ACUSON Maple Diagnostic Ultrasound System, ACUSON Maple Select Diagnostic Ultrasound System
K232617 · Siemens Medical Solutions USA, Inc. · IYN · Sep 26, 2023 · Radiology
Device Facts
Record ID
K232617
Device Name
ACUSON Maple Diagnostic Ultrasound System, ACUSON Maple Select Diagnostic Ultrasound System
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Sep 26, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The ACUSON Maple ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, and Musculoskeletal applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
Device Story
Mobile, software-controlled diagnostic ultrasound system; acquires harmonic ultrasound echo data; displays images in B-Mode, M-Mode, PW/CW Doppler, Color Doppler, Tissue Imaging, and 3D/4D modes. Operated by trained healthcare professionals in clinical settings. System includes Arterial Health Package (AHP) for measuring Intima Media Thickness (IMT) and referencing normative tables. Output displayed on flat panel; used by physicians for clinical diagnosis and patient communication regarding cardiovascular health. Benefits include non-invasive anatomical visualization and quantitative analysis for diagnostic decision-making.
Clinical Evidence
No clinical data required; substantial equivalence supported by non-clinical bench testing, including acoustic output, biocompatibility, cleaning/disinfection, and electrical/mechanical safety testing per recognized standards (IEC 60601 series, ISO 10993-1).
Technological Characteristics
Mobile ultrasound system; harmonic imaging; multiple modes (2D, Doppler, M-mode, 3D/4D). Transducers: C5-2v, 14La/14L4a, 5C1a. Connectivity: On-screen display of thermal/mechanical indices. Safety standards: IEC 62359, AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-1, ISO 14971. Software-controlled.
Indications for Use
Indicated for use by trained healthcare professionals in clinical settings for diagnostic ultrasound imaging (Abdominal, OB/GYN, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, Musculoskeletal). Includes Arterial Health Package for Intima Media Thickness measurement.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
September 26, 2023
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions, USA, Inc. % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114
## Re: K232617
Trade/Device Name: ACUSON Maple Diagnostic Ultrasound System ACUSON Maple Select Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX, OIJ Dated: August 28, 2023 Received: August 28, 2023
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Marjan Nabili - S for
Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K232617
Device Name ACUSON Maple Diagnostic Ultrasound System, ACUSON Maple Select Diagnostic Ultrasound System
#### Indications for Use (Describe) For ACUSON Maple Diagnostic Ultrasound System
The ACUSON Maple ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, and Musculoskeletal applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
For ACUSON Maple Select Diagnostic Ultrasound System
The ACUSON Maple Select ultrasound imaging system is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediat, Urology, Echocardiography and Musculoskeletal applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
Operating Modes 2D-mode - 2D-mode with Harmonic imaging Color flow Doppler - Color (velocity) - Power (energy) Pulsed Wave Doppler - Pulsed Wave Doppler Tissue Imaging - High Pulsed Repetition Frequency Pulsed Wave Doppler Continuous Wave Doppler - Steerable Continuous Wave Doppler for phased array transducers - Auxiliary Continuous Wave Doppler for pencil transducers M-mode - M-mode with Harmonic imaging - Anatomical M-mode
FORM FDA 3881 (6/20)
{3}------------------------------------------------
#### 3D/4D Volume Imaging
Combined Modes 2D-mode with Color 2D-mode with Doppler 2D-mode with Color and Doppler 2D-mode with M-mode 2D-mode with M-mode and Color 2D-mode with Elastography 2D-mode with Contrast Agent Imaging 2D/Anatomical M-mode
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# K232617
| Date: | July 05, 2023 | | | |
|---------------------|------------------------------------------------------------------------------------------------------------------|-----------|--------|--|
| 1. Sponsor: | Siemens Medical Solutions USA, Inc.<br>Ultrasound Division<br>22010 South East 51st Street<br>Issaquah, WA 98029 | | | |
| Contact Person: | Shilpa Rapaka<br>Phone: 512-913-1053 | | | |
| Secondary Contact: | JungIn Choi<br>Phone: +82 10 4695 8694 | | | |
| 2. Device Name(s): | ACUSON Maple Diagnostic Ultrasound System<br>ACUSON Maple Select Diagnostic Ultrasound System | | | |
| Common Name: | Diagnostic Ultrasound System with Accessories | | | |
| Classification: | Regulatory Class: | Class II | | |
| | Classification Panel: | Radiology | | |
| | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYM | |
| | Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO | |
| | Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX | |
| | Biopsy needle guide Kit | 892.1560 | 90-OI | |
| Manufacturing Site: | Siemens Healthcare s.r.o<br>Panattoni Park Kosice Airport ul.<br>Andreja Kyasa 5040 17, Kosice-Barca, Slovakia | | | |
{5}------------------------------------------------
#### 3. Legally Marketed Predicate Devices
The ACUSON Maple Diagnostic Ultrasound System and ACUSON Maple Select Diagnostic Ultrasound System are multi-purpose diagnostic ultrasound systems with proprietary software and optional accessories and are substantially equivalent to the company's own Ultrasound devices.
- -Primary Predicate Device: ACUSON Juniper and Juniper Select (K230207)
- Reference Devices: ACUSON NX2 and NX2 Elite (K173981), ACUSON SC2000 (K211726) -
#### 4. Device Description
The ACUSON Maple Diagnostic Ultrasound System and ACUSON Maple Select Diagnostic Ultrasound System are multi-purpose, mobile, software-controlled, diagnostic ultrasound systems with an onscreen display for thermal and mechanical indices related to potential bio-effect mechanisms. Their function is to acquire harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Color M mode, Doppler Tissue Image mode, Amplitude Doppler Mode, combination modes, Harmonic Imaging and 3D Imaging modes, or Harmonic Imaging and 4D imaging modes on a flat panel display for diagnostic ultrasound imaging.
### 5. Intended Use/Indications for Use
#### ACUSON Maple
The ACUSON Maple Ultrasound Diagnostic Ultrasound System is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Neonatal, Vascular, Urology, Echocardiography, and Musculoskeletal applications using different ultrasound transducers for different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
#### ACUSON Maple Select
The ACUSON Maple Select Diagnostic Ultrasound System is intended to provide images of, or signals from, inside the body by an appropriately trained healthcare professional in a clinical setting for the following applications: Abdominal, Obstetrics, Gynecology, Small Parts, Pediatric, Vascular, Urology, Echocardiography and Musculoskeletal applications using different ultrasound transducers for
{6}------------------------------------------------
different applications. The system also provides the ability to measure anatomical structures and provides analysis packages that provide information used by a physician for clinical diagnostic purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding the state of their cardiovascular system.
### Operating Modes
- . 2D-mode
- । 2D-mode with Harmonic imaging
- Color flow Doppler
- Color (velocity) l
- । Power (energy)
- Pulsed Wave Doppler
- l Pulsed Wave Doppler Tissue Imaging
- । High Pulsed Repetition Frequency Pulsed Wave Doppler
- Continuous Wave Doppler
- Steerable Continuous Wave Doppler for phased array transducers l
- । Auxiliary Continuous Wave Doppler for pencil transducers
- . M-mode
- l M-mode with Harmonic imaging
- Anatomical M-mode l
- 3D/4D Volume Imaging
Combined Modes
- 2D-mode with Color
- 2D-mode with Doppler
- 2D-mode with Color and Doppler
- 2D-mode with M-mode
- 2D-mode with M-mode and Color
- 2D-mode with Elastography
- 2D-mode with Contrast Agent Imaging
- 2D/Anatomical M-mode
#### 6. Substantially Equivalent Devices and Summary of Technological Characteristics
The ACUSON Maple Diagnostic Ultrasound System and ACUSON Maple Select Diagnostic Ultrasound System are substantially equivalent to the company's own Ultrasound devices that are already cleared for distribution in the United States. The predicate and reference devices of ACUSON Maple
{7}------------------------------------------------
and ACUSON Maple Select are ACUSON Juniper Select (K230207), ACUSON NX2 and NX2 Elite (K173981) and ACUSON SC2000 (K211726) with regards to intended use, indications for use, technological characteristics (transducers, accessories, software features), safety, and effectiveness.
The submission device differs from the predicate/reference devices as follows:
#### 1) ACUSON Maple
- The addition of the C5-2v transducer (new part number) which is substantially equivalent to the C5-2v transducer previously cleared on ACUSON NX2 and ACUSON NX2 Elite (K173981).
- The addition of the 14La transducer which is substantially equivalent to the 14L4 transducer previously cleared on ACUSON Juniper and ACUSON Juniper Select (K230207).
- . The addition of the 5C1a transducer, which is substantially equivalent to the 5C1 transducer previously cleared on ACUSON Juniper and ACUSON Juniper Select (K230207).
- . The addition of needle guide kit 14La which is substantially equivalent to the needle guide kit 14L4 previously cleared on ACUSON Juniper and Juniper Select (K230207).
- . The addition of needle guide kit 5C1a which is substantially equivalent to the needle guide kit 5C1 previously cleared on ACUSON Juniper and Juniper Select (K230207).
- . The addition of needle guide kit C5-2v which is substantially equivalent to the needle guide kit C5-2v previously cleared on ACUSON NX2 and NX2 Elite (K173981).
#### 2) ACUSON Maple Select
- The addition of the C5-2v transducer (new part number) which is substantially equivalent to the C5-2v transducer previously cleared on ACUSON NX2 and ACUSON NX2 Elite (K173981).
- . The addition of the 14L4a transducer which is substantially equivalent to the 14L4 transducer previously cleared on ACUSON Juniper and ACUSON Juniper Select (K230207).
- The addition of needle guide kit 1414a which is substantially equivalent to the needle guide kit 14L4 previously cleared on ACUSON Juniper Select (K230207).
- . The addition of needle guide kit C5-2v which is substantially equivalent to the needle guide kit C5-2v previously cleared on ACUSON NX2 and NX2 Elite (K173981).
#### 7. Nonclinical Data
The proposed devices have been evaluated for acoustic output, biocompatibility, cleaning, disinfection, effectiveness as well as thermal, electrical, electromagnetic and mechanical safety and have been found to conform to applicable medical device safety standards. The systems comply with the following voluntary standards:
- . IEC 62359:2010 /A1(2017), Ultrasonic-Field characterization- Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic field / This document and its separate amendments continue to be valid together with the consolidation version.
{8}------------------------------------------------
- AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) and AMD2: 2021 Medical electrical equipment- Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, AMD2: 2021)
- IEC 60601-1:2005/A1(2012)/A2(2020) (Ed. 3.2), Medical electric equipment- Part 1: General requirements for basic safety and essential performance / This document and its separate amendments continue to be valid together with the consolidated version
- IEC 60601-1-2: 2014/A1(2020), Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral Standard: Electromagnetic disturbances- Requirements and tests
- . IEC 60601-2-18 Edition 3.0 2009-08, Medical electrical equipment- Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment
- . IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment- Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- . ISO 10993-1 Fifth edition 2018-08, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process
- . IEC 60601-1-6:2010+A1:2013+A2:2020 Medical Electrical Equipment Part 1-6, General Requirements for Basic Safety and Essential Performance- Collateral standard: Usability
- ANSI AAMI ISO 14971: Medical devices-Applications of risk management to medical devices, 2019
## 8. Clinical Data
The proposed ACUSON Maple and ACUSON Maple Select did not require any clinical studies to support substantial equivalence.
#### 9. Summary
The subject, predicate, reference devices have same indications for use, technological characteristics (transducers, accessories, software features), safety, and effectiveness. The non-clinical data support the safety of the device and demonstrate that the ACUSON Maple and ACUSON Maple Select perform as intended in the specified use conditions. Therefore, it is the opinion of Siemens Medical Solutions USA, Inc. that the ACUSON Maple and ACUSON Maple Select are as safe and effective with substantially equivalent performance as the predicate/reference devices.