NEOPROBE 1000 GEN 1C PORTABLE RADIOISOTOPE DETECTOR & ACCESSORIES
Applicant
Neoprobe Corp.
Product Code
IZD · Radiology
Decision Date
Sep 13, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1320
Device Class
Class 1
Indications for Use
The Neoprobe® Model 1000 GEN 1C Portable Radioisotope Detector is an electronic device intended to detect and quantify gamma radiation. The GEN 1C is indicated for external and intraoperative detection of radioactivity in body tissues or organs in medical procedures where gamma emitting isotopes are administered.
Device Story
Device is a microcomputer-based control unit with a cadmium zinc telluride (CZT) gamma ray detector probe. Used in clinical settings for external and intraoperative detection/quantification of gamma radiation in tissues/organs following administration of gamma-emitting isotopes. System includes control unit, probe, cable, check source (129I), and noise adjustment fixture. Optional accessories include collimators and shields. Physician operates device to locate radioactive tracers; output provides radioactivity counts to assist in surgical or diagnostic localization. Benefits include precise identification of radioactive tissues during procedures.
Clinical Evidence
Bench testing only. Laboratory studies compared 11 mm GEN 1C probes to the predicate 19 mm GEN 1B probe. Testing included counting efficiency comparisons using 129I, counting efficiency vs. distance for 125I, 111In, and 99mTc, and spatial resolution measurements with/without collimation. Results demonstrated functional equivalence between the 11 mm GEN 1C probes and the predicate 19 mm probe.
Technological Characteristics
Microcomputer-based control unit; cadmium zinc telluride (CZT) gamma ray detector; probe diameters 11 mm and 19 mm; aluminum probe cap; check source (129I); noise adjustment fixture (chrome-plated, Teflon-lined lead).
Indications for Use
Indicated for external and intraoperative detection of radioactivity in body tissues or organs in patients undergoing medical procedures involving the administration of gamma-emitting isotopes.
Regulatory Classification
Identification
A nuclear uptake probe is a device intended to measure the amount of radionuclide taken up by a particular organ or body region. This generic type of device may include a single or multiple detector probe, signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Neoprobe Model 1000 GEN 1B Portable Radioisotope Detector and Accessories (K942580)
Related Devices
K983768 — MODIFICATION TO GAMMA GUIDANCE SYSTEM · Radiation Monitoring Devices, Inc. · Jul 2, 1999
K971167 — NEOPROBE MODEL 1500 PORTABLE RADIOISOTOPE DETECTOR AND ACCESSORIES · Neoprobe Corp. · Jun 26, 1997
K093706 — PROXISCAN MODEL 101 · Hybridyne Imaging Technologies, Inc. · Apr 2, 2010
K022342 — EZ-SCOPE AN, EZ-SCOPE LIGHT · Anzai Medical Co., Ltd. · Oct 9, 2002
K982442 — C-TRAK AUTOMATIC ANALYZER (NAME CHANGE FROM ONCOPROBE II TO C-TRAK) · Care Wise Medical Products Corp. · Aug 25, 1998
Submission Summary (Full Text)
{0}
SEP 13 1996
K 9623/9
510(k) SUMMARY
Neoprobe Corporation's Model 1000 GEN 1C
Portable Radioisotope
Detector and Accessories
Submitter's Name, Address, Telephone Number, Contact Person
and Date Prepared
Submitter
Neoprobe Corporation
425 Metro Place North
Suite 400
Dublin, OH 43017-1367
Phone: (614) 793-7500
Facsimile: (614) 793-7520
Contact Person
Susan D. Tiedy-Stevenson
Director of Regulatory Affairs
Neoprobe Corporation
Phone: (301) 620-9416
Facsimile: (301) 620-1438
Date Prepared: June 14, 1996
Name of Device and Name/Address of Sponsor
Neoprobe Model 1000 GEN 1C Portable Radioisotope Detector and Accessories
Neoprobe Corporation
425 Metro Place North
Suite 400
Dublin, OH 43017-1367
Phone: (614) 793-7500
Facsimile: (614) 793-7520
{1}
Common or Usual Name
Radioisotope detector and accessories
Classification Name
Nuclear uptake probe and accessories
Predicate Device
Neoprobe Model 1000 GEN 1B Portable Radioisotope Detector and Accessories (K942580)
Intended Use
The Neoprobe® Model 1000 GEN 1C Portable Radioisotope Detector is an electronic device intended to detect and quantify gamma radiation. The GEN 1C is indicated for external and intraoperative detection of radioactivity in body tissues or organs in medical procedures where gamma emitting isotopes are administered.
Technological Characteristics and Substantial Equivalence
Device Description
The Neoprobe® Model 1000 GEN 1C Portable Radioisotope Detector ("Model 1000") consists of a microcomputer-based control unit, a cadmium zinc telluride gamma ray detector probe, a cable used to connect the probe to the control unit, a check source containing a known quantity of $^{129}\mathrm{I}$, and a chrome-plated Teflon-lined lead noise adjustment fixture used for periodic verification that the system continues to meet factory calibration readings. The detector probe is available in two sizes: (1) a detector probe with a 19 mm external diameter and a detector diameter of 15 mm; and (2) a detector probe with an 11 mm external diameter and a detector diameter of 7 mm. Optional accessories include a detector probe collimator and a detector probe shield.
Safety and Effectiveness
The GEN 1C is substantially equivalent to its predecessor, the Model 1000 GEN 1B, which received premarket clearance in 1995 (K942580). The GEN 1C is identical to the predicate device in general intended use, design, and principle of operation. The only differences between the GEN 1C and the predicate device are the use of a single-piece aluminum probe cap rather than a probe cap comprised of multiple parts, addition of a smaller (11 mm) detector probe, and use of a two millimeter radioactive
2
{2}
encapsulated area (check source) with the existing 19 mm probe and with the 11 mm detector probe. These minor differences raise no new issues of safety or effectiveness because the modified device continues to perform identical functions in the same manner as the predicate device.
## Performance Data
Neoprobe has conducted laboratory studies to demonstrate the functional equivalence of the 11 mm GEN 1C probes to the predicate Model 1000 GEN 1B probe. These tests included: (1) a comparison of the counting efficiencies of six randomly selected 11 mm GEN 1C probes using a calibrated source of $^{129}\mathrm{I}$; (2) measurement of the counting efficiency of a representative probe as a function of probe-to-source distance for $^{125}\mathrm{I}$, $^{111}\mathrm{In}$, and $^{99\mathrm{m}}\mathrm{Tc}$; and (3) measurement of the spatial resolution of the probe with and without collimation for the three nuclides. In summary, the counting efficiency of the 11 mm GEN 1C probes was found to be functionally equivalent to the counting efficiency of the predicate 19 mm Model 1000 GEN 1B probe.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.