Dataset of 1200 subjects, consisting of Alzheimer's patients (AD), mild cognitive impaired patients (MCI) and healthy controls (HC).
—
—
—
Indications for Use
syngo.MR Applications is a syngo based post-acquisition image processing software for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra.
Device Story
syngo.MR Applications is post-acquisition image processing software for MR, MR-PET, CT, PET, and CT-PET images and MR spectra. New feature 'syngo.MR Brain Morphometry' extends MR Neurology workflow; automatically calculates volume properties of brain structures using MPRAGE datasets. Software operates on syngo-based platform; used by clinicians for viewing and evaluating diagnostic images. Output provides volumetric data to assist in neurological assessment; aids clinical decision-making by quantifying brain structure volumes. Benefits include improved workflow efficiency and reduced complexity in imaging procedures.
Post-processing software; conforms to IEC 62304 (software life cycle), ISO 14971 (risk management), IEC 62366-1 (usability), and NEMA/DICOM standards. Operates on syngo-based platform. Features automated volumetric calculation of brain structures from MPRAGE datasets.
Indications for Use
Indicated for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra in patients undergoing diagnostic imaging.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.
Siemens Medical Solutions USA, Inc. % Cordell L. Fields, Esq. Regulatory Affairs Specialist 40 Liberty Blvd., Mailcode 65-1A MALVERN PA 19355
Re: K182904
Trade/Device Name: syngo.MR Applications Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ, LNH Dated: June 5, 2019 Received: June 6, 2019
Dear Mr. Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
July 5, 2019
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented.
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
| 510(k) Number (if known) | K182904 |
| Device Name | syngo.MR Applications |
| Indications for Use (Describe) | syngo.MR Applications is a syngo based post-acquisition image processing software for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra. |
| Type of Use (Select one or both, as applicable) | <span> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </span> <span> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
### Siemens 510(k) Traditional Premarket Notification
{3}------------------------------------------------
## SIEMEN
# Section 5 510(k) Summary
This 510(k) summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
| Company: | Siemens Medical Solutions USA, Inc. |
|----------|-------------------------------------|
|----------|-------------------------------------|
40 Liberty Boulevard, 65-1A Malvern, PA 19355
Date Prepared: October 15, 2018
#### 1. General Information
## Importer/Distributor:
Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, Mail Code 65-1A Malvern, PA 19355, USA Establishment Registration Number: 2240869
## Manufacturing Sites:
Siemens Healthcare GmbH Henkestrasse 127 91052 Erlangen, Germany Establishment Registration Number: 3002808157
#### 2. Contact Person:
Cordell L. Fields, Esq. Regulatory Affairs Technical Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Mail Code 65-1A Malvern, PA 19355, USA Phone: (610) 448-6469 Fax: (610) 448-1787 E-mail: cordell.fields@siemens-healthineers.com
{4}------------------------------------------------
#### 3. Device Name and Classification:
| Device Name | syngo.MR Applications |
|-------------------------|------------------------------------------------------|
| Trade Name | syngo.MR Applications: syngo.MR Brain<br>Morphometry |
| Classification Name: | Picture Archiving and Communication System (PACS) |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR § 892.2050 |
| Device Class: | II |
| Primary Product Code: | LLZ |
| Secondary Product Code: | LNH |
### Legally Marketed Predicate Device: 4.
| Device Name | syngo.MR Applications SMRVB30A |
|-------------------------|---------------------------------------------------|
| 510(k) Number: | K180336, cleared April 19, 2018 |
| Classification Name: | Picture Archiving and Communication System (PACS) |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR § 892.2050 |
| Device Class: | II |
| Primary Product Code: | LLZ |
| Secondary Product Code: | LNH |
#### 5. Intended Use
The indications for use for the subject device are the same as the predicate device:
syngo.MR Applications is a syngo based post-acquisition image processing software for viewing, manipulating, evaluating, and analyzing MR, MR-PET, CT, PET, CT-PET images and MR spectra.
{5}------------------------------------------------
#### 6. Device Description:
The syngo.MR Applications are syngo based post-processing software/applications to be used for viewing and evaluating ' MR images provided by a maqnetic resonance diagnostic device and enabling structured evaluation of MR images. syngo.MR Brain Morphometry extends the MR Neurology workflow and offers a comprehensive package for the automatic calculation of the volume properties of different brain structures using MPRAGE datasets, which are typically acquired for a typical MR examination of the head.
With this premarket submission, the new functionality syngo.MR Brain Morphometry is introduced to extend the MR Neurology workflow that is a part of the formerly cleared medical device syngo.MR Applications (K180336).
#### 7. Technological Characteristics
syngo.MR Applications with SMRVB30A offers a new feature, syngo.MR Brain Morphometry, compared to the predicate device syngo.MR Applications SMRVB30A (K180336; cleared April 19, 2018). While this is a feature that offers additional image viewing and evalution capabilities than the predicate device, the conclusions from all verification and validation data suggest that the feature bears an equivalent safety and performance profile to the predicate device.
The subject device is substantially equivalent to the predicate device with regard to the software, hardware, operational environment, programming language, operating system and performance.
syngo.MR Brain Morphometry for syngo.MR Applications SMRVB30A conforms to the standard for software medical devices (IEC 62304:2006), IEC as well as NEMA standards.
#### 8. Nonclinical Tests
The following performance testing was conducted on the subject device:
- Software verification and validation testing was completed in accordance with i the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005)
The following performance testing was conducted on the subject device:
- Performance of the new feature functionality was demonstrated by quantifying accuracy, repeatability and reproducibility of brain structure volume estimations by syngo.MR Brain Morphometry on a dataset of 1200 subjects, consisting of Alzheimer's patients (AD), mild cognitive impaired patients (MCI) and healthy controls (HC). The accuracy of the volumetric results was validated by comparing the automated results to a reference. Repeatability and reproducibility studies were conducted to demonstrate
<sup>1</sup> While viewing (i.e. assessing) of images from other vendors is always possible; for advanced post-processing applications, some of the post-processing steps may depend on information contained in private DICOM tags; therefore, the evaluation and processing of images can´t be guaranteed for other vendors.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "SIEMENS" in a sans-serif font. The word is in a teal color. The background is white.
the robustness of volume measures estimated by syngo.MR Brain Morphometry with respect to both instrumental and patient noise within a single scanning session on one hand, and to significant changes to instrument, i.e., field strength and vendor as well as patient positioning, on the other hand. Acceptance criteria for performance tests were defined based on a literature review. In all validation experiments, syngo.MR Brain Morphometry passed the acceptance criteria. The correlation between the volumes estimated by syngo.MR Brain Morphometry and a reference device was 0.95 for grey matter, 0.80 for the hippocampus and 0.92 for white matter. The correlation of obtained volumes in the repeatability experiments was 0.96 for grey matter, hippocampus and white matter, 0.99 for the ventricular system. Volume correlation in the reproducibility experiments was 0.97 for grey matter, 0.94 for the hippocampus and 0.98 for white matter.
#### 9. Clinical Tests
No clinical tests were conducted to test the performance and functionality of the new feature syngo.MR Brain Morphometry feature.
Verification and validation and a performance evaluation of the syngo.MR Applications SMR VB30A with the new feature syngo.MR Brain Morphometry has been performed and the new functionality has been validated for its intended use. The data from these activities were used to support the subject device and the substantial equivalence argument. Clinical publications are referenced to provide further information on the use and functionality of the feature.
No animal testing has been performed on the subject device and its new feature.
#### 10. Safety and Effectiveness
syngo.MR Brain Morphometry extends the MR Neurology workflow which was cleared with the medical device syngo.MR Applications with software SMRVB30A (K180336) and is covered by the Intended Use of the synqo.MR Applications.
conclusions from the non-clinical data suggest that the additional The functionality bears an equivalent safety and performance profile to the predicate device (Table 2). Additionally, the indications for use remains the same.
Therefore, this workflow enhancement with syngo.MR Brain Morphometry is considered to be substantially equivalent to the predicate device syngo.MR Applications SMRVB30A.
synqo.MR Applications: synqo.MR Brain Morphometry conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document as stated in the following Table 1.
| Recognition<br>Number | Product Area | Title of Standard | Reference Number<br>and date | Standards<br>Development<br>Organization |
|-----------------------|--------------|-------------------|------------------------------|------------------------------------------|
|-----------------------|--------------|-------------------|------------------------------|------------------------------------------|
### Siemens 510(k) Traditional Premarket Notification
{7}------------------------------------------------
| Recognition<br>Number | Product Area | Title of Standard | Reference Number<br>and date | Standards<br>Development<br>Organization |
|-----------------------|--------------|------------------------------------------------------------------------------------------|----------------------------------------|------------------------------------------|
| 5-40 | General | Medical devices - Application of<br>risk management to medical<br>devices | 14971 Second<br>edition 2007-03-<br>01 | ISO |
| 5-96 | General | Medical devices - Application of<br>usability engineering to medical<br>devices | 62366-1:2015 | AAMI ANSI<br>IEC |
| 13-32 | Software | Medical device software -<br>Software life cycle processes | 62304:2006 | AAMI ANSI<br>IEC |
| 12-300 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology | PS 3.1 - 3.20<br>(2016) | NEMA |
Table 1: Standard requirements for syngo.MR Applications: syngo.MR Brain Morphometry
#### Substantial Equivalence and Conclusion 11.
The syngo.MR Applications with software SMRVB30A including the new feature syngo.MR Brain Morphometry is substantially equivalent to the following predicate device (Table 2).
| Predicate Device | FDA Clearance Number | FDA Clearance Date | Main Product Code |
|-----------------------------------|----------------------|--------------------|-------------------|
| syngo.MR<br>SMRVB30A Applications | K180336 | April 19, 2018 | LLZ, LNH |
Table 2: Predicate device for syngo.MR Brain Morphometry for syngo.MR Applications SMRVB30A
The subject device, syngo.MR Applications SMRVB30A, has been modified to include the new feature syngo.MR Brain Morphometry. No other modifications have been made.
{8}------------------------------------------------
## SIEMENS
syngo.MR Applications: syngo.MR Brain Morphometry has the same intended use and one different technological characteristic compared to the predicate device, syngo.MR Applications with SMRVB30A (K180336), with respect to the software features and functionalities. While the new feature varies with respect to the predicate device with expanded post-processing capabilities for the automatic calculation of the volume properties of different brain structures using MPRAGE datasets, the conclusions from all verification and validation data suggest that the feature bears an equivalent safety and performance profile to the predicate device. The new feature offers the user additional possibilities for the image viewing and evaluation. The modification aims to improve user workflow and reduce the complexity of the imaging procedure and does not change the intended use. The completed performance testing supports the substantial equivalence of the subject device to the predicate device.
In summary, synqo.MR Applications: syngo.MR Brain Morphometry has the same functionalities as the predicate device and, based on the aforementioned information, does not introduce new issues of safety or effectiveness. Therefore, Siemens is of the opinion that syngo.MR Applications: syngo.MR Brain Morphometry is substantially equivalent to the currently marketed device synqo.MR Applications with SMRVB30A (K180336).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.