The HGF is a measuring instrument for preoperative and intraoperative detection of radioactive substances in the energy range between 140 keV and 360 keV (e.g. TC-99M and I-131). It is contraindicated in dosimetric applications, detection of radiation other than gamma and detection of radiopharmaceuticals with radiation energies outside the range of 140keV - 360 keV.
Device Story
Handheld Gamma Finder (HGF) is a nuclear uptake probe for detecting and quantifying gamma radiation (140-360 keV). Device integrates gamma-detector, energy discriminator, amplifier, and display into a single, conical, biocompatible polymer-shielded unit. Operated by physicians during surgery or clinical procedures; held in one hand. Input: gamma radiation from radiopharmaceuticals (e.g., TC-99M, I-131). Processing: internal electronics quantify radiation; output: numerical counts per second (CPS) on LCD, updated every second; optional acoustic/LED signals. Used with disposable sterile sleeves. Battery-powered (non-replaceable). Output assists surgeons in locating radioactive tissues (e.g., sentinel biopsies). Benefits include portability and ease of use for extended procedures.
Clinical Evidence
Bench testing only. Verification testing conducted by a university regarding system and technical evaluations. Electrical, biocompatibility, and sterilization standards testing performed.
Technological Characteristics
Nuclear uptake probe; conical form factor; biocompatible polymer casing. Sensing: gamma-detector, energy discriminator, amplifier. Energy source: internal non-replaceable lithium battery. Connectivity: standalone. Sterilization: compatible with commercial surface disinfectants; probe tip immersible. Control: micro-controller based.
Indications for Use
Indicated for intraoperative or extraoperative detection and quantification of gamma photons in the human body following administration of gamma-emitting radiopharmaceuticals.
Regulatory Classification
Identification
A nuclear uptake probe is a device intended to measure the amount of radionuclide taken up by a particular organ or body region. This generic type of device may include a single or multiple detector probe, signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K013751
Image /page/0/Picture/1 description: The image shows the logo for Silicon Instruments GmbH. The logo features the letters 'SI' in a stylized design, followed by the words 'SILICON INSTRUMENTS' in all caps. Below the logo is the text 'GmbH', which is a German abbreviation for a limited liability company.
Silicon Instruments GmbH Berlin FORESTATION OF SEDERED BEAU ATTERSTAGE OF CHORICAL e M a il : s i @ s i l i c o n - i n s t r u m e n t s . d e
Internet: http//www.silicon-instruments.de
Unternehmen der Silicon Sensor Gruppe
--------
-
Silicon Instruments GmbH Berlin ▪ Ostendstraße 1 ▬ D-12459 Berlin
# 510(k) SUMMARY of SAFETY and EFFECTIVENESS
FEB 0 6 2002
| A. General Information | | | |
|------------------------|----------------------|-----------------------------------------------|--|
| 1. | Submitter's Name: | Silicon Instruments, GmbH | |
| | 2. Address: | Ostendstrasse 1<br>Berlin, D-12459<br>Germany | |
| 3. | Telephone: | 49-3-06-39-92-360 | |
| 4. | Contact Person: | Dr. Thomas Goebel | |
| 5. | Date Prepared: | October 29, 2001 | |
| 6. | Registration Number: | FDA Form 2891 Submitted | |
### B. Device
| 1. | Name: | SI-Handheld Gamma Finder (HGF)® |
|----|----------------------|---------------------------------|
| 2. | Trade Name: | Handheld Gamma Finder |
| 3. | Common Name: | Nuclear Uptake Probe |
| 4. | Classification Name: | Probe, Uptake, Nuclear |
| 5. | Product Code: | IZD |
| 6. | Class: | I |
| 7. | Regulation Number: | 892.1320 |
Silicon Instruments GmbH Ostendstraße 1 = D-12459 Berlin Tel. +49 (0)30/53 01 58 70
Fax +49 (0)30/53 01 58 72
Amtsgericht AG Berlin-Charlottenburg
HRB 68092
Geschäftsführer: Dr. Bernd Kriegel
Dauterha Rank Rorlin
{1}------------------------------------------------
## C. Identification of Legally Marketed Devices
| 1. Name: | Neoprobe Model 1500 |
|------------------|---------------------|
| 2. K Number: | K971167 |
| 3. Date Cleared: | June 26, 1997 |
#### D. Description of the Device
The gamma-detector, energy discriminator, amplifier, controls and energy supply have been assembled into a single hand-held unit SI-Handheld Gamma Finder® (HGF), which is shielded by a biocompatible polymer. The probe's shape is conical; its shaft tapers off towards a tip that holds the collimator, the crystal and the PINdiode while a head of larger diameter houses the display, menu buttons, and other components. The device is held in one hand by the middle of the shaft. Outer diameters and weight are designed so as to allow single-handed use over periods of time longer than those of routine sentinel biopsies. The polymer casing of SI HGF is water resistant and thus allows extended cleansing with commercial surface disinfectants. The probe tip may even be immersed in disinfectants for several hours. This fulfills standard requirements of hospital hygiene. Disinfection of the probe is comparable to other, larger probes that are being used like Gamma Finder in conjunction with disposable sterile plastic covers.
The HGF can be switched on or off by any key. Results are numerically shown in the liquid crystal display in counts per second (CPS), and updated every second. The HGF has only four keys:
- E. Latest Valve Measured
Acoustic Signal ON/OFF in menu mode(+) F.
- LED ON/OFF in menu mode(-) G.
- Menu Mode ON/OFF H.
- Sub-menu dynamics lx
- Sub-menu volume (Six levels) 2x
- 3x Menu Mode off
The currently measured value can be retained in the display for 5 seconds by pressing Key A.
{2}------------------------------------------------
The HGF is packaged and transported in a small suitcase, see Appendix A for a photograph.
The HGF can only be operated by a Sonnenschein Lithium Battery (SL2770/T), with a maximum charging current of 15 mA. It is not user replaceable. The display of the HGF will display an battery symbol, warning the user the battery life is approximately 20 hours from exhaustion. If the battery is exhausted, the display will show "LobAt".
Battery Information is located in Appendix B.
The only accessory is a Sterile Sleeve (No. SI-FOL-101) that accompanies each device, along with the User Manual.
#### E. Intended Use Statement
The HGF is a measuring instrument for preoperative and intraoperative detection of radioactive substances in the energy range between 140 keV and 360 keV (e.g. TC-99M and I-131).
It is contraindicated in dosimetric applications, detection of radiation other than gamma and detection of radiopharmaceuticals with radiation energies outside the range of 140keV - 360 keV.
#### F. Technological Characteristics Summary
The HGF is substantially equivalent to the Neoprobe Model 1500, cleared on June 26, 1997 as K971167.
Each nuclear uptake probe and accessories is compact, operates on an internal battery; is software or micro-controller controlled, and intended to detect and quantify gamma radiation. Both the Neoprobe Model 1500 and HGF are indicated for external and intraoperative detection of radioactivity.
The HGF was tested to the appropriate electrical standards, biocompatibility, and sterilization standards. In addition, verification testing was conducted by a university as to both system and technical evaluations.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 0 6 2002
Dr. Thomas Goebel Head of Senior Technology Silicon Instruments GmbH OstendstraBe 1 D-12459, Berlin GERMANY
Re: K013751
Trade/Device Name: SI-Handheld Gamma Finder® Regulation Number: 21 CFR 892.1320 Regulation Name: Nuclear uptake probe Regulatory Class: I Product Code: 90 IZD Dated: October 29, 2001 Received: November 13, 2001
Dear Dr. Goebel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 510(k) Number: K013751
SI – Handheld Gamma Finder Device Name:
Indications for Use:
- The SI Handheld Gamma Finder is indicated to detect and quantify the gamma t photons in the body intra-operatively or extra-operatively, when gamma emitting radiopharmaceuticals are administered into the human body.
#### PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
OVER-THE-COUNTER USE _________________________________________________________________________________________________________________________________________________________ (optional Form 1-2-96)
David G. Lynam
and Rs 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.