!M1

K170607 · Solutions For Tomorrow AB · IZL · Jul 17, 2017 · Radiology

Device Facts

Record IDK170607
Device Name!M1
ApplicantSolutions For Tomorrow AB
Product CodeIZL · Radiology
Decision DateJul 17, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2
AttributesPediatric

Intended Use

The device is designed to perform general radiography X-ray examinations on all paediatric and all adult patients, in all patient treatment areas. Treatment areas are defined as professional health care facility environments where operators with medical training are continually present during patients' examinations.

Device Story

The !M1 is a mobile digital radiography system consisting of a motorized, battery-powered cart, x-ray generator, x-ray tube, collimator, and touchscreen operator console. It is operated by trained medical personnel in healthcare facilities to perform diagnostic x-ray exams. The system integrates with previously cleared Canon or Konica Minolta flat-panel digital detectors to capture radiographic images. The device transforms x-ray exposure inputs into digital images displayed on the console. The output is used by clinicians for diagnostic decision-making. The system's mobility allows for bedside imaging, providing clinical benefit by enabling examinations outside of stationary x-ray rooms.

Clinical Evidence

No clinical testing was performed. Substantial equivalence was demonstrated through non-clinical bench testing, including usability, battery performance, driving distance, and generator exposure time comparisons. Integration testing was conducted on the flat-panel digital imagers to ensure functional reproducibility.

Technological Characteristics

Mobile x-ray system with motorized battery-powered cart. Components include x-ray generator, tube, collimator, and touchscreen operator console. Utilizes wireless flat-panel digital detectors (Canon/Konica Minolta). Software is classified as moderate level of concern. System is designed for use in professional healthcare environments.

Indications for Use

Indicated for general radiography X-ray examinations in all pediatric and adult patients in professional healthcare facilities.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 17, 2017 Solutions for tomorrow AB % Mr. Martin Yngvesson COO Saxagardsv. 4 Vackelsang, 36251 Kronoberg SWEDEN Re: K170607 Trade/Device Name: !M1 Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL Dated: June 19, 2017 Received: June 19, 2017 Dear Mr. Yngvesson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Image /page/0/Picture/11 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other, with flowing lines beneath them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K170607 Device Name !M1 ( Indications for Use (Describe) The device is designed to perform general radiography x-ray examinations on all pediatric and all adult patients, in all patient treatment areas. Treatment areas are defined as professional health care facility environments where operators with medical training are continually present during patients' examinations. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov 4/7-2017 "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 {3}------------------------------------------------ Image /page/3/Picture/0 description: The image contains the text "SOLUTIONS FOR TOMORROW" in a sans-serif font. To the left of the text is a circular icon with a light blue fill. Inside the circle is an exclamation point, also in white. The overall design is simple and clean. # "510(k) Summary" | 510(k) Owner Name: | Solutions for tomorrow AB | |------------------------|---------------------------------------------------------------------------------| | 510(k) Owner Address: | Saxagårdsvägen 4<br>362 51 Väckelsång | | 510(k) Owner Phone: | +46-(0)10 4564500 | | Contact Person & Info: | Mattias Guldstrand<br>CEO<br>mattias@solutionsfortomorrow.se<br>+46 70 299 6943 | | Date Summary Prepared: | January 31, 2017 | | Device Trade Name: | !M1 | | Device Common Name: | mobile x-ray system | | Classification Name: | System, X-ray, Mobile | | Regulation Number: | 21 CFR 892.1720 | | Device Class: | Class II | | Product Code: | IZL | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the text "SOLUTIONS FOR TOMORROW" in a sans-serif font. To the left of the text is a light blue circle with a white exclamation point inside. The text is dark gray and appears to be the title or heading of a document or presentation. | Predicate Device: | Carestream DRX-Revolution | |----------------------|------------------------------------| | Manufactured by: | Carestream Health, Inc. | | 510(k) Number: | K120062 (January 5, 2011) | | Device Trade Name: | DRX-Revolution Mobile X-ray System | | Device Common Name: | mobile x-ray system | | Classification Name: | System, X-ray, Mobile | | Regulation Number: | 21 CFR 892.1720 | | Device Class: | Class II | | Product Code: | IZL | # Device Description The ModelOne mobile X-ray system is a diagnostic mobile x-ray system utilizing digital radiography technology. The device consists of a self-contained x-ray generator, image receptor(s), imaging display and software for acquiring medical diagnostic images both inside and outside of a standard stationary x-ray room. The ModelOne system incorporates a flat-panel(s) detector that can be used wirelessly for exams as in-bed projections. The system is intended to be marketed with two options with flat-panel digital images from Canon and Konica Minolta. The flat-panel digital imagers have been previously cleared with the following: - K133693 Digital Radiography CXDI-401C Wireless (Canon) - K131106 Digital Radiography CXDI-701C Wireless (Canon) Digital Radiography CXDI-701G Wireless Digital Radiography CXDI-801C Wireless Digital Radiography CXDI-801G Wireless - K102349 AeroDR SYSTEM (Konica Minolta) - K113248 AeroDR SYSTEM with P-21 (Konica Minolta) - K130936 AeroDR SYSTEM with P-31 (Konica Minolta) - K141271 AeroDR SYSTEM 2 (Konica Minolta) - K162504 -SKR 3000 (Konica Minolta) {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the text "SOLUTIONS FOR TOMORROW" next to a blue circle with an exclamation point inside. The text is in a dark gray color and is written in all capital letters. The exclamation point is white. The image is simple and clean, with a focus on the text and the symbol. # Indications for Use / Intended Use: The Indications for Use for the device, as described in its labeling, are: "The device is designed to perform general radiography X-ray examinations on all paediatric and all adult patients, in all patient treatment areas". Note The patient subpopulations are identified by weight. The equipment settings for routinely performed examinations for those populations can be found in User Manual, section Exposure parameters. However, the image post-processing is defined in protocols from Konica Minolta and Canon image software that have been already approved by FDA. # Comparison of Technological Characteristics The !M1 Mobile X-ray System is substantially equivalent to the predicate device, the DRX-Revolution Mobile X-ray System manufactured by Carestream Health, Inc. The !M1 and DRX-Revolution system both consist of an x-ray generator, x-ray tube, collimator and graphical user interface on an operator console with touch screen monitor. These components are mounted on a motorized cart that is battery powered to enable the device to be moved from location to location. Both systems utilize digital flat panel detector for image capture. # Discussion of Testing # Non-clinical testing The performance characteristics and operation of !M1 were evaluated in non-clinical testing. Performance tests include the following: usability test, battery performance, driving distance and generator performance test. The usability test includes the acceptance test performed on the customer site and performance test that has been done at the hospital by professional personal. Battery performance tests consists of beginning of life/end of life test. The driving distance test was performed to verify maximum distance of driving from full to empty battery. The aim of generator performance test was to compare the time of exposure of !M1 and its competitors. Integration test was performed on the previously cleared flat-panel digital imagers in order to demonstrate that all components of the device function in a reproductive way according to the design specifications. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the text "SOLUTIONS FOR TOMORROW" in a sans-serif font. To the left of the text is a light blue circle with a white exclamation point inside. The text is dark gray and appears to be a logo or heading. Performance tests confirm that the device is as effective and performs as well as or better than the predicate device. The software risk is classified as moderate level of concern device according to the Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Technological differences do not raise questions of safety and the device is as safe as legally marketed DRX-Revolution by Carestream. # Clinical testing No clinical testing was performed on the subject device, and performance testing listed in the "Non-Clinical Testing" section was sufficient to demonstrate substantial equivalence.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...