K203504 · Siemens Medical Solutions USA, Inc. · OWB · Dec 22, 2020 · Radiology
Device Facts
Record ID
K203504
Device Name
Cios Flow
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
Dec 22, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Cios Flow is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
Device Story
Mobile fluoroscopic C-arm X-ray system for surgical environments; supports orthopedic and vascular procedures. System comprises C-arm (X-ray source and flat panel detector) and mobile display trolley. Operator (physician/technician) positions C-arm; system acquires X-ray images; processes and displays images on trolley. Features include target pointer, digital cine mode, touch-controlled collimation/brightness/zoom, and dose regulation. Used in OR/clinic; provides real-time visualization to guide interventional procedures. Benefits include improved procedural guidance and diagnostic imaging capabilities.
Clinical Evidence
Bench testing only. Compliance with electrical safety (AAMI ANSI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), and performance standards (IEC 60601-2-43, 60601-2-54) verified. Human factors usability validation performed.
Technological Characteristics
Mobile C-arm fluoroscopic X-ray system. Components: X-ray tube, CMOS flat panel detector (30x30cm or 20x20cm), digital imaging system, collimator, generator. Connectivity: Networked trolley. Software: Windows 10. Features: Wireless foot switch, green laser light localizer, anti-microbial coating. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 62366-1.
Indications for Use
Indicated for X-ray imaging of anatomical structures in patients, including pediatric patients, undergoing interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 22, 2020
Siemens Medical Solutions USA, Inc. % Cordell Fields, Sr. Regulatory Affairs Specialist 40 Liberty Boulevard 65-1A MALVERN PA 19355
Re: K203504
Trade/Device Name: Cios Flow Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, OXO, JAA Dated: November 25, 2020 Received: November 30, 2020
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR
{1}------------------------------------------------
Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-----------------------------------------|----------------------------------|
| Food and Drug Administration | Expiration Date: 06/30/2020 |
| Indications for Use | See PRA Statement on last page. |
| 510(k) Number ( <i>if known</i> ) | K203504 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Device Name | |
| Cios Flow | |
| Indications for Use ( <i>Describe</i> ) | |
| The Cios Flow is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients. | |
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|-----------------------------------------------------------------------------------------------------|------------------------------------|
| <span style="text-align: left;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span><br>Subpart C) | □ Over-The-Counter Use (21 CFR 801 |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. ***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 |
|----------------------|------------------------------------------|
| | PSC Publishing Services (301)443-6740 EF |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
## 510(k) Summary: Cios Flow
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355
#### Date Prepared: November 25, 2020
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
- General Information: 1. lmporter / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
Manufacturing Site: Siemens Healthcare GmbH Röntgenstrasse 19 - 21 95478 Kemnath, Germany Establishment Registration Number: 3002466018
#### 2. Contact Person:
Mr. Cordell Fields Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 306-3167 Email: cordell.fields@siemens-Healthineers.com
#### 3. Device Name and Classification:
Trade Name: Classification Name: Classification Panel: Requlation Number: Device Class: Product Codes: Subsequent Product Code:
Cios Flow Image-intensified fluoroscopic x-ray System Radiology 21 CFR §892.1650 Class II OWB, OXO JAA
#### Legally Marketed Primary Predicate Device 4.
| Trade Name: | Cios Fusion |
|----------------------|---------------------------------------------|
| 510(k) Clearance | K153244 |
| Clearance Date | March 07, 2016 |
| Classification Name: | Image-intensified fluoroscopic x-ray System |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the word "Healthineers" is a graphic of orange dots.
Classification Panel: Requlation Number: Device Class: Product Codes: Subsequent Product Code: Total Product Life Cycle:
Radiology 21 CFR §892.1650 Class II OWB, OXO JAA
All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable issues.
## Legally Marketed Secondary Predicate Device
| Trade Name: | Cios Alpha |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance | K181560 |
| Clearance Date | October 24, 2018 |
| Classification Name: | Image-intensified fluoroscopic x-ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR §892.1650 |
| Device Class: | Class II |
| Product Codes: | OWB, OXO |
| Subsequent Product Codes: | JAA |
| Total Product Life Cycle: | All product Recall incidents are considered during the<br>Design Input phase of development to ensure the latest<br>models will not be affected by any of the applicable |
Legally Marketed Secondary Predicate Device Trade Name: Cios Spin 510(k) Clearance K181550 Clearance Date October 30, 2018 Image-intensified fluoroscopic x-ray System Classification Name: Classification Panel: Radiology Regulation Number: 21 CFR §892.1650 Device Class: Class II Product Code: OWB, OXO Subsequent Product Codes: JAA Total Product Life Cycle: All product Recall incidents are considered during the Design Input phase of development to ensure the latest models will not be affected by any of the applicable
issue.
issues.
#### 5. Device Description:
The Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System is designed for the surgical environment. The Cios Flow provides comprehensive image acquisition modes to support orthopedic and vascular procedures. The system consists of two major components:
- a) The C-arm with X-ray source on one side and the flat panel detector on the opposite side. The c-arm can be angulated in both planes and be lifted vertically, shifted to the side and move forward/backward by an operator.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
- b) The second unit is the image display station with a moveable trolley for t processing and storage system, image display and documentation. Both units are connected to each other with a cable.
The main unit is connected to the main power outlet and the trolley is connected to a data network.
The following modifications were made to the predicate device the Cios Fusion. Siemens Medical Solutions USA, Inc. submits this Special 510(k) to request clearance for the Subject Device Cios Flow (VA30) for the following device modifications made to the Predicate Device (Cios Fusion (VA20).
This 510(k) submission, Subject Device "Cios Flow" with software version VA30" will support the following categories of modifications made to the Subject Device in comparison to Predicate Devices:
- 1) Modified Software:
- Table 1: Overview of Software Modifications supported by Software Version VA30 ●
- 2) Modified Hardware/Software:
- Table 2: Overview of Hardware Modifications supported by Software Version VA30 ●
- 3) Modified Hardware changes
- . Table 3: Overview of Hardware Modifications
## Table 1. Software Modifications for Cios Flow
| Software changes specific to New System Software VA30 | |
|-------------------------------------------------------|-------------------------------------------------------------|
| Device Software Modification | |
| 1. | System Software VA30 software modifications/features |
| | A. Target Pointer |
| | B. Digital Cine Mode (DCM) |
| | C. Cios OpenApps |
| 2. | Interactive User Touch Control (Software component of VA30) |
| | A. Collimation Controls |
| | B. Brightness Contrast Controls |
| | C. Rotate and Flip Controls |
| | D. Zoom and Pan Controls |
| | E. Spot Adapt |
| 3. | Dose Regulation Indicator |
| 4. | New Sound Radiation Delay |
| 5. | New Product Software Security |
| 6. | Upgrade to Windows 10 Operating System |
#### Table 2: Hardware/Software Modifications for Cios Flow
| Software/Hardware changes specific to New System Software VA30 | | |
|----------------------------------------------------------------|---------------------------------------------------------------|--|
| Device Hardware/Software Modifications | | |
| 7. | System Software VA30 software/hardware modifications/features | |
| | A. New CMOS Flat Panel Detector | |
| | B. Wireless Foot Switch | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.
#### Table 3. Hardware Modifications
| Device Hardware Modifications | | |
|-------------------------------|---------------------------------------------|--|
| 8. | Optional Laser Light Localizer green | |
| 9. | Optional Cart for Remote Control Unit | |
| 10. | Anti-microbial Coating on C-Arm and Trolley | |
| Other Device Modifications | | |
| 11. | Update 510(k) Information | |
## 6. Indications for Use:
The Cios Flow is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
## 7. Substantial Equivalence:
The Cios Flow (VA30) system is substantial equivalent to the legally marketed predicates listed in Table 3 below:
| Predicate Device Name and<br>Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties |
|---------------------------------------------|------------------|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Predicate<br>Siemens' Cios Fusion | K153244 | 03/07/2016 | ● Indications for use<br>● System for Image Acquisition<br>● Post-processing Software<br>● Examination Settings<br>● Cyber Security Information |
| Secondary Predicates<br>Siemens' Cios Alpha | K181560 | 10/24/2018 | ● CMOS Flat Panel Detector<br>- Large Detector:<br>Xineos-30 cm x 30 cm<br>- Small detector<br>Xineos-20 cm x 20 cm<br>● Wireless Foot Switch<br>● Cart for Remote Control Unit<br>● Anti-Microbial Coating<br>● Target Pointer<br>● Digital Cine Mode<br>● Cios Open Apps<br>● Windows 10 |
| Secondary Predicates<br>Siemens' Cios Spin | K181550 | 10/30/2018 | ● Laser Light Localizer green |
## Table 3: Predicate Device Comparable Properties for Subject Device Modifications:
#### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The Cios Flow (VA30) System is designed as a set of components (floor stand, C-arm, Xray tube and housing, flat panel detector, digital imaging system, collimator, generator etc.) that is combined to provide a mobile Interventional Fluoroscopic X-ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logos for Siemens and Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
Components used with Cios Flow (VA30) System are either commercially available with current Siemens systems or include updated modifications to existing components. Technological differences between the Subject Device and the Predicate Device is provided in Table 4 below for all modifications.
| Table 4. Summary of Comparison of Technological Characteristics<br>Comparison of Modifications to Predicate Devices | | | |
|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device<br>Cios Flow (VA30<br>Modifications | Primary Predicate Device<br>Cios Fusion<br>K153244 | Comparison<br>Results |
| 1. | System Software VA30 software<br>modifications/features<br>A. Target Pointer<br>B. Digital Cine Mode (DCM)<br>C. Cios OpenApps | Secondary Predicate Device<br>Cios Alpha K181560<br>Target Pointer<br>Digital Cine Mode (DCM)<br>Cios OpenApps | Comparable: 1.A-C have the<br>same functionality as cleared in<br>the Secondary Predicate<br>Device Cios Alpha K181560.<br>The functionality of Target<br>Pointer has not changed<br>except for better visualization<br>of moving k-wires |
| 2. | Interactive User Touch Control<br>(Software components of VA30)<br>A. Collimation Controls<br>B. Brightness Contrast Controls<br>C. Rotate and Flip Controls<br>D. Zoom and Pan Controls<br>E. Spot Adapt | Primary Predicate Device<br>Cios Fusion<br>K153244<br>Collimation Controls<br>Brightness Contrast Controls<br>Rotate and Flip Controls<br>Zoom and Pan Controls<br>Spot Adapt | Comparable: 2.A-E has the<br>same functionality as cleared in<br>the Primary Predicate device,<br>with exception to user interface<br>is with Touch Control. The<br>functionality of these features<br>has not changed from the<br>Primary Predicate device<br>Cios Fusion K153244. |
| 3. | New Dose Regulation Indicator | New Dose Regulation Indicator | Comparable:<br>The Dose regulation is equal to<br>the primary predicate Cios<br>Fusion. While dose regulation<br>remains the same, added is a<br>visual feedback to the operator |
| 4. | New Sound Radiation Delay | New Sound Radiation Delay | Comparable:<br>Comparable to the primary<br>predicate Cios Fusion typically<br>there is a delay after press of<br>button until the radiation starts.<br>While the delay will be similar,<br>there is now an acoustical<br>feedback to the user that the<br>button has been pressed, but<br>radiation has not yet started. |
| 5. | New Product Software Security | New Product Software Security | Comparable: This Product<br>software provides additional<br>cybersecurity feature to protect<br>product security. The software<br>functionality is the same except<br>for inclusions of additional<br>software security features.<br>The updated functionality is the<br>same as cleared in the Primary<br>Predicate device, Cios<br>Fusion K153244. |
| 6. | Upgrade to Windows 10 Operating<br>System | Upgrade to Windows 10 Operating<br>System | Same: The Windows<br>Operating System is the same<br>as cleared in the Primary<br>Predicate Device with |
| Image: Healthineers logo | | | |
| <b>Comparison of Modifications to Predicate Devices</b> | | | |
| <b>7.</b> | System Software VA30<br>software/hardware<br>modifications/features | Secondary Predicate Device<br>Cios Alpha K181560 | exception to an upgraded<br>version Windows 10. The<br>upgraded functionality is the<br>same as cleared in the <b>Primary<br/>Predicate device Cios Fusion<br/>K153244.</b> |
| | A. CMOS Flat Panel Detector | CMOS Flat Panel Detector | Same: The CMOS Flat Panel<br>Detector is the exact same<br>detector cleared in the<br>Secondary Predicate Device<br>Cios Alpha K181560. |
| | B. Wireless Foot Switch | Wireless Foot Switch | Same: The Wireless Foot<br>Switch is the same as in the<br>Secondary Predicate Device<br>Cios Alpha K181560. |
| <b>8</b> | Optional Laser Light Localizer, green | Secondary Predicate Device<br>Cios Spin K181550<br>Optional Laser Light Localizer,<br>green | Same: The Optional Green<br>Laser Light Localizer is the<br>same aimer cleared in the<br>Secondary Predicate Device<br>Cios Spin K181550 |
| <b>9</b> | Optional Cart for Remote Control Unit | Secondary Predicate Device<br>Cios Alpha K181560<br>Optional Cart for Remote Control<br>Unit | Same: The Optional Cart for<br>Remote Control Unit is the<br>same cart cleared in the<br>Secondary Predicate Device<br>Cios Alpha K181560. |
| <b>10</b> | Anti-microbial Coating on C-Arm and<br>Trolley | Anti-microbial Coating on C-Arm<br>and Trolley | Same: The Anti-microbial<br>Coating on C-Arm and Trolley<br>is the same cart cleared in the<br>Secondary Predicate Device<br>Cios Alpha K181560. |
## Table 4: Summary of Comparison of Technological Characteristics
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a pattern.
#### Nonclinical Performance Testing: 9.
Non-clinical tests were conducted for the Cios Flow (VA30) during product development.
The Cios Flow (VA30) was certified by Siemens Healthcare GmbH Corporate Testing Laboratory to comply with the following standards for Electrical safety, performance and Electromagnetic Compatibility:
- AAMI ANSI ES60601-1:2005/(R)2012 ●
- IEC 60601-1-2:2014 ●
- IEC 60601-1-3:2013 ●
- IEC 60601-1-6:2010/A1:2013 ●
- IEC 60825-1:2014 ●
- IEC 62304:2015 ●
- IEC 60601-2-28:2017 ●
- IEC 60601-2-43:2017 ●
- IEC 60601-2-54:2009/A1:2015 ●
- ISO 14971:2014 ●
- . IEC 62366-1:2015
## Table 5: FDA Guidance Documents
{9}------------------------------------------------
# Healthine
| | FDA Guidance Document and Effective Date | |
|-----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 1. | Guidance for Industry and FDA Staff – User Fees and Refunds for Premarket<br>Notification Submissions 510(k)<br>Document issued on October 2, 2017 | |
| 2. | Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept Policy<br>for 510(k)s<br>Document issued on September 13, 2019 | |
| 3. | Guidance for Industry and FDA Staff: The Special 510(k) Program<br>Document issued on September 13, 2019 | |
| 4. | Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a change<br>to an existing device.<br>Document issued on October 25, 2017 | |
| 5. | Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program:<br>Evaluating Substantial Equivalence in Premarket Notifications [510(k)]<br>Document Issued on July 28, 2014 | |
| 6. | Guidance for Industry and FDA Staff: Guidance for the Submission Of 510(k)'s for Solid<br>State X-ray Imaging Devices<br>Document issued on September 1, 2016 | |
| 7. | Guidance for Industry and FDA Staff: Guidance for the Content of Premarket<br>Submission for Software in Medical Devices<br>Document issued on May 11, 2005 | |
| 8. | Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use in<br>Medical Devices<br>Document issued on September 27, 2019 | |
| 9. | Guidance for Industry and FDA Staff: Applying Human Factors and Usability<br>Engineering to Medical Devices.<br>Document issued February 3, 2016 | |
| 10. | Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging Device<br>Premarket Notifications.<br>Document issued on November 28, 2017 | |
| 11. | Guidance for Industry and FDA Staff: Content of Premarket Submissions for<br>Management of Cybersecurity in Medical devices.<br>Document issued on October 2, 2014 | |
| 12. | Guidance for Industry and FDA Staff: Appropriate Use of Voluntary Consensus<br>Standards in Premarket Submission for Medical Devices<br>Document issued on September 14, 2018 | |
| 13. | Guidance for Industry and FDA Staff: Medical Device Accessories - Describing<br>Accessories and Classification Pathways<br>Document issued on December 20, 2017 | |
| 14. | Guidance for Industry and FDA Staff: Recommended Content and Format of Non-<br>Clinical Bench Performance Testing Information in Premarket Submissions<br>Document issued on December 20, 2019 | |
The modifications described in this Premarket Notification are supported with verification and validation testing.
## Verification and Validation:
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Nonclinical tests were conducted on Cios Flow (VA30) during product development.
The Risk analysis was completed, and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
acceptance criteria. Testing for verification and validation for the device acceptable to support the claims of substantial equivalence.
The Cios Flow software (VA30) was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. The Human Factor Usability Validation showed that Human factors are addressed in the system test according to the operator's manual and in clinical use tests with customer report and feedback form. Customer employees are adequately trained in the use of this equipment.
Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. Compliance with IEC 80001-1-2010 is the responsibility of the hospital. Provided in the Software Section is the required cybersecurity information.
#### Summary:
Performance tests were conducted to test the functionality of Cios Flow (VA30) System. These tests have been performed to assess the functionality of the subject device. Results of all conducted testing and clinical assessment were found acceptable and do not raise any new issues of safety or effectiveness.
#### 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
#### 11. Conclusion as to Substantial Equivalence:
The predicate devices were cleared based on non-clinical supportive information and clinical images and data. Similar non-clinical test results demonstrate that the Cios Flow (VA30) System acceptance criteria are adequate for the intended use of the device. The comparison of technological characteristics, non-clinical performance data and software validation data demonstrates that the Subject Device is as safe and effective when compared to the Predicate Devices that is currently marketed for the same intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.