NEOPROBE MODEL 1500 PORTABLE RADIOISOTOPE DETECTOR AND ACCESSORIES
Applicant
Neoprobe Corp.
Product Code
IZD · Radiology
Decision Date
Jun 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1320
Device Class
Class 1
Indications for Use
The Neoprobe® Model 1500 Portable Radioisotope Detector is an electronic device intended to detect and quantify gamma radiation. It is indicated for external and intraoperative detection of radioactivity in body tissues or organs, such as bowel, bone, lymphatics, and red blood cells, where radiopharmaceuticals are administered.
Device Story
Neoprobe Model 1500 is a microcomputer-based portable radioisotope detector; detects/quantifies gamma radiation in body tissues/organs. System includes control unit, reusable steam-sterilizable detector probe tips (12mm collimated/non-collimated, 19mm non-collimated) containing cadmium zinc telluride crystal, disposable sterile probe handle with cable, check source disk, and noise adjustment fixture. Used intraoperatively or externally by physicians to identify areas of increased radionuclide concentration (e.g., ischemic bowel, bone, thyroid, colon, lymphatics). Output provided to clinician to guide surgical decision-making; facilitates localization of tissues labeled with radiopharmaceuticals. Benefits include improved intraoperative identification of target tissues.
Clinical Evidence
Bench testing only. Laboratory studies compared counting efficiency and spatial resolution of new 12mm probe tips against the cleared 11mm probe using 125I, 111In, and 99mTc sources. Results confirmed functional equivalence. Literature previously submitted in K942580 and K864263 supports clinical applications for ischemic bowel, bone, thyroid, and colon; additional literature provided for lymphatic applications.
Technological Characteristics
Microcomputer-based control unit; cadmium zinc telluride (CZT) crystal gamma ray detector; reusable steam-sterilizable probe tips (12mm/19mm); disposable polypropylene-cabled handle. Connectivity: standalone. Energy source: electronic. Calibration verification via chrome-plated Teflon-lined lead noise adjustment fixture.
Indications for Use
Indicated for external and intraoperative detection of radioactivity in body tissues or organs (e.g., bowel, bone, lymphatics, red blood cells) in patients administered radiopharmaceuticals. Prescription use only.
Regulatory Classification
Identification
A nuclear uptake probe is a device intended to measure the amount of radionuclide taken up by a particular organ or body region. This generic type of device may include a single or multiple detector probe, signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Neoprobe Model 1000 GEN 1C Portable Radioisotope Detector and Accessories (K962319)
{0}
K971167
# 510(k) SUMMARY
JUN 26 1997
Neoprobe Corporation's Model 1500
Portable Radioisotope
Detector and Accessories
## Submitter's Name, Address Telephone Number, Contact Person and Date Prepared
**Submitter**
Neoprobe Corporation
425 Metro Place North
Suite 400
Dublin, OH 43017-1367
**Contact Person**
Susan D. Tiedy-Stevenson
Director of Regulatory Affairs
Neoprobe Corporation
Phone: (301) 620-9416
Facsimile: (301) 620-1438
Date Prepared: March 28, 1997
## Name of Device and Name/Address of Sponsor
Neoprobe Model 1500 Portable Radioisotope Detector and Accessories
Neoprobe Corporation
425 Metro Place North
Suite 400
Dublin, OH 43017-1367
Phone: (614) 793-7500
Facsimile: (614) 793-7520
## Common or Usual Name
Radioisotope detector and accessories
## Classification Name
Nuclear uptake probe and accessories
\VV\DC - 61852/2 - 0425053.02
{1}
# Predicate Devices
Neoprobe Model 1000 GEN 1C Portable Radioisotope Detector and Accessories (K962319).
CareWise C-Trak Biopsy System (K922117)
# Intended Use
The Neoprobe® Model 1500 Portable Radioisotope Detector is an electronic device intended to detect and quantify gamma radiation. It is indicated for external and intraoperative detection of radioactivity in body tissues or organs, such as bowel, bone, lymphatics, and red blood cells, where radiopharmaceuticals are administered.
# Device Description
The Neoprobe® Model 1500 Portable Radioisotope Detector (“Model 1500”) consists of (1) a microcomputer-based control unit; (2) a 12 mm collimated detector probe tip, a 12 mm non-collimated detector probe tip, and a 19 mm non-collimated probe tip, all of which are reusable, steam sterilizable and contain a cadmium zinc telluride crystal gamma ray detector; (3) a disposable sterile probe handle with an attached polypropylene cable; (4) a check source disk containing a known quantity of $^{125}\mathrm{I}$; and (5) a chrome-plated Teflon-lined lead noise adjustment fixture used for periodic verification that the system continues to meet factory calibration readings.
# Safety and Effectiveness
The Model 1500 is substantially equivalent to the Model 1000 GEN 1C, which received premarket clearance in 1996 (K92319). This submission covers modifications to the device since the Model 1000 GEN 1C received premarket clearance. These modifications include: (1) a modified control unit, (2) a disposable detector probe handle with attached cable, (3) a 12 mm noncollimated reusable, steam sterilizable detector probe tip, (4) a 12 mm collimated reusable, steam sterilizable reusable detector probe tip, and (5) a 19 mm noncollimated reusable, steam sterilizable detector probe tip. These modifications are intended to facilitate rapid and efficient probe interchangeability and to offer more options to the user.
Neoprobe has conducted laboratory studies to demonstrate the functional equivalence of the Model 1500 12 mm reusable, steam sterilizable non-collimated and collimated detector probe tips to the cleared Model 1000 GEN 1C 11 mm, EtO sterilizable detector probe. The tests performed on the cleared 11 mm and the new 12 mm collimated and non-collimated probe tips included measurement of the counting efficiency as a function of probe to source distance for $^{125}\mathrm{I}$, $^{111}\mathrm{In}$ and
DC 61852/2 0425053 02
{2}
99mTc, and measurement of the spacial resolution of the probe tip with and without collimation for the three nuclides.
The results obtained with the cleared 11 mm probe and the new 12 mm probe tips demonstrate that the two types of detector probes are functionally equivalent except for the change in count rate that is predicted from the change in dimensions of the crystal. In terms of counts per second per unit, area of detector surface were the same for both the 11 mm and the 12 mm detector probe tips. The collimated and non-collimated versions of the reusable, steam sterilizable detector probes can be used interchangeably with the Neoprobe® 1500 control unit with the same disposable handle, without adjustments.
Neoprobe has submitted in previous 510(k) notifications (including K942580 and K864263) literature in support of the indications for use for which the company is seeking clearance in this submission. Specifically, literature has been supplied to FDA on the use of various radiopharmaceutical compounds and detector probes for intraoperative identification of areas with increased radionuclide concentration to support the use of Neoprobe's device during ischemic bowel, bone (osteoid osteoma), thyroid and colon applications. In addition, this 510(k) notice includes literature that supports the use of Neoprobe's Model 1500 for detection of radioactivity in lymphatics.
\DC-61852/2-0425053.02
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Susan Tiedy-Stevenson, M.S.
Director, Regulatory Affairs
Neoprobe Corporation
425 Metro Place North, Suite 400
Dublin, OH 43017-1367
Re: K971167
Neoprobe® Model 1500 Portable
Radioisotope Detector and Accessories
Dated: March 28, 1997
Received: March 31, 1997
Regulatory Class: I
21 CFR 892.1320/Procode: 90 IZD
JUN 26 1997
Dear Ms. Tiedy-Stevenson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, or 594-4613 for Ear, Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Enclosure
{4}
510(k) Number (if known): K971167
Device Name: Neoprobe Model 1500 Portable Radioisotope Detector
Indications For Use:
The Neoprobe Model 1500 Portable Radioisotope Detector is an electronic device intended to detect and quantify gamma radiation. It is indicated for external and intraoperative detection of radioactivity in body tissues or organs, such as bowel, bone, lymphatics, and red blood cells, where radiopharmaceuticals are administered. For proper use and limitations of the Neoprobe Model 1500 refer to the product labeling.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymm
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K971167
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.