MYRIAN

K071000 · Intrasense · LLZ · May 14, 2007 · Radiology

Device Facts

Record IDK071000
Device NameMYRIAN
ApplicantIntrasense
Product CodeLLZ · Radiology
Decision DateMay 14, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

Myrian® with its modules is a software application that is used for reviewing medical images. Myrian® with its modules is designed to run on standard PC hardware. The hardware is all "off-the-shelf' standard computer components and may be purchased independently by the end user. Myrian with its modules receives digital images and data from various sources (including but not limited to CT, MR, US, RF modalities, computed and direct radiographic devices, secondary capture devices, scanners, imaging gateways or imaging sources like PACS). Myrian® and its Modules can be used to communicate, to process, print and display medical images. Users have access to various image processing, Volume of Interest editing and measurement tools to assist them in viewing and quantifying images. Myrian® and its Modules can be integrated with an institution's existing HIS or RIS for a fully integrated electronic patient record. Typical users of Myrian® with its Modules are trained medical professionals, including but not limited to radiologists, technologists and clinicians. When interpreted by a trained physician, filmed or displayed images on the Myrian® and its Modules may be used as a basis for diagnosis, except in the case of mammography images. Myrian® with its Modules does not support the display of mammography images for diagnosis.

Device Story

Myrian is a software suite for medical image review and analysis; runs on off-the-shelf PC hardware. Inputs include DICOM images from modalities (CT, MR, US, RF, PACS). Features include thin/thick MPR, 3D volume rendering, VOI creation, and quantification (volume, surface, density). Used by radiologists, clinicians, and technologists in clinical settings. Output is displayed on monitors for diagnostic review and reporting. Facilitates clinical decision-making by providing visualization and measurement tools. Integrates with HIS/RIS for electronic patient records. Benefits include improved workflow for image interpretation and patient examination follow-up.

Clinical Evidence

Bench testing only. Performance verified against FDA "Guidance of the Content of Pre Market Submissions for Software Contained in Medical Devices." User site testing and benchmarking demonstrate compliance with specifications. No clinical data or adverse effects reported.

Technological Characteristics

Software-only system; runs on standard off-the-shelf PC hardware. Supports DICOM standard for image communication and processing. Features include MPR, 3D volume rendering, and measurement tools. No specific hardware materials or sterilization required as it is a software application.

Indications for Use

Indicated for trained medical professionals (radiologists, technologists, clinicians) to review and analyze medical images (CT, MR, US, RF, CR/DR) for anatomy and pathology assessment. Contraindicated for mammography image interpretation.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## Intrasense PREMARKET NOTIFICATION 510(K) SUBMISSION MYRIAN 060830 . | 5. 510(K) SUMMARY | | | | |----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--| | | [As Required by 21 CFR 807.92]<br>Summary of Safety and Effectiveness | MAY 14 2007 | | | Preparation date | April 6th, 2007 | | | | Submitter | | | | | Name | INTRASENSE | | | | Registration number | TBD | | | | Address | CAP OMEGA, Rond Point Benjamin Franklin CS 39521 34960 Montpellier FR | | | | Tel Fax | (+33) 467 130 130 / (+33) 467 130 132 | | | | Contact Person | Mr Stéphane CHEMOUNY<br>Phone number : (+33) 467 130 130<br>Fax number : (+33) 467 130 132 | | | | Device name | | | | | Common Name | System, Image Processing | | | | Trade Name | MYRIAN | | | | Model number | N/A | | | | Device classification | | | | | Classification name | System, Image Processing, Radiological | | | | Code product | LLZ | | | | Panel | 892 | | | | Regulation number | 892.=2050 | | | | Regulatory class | II | | | | Predicate devices | [K052995] Cleared [November 8, 2005] [General Electric Medical Systems]<br>[Advantage Workstation Version 4.3], manufactured by [General Electric Medical<br>Systems]<br>[K061624] Cleared [June 27, 2006] [Vital Images, Inc.] [Vitrea2 Version 3.9],<br>manufactured by [Vital Images, Inc.] | | | | Description | Myrian® system is a software suite providing the following services :<br>Import of DICOM images from any DICOM modality, workstation or PACS<br>Visualization of DICOM images in thin MPR, thick MPR and full 3D volume rendering<br>Creation of VOI (Volume Of Interest) with dedicated tools<br>Calculation of volumes, surface and of average, minimum and maximum densities of<br>VOI<br>Follow-up of patient examination<br>Generation of medical reports<br>Export of DICOM images to any format, DICOM entity or media | | | | Explanation of how the device operates | Myrian® with its modules is designed to run on standard PC hardware. The hardware<br>is all "off-the-shelf' standard computer components and may be purchased<br>independently by the end user. | | | | Intended use | Myrian® with its modules is a software application that is used for reviewing medical<br>images. Myrian® with its modules is designed to run on standard PC hardware. The<br>hardware is all "off-the-shelf' standard computer components and may be purchased<br>independently by the end user. Myrian with its modules receives digital images and<br>data from various sources (including but not limited to CT, MR, US, RF modalities,<br>computed and direct radiographic devices, secondary capture devices, scanners,<br>imaging gateways or imaging sources like PACS).<br>Myrian® and its Modules can be used to communicate, to process, print and display<br>medical images. Users have access to various image processing, Volume of Interest<br>editing and measurement tools to assist them in viewing and quantifying images.<br>Myrian® and its Modules can be integrated with an institution's existing HIS or RIS for<br>a fully integrated electronic patient record. | | | | | Typical users of Myrian® with its Modules are trained medical professionals, including<br>but not limited to radiologists, technologists and clinicians. | | | | | When interpreted by a trained physician, filmed or displayed images on the Myrian®<br>and its Modules may be used as a basis for diagnosis, except in the case of<br>mammography images.<br>Myrian® with its Modules does not support the display of mammography images for<br>diagnosis. | | | | Performance data | Performance data were verified versus the requirements of the FDA "Guidance of the<br>Content of Pre Market Submissions for Software Contained in Medical Devices" | | | | | User Site Testing, Benchmarking and clinical data analysis demonstrate that MYRIAN<br>meet the required specifications. No adverse affects have been detected. | | | | Substantial equivalence summary | The technological characteristics, features, specifications, materials, mode of<br>operation, and intended use of MYRIAN device are equivalent to those of the<br>predicate devices quoted above. | | | | | MYRIAN is the same as the predicate devices in K061624 and K052995 | | | | | The differences that exist between the devices do not raise new issues of safety or<br>effectiveness regarding MYRIAN Device. | | | | | It is substantially equivalent in terms of safety and effectiveness to the predicate<br>devices. | | | {1}------------------------------------------------ Intrasense PREMARKET NOTIFICATION 510(K) SUBMISSION 060830 {2}------------------------------------------------ Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged in a circular pattern around the symbol. The logo is black and white. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 INTRASENSE % Mr. Nicolas Clary Project Leader CEISO "Les Portes d'Espagne" - Bât. B - 99, route d'Espagne 31100 Toulouse 31100 FRANCE MAY 1 4 2007 Re: K071000 Trade/Device Name: Myrian Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 6, 2007 Received: April 13, 2007 Dear Mr. Clary: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/2/Picture/10 description: The image is a circular logo for the FDA Centennial, celebrating 1906-2006. The logo features the letters "FDA" in a stylized font, with the word "Centennial" written below. The logo is surrounded by a circular border with text and stars. moting Public J {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean I toase or acressed a determination that your device complies with other requirements of the Act that I Dederal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally premainer nedicated device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you use office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under (21 CFF at 80 ! Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincerely yours, Nancy C. Hogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ PREMARKET NOTIFICATION 510(K) SUBMISSION MYRIAN Intrasense 060830 | 4. INDICATIONS FOR USE | | |------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number (if known): | K07/000 | | Device Name: | Myrian | | Indications for Use: | Myrian is a multi modality medical diagnostic device. It is aimed at reviewing and analysing anatomy and pathology. It also includes DICOM communication capabilities and media interchange features (printing, CD burning, storing). It runs on any standard PC including laptops that might be purchased independently by the end user. It provides user a set of tools meant to create and modify volumes of interest. | | | This device is not indicated for mammography use. Lossy compressed mammography images and digitized film screen images must not be used for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 mega pixel resolution and meets other technical specifications approved by the FDA. | | <div style="text-align: left;">Prescription Use</div> <div style="text-align: left;">(Part 21 CFR 801 Subpart D)</div> | <div style="text-align: left;">1.1 AND/OR</div> <div style="text-align: left;">Over-The-Counter Use</div> <div style="text-align: left;">(21 CFR 801 Subpart C)</div> | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) | | | Concurrence of CDRH, Office of Device Evaluation (ODE) | | . . Nancy C Hogdon (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 071000 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...