INNOVA 2100

K050489 · Ge Medical Systems, LLC · OWB · Mar 11, 2005 · Radiology

Device Facts

Record IDK050489
Device NameINNOVA 2100
ApplicantGe Medical Systems, LLC
Product CodeOWB · Radiology
Decision DateMar 11, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The Innova 2100-IQ system is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures, and optionally, rotational angiography procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology diagnostic and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. This device is not intended for mammography applications.

Device Story

Digital Fluoroscopic Imaging System (Innova 2100-IQ) performs fluoroscopic X-ray examinations; utilizes amorphous silicon detector with cesium iodine scintillator. System components: monoplane positioner, vascular/cardiac table, X-ray system, digital detector. Inputs: X-ray radiation; outputs: digital images. Data transmitted via Fiber Channel link to acquisition system; networked via DICOM for post-processing, printing, viewing, archiving. Used in clinical settings (e.g., interventional suites) by trained medical professionals. Hardware/software redundancies prevent single-point failures causing unintended motion. Benefits: replaces image intensifier technology with digital imaging capabilities for vascular and cardiac procedures.

Clinical Evidence

No clinical data provided. Equivalence is based on comparison to previously cleared GE digital imaging products (LCV+ Version 2 and Innova 4100) regarding image quality and diagnostic capabilities. Bench testing and risk management documentation were used to support the safety and effectiveness profile.

Technological Characteristics

Amorphous silicon detector with cesium iodine scintillator. 480 VAC 50/60Hz power. Monoplane positioner, vascular/cardiac table. Connectivity via Fiber Channel and DICOM network. Materials compliant with UL 187 and IEC 60601-1. Hardware and software redundancies for motion control.

Indications for Use

Indicated for human patients requiring vascular or cardiac diagnostic and interventional fluoroscopic imaging, including rotational angiography. Not for mammography.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Tage 1 of 3 MAR í í 2005 Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular shape. The logo is black and white and has a vintage look. GE Healthcare P.O. Box 414, W-400 Milwaukee, WI 53201 USA ## 4050489 ## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h). | Identification of Submitter: | Larry A. Kroger, Ph.D.<br>Senior Regulatory Programs Manager<br>GE Healthcare<br>Tel. (262) 544-3894<br>Summary prepared: 07 February 2005 | |---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Identification of Product:<br>Classification Name:<br>Manufacturer: | Digital Fluoroscopic Imaging System<br>Fluoroscopic X-ray System<br>GE Medical Systems SCS.<br>283, rue de la Minière | | Distributed by: | 78530 Buc Cedex, France<br>GE Medical Systems, LLC, Milwaukee, WI | | Marketed Devices: | The Innova 2100-IQ is substantially equivalent to the currently<br>marketed Vascular Angiographic system Innova 3100 (K031637)<br>and complies with the same or equivalent standards.<br>Optional feature InnovaSpin and the cardiac imaging modes in<br>the Innova 2100-IQ are substantially equivalent to the Innova<br>2000/Innova 2000S systems cleared under K022322 and LCV+<br>Version 2 systems cleared under K993037. | | Device Description: | The Digital Fluoroscopic Imaging System is designed to perform<br>fluoroscopic x-ray examinations. The detector is comprised of<br>amorphous silicon with a cesium iodine scintillator. The resulting<br>digital image can be sent through a Fiber Channel link to an<br>acquisition system then to network (in using DICOM) for<br>applications such as post-processing, printing, viewing and<br>archiving. Digital Fluoroscopic Imaging System consists of an a<br>monoplane positioner, a vascular or cardiac table, an X-RAY<br>system and a digital detector. | | Materials: | All construction and materials are compliant with UL 187 for the<br>existing parts of the product and with IEC 60601-1 for the new<br>parts. | {1}------------------------------------------------ ygcoiz K050489 There are hardware and software redundancies to prevent single Design: point failures that could cause unintended motion. 480 VAC 50/60Hz. Energy Source: The Innova 2100-IQ system is indicated for use in generating Indications for Use: fluoroscopic images of human anatomy for vascular angiography diagnostic and interventional procedures, and optionally, rotational angiography procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology diagnostic and interventional procedures. > It is intended to replace fluoroscopic images obtained through image intensifier technology. This device is not intended for mammography applications. The Innova 2100-IQ is substantially equivalent to the currently Comparison with marketed Vascular Angiographic system Innova 3100 (K031637) Predicate: and complies with the same or equivalent standards. The optional feature InnovaSpin and cardiac imaging modes in the Innova 2100-IQ are substantially equivalent to the Innova 2000/Innova 2000S systems cleared under K022322 and LCV+ Version 2 systems cleared under K993037. > Innova 2100-IQ combines features from Innova 2000/LCV+ Version 2 and Innova 3100 to enable cardiac and vascular procedures. The indications of use for the predicative devices are given below: - The Innova 3100 System is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, and optionally, rotational angiography procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. - Innova 2000 / Innova 2000S: The Digital Fluoroscopic Imaging Systems are indicated for use in generating fluoroscopic images of human anatomy for diagnostic and interventional cardiac angiography, and optionally, rotational cardiac angiography procedures. It is intended to replace fluoroscopic images obtained through the image intensifier technology. - The LCV+ Version 2 is indicated for use in generating fluoroscopic images of human anatomy for cardiology diagnostic / interventional procedures. It is intended to replace fluoroscopic images obtained through the image intensifier technology. {2}------------------------------------------------ Yage 3rd? KOSO48 References in term of clinical data have been submitted for first Summary of the Studies: digital products in the family introduced by GE Medical Systems-Europe ie LCV+ Version 2 for cardiology diagnostic and interventional procedures, and in Innova 4100 for vascular angiography diagnostic and interventional procedures. As Innova 2100-IQ is considered substantially equivalent to Innova 3100 (predicated by Innova 4100), and LCV+ Version 2 in terms of image quality and diagnostic capabilities, it is not applicable. Conclusions: GE considers the 20 cm Digital Fluoroscopic Imaging System to be equivalent with the predicate devices. The potential hazards, e.q., wrong measurements and misdiagnosis, are controlled by a risk management plan including: - A hazard identification (Attachment 8) . - A risk evaluation (Attachment 8) 0 - A Software Development and Validation Process . (Attachment 7) {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service JUL 30 2012 Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 GE Medical Systems, LLC % Mr. Daniel W. Lehtonen Third Party Official Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719 Re: K050489 Trade/Device Name: Digital Fluoroscopic Imaging System-Innova 2100-IQ Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB and JAA Dated: February 23, 2005 Received: February 25, 2005 Dear Mr. Lehtonen: This letter corrects our substantially equivalent letter of March 11, 2005. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: K050489 Digital Fluoroscopic Imaging System - Innova 2100-IQ Indications for Use: The Innova 2100-IQ system is indicated for use in generating fluoroscopic images of human The innova 2 foo Ta System to the diagnostic and interventional procedures, and optionally. rotational angiography procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology diagnostic and interventional procedures. It is intended to replace fluoroscopic images oblained through image intensifier technology. This device is not intended for mammography applications. Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter-Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED) AND/OR Concurrence of CDRH, Office of Device Evaluation (ODE) David h. Lynn X
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...