SYNGO US WORKPLACE

K091286 · Siemens Medical Solutions USA, Inc. · LLZ · May 15, 2009 · Radiology

Device Facts

Record IDK091286
Device NameSYNGO US WORKPLACE
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeLLZ · Radiology
Decision DateMay 15, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric, 3rd-Party Reviewed

Intended Use

Syngo US Workplace is used to view post processed ultrasound images, to store and print images, and to manipulate and make measurements on images using a personal computer or a compatible DICOM compliant PACS system. The Siemens ultrasound imaging systems are intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".

Device Story

Syngo US Workplace is a software-based image processing system for ultrasound data. It operates on a personal computer or DICOM-compliant PACS. Physicians use the device to view, store, print, and manipulate ultrasound images. It includes analysis packages for measuring anatomical structures, specifically the Arterial Health Package (AHP) for Intima Media Thickness (IMT) measurements. AHP allows comparison against validated normative tables to assist in communicating cardiovascular risk to patients. The device supports clinical diagnosis by providing quantitative data derived from ultrasound images. It is intended for use by clinicians in various medical specialties to facilitate patient communication and clinical decision-making regarding vascular health.

Clinical Evidence

No clinical data provided. The device relies on established clinical consensus statements (ASE Consensus Statement) for the use of Carotid Intima-Media Thickness measurements in identifying subclinical vascular disease.

Technological Characteristics

Software-based image processing system; DICOM-compliant; operates on standard personal computers or PACS environments. Features include measurement tools and analysis packages for ultrasound imaging. No specific hardware materials or energy sources are described as the device is a software application.

Indications for Use

Indicated for patients requiring ultrasound imaging analysis across general radiology, fetal, abdominal, intraoperative, pediatric, small parts, neonatal/adult cephalic, cardiac, transesophageal, pelvic, transcranial, OB/GYN, urology, vascular, and musculoskeletal applications. Includes Arterial Health Package for measuring Intima Media Thickness to assess cardiovascular system state.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary Prepared April 27, 2009 K091286 Page 1 of 2 | Sponsor: | Siemens Medical Solutions, Inc.,<br>Ultrasound Division<br>1230 Shorebird Way<br>Mountain View, California 94043 | |-----------------|------------------------------------------------------------------------------------------------------------------| | Contact Person: | Shelly Beere | MAY 15 2009 - Contact Person: Shelly Pearce Telephone: (650) 694-5988 (650) 694-5580 Fax: - Submission Date: April 27, 2009 - Device Name: syngo US Workplace - Common Name: System, Image Processing, Radiological # Classification: Regulatory Class: ll Review Category: Tier II Classification Panel: Radiology Picture Archiving and Communications System FR # 892.2050 Product Code 90-LLZ # A. Legally Marketed Predicate Devices syngo US Workplace, K060992, K072090; syngo AHP, K083149. # B. Device Description: The modified software feature provides for the measurement of anatomical structures and for analysis packages that provide information. {1}------------------------------------------------ # C. Intended Use KC91286 Page 2 of 2 Syngo US Workplace is used to view post processed ultrasound images, to store and print images, and to manipulate and make measurements on images using a personal computer or a compatible DICOM compliant PACS system. The Siemens ultrasound imaging systems are intended for the following applications: General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging". {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure. MAY 15 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Siemens Medical Solutions USA, Inc. c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 2500 Street, N.W. BUFFALO MN. 55313 K091286 Re: Trade/Device Name: Syngo US Workplace Regulation Number: 21 CFR §892.2050 Regulation Name: Picture archiving and communications system Product Code: LLZ Dated: April 30, 2009 Received: May 1, 2009 Dear Mr. Job: . . We have reviewed your Section -510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclobate) to tegains at date of the Medical Device Amendments, or to Connine.co pror to May 20, 2976, the encordance with the provisions of the Federal Food, Drug, devices that have occh recision of require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, morelors, mains of the Act include requirements for annual registration, listing of gonomics profices proficturing practice, labeling, and prohibitions against misbranding and adulteration. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I odotar buttor and registements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set do relator autocres (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you don't be Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. (Gastroenterology/Renal/Urology) (240) 276-0115 21 CFR 876.xxx (240) 276-0115 (Obstetrics/Gynecology) . 21 CFR 884.xxx (240) 276-0120 21 CFR 892.xxx (Radiology) (240) 276-0100 Other Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, [signature] Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### 1.3 Indications for Use K091286 510(k) Number (if known): Device Name: Syngo US Workplace Indications for Use: Syngo US Workplace is used to view post processed ultrasound images, to store and print images, and to manipulate and make measurements on images using a personal computer or a compatible DICOM compliant PACS system. The Siemens ultrasound imaging systems are intended for the following applications. General Radiology, Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Pelvic, Transcranial, OB/GYN, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imagina". Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Vorzu In Whang Division of Reproductive, Abdominal an Radiologica 510(k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...