The device performance was validated by comparing software-generated segmentations and measurements against a ground-truth dataset derived from 238 retrospective clinical MRI studies collected from five U.S. clinical sites.
MSKai is an image identification, post-processing, measurement, and reporting software tool that provides qualitative viewing and quantitative spine measurements from previously-acquired T2 weighted DICOM lumbar spine Magnetic Resonance Imaging (MRI) images for users' review, evaluation and analysis. It provides the following functionality to assist users in identifying, observing, measuring and reporting measurements: - Anatomy segmentation; - Anatomy labeling; - Anatomy measurement; and - Export of measurement results to a qualitative and quantitative report for user's evaluation, amendment and authorization MSKai does not serve as a diagnostic device by providing or recommending any type of medical diagnosis or treatment. MSKai simply provides users the ability to access objective and repeatable identification, segmentation, measurement and reported measurements of the Lumbar spine. The user is responsible for the indications of preferences and settings, confirming the software-generated measurements, and reviewing, confirming and approving draft reports based on their medical training. The device is intended to be used only by physicians, radiologist, hospitals and other medical institutions. Only T2 weighted DICOM images of MRI acquired from lumbar spine exams of patients aged 18 and above are acceptable input. MSKai does not support DICOM images of patients that are pregnant, have post-operational complications, tumors, infections, or complex hardware.
Device Story
MSKai is a SaMD tool for lumbar spine MRI analysis; inputs T2-weighted DICOM MRI images; uses Mask R-CNN algorithm for anatomy segmentation, labeling, and measurement; outputs quantitative measurements and draft reports for clinician review. Used in hospitals/clinics by radiologists and spine surgeons. Clinician reviews, confirms, or rejects software-generated segmentations and measurements before report finalization. Device serves as a workflow aid; does not provide automated diagnosis. Benefits include objective, repeatable measurements to assist clinical decision-making.
Clinical Evidence
No human clinical study conducted. Standalone software performance assessment study used 238 independent MR image studies (5 sites, U.S.). Ground truth established by 5 experts (3 neurosurgeons, 1 interventional radiologist, 1 PhD). Primary endpoints (MAE) and secondary endpoints (Mean Dice Coefficient) met pre-defined limits. MDC results ranged 0.911–0.993 across anatomical structures. MAE results for measurements were within 2mm or 6° limits. Accuracy confirmed across MRI manufacturers (GE, Philips, Siemens), age, gender, and race.
Technological Characteristics
SaMD; Mask Region-based Convolutional Neural Network (Mask R-CNN); processes T2-weighted DICOM MRI; standalone software; conforms to IEC 62304, ISO 14971, IEC 62366-1, ISO 15223-1, and NEMA PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for patients aged 18+ undergoing lumbar spine MRI. Excludes patients who are pregnant, have post-operational complications, tumors, infections, or complex hardware.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
MSKai
Wade Lloyd
Chief Operating Officer
6075 Poplar
Suite 221
Memphis, TN 38119
December 16, 2024
Re: K240793
Trade/Device Name: MSKai
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: November 10, 2024
Received: November 15, 2024
Dear Wade Lloyd:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
2K240793 - Wade Lloyd
Page
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
3K240793 - Wade Lloyd
Page
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb
Assistant Director
Imaging Software Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K240793
Device Name
MSKai
Indications for Use (Describe)
MSKai is an image identification, post-processing, measurement, and reporting software tool that provides qualitative viewing and quantitative spine measurements from previously-acquired T2 weighted DICOM lumbar spine Magnetic Resonance Imaging (MRI) images for users' review, evaluation and analysis. It provides the following functionality to assist users in identifying, observing, measuring and reporting measurements:
- Anatomy segmentation;
- Anatomy labeling;
- Anatomy measurement; and
- Export of measurement results to a qualitative and quantitative report for user's evaluation, amendment and authorization
MSKai does not serve as a diagnostic device by providing or recommending any type of medical diagnosis or treatment. MSKai simply provides users the ability to access objective and repeatable identification, segmentation, measurement and reported measurements of the Lumbar spine. The user is responsible for the indications of preferences and settings, confirming the software-generated measurements, and reviewing, confirming and approving draft reports based on their medical training.
The device is intended to be used only by physicians, radiologist, hospitals and other medical institutions. Only T2 weighted DICOM images of MRI acquired from lumbar spine exams of patients aged 18 and above are acceptable input. MSKai does not support DICOM images of patients that are pregnant, have post-operational complications, tumors, infections, or complex hardware.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K240793
MSKai
510k Summary
1. Submitter
MSKai
6075 POPLAR SUITE 221
Memphis TN 38119
United States
Phone: (901) 498-6655
Contact Person: Dr. Lloyd Wade, Ph.D.
Position: Chief Operating Officer
2. Device
Common or Usual Name: MSKai
Classification Name: Automated Radiological Image Processing Software
Regulatory Class: 892.2050
Product Code: QIH
3. Predicate Devices
Device Name: CoLumbo
Manufacturer: Smart Soft Healthcare AD
Classification Panel: Radiology
Classification Name: Medical image management and processing system (21 CFR 892.2050)
Product Code: QIH
Device Class: Class II 510(k)
Number: K220497 (cleared on June 23, 2022)
Device Name: AI-Rad Companion Brain MR
Manufacturer: Siemens Healthcare GmbH
Classification Name: Medical image management and processing system (21 CFR 892.2050)
Secondary Classification Name: Magnetic resonance diagnostic device
Classification Product Code: LLZ Subsequent
Product Code: LNH Classification Panel: Radiology
Device Class: Class II 510(k)
Number: K193290 cleared July 5, 2019
Device Name: AI-Rad Companion (Cardiovascular)
Manufacturer: Siemens Medical Solutions USA, Inc.
Classification Name: Computed tomography x-ray system
Regulation Number: 21 CFR 892.1750
Classification Product Code: JAK
Subsequent Product Code: LLZ
Classification Panel: Radiology Device Class: Class II
{5}
4. Device Description
MSKai is a medical device (software) for inspecting and evaluating T2-weighted magnetic resonance imaging (MRI) of the lumbar spine. The software is an imaging interpretation tool that assists radiologists and neuro/ortho spine surgeons ("users") to identify and measure lumbar spine features in medical images and document their interpretations in a report. The segmentation and measurements are classified using "alerts" based on rule-based algorithms. The user also identifies and classifies any other observations that the software may not annotate.
The purpose of MSKai is to provide information regarding common spine measurements confirmed by the user. Every feature segmented, labeled, and measured by the software, based on the user-defined settings, must be reviewed and affirmed by the radiologist before the measurements of these features can be stored and reported. The software initiates adjustable measurements resulting from semi-automatic segmentation. If the user rejects a measurement the corresponding segmentation is rejected too. Segmentations are not intended to be a final output but serve the purpose of interpretation and calculating measurements. The device outputs are intended to be a starting point for a clinical workflow and should not be interpreted or used as a diagnosis. The user is responsible for confirming segmentation and all measurement outputs. The output is an aid to the clinical workflow of measuring patient anatomy and should not be misused as a diagnosis tool.
User-confirmed/defined settings control the sensitivity of the software for labelling measurements in an image. The user (not the software) controls identifying out-of-range measurements, and, in every case once an out-of-range measurement is identified, the user must confirm or reject its presence.
The software facilitates this process by labeling and mask segmenting around features of the relevant anatomy and displaying measurements based on these masks. The user maintains control of the process by inspecting the segmentation, measurements and labels upon which the measurements are based. The user may also examine other features of the imaging not annotated by the software to form a complete impression and diagnostic judgment of the overall state of disease, disorder, or trauma.
5. Indications for Use
MSKai is an image identification, post-processing, measurement, and reporting software tool that provides qualitative viewing and quantitative spine measurements from previously-acquired T2 weighted DICOM lumbar spine Magnetic Resonance Imaging (MRI) images for users' review, evaluation and analysis. It provides the following functionality to assist users in identifying, observing, measuring and reporting measurements:
- Anatomy segmentation;
- Anatomy labeling;
{6}
- Anatomy measurement; and
- Export of measurement results to a qualitative and quantitative report for user's evaluation, amendment and authorization
MSKai does not serve as a diagnostic device by providing or recommending any type of medical diagnosis or treatment. MSKai simply provides users the ability to access objective and repeatable identification, segmentation, measurement and reported measurements of the Lumbar spine. The user is responsible for the indications of preferences and settings, confirming the software-generated measurements, and reviewing, confirming and approving draft reports based on their medical training.
The device is intended to be used only by physicians, radiologist, hospitals and other medical institutions. Only T2 weighted DICOM images of MRI acquired from lumbar spine exams of patients aged 18 and above are acceptable input. MSKai does not support DICOM images of patients that are pregnant, have post-operational complications, tumors, infections, or complex hardware.
## 6. Comparison of the Technological Characteristics with the Predicate Devices
In comparison to the Predicate Device and the Reference Devices, the Subject Device provides comparable outputs in terms of segmentation, measurement and reporting. A tabular high-level comparison of the Subject Device and the Predicate Device is provided in the table below.
{7}
| Comparison of Technology Characteristics with Predicate Device | | | |
| --- | --- | --- | --- |
| | Predicate Device | Subject Device | Remarks Discussion |
| | Columbo
(K220497) | MSKai | |
| Intended User | · Radiologist
· Neurosurgeons and Ortho/Spine Surgeons | · Radiologist
· Neurosurgeons and Ortho/Spine Surgeons | Largely Similar |
| Intended Patient Population | Not subject to restrictions other than above the age of 18, not pregnant, without postoperative complication, scoliosis, tumors, infections, fractures | Not subject to restrictions other than above the age of 18, not pregnant, without postoperative complication, tumors, infections, complex hardware | |
| Supported Body Part | Lumbar Spine | Lumbar Spine | · Similar to Predicate Device and Reference Device 2
· Different from Reference Device 1 |
| Segmentation | Yes, Segmentation and Quantitative Analysis | Yes, Segmentation and Quantitative Analysis | Same |
| Measurement | Yes, Quantitative comparison of structure with normative data or user set thresholds | Yes, Quantitative comparison of structure with normative data. | Same |
| Reporting | Yes, Exploration of results with the findings for further reporting. | Yes, Exploration of results with the findings for further reporting. | Same. Reports radiologists of physicians evaluate, authorized and modify reports |
| SaMD | Yes | Yes | Same |
| Algorithm | Deep Convolutional Image to Image Neural Network | Mask Region-based Convolutional Neural Network | Similar |
| Supported Modality | MR | MR | Same as Predicate device, similar to reference devices. |
{8}
| Comparison of Technology Characteristics with Reference Device (1) | | | |
| --- | --- | --- | --- |
| | Predicate Device | Subject Device | Remarks Discussion |
| | Al-Rad Companion Brain MR
(K193290) | MSKai | |
| Intended User | · Radiologist | · Radiologist
· Neurosurgeons and Ortho/Spine Surgeons | Largely Similar |
| Intended Patient Population | Patient population above the age of 22 years old. | Not subject to restrictions other than above the age of 18, not pregnant, without postoperative complication, tumors, infections, complex hardware | Largely Similar |
| Supported Body Part | Brain | Lumbar Spine | · Similar to Predicate Device and Reference Device 2
· Different from Reference Device 1 |
| Segmentation | Yes, Segmentation and Quantitative Analysis | Yes, Segmentation and Quantitative Analysis | Same |
| Measurement | Yes, Quantitative comparison of structure with normative data or user set thresholds. | Yes, Quantitative comparison of structure with normative data. | Same. |
| Reporting | Yes, Exploration of results with the findings for further reporting. | Yes, Exploration of results with the findings for further reporting. | Same. Reports radiologists of physicians evaluate, authorized and modify reports |
| SaMD | Yes | Yes | Same |
| Algorithm | Unclear | Mask Region-based Convolutional Neural Network | Similar |
| Supported Modality | CT | MR | Same as Predicate device, similar to reference devices. |
{9}
| Comparison of Technology Characteristics with Reference Device (2) | | | |
| --- | --- | --- | --- |
| | Predicate Device | Subject Device | Remarks Discussion |
| | Al-Rad Companion Musculoskeletal (K193267) | MSKai | |
| Intended User | ·Radiologist
·Primary Care Physicians
·Specialty/Urgent Care Physicians | ·Radiologist
·Neurosurgeons and Ortho/Spine Surgeons | Largely Similar |
| Intended Patient Population | Not subject to restrictions other than above the age of 22. | Not subject to restrictions other than above the age of 18, not pregnant, without postoperative complication, tumors, infections, complex hardware | Largely Similar |
| Supported Body Part | Thorax and Thoracic Spine | Lumbar Spine | ·Similar to Predicate Device and Reference Device 2
·Different from Reference Device 1 |
| Segmentation | Yes, Segmentation and Quantitative Analysis | Yes, Segmentation and Quantitative Analysis | Same |
| Measurement | Yes, Quantitative comparison of structure with normative data or user set thresholds | Yes, Quantitative comparison of structure with normative data. | Same. |
| Reporting | Yes, Exploration of results with the findings for further reporting. | Yes, Exploration of results with the findings for further reporting. | Same. Reports radiologists of physicians evaluate, authorized and modify reports |
| SaMD | Yes | Yes | Same |
| Algorithm | 3D Image to Image Network | Mask Region-based Convolutional Neural Network | Similar |
| Supported Modality | CT | MR | Same as Predicate device, similar to reference devices. |
{10}
The Subject Device is substantially equivalent in comparison to the Predicate Device. The information regarding the Subject Device does not raise new questions about safety and effectiveness and demonstrates that MSKai is at least as safe and effective as the legally marketed devices.
## 7. Performance Data
7.1. Biocompatibility Testing
Not applicable.
## 7.2. Electrical Safety and Electromagnetic Compatibility (EMC)
Not applicable.
## 7.3. Animal Study
Not applicable.
## 7.4. Voluntary Conformance Standards performance testing demonstrated that MSkai complies with the following voluntary FDA recognized Consensus Standards listed in the table below.
| Recognition # | Standard |
| --- | --- |
| 13-79 | IEC 62304:2006/AMD 1:2015 Medical device software — Software life cycle processes — Amendment 1 |
| 5-125 | ISO 14971:2019 Medical devices — Application of risk management to medical devices |
| 5-129 | IEC 62366-1:2015+AMD1:2020 Medical devices — Part 1: Application of usability engineering to medical devices |
| 5-117 | ISO 15223-1:2016 Medical devices — Symbols to be used with medical device labels, labelling and information to be supplied — Part 1: General requirements |
| 12-300 | NEMA PS 3.1 - 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set |
## 7.5 Non Clinical Tests
MSKai has performed software design verification testing and a standalone performance assessment study, in accordance with the FDA guidance, General Principles of Software Validation; Final Guidance for Industry and FDA Staff, issued on January 11, 2002. All software requirements and risk analysis have been successfully verified and traced. The performance data demonstrates continued conformance with special controls for medical devices containing software.
Software documentation for Basic Documentation level, per FDA Guidance for Industry and Food and Drug Administration Staff, Content of Premarket Submissions for Device Software Functions, issued on June 14, 2023, were provided. Remedy Logic conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient, per FDA Guidance for Industry and Food and Drug Administration Staff, Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued on September 27, 2023, as well as FDA Guidance for Industry and Food and Drug Administration Staff, Postmarket Management of
{11}
Cybersecurity in Medical Devices, issued on December 28, 2016. The vulnerability assessment and penetration testing demonstrated satisfactory security performance.
The nonclinical test data demonstrated conformance with special controls and substantial equivalence to predicate devices' performance.
## Standalone Software Performance Study
To validate the MSKai software from a clinical perspective, a clinical data based standalone software performance assessment study was conducted in the U.S. Three blind independent data sets were used to train, test and measure within the model. Each data set was independent with no overlap of images across datasets.
- Ground Truth dataset: 255 patient images
- Pass/Fail Criteria dataset: 575 patient images
- Testing dataset: 238 patient images
The standalone software performance assessment study of MSKai included 238 MR image studies for 238 patients of different ages and racial groups, collected from five sites across the U.S (Table 7). The standalone software performance assessment study compared the MSKai software outputs without any editing by a radiologist to the ground truth defined by 3 neurosurgeons, 1 interventional radiologist, and 1 PhD in Biomechanics on segmentations and measurements. Measurement analysis was performed by 4 separate and independent experts (2 neurosurgeons.2 radiologist) Additionally, a confusion matrix analysis was conducted for pass/fail analysis of anatomical segmentation. Inter-ratter reliabilities were conducted in the experts who performance the training/testing measurements.
{12}
Table 7
Participant Demographics
| | Number of Subjects | Percent Total |
| --- | --- | --- |
| Total Number of Subjects | 238 | 100.0% |
| Gender-Male | 125 | 52.5% |
| Gender-Female | 113 | 47.5% |
| Age-18 through 21 | 51 | 21.4% |
| Age-22 through 50 | 132 | 55.4% |
| Age-51 and above | 55 | 23.2% |
| Racial-Caucasian | 146 | 61.3% |
| Racial-Black/African American | 55 | 23.1% |
| Racial-Hispanic | 28 | 11.7% |
| Racial-American Indian | 3 | 1.2% |
| Racial-Other | 6 | 2.5% |
# Imaging Systems
The 238 MR studies were acquired on MRI imaging systems made by three (3) manufacturers.
Table 8
Image Systems
| Manufacturer | Number of MRI Exams Collected | Percent total |
| --- | --- | --- |
| GE (1.5 & 3.0T) | 116 | 48.7% |
| Philips (1.5 & 3.0T) | 80 | 33.6% |
| Siemens (1.5 & 3.0T) | 42 | 17.6% |
| Total | 238 | 100.0% |
# Ground Truth
To support model accuracy, error analysis is performed before and after applying data augmentation and pre-trained weights. Ground truth data, curated by five experts in a two-phase process, underpins model training. In order to have a blind tested model a dataset 255 images were utilized in the ground truth development with a second independent dataset of 238 patients being utilized for testing being measured by an independent group of 4 experts. A third dataset represented a pass/fail analysis for anatomy segmentation identification. This comprehensive approach supports MSKai's aim of accurate, reliable spine image analysis in a clinical setting.
# Acceptance Criteria and Study Results
Primary Endpoints: - For measurements, the maximum Mean Absolute Error (MAE) as defined as the upper limit of the $95\%$ confidence interval for MAE is below a
{13}
predetermined allowable error limit (MAE limit) for each measurement listed below. Table 10.
Secondary Endpoints: - For segmentations of anatomical structures: the minimum Mean Dice Coefficient, defined as the lower limit of the $95\%$ confidence interval for MDC, is above a predetermined allowable limit (MDC Limit) for each segmentation listed below. Table 9a.
Table 9a Mean Dice Coefficient Results
| Anatomy Segmentation | View | Mean Dice Coefficient (MDC) | 95% Confidence Interval (CI) | MDCLimit |
| --- | --- | --- | --- | --- |
| Vertebral Body (L1) | Sagittal | 0.968 | 0.92-0.98 | 0.8 |
| Vertebral Body (L2) | Sagittal | 0.977 | 0.93-0.98 | 0.8 |
| Vertebral Body (L3) | Sagittal | 0.981 | 0.94-0.99 | 0.8 |
| Vertebral Body (L4) | Sagittal | 0.963 | 0.92-0.98 | 0.8 |
| Vertebral Body (L5) | Sagittal | 0.985 | 0.91-0.98 | 0.8 |
| Vertebral Body (S1) | Sagittal | 0.945 | 0.93-0.99 | 0.8 |
| L5/S1 Disc | Sagittal | 0.993 | 0.91-0.99 | 0.8 |
| L4/L5 Disc | Sagittal | 0.991 | 0.93-0.99 | 0.8 |
| L3/L4 Disc | Sagittal | 0.992 | 0.93-0.99 | 0.8 |
| L2/L3 Disc | Sagittal | 0.989 | 0.91-0.99 | 0.8 |
| L1/L2 Disc | Sagittal | 0.986 | 0.94-0.99 | 0.8 |
| Cord Canal | Sagittal | 0.983 | 0.93-0.99 | 0.8 |
| Axial Disc | Axial | 0.984 | 0.89-0.97 | 0.8 |
| Vertebral Body | Axial | 0.991 | 0.93-0.99 | 0.8 |
| Dural Sac | Axial | 0.978 | 0.90-0.98 | 0.8 |
| Nerve Root | Axial | 0.952 | 0.90-0.95 | 0.8 |
| Posterior Arch | Axial | 0.911 | 0.90-0.96 | 0.8 |
{14}
Table 10 Maximum Mean Absolute Error Results
| Structural Measurements | View | Mean Absolute Error (MAE) | 95% Confidence interval | MAELimit |
| --- | --- | --- | --- | --- |
| Protruding Disc Material (L5/S1) | Sagittal | 1.19mm | 1.11 - 1.68mm | 2mm |
| Protruding Disc Material (L4/L5) | Sagittal | 1.22mm | 1.12 - 1.71mm | 2mm |
| Protruding Disc Material (L3/L4) | Sagittal | 1.23mm | 1.14 - 1.65mm | 2mm |
| Protruding Disc Material (L2/L3) | Sagittal | 1.19mm | 1.07 - 1.61mm | 2mm |
| Protruding Disc Material (L1/L2) | Sagittal | 1.21mm | 1.11 - 1.63mm | 2mm |
| Intervertebral Angle (L5/S1) | Sagittal | 2.6° | 1.58 - 2.45° | 6° |
| Intervertebral Angle (L4/L5) | Sagittal | 2.7° | 1.61 - 2.59° | 6° |
| Intervertebral Angle (L3/L4) | Sagittal | 2.7° | 1.57 - 2.54° | 6° |
| Intervertebral Angle (L2/L3) | Sagittal | 2.9° | 1.64 - 2.62° | 6° |
| Intervertebral Angle (L1/L2) | Sagittal | 2.4° | 1.66 - 2.48° | 6° |
| Vertebral Body Height (Anterior) (L1) | Sagittal | 0.66mm | 0.62 - 0.91mm | 2mm |
| Vertebral Body Height (Anterior) (L2) | Sagittal | 0.68mm | 0.61 - 0.88mm | 2mm |
| Vertebral Body Height (Anterior) (L3) | Sagittal | 0.69mm | 0.61 - 0.93mm | 2mm |
| Vertebral Body Height (Anterior) (L4) | Sagittal | 0.64mm | 0.58 - 0.91mm | 2mm |
{15}
| Vertebral Body Height (Anterior) (L5) | Sagittal | 0.67mm | 0.61 - 0.91mm | 2mm |
| --- | --- | --- | --- | --- |
| Vertebral Body Height (Midline) (L1) | Sagittal | 0.94mm | 0.62 - 0.87mm | 2mm |
| Vertebral Body Height (Midline) (L2) | Sagittal | 0.93mm | 0.54 - 1.03mm | 2mm |
| Vertebral Body Height (Midline) (L3) | Sagittal | 0.96mm | 0.61 - 0.1.01mm | 2mm |
| Vertebral Body Height (Midline) (L4) | Sagittal | 0.97mm | 0.57 - 1.13mm | 2mm |
| Vertebral Body Height (Midline) (L5) | Sagittal | 0.94mm | 0.57 - 0.99mm | 2mm |
| Vertebral Body Height (Posterior) (L1) | Sagittal | 0.92mm | 0.67 - 0.99mm | 2mm |
| Vertebral Body Height (Posterior) (L2) | Sagittal | 0.93mm | 0.61 - 1.01mm | 2mm |
| Vertebral Body Height (Posterior) (L3) | Sagittal | 0.91mm | 0.68 - 0.99mm | 2mm |
| Vertebral Body Height (Posterior) (L4) | Sagittal | 0.92mm | 0.71 - 1.06mm | 2mm |
| Vertebral Body Height (Posterior) (L5) | Sagittal | 0.93mm | 0.68 - 1.09mm | 2mm |
| Disc Height (Anterior) (L5/S1) | Sagittal | 0.91mm | 0.67 - 0.99mm | 2mm |
| Disc Height (Anterior) (L4/L5) | Sagittal | 0.90mm | 0.57 - 1.06mm | 2mm |
| Disc Height (Anterior) (L3/L4) | Sagittal | 0.87mm | 0.62 - 1.03mm | 2mm |
{16}
| Disc Height (Anterior) (L2/L3) | Sagittal | 0.89mm | 0.78 - 1.06mm | 2mm |
| --- | --- | --- | --- | --- |
| Disc Height (Anterior) (L1/L2) | Sagittal | 0.93mm | 0.71 - 1.23mm | 2mm |
| Disc Height (Midline) (L5/S1) | Sagittal | 0.93mm | 0.73 - 1.12mm | 2mm |
| Disc Height (Midline) (L4/L5) | Sagittal | 0.90mm | 0.68 - 1.01mm | 2mm |
| Disc Height (Midline) (L3/L4) | Sagittal | 0.89mm | 0.71 - 1.13mm | 2mm |
| Disc Height (Midline) (L2/L3) | Sagittal | 0.91mm | 0.64 - 1.03mm | 2mm |
| Disc Height (Midline) (L1/L2) | Sagittal | 0.92mm | 0.69 - 1.11mm | 2mm |
| Disc Height (Posterior) (L5/S1) | Sagittal | 0.87mm | 0.58 - 1.03mm | 2mm |
| Disc Height (Posterior) (L4/L5) | Sagittal | 0.93mm | 0.67 - 0.99mm | 2mm |
| Disc Height (Posterior) (L3/L4) | Sagittal | 0.87mm | 0.66 - 1.07mm | 2mm |
| Disc Height (Posterior) (L2/L3) | Sagittal | 0.93mm | 0.72 - 1.21mm | 2mm |
| Disc Height (Posterior) (L1/L2) | Sagittal | 0.89mm | 0.58 - 0.91mm | 2mm |
| Anterio-Lithesis (L5/S1) | Sagittal | 1.04mm | 0.81 - 1.43mm | 2mm |
| Anterio-Lithesis (L4/L5) | Sagittal | 1.02mm | 0.77 - 1.52mm | 2mm |
| Anterio-Lithesis (L3/L4) | Sagittal | 1.05mm | 0.88 - 1.61mm | 2mm |
| Anterio-Lithesis(L2/L3) | Sagittal | 1.07mm | 0.84 - 1.43mm | 2mm |
| Anterio-Lithesis (L1/L2) | Sagittal | 1.02mm | 0.79 - 1.33mm | 2mm |
| Retro-Lithesis (L5/S1) | Sagittal | 1.07mm | 0.82 - 1.51mm | 2mm |
{17}
| Retro-Lithesis (L4/L5) | Sagittal | 1.049mm | 0.78 - 1.42mm | 2mm |
| --- | --- | --- | --- | --- |
| Retro-Lithesis (L3/L4) | Sagittal | 1.01mm | 0.81 - 1.29mm | 2mm |
| Retro-Lithesis (L2/L3) | Sagittal | 1.05mm | 0.77 - 1.34mm | 2mm |
| Retro-Lithesis (L1/L2) | Sagittal | 1.08mm | 0.83 - 1.27mm | 2mm |
| Lordotic Angle | Sagittal | 2.99° | 2.01 - 3.62° | 6° |
| Protruding Disc Material | Axial | 0.97 mm | 0.72 - 1.42mm | 2mm |
| Dural Sac Diameter | Axial | 1.3 mm | 0.87 - 1.39mm | 2mm |
The MSKai software was shown to produce segmentations and measurements accurate to within a prospectively defined margin of error around the Ground Truth. This accuracy was preserved for all critical subgroups, including MRI scanner manufacturer, patient age, gender and race.
## Training, Testing and Validation Data Independence:
The MSKai software machine learning algorithm training and testing data used during the algorithm development, as well as validation data used in the U.S. standalone software performance assessment study were all independent data sets.
## 7.6 Clinical Validation Study
No human clinical study was conducted to support the pre-market clearance.
## 8.0 Conclusions
The MSKai software is as safe and effective as the predicate device. The subject device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences between subject and predicate device in indications do not alter the intended use.
The software verification and validation testing data, including the standalone software performance assessment study data, supports the safety of the devices and demonstrates that the MSKai software performs as intended in the specified use conditions.
Therefore, the MSKai software demonstrates substantial equivalence to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.