Custom implants | Yes | Yes |
| Database | No | Yes |
| Case sharing | No | Yes |
| Human Intervention<br>for interpretation and<br>manipulation of<br>images | Required | Required |
| Web content | Yes | Yes |
# Table 1: Comparison of features and technological characteristics of UNiD Spine Analyzer and its predicate
{6}------------------------------------------------
#### 7. Non-clinical test summary
Performance data for the UNiD Spine Analyzer consisted of verification and validation activities. Software verification and validation testing was conducted and documentation was provided as recommended by FDA's guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained on Medical Devices". Agency recommendations in the guidance entitled "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" was followed to validate the UNiD Spine Analyzer. The software for this device was considered to be of moderate level of concern.
For basic measurement testing (angles and distances), random lines and angles have been drawn and measured by two different tools: Surgimap, which is the claimed predicate device, and the subject device, and the two values have been compared. Several configurations and values were tested. The mean error and the standard deviation (for distances and angles) were calculated to establish substantial equivalence. Mean error obtained for distance was 0.23mm and standard deviation was 0.42mm. For angles, mean error obtained was 0.2° and standard deviation was 0.4°.
For surgical tools (wedge and cage), sets of images were created with the wedge(s) or cage(s) to apply. For each image, the tool(s) was(were) applied, and the obtained images (with UNiD Spine Analyzer and with Surgimap) were superimposed to confirm that the measurements that moved because of the wedge(s) are at the same place using the twodifferent software devices. Possible deviation (angle) was measured. Mean error and standard deviation was calculated to establish substantial equivalence. Mean error obtained for wedge tool was 0.25° and standard deviation was 0.44°. For cages, mean error obtained was 0.4° and standard deviation was 0.5°.
All performance testing demonstrates that the subject device is substantially equivalent to the predicate device.
#### 8. Conclusion
The UNiD Spine Analyzer is substantially equivalent to the legally marketed predicate device as demonstrate by the completed performance testing.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.