Retrospective, multicenter study using imaging equipment from Philips, GE, and Siemens.
3 (cardiologists)
Indications for Use
InVision Precision LVEF is used to process previously acquired transthoracic cardiac ultrasound images, store images, and manipulate and make measurements on images using an ultrasound device, personal computer, or a compatible DICOM-compliant PACS system to provide an automated estimation of LVEF. This measurement can be used to assist the clinician in a cardiac evaluation. InVision Precision is indicated for use in patients 22 years and older by sonographers and physicians evaluating cardiac ultrasound.
Device Story
SaMD for echocardiographic data analysis; processes DICOM transthoracic cardiac ultrasound images. Inputs: echocardiogram video clips. Processing: machine learning algorithm selects appropriate views; performs semi-automatic segmentation of left ventricle; calculates LVEF. Output: numerical LVEF estimate visualized as adjustable annotations within clinician's PACS interface. Used in clinical settings by sonographers/physicians. Clinician reviews, modifies segmentations, and finalizes LVEF measurement. Real-time updates provided during user modification. Benefits: assists clinicians in cardiac evaluation by providing automated, reviewable LVEF measurements.
Clinical Evidence
Retrospective, multicenter study using Philips, GE, and Siemens imaging data. Evaluated LVEF calculation accuracy against ground truth. Results: Biplane RMSD 6.06, A4C RMSD 6.17, A2C RMSD 7.12. Dice scores: A4C 0.89, A2C 0.90. Subgroup analysis performed for race, age, gender, BMI, and equipment. Device met all endpoints.
Technological Characteristics
Software as a Medical Device (SaMD). Machine learning-based algorithm using spatiotemporal pooling network architecture. Operates on DICOM-compliant echocardiographic video clips. Supports single-view or biplane EF calculation. Integrated into existing PACS workflow.
Indications for Use
Indicated for patients 22 years and older. Used by sonographers and physicians to process transthoracic cardiac ultrasound images and provide automated LVEF estimation to assist in cardiac evaluation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
InVision Medical Technology Corporation Golnaz Moeini Regulatory and Strategy Consultant 9702 Cisco St. Los Angeles, CA 90034
Re: K232331
April 25, 2024
Trade/Device Name: InVision Precision LVEF (LVEF) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: March 27, 2024 Received: March 27, 2024
Dear Golnaz Moeini:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K232331
Device Name InVision Precision LVEF (LVEF)
#### Indications for Use (Describe)
In Vision Precision LVEF is used to process previously acquired trans thoracic cardiac ultrasound images, and manipulate and make measurements on images using an ultrasound device, personal computer, or a compatible DICOM compliant PACS system to provide an automated estimation of LVEF. This measurement can be used to assist the clinician in a cardiac evaluation. In Vision Precision is indicated for use in patients 22 years and older by sonographers and physicians evaluating cardiac ultrasound.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for InVision Medical Technology. The logo features a stylized heart shape at the top, formed by interconnected lines in a teal color, resembling a circuit or network. Below the heart shape, the word "InVision" is written in a bold, blue font. Underneath "InVision", the words "Medical Technology" are written in a smaller, lighter font, also in blue.
# 510(k) Summary
### Correspondent Information
InVision Medical Technology Corporation % Golnaz Moeini MedEdge Consulting Email: golnaz.moeini@gmail.com Phone: (408) 504-3187 Date prepared: March 27, 2024
# Device
| Proprietary name: | InVision Precision LVEF (LVEF) |
|----------------------|------------------------------------------------------|
| Common name: | InVision Precision |
| Classification name: | Medical image management and processing system |
| Regulation: | 21 CFR 892.2050 |
| Regulatory class: | II |
| Product code: | QIH Automated Radiological Image Processing Software |
### Predicate Device
Caption Interpretation Automated Ejection Fraction, K210747.
# Device Description
InVision Precision LVEF is a software as a medical device (SaMD), manufactured by InVision Medical Technology Corporation, intended as an aid in diagnostic review and analysis of echocardiographic data, including the evaluation of left ventricular ejection fraction (LVEF) in cardiovascular ultrasound images in DICOM format.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for InVision Medical Technology. The logo features a stylized heart shape constructed from interconnected teal lines, suggesting a modern and technological approach to healthcare. Below the heart is the company name, "InVision," in a bold, sans-serif font, with the words "Medical Technology" in a smaller font size underneath.
The software interfaces with data files uploaded to a PACS by any ultrasound or data collection equipment. It selects a set of echocardiogram videos of the correct view and generates semi-automatic segmentations of the left ventricle using a machine learning algorithm to form the basis for the calculator of the LVEF output. The analysis results are visualized by the clinician's integrated image view application as adjustable annotations. The user has the option to modify the semi-automatic segmentations suggested by the software. The EF calculation is updated in real-time with the user's modification of the segmentation. A cardiologist can adjust the annotations and the downstream measurement of LVEF prior to finalization.
#### Intended Use/ Indications for Use
InVision Precision LVEF is used to process previously acquired transthoracic cardiac ultrasound images, store images, and manipulate and make measurements on images using an ultrasound device, personal computer, or a compatible DICOM-compliant PACS system to provide an automated estimation of LVEF. This measurement can be used to assist the clinician in a cardiac evaluation. InVision Precision is indicated for use in patients 22 years and older by sonographers and physicians evaluating cardiac ultrasound.
# Summary of Technological Characteristics
A comparison of technological similarities and differences between the subject and predicate devices is as follows:
| Feature/<br>Function | Subject Device:<br>InVision Precision | Predicate Device:<br>Caption Interpretation<br>Automated Ejection<br>Fraction Software<br>K210747 |
|--------------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------------|
| General Principles of Operation | | |
| Machine Learning-<br>Based Algorithm | Yes | Yes |
| Operates on DICOM<br>clips | Yes | Yes |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for InVision Medical Technology. The logo features a stylized heart shape made of teal lines, with a small circle connected to the lines on the left side. Below the heart shape, the word "InVision" is written in a bold, dark blue font. Underneath "InVision", the words "Medical Technology" are written in a smaller, lighter blue font.
| Feature/<br>Function | Subject Device:<br>InVision Precision | Predicate Device:<br>Caption Interpretation<br>Automated Ejection<br>Fraction Software<br>K210747 |
|--------------------------------------------------------------------------|-----------------------------------------------|---------------------------------------------------------------------------------------------------|
| Automation level | Semi automated | Fully automated |
| Automated View<br>Selection | Yes | Yes |
| Views for EF Calculation | Single view or Biplane | Single view or Biplane |
| Automated Ejection<br>Fraction Calculation | Yes | Yes |
| Ejection Fraction<br>reported | Numerical estimate | Numerical estimate |
| User Interface | Uses the interface of<br>approved PACS system | Includes own user<br>interface |
| Quantitative feedback to<br>enable clinician to<br>assess EF calculation | Adjustable annotation | Confidence Metric |
| EF result shown with<br>video clip | Yes, with adjustable<br>annotation | Yes |
| User confirmation/<br>rejection of result | Yes | Yes |
| Technological Characteristics | | |
| Network Architecture | Spatiotemporal Pooling | Simple Pooling |
Both subject and predicate devices have similar technological characteristics and principles of operation. Any differences do not raise new concerns of safety and efficacy.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for InVision Medical Technology. The logo features a stylized heart shape made of teal lines, with a small circle connected to one of the lines, resembling a circuit or connection point. Below the heart shape, the word "InVision" is written in a bold, dark blue font, and below that, the words "Medical Technology" are written in a smaller, lighter blue font.
### Performance Data
The InVision Precision software was developed and tested in accordance with InVision Medical Technology's Design Control processes and has been subjected to extensive safety and performance testing. Verification and validation testing was conducted to demonstrate the substantial equivalence of the subject device to the predicate. The primary success criterion was that the subject device would produce an ejection fraction number with a Root Mean Square Deviation below a set threshold as compared to the reference ground truth EF as well as Dice score above a set threshold compared to the consensus annotation of three cardiologists.
Extensive algorithm development and software verification and clinical validation testing, assessed the performance of the software's image video clip selection function and performance characteristics of the algorithm, including accuracy and overall functional performance. Images and cases used for verification and validation testing were separate and carefully segregated from training datasets. Non-clinical verification and validation test results established that the device meets its design requirements and intended use.
#### Machine Learning Performance Evaluation Summary
A retrospective, multicenter study was designed to evaluate the capability of the Precision machine learning model in calculating LVEF against ground truth. A variety of imaging equipment manufacturers included Philips, GE and Siemens were used for the input data. The average RMSD (Root Mean Square Deviation) for the device's biplane view was around 6.06, for A4C view was 6.17, for A2C view was 7.12. The Dice score for A4C segmentation was 0.89 and for A2C segmentation was 0.90. Data analysis of subgroups included race/ ethnicity, age, gender, BMI, imaging site and equipment, and ejection fraction values. The Precision device met all endpoints.
#### Conclusions
Performance testing demonstrated that the InVision Precision software performs as expected and in a manner that is substantially equivalent to the predicate device. The InVision Precision software has the same intended use, and similar indications for use, principles of operation and technological characteristics as the predicate device. The
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains a logo for InVision Medical Technology. The logo features a stylized, geometric heart shape in teal, positioned above the company name. The text "InVision" is written in a bold, dark blue font, and below it, "Medical Technology" is written in a smaller, lighter font.
differences do not raise new concerns of safety and effectiveness, therefore the InVision Precision software is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.