ACUSON P500 Ultrasound System

K163396 · Siemens Medical Solutions USA, Inc. · IYN · Jan 4, 2017 · Radiology

Device Facts

Record IDK163396
Device NameACUSON P500 Ultrasound System
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeIYN · Radiology
Decision DateJan 4, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The ACUSON P500 ultrasound imaging system is intended for the following applications: Fetal, Abdominal (including liver), Pediatric, Small Parts, Transcranial, OB/GYN(useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Pelvic, Neonatal, Cardiac, IntraCardiac, Vascular(including Peripheral Vessel), Musculoskeletal, Superficial Musculoskeletal, and Urology applications. The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Society of Echocardiography Carotid Intima-Media Thickness Task Force. Endorsed by the Society for Vascular Imaging." The ACUSON Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients.

Device Story

The ACUSON P500 is a mobile, software-controlled diagnostic ultrasound system. It acquires harmonic ultrasound echo data via transducers and displays it in various modes (B-Mode, PW/CW Doppler, Color Doppler, Color M, Tissue Doppler, Power Doppler) on a flat-panel display. Operated by clinicians in clinical settings, it provides anatomical measurements and calculation packages to assist in diagnosis. The Arterial Health Package (AHP) measures Intima-Media Thickness (IMT) to communicate cardiovascular status to patients. The system also supports the Acunav catheter for intracardiac/intraluminal visualization. Output is used by physicians to inform clinical decision-making and patient communication. The device benefits patients by providing non-invasive diagnostic imaging and anatomical quantification.

Clinical Evidence

No clinical data was required or submitted. Substantial equivalence is based on technological characteristics, non-clinical bench testing, and the long history of safe and effective performance of diagnostic ultrasound systems.

Technological Characteristics

Mobile ultrasound system; 64-channel digital architecture; supports B, M, PW/CW Doppler, Color Doppler, Power Doppler, and harmonic imaging modes. Patient contact materials tested to ISO 10993-1. Complies with ANSI/AAMI ES 60601-1, IEC 60601-1-2, and AIUM/NEMA UD-2/UD-3 standards. Connectivity includes wireless and DICOM SR support.

Indications for Use

Indicated for fetal, abdominal, pediatric, small parts, transcranial, OB/GYN, pelvic, neonatal, cardiac, intracardiac, vascular, musculoskeletal, and urology diagnostic imaging in adult and pediatric patients.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 4, 2017 SIEMENS MEDICAL SOLUTIONS USA, INC. 0 Mr. MARK JOB OFFICIAL CORRESPONDENT 1394 25TH STREET, NW BUFFALO MN 55313 Re: K163396 Trade/Device Name: Acuson P500 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, OBJ Dated: December 1, 2016 Received: December 5, 2016 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D'Hara For Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | Form Approved: OMB No. | 0910-0120 | |------------------------|--------------------------| | Expiration Date: | January 31, 2017 | | | See PRA Statement below. | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration **Indications for Use** | 510(k) Number (if known) | K163396 | |--------------------------------|-------------------------------| | Device Name | ACUSON P500 Ultrasound System | | Indications for Use (Describe) | | The ACUSON P500 ultrasound imaging system is intended for the following applications: Fetal, Abdominal (including liver), Pediatric, Small Parts, Transcranial, OB/GYN(useful for visualization of the ovaries, follicles, uterus and other pelvic structures), Pelvic, Neonatal, Cardiac, IntraCardiac, Vascular(including Peripheral Vessel), Musculoskeletal, Superficial Musculoskeletal, and Urology applications. The system also provides the ability to measure anatomical structures fetal, abdominal, small organ, transrectal, transvaginal, cardiovascular, peripheral vessel, musculoskeletal (conventional), and musculoskeletal (superficial) and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Society of Echocardiography Carotid Intima-Media Thickness Task Force. Endorsed by the Society for Vascular Imaging." The ACUSON Acunav Ultrasound Catheter is intended for intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology, as well as visualization of other devices in the heart of adult and pediatric patients. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF in {3}------------------------------------------------ : (umoux ili) vədmiş (x) (xəstəsinin (xidmətindən (xidmətindən əlaqəsində (xalq və qalında qalınmışdır. Bu mən ### ​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​ | sinə aid bitki növü. İstinadlar Respublikası Şaman Şaman Şamanın Şaman Şaman Şaman Şaman Şaman Şaman Şaman Şamaları Şaman Şaman Şamaları Şaman Şamaları<br>tosibleට-ຍາງ<br>(sinfinə aid bitki növü. İstinadlar Respublikasının filmləri Respublikasının filmləri Sayanın Filmin Filmin Filmin Filmin Filmin Filmin Filmin Filmin Film<br>səs və birlərinin mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənas<br>P<br>N<br>N<br>P<br>P<br>N<br>N<br>N<br>P<br>N | | | อยู่ในปี (วิธีที่ตัว | (วยเขตเธอร์เกมส์) (วัยเขายา) (วัยเขายา) (วัยเขาเธอ | isibles<br>ากย่องๆ วัยเขาย<br>N<br>N<br>N<br>N<br>N | lubA วิธเมริ<br>P<br>P<br>P<br>P<br>P | (Misədə) | ອງກວຣຍາ- ຂາມ | əxs-olunarın<br>P<br>P<br>P<br>P | əxs-olunarların əvvəl əmələ qalınmışdır. Bu məşğul olunmuşdur. Bu məşğul olunmuşdur. Bu mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənas<br>P<br>P<br>P<br>P | 10089-2001 | ərin tərəfindən sinfinin fəsiləsinin cinsinə aid bitki növü. İstinadlar Respublikasının Filmin Film Alanmada Şamanın Şamanın Şamanın Şamanın Şamarın Şamanın | euifen-sue.<br>P<br>P<br>P<br>P | ยุวอน-sue<br>P<br>P<br>P<br>P | əli qalında qalında bir mənasında qalınmışdır. Bu mənist<br>N | larında qalınmışdır. Bu da bir mənist<br>N | ranipemi<br>(1 agon) uebuolleuns<br>P<br>P<br>P<br>P | late a<br>ərin fəsiləsinə cinsinə aid bitki növü. İstinadlar Respublikasının fəsiləsinə cinsinə aid bitki növü. İstinadlar Alanmada Şəhər Qaran mənist Qaran Qaran Qaran | oidoosousda | əsinə aid bitki növü. İstinadlar | əviləsinə qo-sun (ölətov | euimopq<br>P<br>P<br>P<br>P<br>P | lıqla qalan mənasını qalınmışdır. Bu mənist və qalınmışdır. Bu mənist və bir mənist mənist mənist mənist mənist mənist mənasının mənist mənist mənist mənist mənasının mənası<br>P<br>P<br>P<br>P<br>P | նայքվեկն<br>քյալներիկն | Other Track1 Only)<br>ാനിരാമര കഥ ജോലിT<br>B<br>M<br>awa<br>amo<br>alddoc | oitesildqA ไมวิเทพีเว<br>and the of Open of the store | əmin təqvimlərinə aid bitki növü. İstinadlar Rəsmin Rəssamlar Rəsmin Filmin Filmin Filmin Film Sənədlər Rəsmin Filmin Film Sənədli Film Sənədlər Film<br>າໄປ ສະ ປຸ່ນປຸ ແຍກປ່າ ອປີ ໄດ້ ຄ່າ ເປັນ ອ້າງອີງ ສາມາດອະສາລັບ ພົມບອສ MOSU | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------------------|----------------------|----------------------------------------------------|-----------------------------------------------------|---------------------------------------|----------|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|-------------------------------|---------------------------------------------------------------|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|----------------------------------|--------------------------|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | P | N | | | N | P | | | P | P | | | P | P | | | P | | | | | P | P | | əlddən sahəsində aid bitki növü. İstinadlar Rəmaniya Rəmaniya Rəmaniya Rəmaniya Rəmaniya Rəmaniya Rəsmin Filmin Filmin Filmin Filmin Filmin Filmin Filmi | | | | | одиа | одма | | | одив | одма | | | одма | одма | | | одиц | одиц | | | odwa | | | | | осма | осма | | (Alisədə) | | | | | 'ZI, 'UL '6, 'A, 9, 'III, (2,<br>Eレ<br>の | ore 2, 4, 10, 10, 12, 14,<br>ู้จัก<br>01 | | | state 2, 4, 6, 7, 00, 12, | ote 2, 4, 0, 7, 8,<br>の<br>こし<br>ರಿಗ | | | otes 2, 8, 8, 9,<br>, 2017, 2017, 10:22, 10, 10, 2017, 10:20, 10, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017, 10, 2017<br>દા | in filmin fəsiləsinə cinsinə aid bitki növü. İstinadlar Azərbaycan Azərbaycan Azərbaycan Qarşıq Qarşıq Qarşıq Qarşıq Qarşıq Qarşıq Qarşıq Qarşıq Qarşıq Qarş<br>9,<br>、「いい<br>દા | | | 'µr'Þ'£ 'Z eig | ote 2, 3, 4, 11, 12, | | | də qalınmışdır. Bu mənist və bir mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının mənasının<br>ャ<br>'乙レ'レレ '6 ' | | | | | 2, 2, 8, 4, 11, | ������������������������������������������������������������������������������������������������������������������������������������������������������������������������������<br>ู้อ้า | | riseds ( | | | | | | | | | | ətən ərazi və bir və qalında çəkilməsinə aid bitki növü. İstinadlar Azərbaycan Respublikasının fəsiləsinin fəsiləsinin cinsinə aid bitki növü. İstinadlar Azərbaycan<br>ת > | n qalan və qalında və qalında və qalında və qalında və qalında və qalında və qalınmışdır. Bu və qalınmış və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və qalınmışdır. | consisted on the status of the status of the status of the status of the station of the station of the station of the station of the station of the station of the station of | ə ələnin sənələrin əsasən ərazi və bir və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və mərkəzi və | ətən ərazi və qalında və qalınmışdır. Bu mənist və qalınmışdır. Bu mənist və bir və qalınmışdır. Bu mənist və bir mənist və bir mənist və bir mənist və bir mənist və qalınmı | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------| | | | | | | | əsinə aid bitki növü. İstinadlar Şəhriman Şəhrimanın Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Qaraqların | ə Sənlər Sənəyən Şərqlər (Azərbaycan filmləri (Azərbaycan filmləri) və Azərbaycan (Azərbaycan filmləri (Azərbaycan filmləri (Azərbaycan filmləri (Azərbaycan filmləri (Azərba | vanced SieClea | | ə şərərsən qalınmışdır. | ക്കുടെ ഗ്രട്രാസ്ത്രാമ = 9 : വാദ്യഭാഗ്മാവ് സമ്പ | | ətələrin əsasən bir və bir və qalında və bir və qalında və bir və qalında və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və bir və bir və bir və bir və bir və qalınmışd | ətələrin əsasən ərazi vardır. Bu mərkəzi və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir<br>.5 | ətələrin əsasən ərazi və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və b<br>1 | ətələrin əsasən bir və bir və qalınmışdır. Bu mərkəzi və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və qalınmışdır. Bu mə<br>દા | ətələrin əsasən bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir mənist və bir mənist | ətələrin əsasən ərazi vahidində kənd. Bu mərkəzi və bir mənist və bir mənist<br>ー | ilə qalınmışdır. Bu mərkəzi və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist | ə ətəl | ətələrin əsasən ərazi və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və b | ətəli qalında çəkilməsinə aid bitki növü. İstinadlar Respublikasının fəsiləsinə filmlərin fəsiləsinə cinsinə aid bitki növü. İstinadlar Azərbaycan Azərbaycan Filmi | əşən ərazi vardır. Bu mərkəzi və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist<br>6 | | | rodde Save | COSOUND Communica | | ilər | | ്യജാടങ്ങി (മ്വന്ത്രണി | ra | ЯНА | ടാട്ടല്പടക | ənissəslərin məşğul olunmuşdur. Bu mənist | odo Eresson | | (a3d33n il 30a9 gahtona no aunit1000-an11 siht woj38 atti9W tow od 32a319) ALAMEDA COUNTY DISTRICT ATTORNEY NANCY E. O'MALLEY DISTRICT ATTORNEY (ອາມາຣունչ) (HAQO) ຕັ້ງເອກ ໂຣວເຄດໄທນັກ ເຊິ່ງເທດ ການປະກວດນາງງາມ ການປະກວດນາງງາມ ອວຖອງພວກເວທີ ຂາພາດ​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​ (x)ove (ອານາຄຸນນິຣ) (HAQO) ຕໍ່ປຸເຈຍຕ ໂຣວເຄີຍໄດ້ເນືອນທີ່ 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 10, 1 {4}------------------------------------------------ ิ(ฟ)ุ่มมรร (ອາມາຣունչ) (НЯСС) ຕຳແໜ່ ແລະເປັນອຸປິດເປັນສົມ ໂດຍ ຂອງທອງ 10 ກອງກອງການວາດວົ (ອາກາຣແນ້ຣ) (HAQO) ຕຳໂຣອH ໂຂວ່ງໂດຣປວເທຣສ ອນທອງ 10 ກອນອງ 10 ອວບອາການວາດວ (a3a3an al asaa gahtona no aunitinoo-aniji siht wojaa atiraw ton oa asaajin | 2005) Xarici keçidlər Respublikası Respublikası Respublikası<br>(Alian Azərbaycan Respublikasının Azərbaycan Respublikasının Azərbaycan Respublikasının Azərbaycan Respublikasının Azərbaycan Respublikasının Azərbaycan Respublikasının Azər | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | l ətən<br>ərbaycan bioxidir. Sənəd şəhəri şəhərin qalınmışdır. Vilan<br>g ətəl<br>teress seen o | | n bir və qalında və bir və qalında və bir və qalınmışdır. Bu və bir və bir və bir və bir və bir və bir və bir və bir və mərkəzi və bir və mərkəzi və mərkəzi və mərkəzi və mə<br>ətələrin əsasən ərazi vahidində kənd. Bu mənist və bir mənist<br>ənissəsMəişə | | ç ətəli<br>ətələrin əsasən bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir<br>າຣວຣິອເຣສ | | n ətənin əsasən və bir və qalında və bir və qalında və bir və qalında və bir və qalında və bir və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və bir və bir və bir və bi<br>ə qalınmış və qalınmışdır. Bu məşğul olunmuşdur. Bu mənşənin baş və bir və bir mənist<br>də qalında və bir və qalında və bir və qalında və bir və qalınmışdır. Bu mərkəzi və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir v<br>ЧНА | | Notote 5<br>ə mərkəzi və qalınmışdır. İstinadlar Şəhriman Şəhrimanın Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin Şəhrin<br>ətələr Bakı Şamar Şamar Şamar Şamar Şamar Şamar Şamar Şamar Şamar Şamalar Şamar Şamalar Şamalar Şamalar Şamalar Şamalar Şamalar Şamalar<br>ITC | | ətələr Bakı Şamar Şamar Şamar Şamar Şamar Şamar Şamar Şamar Şamalar Şamar Şamalar Şamalar Şamalar Şamalar Şamalar Şamalar Şamalar Şamal<br>ഓട്ടേഴ്ട) ഇവിടുടെയി വ്യാപ്പെടു | | ətələrin əsasən bir və qalınmışdır. Bu mənist<br>Clarity VE | | ətələrin əsasən ərazi vahidində kənd. Bu mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist və bir mənist<br>uotezilensiy əlbəə | | ətələrin əsasən bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir<br>レレ | | ətələrin əsasən bir və qalınmışdır. Bu mənist və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və qalınmışdır. Bu mər<br>ുന്നുണ്ടുന്നു (ヨコ) | | ətən ələn bir və qalında və bir və qalında və qalında və qalında və qalınmışdır. Bu və bir və qalınmışdır. Bu və bir və bir və bir və bir və bir və bir və bir və bir və bir<br>rəvan ədədir. | | əsində aid bitki növü. İstinadlar Azərbaycan Respublikasının fəsiləsinin fəsiləsinin fəsiləsinin fəsiləsinin cinsinə aid bitki növü. İstinadlar Azərbaycan Azərbayc | | | | | | | wollot se ypod nemny ay to sisvleur woll binit 10 puipemi oitsourge | | | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---|---|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|---|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | oriesilqqA lвзініЮ | | | | and the sports of the provinsition of the provinsition of the provinsition of the provinsition of the provinsition of the provinsition of the provinsition of the provinsition | | | | | | | Other Track1 Only | ົວຫຼາຍວັນອາໄປສອງປີ ເປັນສາມາ | B | M | awa | ന്നും അവ | | i əsasən fəsiləsinə cinsinə aid bitki növü. İstinadlar Rəmaniya Rəsmin Filmin Filmin Filmin Filmin Filmin Filmin Filmin Filmin Filmin Film | (Ayısədə) (Ayısadə) (Ağıqanın (Azərbaycan Respublikası) (Azərbaycan (Azərbaycan Respublikası) (Azərbaycan (Azərbaycan Respublikası) (Azərbaycan (Azərbaycan Respublikası) (A | vised o | | തിയിലും പ്രാമ്യൂർ വ | വന്തിന്റെ വിശ്വാസ്ത്രിക്കുന്നു. അവലംബം കേരള സ്വാതന്ത്രിക്കുന്നു | | | | | | | | | | | lətə qalınmışdır. Bu mənist və bir mənist və bir mənist mənist mənist mənist mənist mənasını qalınmışdır. Bu mənist mənist mənist mənist mənisti mənasının mənasının mənasını | P | P | P | P | P | P | одма | ������������������������������������������������������������������������������������������������������������������������������������������������������������������������������ | | | suimobia | P | ਹ | P | P | P | P | одиа | lote 2, 4, 72, 16, | | | əvizsi və qalında və qalında və qalında və qalında qalınmışdır. Bu mərkəzi və qalınmışdır. Bu mərkəzi və qalınmışdır. Bu mərkəzi və qalınmışdır. Bu mərkəzi və qalınmışdır. B | | | | | | | | | | | əvillər | | | | | | | | | | | idossonsda | | | | | | | | | | lsteally | ərin fəsiləsinə cinsinə aid bitki növü. İstinadlar Respublikasının fəsiləsinə cinsinə aid bitki növü. İstinadlar Alanmada Şamanın Qaran mənasının Qaran mənist | | | | | | | | | | ənin cinsinə aid bitki növü. İstinadlar Şərqi Qurğunu Şəxsi Şəhrin Qaranın Qaranın Qaran Qaran Qaran Qaran Qaran Qaran Qaran Qaran Qaran Qaran Qaran Qara | (1 agon) | | | | | | | | | | | ərin fəsiləsinə cinsinə aid bitki növü. İstinadlar Respublikasının fəsiləsinə filmlərin fəsiləsinə filmlərin fəsiləsinə filmlərin fəsiləsinə filmlərin fəsiləsin | | | | | | | | | | | ərin müharibəsi və qalınması və qalınması və qalınmışdır. Bu mənist | | | | | | | | | | | letoer-suer | | | | | | | | | | | leuipev-sueu | | | | | | | | | | | ın salınmışdır. Bu yarımları və bir sahil olunması və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və bir və mərkəzi və bir və mərkəzi və bir və | | | | | | | | | | | (വാലോ-ബേബ) പ്രശസ്തം (പാലം-ചെയ് | | | | | | | | | | | อฟร-อนทวรทุก<br>("quevuo") | | | | | | | | | | | əyə (olmadına) və qalınmışdır. Bu məşğulları və qalınmışdır. Bu məşğulları və qalınmışdır. Bu məşğulları və qalınmışdır. Bu məşğulları və başların mərkəzi və başların mərkəz | | | | | | | | | | | າຣໄນວຣຣັv-ຣາມຸນ | | | | | | | | | | | (Alisədə) | | | | | | | | | | | ป็นปริสต์ศตวรรษที่มีความคริสต์ (Anglassical) (Anglassical Property) (Anglassical Property) (Anglassical Property) (Anglassical Property) (Anglassical (Anglession) (Angless ( | P | P | P | P | P | P | одив | ote 2, 4, 6, 7, 8, 10, 12,<br>ಾ | | ട്രിമ്പരാ | intsibe9 วิธเขาธิ | | | | | | | | | | | ายไมวยอง-ธานา (วัยเขายว) | | | | | | | | | | | ട്ടാട്ടന്‍റ്റ് (ാങ്ങ്ങളുടെ (വൈ) | | | | | | | | | | | ວຣibrຣວ-ຣາມຸນ | | | | | | | | | | | rin fəs…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...