K193290 · Siemens Medical Solutions USA, Inc. · LLZ · Jun 17, 2020 · Radiology
Device Facts
Record ID
K193290
Device Name
AI-Rad Companion Brain MR
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Brain structure segmentation and volumetric analysis
—
—
—
—
—
—
—
Indications for Use
AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing and evaluating MR brain images. AI-Rad Companion Brain MR provides the following functionalities: - Automatic segmentation and quantitative analysis of individual brain structures - Quantitative comparison of each brain structure with normative data from a healthy population - Presentation of results for reporting that includes all numerical values as well as visualization of these results
Device Story
AI-Rad Companion Brain MR is post-processing software for MR brain images; inputs are T1 MPRAGE datasets acquired during standard head MR. Device performs automatic segmentation and quantitative analysis of 30 brain structures (e.g., hippocampus, caudate, frontal grey matter); calculates absolute and normalized volumes relative to total intracranial volume. Normalized values are compared against age-matched normative data from healthy subjects. Output includes numerical values and visualization of results (including a customizable deviation map) presented for radiologist reporting. Used in clinical settings by radiologists/clinicians; results are archived in PACS. Enhancements include deployment via AI-Rad Companion Engine platform in a cloud-based environment. Device assists clinicians in evaluating volumetric properties, potentially aiding in clinical decision-making by providing quantitative assessment of brain structures.
Clinical Evidence
No clinical tests or animal testing were conducted. Substantial equivalence is supported by software verification, system integration testing, and usability validation. Bench testing confirmed compliance with FDA-recognized consensus standards (IEC 62366-1, ISO 14971, AAMI/ANSI/IEC 62304, DICOM PS 3.1-3.20).
Technological Characteristics
Software-based post-processing tool. Operates on T1 MPRAGE MR datasets. Features automatic segmentation and volumetric quantification algorithms. Deployed via cloud-based AI-Rad Companion Engine. Complies with DICOM standards for interoperability. Software life cycle managed per IEC 62304. Cybersecurity controls implemented per FDA guidance. Moderate level of concern software.
Indications for Use
Indicated for clinicians to assist in viewing, analyzing, and evaluating MR brain images. Intended for use by healthcare professionals familiar with MR image post-processing.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
June 17, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions USA, Inc. % Ms. Lauren Bentley Senior Manager, Regulatory Affairs 40 Liberty Blvd., Mail Code 65-3 MALVERN PA 19355
Re: K193290
Trade/Device Name: AI-Rad Companion Brain MR Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: May 15, 2020 Received: May 19, 2020
Dear Ms. Bentley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
Device Name AI-Rad Companion Brain MR
Indications for Use (Describe)
AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing and evaluating MR brain images.
AI-Rad Companion Brain MR provides the following functionalities:
- Automatic segmentation and quantitative analysis of individual brain structures
- Quantitative comparison of each brain structure with normative data from a healthy population
- Presentation of results for reporting that includes all numerical values as well as visualization of these results
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words are a series of orange dots arranged in a pattern.
# 510(k) SUMMARY FOR AI-RAD COMPANION BRAIN MR
Submitted by: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355 Date Prepared: November 25, 2019
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Devices Act of 1990 and 21 CFR §807.92.
#### 1. Submitter
| Importer/Distributor | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Malvern, PA 19355<br>Mail Code: 65-1A<br>Registration Number: 2240869 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturing Site | Siemens Healthcare GmbH<br>Henkestrasse 127<br>Erlangen, Germany 91052<br>Registration Number: 3002808157 |
#### 2. Contact Person
Lauren Bentley Senior Manager, Regulatory Affairs Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Mail Code: 65-3 Malvern, PA 19335 Phone: +1 (610) 241 - 6736 Email: lauren.bentley@siemens-healthineers.com
# 3. Device Name and Classification
| Product Name: | AI-Rad Companion Brain MR |
|----------------------|--------------------------------------------|
| Trade Name: | AI-Rad Companion Brain MR |
| Classification Name: | Picture Archiving and Communication System |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots arranged in a circular pattern.
| Classification Panel: | Radiology |
|-------------------------|------------------|
| CFR Section: | 21 CFR §892.2050 |
| Secondary CFR Section: | 21 CFR §892.1000 |
| Device Class: | Class II |
| Product Code: | LLZ |
| Secondary Product Code: | LNH |
#### 4. Predicate Device
| Product Name: | syngo.MR Applications |
|-------------------------|--------------------------------------|
| Propriety Trade Name: | syngo.MR Applications |
| 510(k) Number: | K182904 |
| Clearance Date: | July 5, 2019 |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.2050 |
| Secondary CFR Section: | 21 CFR §892.1000 |
| Device Class: | Class II |
| Primary Product Code: | LLZ |
| Secondary Product Code: | LNH |
| Recall Information: | N/A |
# 5. Intended Use
AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing and evaluating MR brain images.
AI-Rad Companion Brain MR provides the following functionalities:
- Automatic segmentation and quantitative analysis of individual brain structures ●
- Quantitative comparison of each brain structure with normative data from a healthy ● population
- Presentation of results for reporting that includes all numerical values as well as visualization of these results.
### 6. Device Description
AI-Rad Companion Brain MR is an enhancement to the predicate, syngo.MR Application (K182904). Just as in the predicate, AI-Rad Companion Brain MR addresses the automatic quantification and visual assessment of the volumetric properties of various brain structures based on T1 MPRAGE datasets. These datasets are acquired as part of a typical head MR acquisition. The results are directly archived in PACS as this is the standard location for reading by radiologist. From a predefined list of 30 structures (e.g. Hippocampus, Caudate, Left Frontal Grey Matter, etc.), volumetric properties are calculated as absolute and normalized volumes with respect to the total intercranial volume. The normalized values for a given patient are compared
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is an orange graphic of several dots arranged in a circular pattern.
against age-matched mean and standard deviations obtained from a population of healthy reference subjects.
As an update to the previously cleared device, the following modifications have been made:
- 1. Modified Intended Use Statement
- 2. Addition of a customizable deviation map
- 3. Architectural enhancement for the clinical extension to be deployed with the AI-Rad Companion Engine platform in a cloud-based environment
### 7. Technological Characteristics
The subject device, AI-Rad Companion Brain MR is substantially equivalent with regards to software, programming language, operating system, performance and technology (including algorithms). AI-Rad Companion Brain MR offers enhancements and improvements to the existing predicate device, syngo.MR Applications: Brain Morphometry (K182904). While these enhancements offer additional visualization capabilities, compared to the predicate device, the conclusions from all verification and validation data suggest that these modifications do not adversely affect the safety and effectiveness of the predicate device.
#### 8. Nonclinical Tests
Non-clinical tests were conducted to test the functionality of AI-Rad Companion Brain MR. Software validation and bench testing have been conducted to assess the performance claims as well as the claim of substantial equivalence to the predicate device.
AI-Rad Companion has been tested to meet the requirements of conformity to multiple industry standards. Non-clinical performance testing demonstrates that AI-Rad Companion Brain MR complies with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005) as well as with the following voluntary FDA recognized Consensus Standards listed in Table 1 below.
| Recognition<br>Number | Product Area | Title of Standard | Reference<br>Number and<br>Date | Standards<br>Development<br>Organization |
|-----------------------|--------------------------|------------------------------------------------------------------------------------------------------------------------|---------------------------------|------------------------------------------|
| 5-114 | General | Medical Devices -<br>Application of usability<br>engineering to medical<br>devices [including<br>Corrigendum 1 (2016)] | 62366-1:<br>2015-02 | IEC |
| 5-40 | General | Medical Devices -<br>application of risk<br>management to medical<br>devices | 14971:2007 | ISO |
| 13-79 | Software/<br>Informatics | Medical device software -<br>software life cycle | 62304:<br>2006/A1:2016 | AAMI<br>ANSI<br>IEC |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Siemens Healthineers logo. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots arranged in a circular pattern.
| Recognition<br>Number | Product Area | Title of Standard | Reference<br>Number and<br>Date | Standards<br>Development<br>Organization |
|-----------------------|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|------------------------------------------|
| | | processes [Including<br>Amendment 1 (2016)] | | |
| 12-300 | Radiology | Digital Imaging and<br>Communications in<br>Medicine (DICOM) Set | PS 3.1 – 3.20<br>(2016) | NEMA |
| 12-261 | Radiology | Information Technology –<br>Digital Compression and<br>coding of continuous -tone<br>still images: Requirements<br>and Guidelines [including:<br>Technical Corrigendum<br>1(2005)] | 10918-1<br>1994-02-15 | ISO<br>IEC |
Table 1: Voluntary Conformance Standards
#### Verification and Validation
Software documentation for a Moderate Level of Concern software, per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005, is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the subject device during product development.
Software "bench" testing in the form of Unit. System and Integration tests were performed to evaluate the performance and functionality of the new features and software updates. All testable requirements in the Requirement Specifications and the Risk Analysis have been successfully verified and traced in accordance with the Siemens Healthineers DH product development (lifecvcle) process. Human factor usability validation is addressed in system testing and usability validation test records. Software verification and regression testing have been performed successfully to meet their previously determined acceptance criteria as stated in the test plans.
Siemens Healthineers adheres to the cybersecurity requirements as defined the FDA Guidance "Content of Premarket Submissions for Management for Cybersecurity in Medical Devices," issued October 2, 2014 by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.
### 9. Clinical Tests
No clinical tests were conducted to test the performance and functionality of the modifications introduced within AI-Rad Companion Brain MR. Verification and validation of the enhancements and improvements have been performed and these modifications have been validated for their intended use. The data from these activities were used to support the subject device and the substantial equivalence argument.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the word "Healthineers" is a graphic of orange dots.
No animal testing has been performed on the subject device.
### 10. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to ensure safe and effective use of the device.
Risk management is ensured via ISO 14971:2007 compliance to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the development of the product. These risks are controlled via measures realized during software development, testing and product labeling.
Furthermore, the device is intended for healthcare professionals familiar with the post processing of magnetic resonance images.
# 11. Substantial Equivalence and Conclusion
AI-Rad Companion Brain MR is substantially equivalent to the follow predicate device (Table 2):
| Predicate Device | FDA Clearance<br>Number | FDA Clearance<br>Date | Main Product Code |
|-----------------------|-------------------------|-----------------------|-------------------|
| syngo.MR Applications | K182904 | July 5, 2019 | LLZ |
Table 2: Predicate device for AI-Rad Companion Brain MR
AI-Rad Companion Brain MR has the same intended use and technical characteristics compared to the predicate device, syngo.MR Applications (K182904), with respect to the software features, functionalities and core algorithms. The enhancements and improvements provided in AI-Rad Companion Brain MR increase the usability and reduce the complexity of the imaging workflow for the clinical user. The conclusions from all verification data suggest that these enhancements are equivalent with respect to safety and effectiveness of the predicate device. These modifications do not change the intended use of the product. Siemens is of the opinion that AI-Rad Companion Brain MR is substantially equivalent to the currently marketed device, syngo.MR. Applications (K182904).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.