K231855 · Quanta Computer, Inc. · QKB · Feb 13, 2024 · Radiology
Device Facts
Record ID
K231855
Device Name
QOCA® image Smart RT Contouring System
Applicant
Quanta Computer, Inc.
Product Code
QKB · Radiology
Decision Date
Feb 13, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K231855 · Feb 13, 2024
QOCA® image Smart RT Contouring System
Quanta Computer, Inc.
Retrospective hospital clinical CT imaging datasets; The Cancer Imaging Archive (TCIA) public dataset; Gold Atlas - Male Pelvis (GA) public dataset
Retrospective clinical CT data was used to train, validate, and test the deep-learning algorithm's ability to perform automated organ-at-risk (OAR) contouring for radiation therapy planning.
Retrospective data; Deep learning validation; Performance evaluation; DICOM CT images
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective performance study on CT data; Retrospective performance validation; Study Period: 2000-2021 (data collection period)
Adult patients (head, neck, and pelvis regions); Sample Size: 220 cases (110 head and neck, 110 pelvis); Number of Sites: Multiple (Taiwan hospitals and US public datasets)
Not applicable for this study
DICE similarity coefficient (DSC) and Hausdorff distance (HD95)
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Organ at Risk (OAR) Contouring
Deep learning-based algorithms
DSC > 0.66 to 0.93
DSC 0.805 to 0.956
Training: 317 head and neck cases; Validation: 317 head and neck cases; Test: 317 head and neck cases (8:1:1 ratio).
—
Retrospective performance study: 110 head and neck CT images (50 from Taiwan, 60 from TCIA).
Retrospective performance study: 110 pelvis CT images (50 from Taiwan, 60 from TCIA).
—
Indications for Use
QOCA® image Smart RT Contouring System is a post-processing software intended to automatically contour DICOM CT imaging data using deep-learning-based algorithms. Contours that are generated by QOCA® image Smart RT Contouring System may be used as input for clinical workflows including external beam radiation therapy treatment planning. QOCA® image Smart RT Contouring System must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by QOCA® image Smart RT Contouring System. The output of QOCA® image Smart RT Contouring System in the format of RTSTRUCT objects are intended to be used by radiation oncology department. QOCA® image Smart RT Contouring System does not provide a user interface for data visualization. System settings, user settings, progress status, and other functionalities are managed via a web-based interface. The software is not intended to automatically detect or contour lesions. Only DICOM images of adult patients are considered to be valid input.
Device Story
Post-processing software; automates contouring of organs at risk (OAR) on DICOM CT images for radiation therapy planning. Uses deep-learning algorithms to generate RTSTRUCT objects. Operated by radiation oncology staff; requires integration with external Treatment Planning Systems (TPS) or interactive contouring software for review, editing, and final acceptance. Web-based interface manages system settings and status. Does not visualize data or detect lesions. Benefits include standardized, automated anatomical segmentation to support efficient, precise radiation therapy treatment planning.
Clinical Evidence
No clinical data. Bench-only retrospective performance study using 220 independent CT cases (110 head/neck, 110 pelvis) from Taiwan and US (TCIA). Ground truth established via consensus guidelines. Primary endpoint: Dice Similarity Coefficient (DSC) and Hausdorff Distance (HD95). Results showed median DSC > 0.80 across validated OARs, confirming performance consistency across contrast and non-contrast CT images.
Technological Characteristics
Software-only device; deep-learning-based segmentation algorithm. Compatible with DICOM CT images (contrast/non-contrast). Outputs RTSTRUCT objects. Web-based management interface. Operates on Windows. Complies with IEC 62304:2006/A1:2016 for software lifecycle and cybersecurity standards.
Indications for Use
Indicated for adult patients undergoing radiation therapy requiring identification of organs at risk (OAR) in head, neck, and pelvis regions using DICOM CT imaging data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
February 13, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Quanta Computer Inc. % Joe Wang, Research Specialist No. 188, Wenhua 2nd Rd Guishan Dist. Taoyuan City, 33383 TAIWAN
## Re: K231855
Trade/Device Name: QOCA® image Smart RT Contouring System Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QKB Dated: January 8, 2024 Received: January 8, 2024
Dear Joe Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming
{1}------------------------------------------------
product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lora Weidner
Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231855
#### Device Name
QOCA® image Smart RT Contouring System
#### Indications for Use (Describe)
QOCA® image Smart RT Contouring System is a post-processing software intended to automatically contour DICOM CT imaging data using deep-learning-based algorithms.
Contours that are generated by QOCA® image Smart RT Contouring System may be used as input for clinical workflows including external beam radiation therapy treatment planning. QOCA® image Smart RT Contouring System must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by OOCA® image Smart RT Contouring System. The output of QOCA® image Smart RT Contouring System in the format of RTSTRUCT objects are intended to be used
by radiation oncology department.
QOCA® image Smart RT Contouring System does not provide a user interface for data visualization. System settings, user settings, progress status, and other functionalities are managed via a web-based interface.
The software is not intended to automatically detect or contour lesions. Only DICOM images of adult patients are considered to be valid input.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| <b>5.1. Type of Submission:</b> | Traditional |
|---------------------------------|------------------------------------------------------------------------------|
| <b>5.2. Date of Summary:</b> | 01/08/2024 |
| <b>5.3. Submitter:</b> | Quanta Computer Inc. |
| <b>Address:</b> | No. 188, Wenhua 2nd Rd., Guishan Dist. Taoyuan<br>City 33383, Taiwan (R.O.C) |
| <b>Phone:</b> | +886-3-327-2345 |
| <b>Contact:</b> | Joe Wang<br>joe_wang@quantatw.com |
## 5.4. Identification of the Device:
| Proprietary/Trade Name: | QOCA® image Smart RT Contouring System |
|-------------------------|------------------------------------------------|
| Model Number: | ZSWR901 |
| Review Panel: | Radiology |
| Regulation Name: | Medical Image Management and Processing System |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | QKB |
| Device Class: | II |
# 5.5. Identification of the Predicate Device:
| Predicate Device Name: | AccuContour™ |
|------------------------|--------------------------------|
| Model Number: | -- |
| 510(k) Number: | K191928 |
| Manufacturer: | Xiamen Manteia Technology LTD. |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | QKB |
| Device Class: | II |
{4}------------------------------------------------
### 5.6. Intended Use/Indications for Use of the Device
QOCA® image Smart RT Contouring System is a post-processing software intended to automatically contour DICOM CT imaging data using deep-learning-based algorithms.
Contours that are generated by QOCA® image Smart RT Contouring System may be used as input for clinical workflows including external beam radiation therapy treatment planning. QOCA® image Smart RT Contouring System must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by QOCA® image Smart RT Contouring System.
The output of QOCA® image Smart RT Contouring System in the format of RTSTRUCT objects are intended to be used by radiation oncology department.
QOCA® image Smart RT Contouring System does not provide a user interface for data visualization. System settings, user settings, progress status, and other functionalities are managed via a web-based interface.
The software is not intended to automatically detect or contour lesions. Only DICOM images of adult patients are considered to be valid input.
## 5.7. Device Description
QOCA® image Smart RT Contouring System is a post-processing software used to automatically contour DICOM CT imaging data using deep-learning-based algorithms. OOCA® image Smart RT Contouring System contouring workflow supports CT inout data and produces RTSTRUCT outputs. Contours that are generated by QOCA® image Smart RT Contouring System may be used as input for clinical workflows including external beam radiation therapy treatment planning.
The output of QOCA® image Smart RT Contouring System, in the form of RTSTRUCT objects, are intended to be used by radiation oncology department. The output of QOCA® image Smart RT Contouring System must be used in conjunction with appropriate software such as Treatment Planning Systems and Interactive Contouring applications, to review, edit, and accept contours generated by QOCA® image Smart RT Contouring System.
QOCA® image Smart RT Contouring System includes the following functionality:
- Automated contouring of organs at risk (OAR) workflow ●
- Input - DICOM CT
- Output - DICOM CT (Original), DICOM RTSTRUCT
{5}------------------------------------------------
- Web-based interface of system settings, user settings, and checking progress status
QOCA® image Smart RT Contouring System is intended to be used on adults undergoing treatment that requires the identification of anatomical structures in the body considered to be OAR. QOCA® image Smart RT Contouring System is intended to be used in the head, neck, and pelvis regions.
## 5.8. Comparison of Technological Characteristics with the Predicate Device
QOCA® image Smart RT Contouring System submitted in this 510(k) file is substantially equivalent in intended use, safety and performance to the cleared AccuContour™ (K191928). Differences between the devices cited in this section do not raise any new issue of substantial equivalence.
| Item | Subject Device | Predicate Device | Substantial<br>Equivalence<br>Determination | |
|----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K231855 | K191928 | -- | |
| Proprietary<br>Name | QOCA® image Smart<br>RT Contouring<br>System | AccuContour™ | -- | |
| Manufacturer | Quanta Computer Inc. | Xiamen Manteia<br>Technology LTD. | -- | |
| Regulation<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same | |
| Product Code | QKB | QKB | Same | |
| Classification | Class II | Class II | Same | |
| Intended<br>Use/Indication<br>for Use | QOCA® image Smart<br>RT Contouring<br>System is a post-<br>processing software<br>intended to<br>automatically contour<br>DICOM CT imaging<br>data using deep-<br>learning-based<br>algorithms. | It is used by<br>radiation oncology<br>department to<br>register<br>multimodality<br>images and<br>segment (non-<br>contrast) CT<br>images, to generate<br>needed information | Similar<br>Both devices utilize<br>artificial<br>intelligence<br>algorithms to<br>automatically<br>contour organs at<br>risk, including the<br>head and neck, as<br>well as the pelvis. | |
| Contours that are | for treatment | for radiation | | |
| generated by QOCA® | planning, treatment | treatment planning. | | |
| image Smart RT | evaluation and | | | |
| Contouring System | treatment | | | |
| may be used as input | adaptation. | | | |
| for clinical workflows | | | | |
| including external | | | | |
| beam radiation | | | | |
| therapy treatment | | | | |
| planning. QOCA® | | | | |
| image Smart RT | | | | |
| Contouring System | | | | |
| must be used in | | | | |
| conjunction with | | | | |
| appropriate software | | | | |
| such as Treatment | | | | |
| Planning Systems and | | | | |
| Interactive Contouring | | | | |
| applications, to | | | | |
| review, edit, and | | | | |
| accept contours | | | | |
| generated by QOCA® | | | | |
| image Smart RT | | | | |
| Contouring System. | | | | |
| The output of QOCA® | | | | |
| image Smart RT | | | | |
| Contouring System in | | | | |
| the format of | | | | |
| RTSTRUCT objects | | | | |
| are intended to be | | | | |
| used by radiation | | | | |
| oncology department. | | | | |
| QOCA® image Smart | | | | |
| RT Contouring | | | | |
| System does not | | | | |
| | provide a user<br>interface for data<br>visualization. System<br>settings, user settings,<br>progress status, and<br>other functionalities<br>are managed via a<br>web-based interface.<br>The software is not<br>intended to<br>automatically detect or<br>contour lesions. Only<br>DICOM images of<br>adult patients are<br>considered to be valid<br>input. | | | |
| Operating System | Windows | Windows | Same | |
| Algorithm | Deep Learning | Deep Learning | Same | |
| Segmentation of<br>Organ at Risk in<br>the Anatomic<br>Regions | Brain stem,<br>Esophagus, Mandible,<br>Pharyngeal,<br>Constrictor Muscle<br>(PCM), Spinal cord,<br>Thyroid, Right eye,<br>Light eye, Right lens,<br>Left lens, Right optic<br>nerve, Left optic<br>nerve, Right parotid,<br>Left parotid,<br>Anorectum, Bladder,<br>Bowel bag, Lumbar<br>spine L5, Bilateral<br>seminal vesicles,<br>Right iliac, Left iliac. | Head and Neck,<br>Thorax, Abdomen,<br>and Pelvis | <i>Similar</i><br>The subject device<br>contains head and<br>neck, and pelvis. | |
| Left proximal femur | | | | |
| Compatible<br>Modality | CT Images | Non-Contrast CT<br>Images | Similar<br>The subject device<br>and the predicate<br>device are both<br>compatible only<br>with CT images for<br>the segmentation<br>feature. The<br>predicate device<br>claims to handle<br>only non-contrast<br>CT images, while<br>the subject device<br>can be used with<br>both contrast and<br>non-contrast CT<br>images. | |
| Compatible<br>Scanner Models | No specific<br>requirement for the<br>scanner model, it is<br>recommended to use<br>multi-detector CT<br>(MDCT) equipment<br>with more than 16<br>slices for Radiation<br>Therapy Simulation<br>CT, and DICOM<br>compliance required. | No Limitation on<br>scanner model,<br>DICOM 3.0<br>compliance<br>required. | Similar<br>The Subject Device<br>and Predicate<br>Device are<br>substantially<br>equivalent in their<br>requirements for<br>DICOM<br>compliance,<br>ensuring<br>interoperability and<br>standardization in<br>medical imaging<br>data. The Subject<br>Device<br>recommends a | |
| | | | scanner with more | |
| | | | than 16 slices | |
| | | | specifically for | |
| | | | Radiation Therapy | |
| | | | Simulation CT to | |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.