Intended for use by a qualified/trained medical professionals, who have full understanding of the safety information and emergency procedures as well of capabilities and function of the device. The device provides radiographic, multiradiographic and fluoroscopic imaging and is used for guidance and visualization during diagnostic radiographic, surgery, and interventional procedures. The device is to be used in healthcare facilities both inside and outside of hospital, in a variety of procedures of the skull, spinal column, chest, abdomen, and extremities. The device may be used for other imaging applications on all patients except pediatrics within the limits of the device. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. The system is not intended for mammography applications. (RX Only)
Device Story
MobileRay Pulse SE is a mobile digital X-ray system for radiographic and fluoroscopic imaging. Inputs: X-ray radiation captured by a portable flat panel detector (wired/wireless). System components: X-ray generator, tube, collimator, mobile cart, workstation software, and laser SSD measuring system. Operation: Physician/technician positions system; electronic alignment system ensures X-ray source and detector are aligned; radiation inhibit circuit prevents emission if misaligned. Software calculates radiation skin dose based on KVp, mAs, and SSD. Output: Digital images displayed on workstation monitor for clinical review. Used in hospitals/clinics for diagnostic, surgical, and interventional guidance. Benefits: Provides mobile, real-time visualization for procedures; electronic alignment replaces mechanical C-arm constraints; digital panel improves image acquisition over traditional intensifiers.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, electrical safety/EMC testing (EN 60601-1-2, IEC 60601-1), and software validation. Digital panels were previously cleared (K161942, K161966).
Technological Characteristics
Mobile digital X-ray system. Materials: Amorphous Silicon Direct deposition Csl:TI flat panel detector. Energy: 110-120V (4kW) or 220-230V (8kW) high-frequency inverter generator. Connectivity: Ethernet/Wi-Fi, DICOM compliant. Alignment: Electronic X-ray source/detector alignment system with radiation inhibit circuit. Software: Workstation-based image acquisition/processing. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for radiographic, multiradiographic, and fluoroscopic imaging for guidance and visualization during diagnostic, surgical, and interventional procedures of the skull, spinal column, chest, abdomen, and extremities. For use on adult patients in healthcare facilities. Contraindicated for pediatrics and mammography applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
AnyView 500R Fluoroscopic Mobile X-ray System (K160279)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Portavision Medical LLC % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Court NAPLES, FL 34114
Re: K191503
Trade/Device Name: MobileRay Pulse SE Digital Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: Mav 29, 2019 Received: June 6, 2019
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel M. Burk, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191503
Device Name MobileRay Pulse SE Digital Imaging System
#### Indications for Use (Describe)
Intended for use by a qualified/trained medical professionals, who have full understanding of the safety information and emergency procedures as well of capabilities and function of the device provides radiographic, multiradiographic and fluoroscopic imaging and is used for guidance and visualization during diagnostic radiographic, surgery, and interventional procedures. The device is to be used in healthcare facilities both inside of hospital, in a variety of procedures of the skull, spinal column, chest, abdomen, extremities. The device may be used for other imaging applications on all patients except pedian the limits of the device. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. The system is not intended for mammography applications. (RX Only)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(K) Summary, K191503
Image /page/3/Picture/1 description: The image shows the logo for PortaVision Medical. The logo consists of three blue circles with the letters P, V, and M in white. Below the circles, the words "PortaVision Medical" are written in a simple, sans-serif font.
PortaVision Medical LLC 800 Central Avenue Jefferson, Louisiana 70121 Phone: 866-783-4133 email: info@pvmed.net www.portavisionmedical.com Contact: Terry Ancar, Predsident Date Prepared: August 3, 2019
#### 1. Identification of the Device:
Trade/Device Names: MobileRay Pulse SE Digital Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image intensified fluoroscopic x-ray system Regulatory Class: II Product Codes: OWB, JAA, OXO Common/Usual Name: Mobile Fluoroscopic System
#### 2. Equivalent legally marketed device: K160279
Trade/Device Name: AnyView 500R Fluoroscopic Mobile X-ray System Manufacturer: EcoTron Regulation Number: 21 CFR 892.1650 Regulation Name: Image intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA, OXO Common/Usual Name: Mobile Fluoroscopic System
- 3. Reference Devices: (Used with this device, customer picks one of the units below. They differ only in size)
## K161942
Trade/Device Name: XRpad2 3025 HWC-M Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB K161966 Trade/Device Name: XRpad2 4336 HWC-M Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB
{5}------------------------------------------------
- 4. Indications for Use Intended for use by a qualified/trained medical professionals, who have full understanding of the safety information and emergency procedures as well of capabilities and function of the device. The device provides radiographic and fluoroscopic imaging and is used for guidance and visualization during diagnostic radiographic, surgery, and interventional procedures. The device is to be used in healthcare facilities both inside and outside of hospital, in a variety of procedures of the skull, spinal column, chest, abdomen, and extremities. The device may be used for other imaging applications on all patients except pediatrics within the limits of the device. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position. The system is not intended for mammography applications. (RX Only)
- 5. Description of the Device: The MobileRay Pulse SE system is a radiographic multi-rad, and fluoroscopic digital mobile X-Ray system that is designed for all pediatric and adult patients during diagnostic, surgery, and interventional procedures. The systems include (1) an x-ray source consisting of a X-Ray generator, X-Ray tube, and x-ray collimator, (2) a portable digital imaging system consisting of a wired and/or wireless flat panel digital detector, workstation software to acquire and process digital images in radiographic, multi-rad, and fluoroscopic modes, a laser SSD measuring system to accurately measure the distance between the X-ray tube focal spot and the patient, software which calculate radiation skin dose for KVp, mAs and SSD selected for the patient weight and body part being examined, a computer, mouse, keyboard, and image display monitor, or a laptop computer and (3) a mobile cart. The device does not incorporate a mechanically C-Arm or other mechanically device to ensure the x-ray source and flat panel detector are aligned. The device incorporates an electronic X-ray source and flat panel detector alignment system to comply with FDA regulation 21 CFR 1020.31 (g) (1). The system incorporates an automatic X-ray beam limited device to comply with FDA regulation 21 CFR 1020.32 (ii) (ii). There are several advance motion sensing technologies available to accurately determine if the Xray source is aligned to the flat panel detector. The device incorporates a radiation inhibit safety circuit to prevent the emission of radiation if the
X-ray source is not aligned to the detector and to immediately shut down emission of radiation if for whatever reason the X-ray source is no longer aligned to the detector within the predetermine tolerance of FDA regulation 21 CFR 1020.31 (g) (1). The portable x-ray generator employed is a slightly modified version of the one cleared in K121410.
- 6. Safety and Effectiveness, comparison to predicate device. This device has similar indications for use and similar technological characteristics as the predicate device, and employs already 510(k) cleared digital panels. The chief differences are: The predicate uses a fixed Source to Image distance dictated by a "C-Arm," and uses image intensifier technology whereas the proposed new device uses a unique source to image receptor location technology and the image receptor is a digital panel. Otherwise the two systems have the same functionality and uses.
#### 7. Substantial Equivalence Chart: (Next Page)
{6}------------------------------------------------
| Characteristic | EcoTron Anyview 500R K160279 | MobileRay Pulse SE Digital Imaging<br>System |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use: | Anyview-500R mobile C arm, fluoroscopic x-<br>ray system, is radiation medical equipment<br>only used by professional radiologists. This<br>product is designed to provide fluoroscopic<br>and spot film images of the patient during<br>diagnostic and interventional procedures.<br>This system can be applied in emergency<br>room, operation room, cast room or etc. of<br>hospital | Intended for use by a qualified/trained<br>medical professionals, who have full<br>understanding of the safety information and<br>emergency procedures as well of capabilities<br>and function of the device. The device<br>provides radiographic, multiradiographic and<br>fluoroscopic imaging and is used for guidance<br>and visualization during diagnostic<br>radiographic, surgery, and interventional<br>procedures. The device is to be used in<br>healthcare facilities both inside and outside<br>of hospital, in a variety of procedures of the<br>skull, spinal column, chest, abdomen, and<br>extremities. The device may be used for<br>other imaging applications on all patients<br>except pediatrics within the limits of the<br>device. Applications can be performed with<br>the patient sitting, standing, or lying in the<br>prone or supine position. The system is not<br>intended for mammography applications |
| Energy Source | 220V-230V, Single 50-60 Hz | 110V-120V, Single 50-60Hz (4 kW) or<br>220-230V Single 50-60Hz (8 kW) |
| System<br>Weight and<br>Size | 200 lbs.<br>35.75" x 28" x 72.5" | 136 lbs.<br>65" x 26" x 87" |
| Generator<br>Type | High frequency Inverter type | High frequency Inverter type |
| Output Power | 5KW | 4 kW (120v) or 8KW (220v) |
| Fluoroscopy | | |
| Pulse | 0.5 – 20 mA | 0.5 – 16mA |
| Boost | 0.5 - 30 mA | N/A |
| Radiography | | |
| KV range | 40 – 125 KV | 40 – 125 KV |
| mA range | 20 – 100 mA | 20 - 100 mA |
| mAs range | 0.8 – 200 mAs | 0.8 – 200mAs |
| Pulse width | N/A | 10 ms |
| Pulse rate | 8, 15, 30 | 1 - 7.5 fps but see key advantage<br>immediately below: |
| Weight | 552 lbs (Fairly heavy) | 139 lbs. Much lighter and easier to<br>maneuver. |
| X-ray Tube | Rotating anode | Stationary anode |
| Characteristic | EcoTron Anyview 500R K160279 | MobileRay Pulse SE Digital Imaging<br>System |
| Indicators | Display on workstation monitor | Same as predicate |
| Collimator | Multi-leaf adjustable motorized | Multi-leaf adjustable motorized |
| Digital Panel<br>Specification | N/A, not digital | XRpad2 3025 100 μ, 3004 x 2508 pixels<br>XRpad2 4336 100 μ, 4288 x 3524 pixels |
| Pixel Pitch | N/A not digital | 100 μ |
| Pixel Matrix | N/A not digital | 3004 x 2508 or 4288 x 3524 |
| AD<br>Conversion | N/A not digital | 16 bits |
| DQE | N/A | 75% (@ 0 lp/mm) |
| MTF | N/A | 70% (1 lp/mm) |
| Image<br>acquisition | Image Intensifier 9/6/4.5 in<br>E5830SD-P4A (TOSHIBA) | Amorphous Silicon Direct deposition<br>Csl:TI |
| Connection | Ethernet or Wi-Fi | Same as predicate |
| DICOM | Yes | Same as predicate |
| Electrical<br>Safety | IEC60601-1:2005 + A1 (2012)<br>IEC60601-1-2:2007 | Same as predicate, See bench test<br>information below. |
| Photo | Image: EcoTron Anyview 500R K160279 | Image: MobileRay Pulse SE Digital Imaging System |
{7}------------------------------------------------
The following table compares The MobileRay Pulse SE software to the predicate software.
| Feature | Predicate Device<br>EcoTron Anyview K160279 | Proposed Device<br>MobileRay Pulse SE |
|-------------------------------|---------------------------------------------|---------------------------------------|
| Acquiring image from detector | Yes | Yes |
| Viewing image | Yes | Yes |
| Change window/level | Yes | Yes |
| Invert | Yes | Yes |
| Lookup Table | Yes | Yes |
| Zoom | Yes | Yes |
{8}------------------------------------------------
| Feature | Predicate Device<br>EcoTron Anyview K160279 | Proposed Device<br>MobileRay Pulse SE |
|-------------------------|---------------------------------------------|---------------------------------------|
| Pan | Yes | Yes |
| Noise Reduction | Yes | Yes |
| Patient Information | Yes | Yes |
| Annotation | Yes | Yes |
| Image rotation | Yes | Yes |
| X-Ray generator control | Yes | Yes |
| DICOM worklist and Send | Yes | Yes |
- 8. Summary of non-clinical testing: Bench testing was performed to assess the device safety and effectiveness. Electrical safety and EMC testing was performed on the unit. The standards employed were: EN 60601-1-2 (2015): Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests EN 301 489-1 V2.2.0 (2017): Electromagnetic compatibility and Radio spectrum Matters (ERM); ElectroMagnetic Compatibility (EMC) standard for radio equipment and services; Part 1: Common technical requirements & EN 301 489-17 V3.2.0 (2017): Electromagnetic compatibility and Radio spectrum Matters (ERM); ElectroMagnetic Compatibility (EMC) standard for radio equipment; Part 17: Specific conditions for Broadband Data Transmission Systems AND: IEC 60601-1: 2005 + Corr.1: 2006 + A1: 2012 EN 60601-1: 2006 + A11: 2011 + A1: 2013 + AC:2014 + A12:2014 UNE-EN 60601-1: 2008 + Corr.: 2010 + A11: 2010 + A11: 2012 + AC:2014 + A12:2015 POSE000 14 (General procedure of Safety Lab) EMC and Electrical Safety performance for the digital receptor panels had previously been submitted to FDA in K161942 and in K161966. Software has been written and validated according to the FDA Software Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document issued on: May 11, 2005 Cybersecurity concerns have been addressed in accordance with: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff (October 2, 2014). In order to be able to perform fluoroscopic exposures the x-ray generator (cleared in K121410) was modified to include cooling fans and to allow for repetitive exposures, i.e. fluoro/pulsed mode. The internal temperature rise in the generator was measured during operation and found to be acceptable.
- 9. Summary of clinical testing: No clinical data is necessary to evaluate safety or effectiveness for purposes of determining substantial equivalence of the proposed modification. The digital panels both received previous 510(k) clearances
- 10. Conclusion: After analyzing software integration validation, safety testing data, and bench test images, it is the conclusion of PortaVision Medical LLC that the MobileRay Pulse SE Digital Imaging System is as safe and effective as the predicate device, has insignificant technological differences, and has identical indications for use, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.