GE INNOVA SOLID STATE XRAY-IMAGER FLUOROSCOPIC X-RAY SYSTEM, MODEL: 4100IQ, 3100IQ, 2100IQ
K092004 · GE Healthcare · OWB · Aug 25, 2009 · Radiology
Device Facts
Record ID
K092004
Device Name
GE INNOVA SOLID STATE XRAY-IMAGER FLUOROSCOPIC X-RAY SYSTEM, MODEL: 4100IQ, 3100IQ, 2100IQ
Applicant
GE Healthcare
Product Code
OWB · Radiology
Decision Date
Aug 25, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. These devices are not intended for mammography applications.
Device Story
Innova 4100 IQ, 3100 IQ, and 2100 IQ systems are x-ray angiographic imaging devices modified with an optional software feature, StentViz. StentViz processes video image sequences recorded during fluoroscopic guidance to produce an enhanced static image of a stent. The device is used in clinical settings (e.g., interventional suites) by physicians to improve stent visibility during procedures. StentViz does not provide real-time guidance; it generates a post-processed static image. By enhancing stent visualization, the device assists clinicians in evaluating stent placement, potentially improving procedural outcomes. The system operates as an integrated component of the existing Innova fluoroscopy platform.
Clinical Evidence
No clinical studies were required. Substantial equivalence was supported by bench testing using a library of clinical images to compare stent visibility enhancement between the subject device and the predicate StentOp feature.
Technological Characteristics
X-ray angiographic system with optional StentViz software. StentViz is a non-real-time image processing feature that derives enhanced static images from fluoroscopic video sequences. The system complies with voluntary standards for safety and performance. Development followed quality assurance measures including risk analysis, module verification, system integration testing, and final validation.
Indications for Use
Indicated for patients undergoing vascular angiography, cardiology, diagnostic, and interventional procedures requiring fluoroscopic imaging. Not for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K092004
P. 1 of 3
AUG 2 5 2009
GE Healthcare
510(k) Premarket Notification Submission
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
1 st July 2009 Date:
Submitter: GE Healthcare {GE Medical Systems, LLC] Doing business with GE Healthcare 3000 N. Grandview Blvd Waukesha, WI 53188
Primary C
Device: Commo
| Primary Contact Person: | Alan Totah<br>Regulatory Affairs Director, Pre-market<br>GE Healthcare<br>(262) -544-3424<br>(262) -544-3202 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Secondary Contact Person: | Philip Malca<br>Interventional Regulatory Affairs Director<br>GE Healthcare GE Medical Systems SCS.<br>33 1 30 70 42 07<br>33 1 30 70 43 99 |
| Device: Trade Name: | Innova 4100 <sup>IQ</sup> , 3100 <sup>IQ</sup> , 2100 <sup>IQ</sup> with StentViz |
| Common/Usual Name: | Innova 4100 <sup>IQ</sup> , 3100 <sup>IQ</sup> , 2100 <sup>IQ</sup> with StentViz |
| Classification Names: | System X-Ray, Angiographic |
| Product Code: | OWB, JAA, IZI |
| Predicate Device(s): | Innova 4100 <sup>IQ</sup> , 3100 <sup>IQ</sup> , 2100 <sup>IQ</sup><br>K052412, K050489, K060259, K061163<br>IC-PRO device (K083745) featuring StentOp<br>Philips Fresco release 1 (featuring StentBoost) (K031836) |
| Device Description: | The Innova 4100 <sup>IQ</sup> , 3100 <sup>IQ</sup> , 2100 <sup>IQ</sup> Systems are modified with an<br>optional software feature called StentViz. The StentViz feature<br>enhances the visibility of stents in the x-ray images produced by<br>the Innova systems. Specifically, StentViz provides an enhanced<br>static image of the stent that is derived from the video image<br>sequence as recorded during fluoroscopic guidance. It does not |
- The Innova systems are indicated for use in generating Intended Use: fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to
provide real-time guidance.
- replace fluoroscopic images obtained through image intensifier technology. These devices are not intended for mammography applications.
optionally,
{1}------------------------------------------------
K092004
?. 2 of 3
# GE Healthcare
### 510(k) Premarket Notification Submission
#### Technology:
Determination of Substantial Equivalence:
The Innova 4100'0, 3100'0, 2100'0 with StentViz employs the same fundamental scientific technology as its predicate devices.
The subject device is of a comparable type and substantially equivalent to the unmodified Innova 41000, Innova 310019 Innova 2100 10 devices. For the purpose of comparison, the modified and unmodified devices are identical except for the additional software option (StentViz). When StentViz is used, the image quality and visibility of the stent is improved. This improvement does not adversely impact safety or effectiveness.
Two other predicate devices contain an equivalent feature to the StentViz:
- Paieon StentOp feature of the IC-PRO device (K083745)
- and Philips Fresco (also known as StentBoost) (K031836).
StentOp is a stand-alone software application that can be used on any fluoroscopy system while StentBoost is a feature embedded in the Philips Fresco fluoroscopy system. Both the StentOp and StentBoost offer the ability to provide an enhanced static image of the stent that is derived from the video image sequence as recorded during fluoroscopic guidance. As these predicate devices. StentViz does not provide real-time quidance.
The image quality of the stent is enhanced in a comparable way with StentViz than with StentOp and StentBoost. From a design validation performance standpoint, bench tests were performed based on a library of clinical images. This library was used to assess the enhancement of stent visibility and to compare the performance of the Innova with StentViz to:
- Innova without StentViz applied on the video image sequence as recorded
- the performance of the similar feature contained in the predicate device StentOp. The image quality of the stent is enhanced in a comparable way with StentOp.
The Innova 410019, 310019, 2100'9 with StentViz and its applications comply with voluntary standards as detailed in Sections 9 and 17 of this premarket submission. The following quality assurance measures were applied to the development of the svstem:
- Risk Analysis .
- . Requirements Reviews
- Design Reviews .
- . Testing on unit level (Module verification)
- . Integration testing (System verification)
- . Final acceptance testing (Validation)
{2}------------------------------------------------
## K092004 P. 3 of 3
## GE Healthcare
510(k) Premarket Notification Submission
- Performance testing (Verification) ●
- Safety testing (Verification) ●
#### Summary of Clinical studies:
The subject of this premarket submission, Innova 41000, 310000, 210019 with StentViz, did not require clinical studies to support substantial equivalence.
Conclusion:
GE Healthcare considers the Innova 41000, 31000, 21000 with StentViz to be as safe and as effective as the predicate devices, and its performance is substantially equivalent to the predicate devices.
Image /page/2/Picture/9 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' in a stylized, cursive font, enclosed within a circular border. The letters and the border are solid black, creating a strong contrast against the white background. The logo is a well-known symbol associated with the multinational conglomerate company, General Electric.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Alan Totah Director, RA, PreMarket GE Healthcare Systems 3000 N. Grandview Blvd. WAUKESHA WI 53188
1111 30 2012
Re: K092004
Trade/Device Name: Innova 410010. 310010. 210010 with StentViz Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA and IZI Dated: July 1, 2009 Received: July 2, 2009
Dear Mr. Totah:
This letter corrects our substantially equivalent letter of August 25, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. . Code of Federal Regulations (CFR), Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined inside a circle. The letters are stylized and appear to be hand-drawn. The logo is black and white.
GE Healthcare 510lk) Premarket Notification Submission
510(k) Number (if known) : _ K09 2004
Innova 4100'0, 3100'0 , 2100'0 with StentViz Device Name :
Indications for Use:
The Innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally, rotational imaging procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. They are intended to replace fluoroscopic images obtained through image intensifier technology. These devices are not intended for mammography applications.
Prescription Use_ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Reproductive, Abdominal and Radiological Devic
510(k) Number K072004
17/138
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.