Precise Position

K203514 · Philips Healthcare (Suzhou) Co., Ltd. · JAK · Jun 17, 2021 · Radiology

Device Facts

Record IDK203514
Device NamePrecise Position
ApplicantPhilips Healthcare (Suzhou) Co., Ltd.
Product CodeJAK · Radiology
Decision DateJun 17, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
AttributesAI/ML, Pediatric

Intended Use

The Precise Position is intended for use with Philips Incisive CT systems. The following guided workflow. - Patient orientation identification - · Surview range recommendation - · Automatic centering the patient anatomy - · Provide visual images of patient on the table Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies. Patient population limitation: Patient younger than 16 years are not supported.

Device Story

Precise Position is an optional feature for Philips Incisive CT systems to assist patient positioning. A ceiling-mounted camera with color and depth sensing captures images of the patient on the table. Data is transmitted via high-speed fiber/copper hybrid USB cable to the system. An AI algorithm (Convolutional Neural Network) processes input to detect patient joints, orientation, and anatomy center, and recommends surview start/end positions. The system displays visual images of the patient to the operator. Used in hospitals and clinics by radiology staff, the device automates positioning, reducing setup time and improving vertical/horizontal centering accuracy and consistency. It benefits patients by streamlining the workflow and enhancing positioning precision before CT scans.

Clinical Evidence

Clinical evaluation involved 80 scan positions (40 with Precise Position, 40 without) assessed by 5 experts. Results showed up to 23% reduction in patient positioning time, up to 50% increase in vertical (iso)center accuracy, and up to 70% increase in vertical/horizontal positioning consistency. No radiation was delivered to volunteers. Study concluded the device performs as intended with a favorable benefit-risk ratio.

Technological Characteristics

Ceiling-mounted color/depth camera; high-speed fiber/copper hybrid USB connectivity; powered by CT gantry. Software utilizes a Convolutional Neural Network (CNN) for joint detection and anatomy localization. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 14971, and IEC 62304. Operates as an integrated feature of the Philips Incisive CT system.

Indications for Use

Indicated for CT imaging of head, chest, abdomen, pelvis, and combinations thereof. Patient population: adults and children 16 years and older. Contraindications: patients younger than 16 years; decubitus orientations.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Philips Healthcare (Suzhou) Co., Ltd. % Shiguang An Advanced Regulatory Engineer No. 258, Zhong Yuan Road, Suzhou Industrial Park Suzhou, Jiangsu 215024 CHINA Re: K203514 Trade/Device Name: Precise Position Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: May 19, 2021 Received: May 19, 2021 Dear Shiguang An: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for June 17, 2021 {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K203514 Device Name Precise Position Indications for Use (Describe) The Precise Position is intended for use with Philips Incisive CT systems. The following guided workflow. - Patient orientation identification - · Surview range recommendation - · Automatic centering the patient anatomy - · Provide visual images of patient on the table Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies. Patient population limitation: Patient younger than 16 years are not supported. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"> <span>☒</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div style="display:flex; align-items:center;"> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the collection of information is estimated to average 79 hours p time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." EF PSC Publishing Services (301) 443-6740 {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text. ## 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92. | Date Prepared: | November 18, 2020 | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, ZhongYuan Road, Suzhou Industrial Park, 215024<br>Suzhou, Jiangsu Province, PEOPLE'S REPUBLIC OF CHINA<br>Establishment Registration Number: 3009529630 | | Primary Contact<br>Person: | Shiguang An<br>Advanced Regulatory Engineer<br>Phone: +86-13940106467<br>E-mail: shiguang.an@philips.com | | Secondary Contact<br>Person: | Erhong Wang<br>Senior Regulatory Affairs Manager<br>Phone: +86-13021019589<br>E-mail: erhong.wang@philips.com | | Device Name: | Precise position | | Classification: | Classification Name<br>Computed tomography x-ray system<br>Classification Regulation:<br>21CFR §892.1750<br>Classification Panel:<br>Radiology<br>Device Class:<br>Class II<br>Primary product code:<br>JAK | | Predicate Device: | Trade Name:<br>Philips Incisive CT<br>Manufacturer:<br>Philips Healthcare (Suzhou) Co., Ltd.<br>510(k) Clearance:<br>K180015-March 20, 2018<br>Classification Regulation:<br>21 CFR, Part 892.1750<br>Classification Name:<br>Computed tomography x-ray system<br>Classification Panel:<br>Radiology<br>Device Class:<br>Class II<br>Product Code<br>JAK | | Reference Device: | Manufacturer:<br>Auto Positioning | | | GE Hangwei Medical System<br>Co., Ltd. | | | | | 510(k) Clearance: | K192956 (January 16, 2020) | | Classification Regulation: | 21 CFR, Part 892.1750 | | Classification Name: | Computed tomography x-ray<br>system | | Classification Panel: | Radiology | | Device Class: | Class II | | Product Code | JAK | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. The background is white. | Device<br>description: | Precise Position is an optional feature to assist user for position the<br>patient before the body examination such as CT scan. The purpose<br>of this feature is to reduce the patient position time via the camera<br>detection and calculation result. It includes automatic detect patient<br>orientation, patient anatomy scan range and center of patient<br>anatomy. | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Precise Position including a camera with both color and depth<br>function is installed in the ceiling of the scan room, in such a way<br>to cover the entire patient on the patient table. The camera control<br>and image data transmit via the high speed fiber and copper hybrid<br>USB cable. The power supply of the camera is from the gantry.<br>Precise position adopts the AI algorithm (Convolution Neural<br>Network) to detect the joints of the patient body, and then identify<br>surview start/end position and patient orientation. The algorithm<br>can also support detect center of patient anatomy. | | | Limitation for Precise Position<br>There is no limitation for Precise Position except below items:<br>• Patients below the age of 16 are not supported.<br>• Decubitus orientations are not supported. | | | The Precise Position display results may get affected by the<br>following conditions:<br>• When the patient is covered by sheet, blanket etc.,<br>• When the patient is not completely covered by the ceiling camera<br>view, e.g. blocked by the gantry or out of camera's FOV etc.<br>• When the patient is wearing clothes that reflects light, e.g. plastic-<br>like clothes.<br>• When the patient is wearing black clothes.<br>• When the patient is wearing thick clothes.<br>• When there are other people around the patient. | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the words "Traditional 510(k)". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Traditional" appearing first, followed by "510(k)". Image /page/5/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame. | Indications for Use: | The Precise Position is intended for use with Philips Incisive CT systems. The device provides the following guided workflow.<br>Patient orientation identification Surview range recommendation Automatic centering the patient anatomy Provide visual images of patient on the table Precise position is indicated for use for CT imaging of the head, chest, abdomen, pelvis, and combination of those anatomies.<br>Patient population limitation: Patient younger than 16 years are not supported. | |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is the primary focus of the image and dominates the visual space. | Fundamental<br>scientific<br>technology: | Based on the information provided above, the precise position is<br>considered substantially equivalent to the primary currently<br>marketed and predicate device Philips Incisive CT (K180015,<br>20/March/2018) in terms of fundamental scientific technology. | | | |------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Attribute | Predicate Device<br>Philips Incisive CT<br>(K180015) | Proposed Device Precise<br>Position | | | Patient<br>positioning<br>workflow | Manually position<br>the patient via the<br>couch motion button<br>and laser. Normally,<br>need several round<br>adjustments by user. | Precise position provides<br>the auto workflow to set<br>the surview start/end<br>position, center of patient<br>anatomy, and patient<br>orientation. | | | Detection<br>algorithm | The manual<br>workflow does not<br>need special<br>algorithm. | Precise position adopt the<br>AI algorithm (Convolution<br>Neural Network) to detect<br>the joints of the patient<br>body automatically, and<br>then automatic detect<br>surview start/end position,<br>center of patient anatomy<br>and patient orientation. | | | Hardware<br>need to<br>support the<br>patient<br>positioning | Manual patient<br>positioning does not<br>need any unique<br>hardware except<br>traditional exiting<br>hardware on the CT<br>or other system.<br>Exiting hardware<br>includes table<br>motion button<br>(in/out, up/down) on<br>gantry panel or<br>CTBOX, lasers and<br>etc. | Precision position feature<br>need color and depth<br>camera, high speed USB<br>transmission cable, and<br>power cable, and relative<br>the enhanced workflow.<br>The tradition exiting<br>patient positioning<br>hardware still be available<br>for user all the time, it is<br>convenient to switch<br>between the camera<br>automatically detecting<br>mode and manual mode. | | | Patient<br>population | All ages | More than 16 years old. | | | Environment<br>of use | Hospitals, outpatient<br>clinics, research<br>institutions, and<br>other clinical<br>facilities. | Hospitals, outpatient<br>clinics, research<br>institutions, and other<br>clinical facilities. | | Summary of Non-<br>Clinical | The Precise Position complies with the following international and<br>FDA-recognized consensus standards: | | | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text. | Performance data: | • AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R) 2012 and A2:2010/(R) 2012<br>(Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements. For Basic Safety and Essential Performance (IEC 60601-1:2012, MOD).<br>FDA/CDRH recognition number 19-4<br>• IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance -- Collateral standard: Electromagnetic disturbances - Requirements and tests.<br>FDA/CDRH recognition number 19-8<br>• ISO 14971 Medical devices – Application of risk management to medical devices. FDA/CDRH recognition number 5-40.<br>• IEC 62304:2015, Medical device software -- Software life cycle processes<br>FDA/CDRH recognition number 13-79<br>There are no risks identified in risk management documentation that require clinical data for the purpose of clinical evaluation; Risk Management Plan as Appendix 001, Risk Management Report as Appendix 002, and Risk Management Matrix as Appendix 003 of Precise Position.<br>There are no clinical risks identified by the evaluated clinical data.<br>Sufficient evidence is available to demonstrate the ability of Precise Position achieve the intended performances during normal condition of use.<br>Full consistency exists between the state-of-the-art, the evaluated data, the risk management documentation and the information materials supplied.<br>Therefore, the Precise Position is substantially equivalent to the primary currently marketed and predicate device (K180015, 20/March/2018) in terms of safety and effectiveness. | |------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Summary of<br>Clinical Data: | Precise Position is evaluation covered total 80 clinical scan positions in which 40 cases used with Precise Position and another 40 cases without the usage of Precise Position in order to meet the sample size calculation of 40 cases performed average on a CT scanner per day.<br>The thorough clinical evaluation of this feature is done by 5 Clinical experts.<br>The testing did not deliver radiation to volunteers as Precise Position is only for supporting the positioning of the patient for the localization radiograph and therefore “radiation" was not required for those volunteers. | | | | | | The clinical evaluation was done with 3 major objectives as follows, | | | To calculate the time saved per surview planning with and without | | | Precise Position. | | | To calculate the accuracy of vertical (iso)center positioning and | | | Surview (horizontal) start /end position with and without Precise<br>Position | | | Intra operator consistency in positioning the patient and surview scan<br>range | | | The summary of clinical evaluation testing was clearly demonstrating<br>the objectives in terms of user benefits as shown below<br>Average Time taken to position without Precise Position and with<br>Precise Position is recorded. Time at user select the patient from<br>Gantry Panel, to time when user pressed Go button from Gantry Panel<br>is measured and concluded that up to 23% time reduction in patient<br>positioning achieved with “Precise Positioning workflow”. | | | The average offset in mm for vertical (iso) center position among 5<br>operators without Precise Position and with Precise Position is<br>recorded and results shown that with "Precise Position" the vertical<br>position accuracy is increased up to 50%. | | | Standard deviation in mm for vertical (iso)center positioning &<br>Surview (horizontal) start position among 5 operators without Precise<br>Position and with Precise Position is recorded and results shown up to<br>70% increase in Vertical and horizontal position consistency with<br>Precise Position. | | | Overall, the Precise Position Clinical Review Report. (Appendix_007)<br>concluded that Philips Incisive CT systems with Precise Position,<br>under normal condition of use, perform as intended, are safe for its<br>intended use and have a favorable benefit-risk ratio. Further clinical<br>investigations are not necessary, as sufficient evidence exists to<br>support these conclusions. | | Substantial | The Precise Position is substantially equivalent to the primary | | Equivalence | currently marketed and predicate device (K180015, 20/March/2018) in | | Conclusion: | terms of design features, fundamental scientific technology, | | | indications for use, and safety and effectiveness. Additionally, | | | substantial equivalence was demonstrated with non-clinical | | | performance tests, which complied with the requirements specified in | | | the international and FDA-recognized consensus standards, IEC 62304 | | | and ISO 14971. The results of these tests demonstrate that Precise | | | Position met the acceptance criteria and is adequate for this intended | | | use. | {8}------------------------------------------------ ## Traditional 510(k) Image /page/8/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...