MD301C4 29.8 DIAGNOSTIC IMAGING LCD MONITOR, L309TY

K111237 · Nec Display Solutions , Ltd. · LLZ · Jun 15, 2011 · Radiology

Device Facts

Record IDK111237
Device NameMD301C4 29.8 DIAGNOSTIC IMAGING LCD MONITOR, L309TY
ApplicantNec Display Solutions , Ltd.
Product CodeLLZ · Radiology
Decision DateJun 15, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

MD301C4 is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians. Caution: MD301C4 cannot be used for a life-support system. This unit is designed as component of a final system which is compliance to IEC60601-1-1 requirements. MD301C4 must not be used in digital mammography.

Device Story

The MD301C4 is a 29.8-inch color LCD monitor providing 4-megapixel (2560x1600) resolution for medical imaging. It features adjustable gamma grayscale settings to support diagnostic review in CT, MRI, HIS, and PACS environments. The device is intended for use by trained physicians in clinical settings as a component of a larger medical system. It conforms to DICOM standards for image display. The monitor receives digital image inputs and renders them for clinical interpretation, aiding physicians in diagnostic decision-making. It is not intended for life-support applications or digital mammography.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

29.8" color LCD monitor; 4MP (2560x1600) resolution; adjustable gamma grayscale; DICOM compliant. Safety standards: EN60601-1-2 (radiation), UL60601-1 (electrical).

Indications for Use

Indicated for trained physicians to display and view digital medical images for diagnostic purposes. Contraindicated for use in life-support systems and digital mammography.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K111237 JUN 1 5 2018 # 510(K) Summary of Safety and Effectiveness As required by 807.92 #### DEVICE ESTABLISHMENT AND CONTACT PERSON 1. Mr. Shuichi Kino Manager NEC Display Solutions Ltd. 4-13-23 Shibaura, Minato-ku, Tokyo, 108-0023 Japan Ph: +81-465-85-2376 Fax: +81-465-85-2378 ### 2. COMPANY REISTRATION NUMBER 3003623028 .............................................................................................................................................................................. #### 3. DATE SUMMARY PREPARED 09 February 2011 #### 4. DEVICE NAME | Trade Name: | MD301C4 29.8" Diagnostic Imaging LCD monitor | |----------------------|----------------------------------------------------------------| | Model Name: | L309TY | | Common Name: | Color LCD Monitor, Color Diagnostic Display, etc. | | Classification Name: | System, Image Processing, Radiological (CLASS II CFR 892.2050) | #### 4. PREDICATE DEVICE MD304MC 29.8" 4MP Color LCD Monitor by NEC Display Solutions Ltd. (K083916). Trade Name: MD304MC 29.8" Diagnostic Imaging LCD monitor L307TD Model Name: #### DEVICE DESCRIPTION క. Medical Display, MD301C4 is a 29.8" Color LCD monitor that displays image for medical use. It provides 4 mega pixel (2560*1600) resolution with adjustable gamma gray scale for more precise diagnose use in CT, MRI, HIS, and PACS. MD301C4 conforms with DICOM. {1}------------------------------------------------ #### INDICATION OF USE 6. .. MD301C4 is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians. Caution: MD301C4 cannot be used for a life-support system. This unit is designed as component of a final system which is compliance to IEC60601-1-1 requirements. MD301C4 must not be used in digital mammography. #### 7. CONCLUSION Device with trade name MD301C4 (model name is L309TY) and predicate device with trade name MD304MC (model name is L307TD) have the same target population of trained practitioner in hospital; it shares the same design, same performance and is the same in radiation safety (EN60601-1-2), mechanical safety, electrical safety (UL60601-1) and human factors. It uses similar material, and has same compatibility with environment and other devices. The differences between the two devices are display colors (MD304MC can display more colors), input signals, input terminals, and USB input, power consumes in power save mode (MD301C4 consumes less power in power save mode), power consumption (MD301C4 consumes more power), however these do not effect the safety and effectiveness of the MD301C4 to be substantially equivalent to the predicate device. These two devices also have the same intended use; Therefore we concluded that it is substantially equivalent to MD304MC by NEC Display Solutions Ltd. (K083916). {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 NEC Display Solutions Ltd. % Mr. Marc Mouser Engineering Leader/ FDA Office Coordinator Underwriters Laboratories, Inc. 2600 N.W. Lake Road CAMAS WA 98607-8542 JUN 1 5 2011 Re: K111237 Trade/Device Name: Medical Display, MD301C4 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 25, 2011 Received: June 3, 2011 Dear Mr. Mouser: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours. Mary Patel Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 5:10(k) Number (if known): Device Name: Medical Display, MD301C4 Indications For Use: MD301C4 is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians. Caution: MD301C4 cannot be used for a life-support system. This unit is designed as component of a final system which is compliance to IEC60601-1-1 requirements. MD301C4 must not be used in digital mammography. Prescription Use Over-The-Counter Use X (Part 21 CFR 801 Subpart D) AND/OR (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) Mary S. Postel Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K///237 Page 1 of __ ース i
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...