Rapid ASPECTS (v3)

K232156 · Ischemaview, Inc. · POK · Jan 19, 2024 · Radiology

Device Facts

Record IDK232156
Device NameRapid ASPECTS (v3)
ApplicantIschemaview, Inc.
Product CodePOK · Radiology
Decision DateJan 19, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2060
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
ASPECT ScoreRandom forest machine learning technique—82.8% percent agreement at the region and scan level——Standalone performance study: 88 scans. Reader improvement study: 102 scans.6 (typical readers)

Indications for Use

Rapid ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data. The Software automatically registers images and segments and analyzes ASPECTS Regions of Interest (ROIs). Rapid ASPECTS extracts image data for the ROI(s) to provide analysis and computer analytics based on morphological characteristics. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECT (Alberta Stroke Program Early CT) Score. Rapid ASPECTS is indicated for evaluation of adult patients presenting for diagnostic imaging workup, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. This device provides information that may be useful in the characterization of early ischemic brain tissue injury for ischemic stroke patient (typically < 24 hours since last known well) during image interpretation following the standard of care. Rapid ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment using the ASPECTS atlas definitions and atlas display including highlighted ROIs and numerical scoring. Rapid ASPECTS presents the original and annotated images for concurrent reads.

Device Story

Rapid ASPECTS (v3) is a CADx software module within the Rapid Platform, used by clinicians (neuroradiologists/physicians) to assist in assessing brain tissue abnormalities on non-contrast CT (NCCT) images. The device performs rigid registration/alignment of input images to an anatomical template, delineates pre-defined ASPECTS regions of interest (ROIs), and extracts morphological features. A random forest machine learning algorithm synthesizes these features into a numerical ASPECT score. The system outputs annotated images with highlighted ROIs and the calculated score for physician review. The clinician may adjust the score based on clinical judgment before finalizing the report in the PACS. The device is intended to provide standardized, consistent reads to aid in stroke workup, particularly for non-expert readers in community hospitals, potentially improving diagnostic accuracy and timeliness. It does not replace clinical judgment or primary interpretation.

Clinical Evidence

Clinical validation included a reader improvement study (N=102) and standalone performance testing (N=88). Standalone percent agreement with reference was 82.8%. Reader improvement study showed that using Rapid ASPECTS increased agreement with the reference standard by 0.02 on average (from 82% to 84% for all readers; 80.4% to 83.3% excluding experts). Supplemental assessment of 115 reads on confounder/mimic cases (e.g., abscess, hydrocephalus) showed minimal impact (2.6% change) on performance.

Technological Characteristics

Software as a Medical Device (SaMD) utilizing random forest machine learning. Operates on NCCT brain image data. Complies with ISO 14971:2019, IEC 62304:2015, IEC 62366:2015, and DICOM (NEMA PS 3.1-3.20) standards. Deployed via server or virtual system. Cybersecurity controls include vulnerability assessments, SBOMs, and penetration testing.

Indications for Use

Indicated for adult patients presenting for diagnostic imaging workup for evaluation of extent of disease (number of ASPECTS regions affected) in early ischemic brain tissue injury (typically < 24 hours since last known well). Contraindications: excessive patient motion, hemorrhagic transformation, hematoma, and very thin or no ventricles.

Regulatory Classification

Identification

A radiological computer-assisted diagnostic software for lesions suspicious of cancer is an image processing prescription device intended to aid in the characterization of lesions as suspicious for cancer identified on acquired medical images such as magnetic resonance, mammography, radiography, or computed tomography. The device characterizes lesions based on features or information extracted from the images and provides information about the lesion(s) to the user. Diagnostic and patient management decisions are made by the clinical user.

Special Controls

A radiological computer-assisted diagnostic (CADx) software for lesions suspicious for cancer must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, and algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will improve reader performance as intended. iii. Results from performance testing protocols that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, results, and cybersecurity). 2. Labeling must include: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended reading protocol. iii. A detailed description of the intended user and recommended user training. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations, including situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. Detailed instructions for use. viii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders (e.g., lesion and organ characteristics, disease stages, and imaging equipment).

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include: (i) A detailed description of the image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, and algorithm limitations. (ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will improve reader performance as intended. (iii) Results from performance testing protocols that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures ( *e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain sufficient numbers of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Standalone performance testing protocols and results of the device. (v) Appropriate software documentation ( *e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; and description of verification and validation activities including system level test protocol, pass/fail criteria, results, and cybersecurity).(2) Labeling must include: (i) A detailed description of the patient population for which the device is indicated for use. (ii) A detailed description of the intended reading protocol. (iii) A detailed description of the intended user and recommended user training. (iv) A detailed description of the device inputs and outputs. (v) A detailed description of compatible imaging hardware and imaging protocols. (vi) Warnings, precautions, and limitations, including situations in which the device may fail or may not operate at its expected performance level ( *e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Detailed instructions for use. (viii) A detailed summary of the performance testing, including: Test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders ( *e.g.,* lesion and organ characteristics, disease stages, and imaging equipment).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. iSchemaView, Inc. Jim Rosa SVP Regulatory and Quality 1120 Washington Ave. Suite 200 Golden, Colorado 80401 January 19, 2024 #### Re: K232156 Trade/Device Name: Rapid ASPECTS (v3) Regulation Number: 21 CFR 892.2060 Regulation Name: Radiological computer-assisted diagnostic software for lesions suspicious of cancer Regulatory Class: Class II Product Code: POK Dated: December 22, 2023 Received: December 26, 2023 Dear Jim Rosa: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Jessica Lamb Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT 8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K232156 Device Name Rapid ASPECTS (v3) Indications for Use (Describe) Rapid ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data. The Software automatically registers images and segments and analyzes ASPECTS Regions of Interest (ROIs). Rapid ASPECTS extracts image data for the ROI(s) to provide analysis and computer analytics based on morphological characteristics. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECT (Alberta Stroke Program Early CT) Score. Rapid ASPECTS is indicated for evaluation of adult patients presenting for diagnostic imaging workup, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. This device provides information that may be useful in the characterization of early ischemic brain tissue injury for ischemic stroke patient (typically < 24 hours since last known well) during image interpretation following the standard of care. Rapid ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment using the ASPECTS atlas definitions and atlas display including highlighted ROIs and numerical scoring. Rapid ASPECTS presents the original and annotated images for concurrent reads. Limitations: 1. The ASPECTS score should be only used for ischemic stroke patients following the standard of care. 2. Rapid ASPECTS is an adjunct tool and is not intended to replace a clinicians review of the original imaging or their clinical judgement. 3. Physicians should not use the CAD generated output as the primary interpretation without their concurrence. Contraindications/Exclusions/Cautions: - · Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. - · Hemorrhagic Transformation, Hematoma - · Very thin or no Ventricles Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {4}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ ### iSchemaView, Inc.'s Rapid ASPECTS (v3) (K232156) This document contains the 510(k) summary for the iSchemaView Rapid ASPECTS (v3). The content of this summary is based on the requirements of 21 CFR Section 807.92(c). #### Applicant Name and Address: | Name: | iSchemaView, Inc. | |-------------------|------------------------------------------------------------------| | Address: | 1120 Washington Ave<br>Ste. 200<br>Golden, CO 80401 | | Official Contact: | Jim Rosa<br>Phone: (303) 704-3374<br>Email: rosa@ischemaview.com | Summary Preparation Date: 9 January 2024 ### Device Name and Classification: | Trade Name: | Rapid ASPECTS (v3) | |--------------------------|---------------------| | Common Name: | CADx | | Classification: | II | | Product Code: | POK | | Regulation No: | 21 C.F.R. §892.2060 | | Classification<br>Panel: | Radiology Devices | ### Predicate Devices: The iSchemaView Rapid is claimed to be substantially equivalent to the following legally marketed predicate device: Rapid ASPECTS (K200760) ### Device Description: #### Rapid Platform The Rapid platform is Software as a Medical Device (SaMD), which provides for the visualization and study of changes in tissue and vasculature using digital images captured by diagnostic imaging systems including CT (Computed Tomography), CTA (CT Angiography), MRI (Magnetic Resonance Imaging) and MRA (MR Angiography) as an aid to physician diagnosis. Rapid can be installed on a customer's Server or it can be accessed online as a virtual system. It provides viewing, quantification, analysis, and reporting capabilities. The Rapid platform has multiple modules a clinician may elect to run and provide analysis for decision making. {6}------------------------------------------------ ### Rapid ASPECTS Module Rapid ASPECTS provides an automatic ASPECT Score based on the case input file for the physician. The score includes which ASPECT regions are identified based on regional imaging features derived from Non-Contrast Computed Tomography (NCCT) brain image data. The results are generated based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines and provided to the clinician for review and verification. At the discretion of the clinician, the scores may be adjusted based on other clinical factors the clinician may integrate though the Rapid Platform Interface. The ASPECTS software module processing pipeline performs four major tasks: - . Orientation and spatial normalization of the input imaging data (rigid registration/alignment with anatomical template). - . Delineation of pre-defined regions of interest on the normalized input data and computing numerical values characterizing underlying voxel values within those regions. - Identification and highlighting previous/old stroke areas along with areas of early ischemic change; and - Labeling of these delineated regions and providing a summary score ● reflecting the number of regions with early ischemic change as per ASPECTS guidelines. Subsequently. the system notifies the physician of the availability of the ASPECT Score with an overlayed atlas. The ASPECTS information is then available for the physician to review and edit prior to sending the data to a PACS or Workstation. The final summary score together with the regions selected and underlying voxel values are then sent to the Picture Archiving and Communication System (PACS) to become a part of the permanent patient medical record. ### Clinical Characteristics The primary users of Rapid Platform software are medical imaging professionals. The images generated by Rapid ASPECTS provide additional diagnostic information, which is derived from the temporal/diffusion/density features of the native CT images. The following figure provides a general layout of the ASPECTS display image as provided from Rapid ASPECTS. {7}------------------------------------------------ 510(k) Summary Image /page/6/Figure/2 description: The image shows a series of brain scans with corresponding Hounsfield Unit measurements. The scans are segmented into different regions, labeled C, IC, L, I, M1-M6, with numerical values for both the right and left hemispheres. The left hemisphere shows lower Hounsfield Unit values in several regions, particularly M3, M5, and M6, which are marked with an asterisk. The ASPECTS score is displayed as 9. ### Indications for Use: Rapid ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data. The Software automatically registers images and segments and analyzes ASPECTS Regions of Interest (ROIs). Rapid ASPECTS extracts image data for the ROI(s) to provide analysis and computer analytics based on morphological characteristics. The imaging features are then synthesized by an artificial intelligence algorithm into a single ASPECT (Alberta Stroke Program Early CT) Score. Rapid ASPECTS is indicated for evaluation of adult patients presenting for diagnostic imaging workup, for evaluation of extent of disease. Extent of disease refers to the number of ASPECTS regions affected which is reflected in the total score. This device provides information that may be useful in the characterization of early ischemic brain tissue injury for ischemic stroke patient (typically < 24 hours since last known well) during image interpretation following the standard of care. Rapid ASPECTS provides a comparative analysis to the ASPECTS standard of care radiologist assessment using the ASPECTS atlas definitions and atlas display including highlighted ROIs and numerical scoring. Rapid ASPECTS presents the original and annotated images for concurrent reads. {8}------------------------------------------------ #### 510(k) Summary Limitations: - 1. The ASPECTS score should be only used for ischemic stroke patients following the standard of care. - 2. Rapid ASPECTS is an adjunct tool and is not intended to replace a clinicians review of the original imaging or their clinical judgement. - Physicians should not use the CAD generated output as the primary interpretation without their 3. concurrence. Contraindications/Exclusions/Cautions: - Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. ● - Hemorrhagic Transformation, Hematoma ● - Very thin or no Ventricles ● #### Technological Characteristics: Rapid ASPECTS is a machine learning implementation. Rapid ASPECTS provides an automatic ASPECT score based on the case input file for the physician. The score includes which ASPECT regions are identified based on regional imaging features derived from non-contrast computed tomography (NCCT) brain image data based on the random forest machine learning technique. The results are generated based on the Alberta Stroke Program Early CT Score (ASPECTS) guidelines and provided to the clinician for review and verification. At the discretion of the clinician, the scores may be adjusted based on the clinician's judgment as well as other factors the clinician may integrate though the Rapid Platform User Interface. #### Performance Standards: Rapid has been developed in conformance with the following standards, as applicable: | ISO 14971:2019 | Application of Risk Management to Medical Devices | |--------------------|---------------------------------------------------------| | IEC 62304:2015 | Medical device software – Software lifecycle processes | | IEC 62366:2015 | Application of Usability Engineering to Medical Devices | | NEMA PS 3.1 - 3.20 | Digital Imaging and Communications in Medicine (DICOM) | Rapid has been designed to meet cybersecurity requirements using design Vulnerability Assessments, SBOM's, and PEN Testing. ### Performance Data: Rapid complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20. iSchemaView conducted extensive performance validation testing and software verification and validation testing of the Rapid ASPECTS module as standalone software and as integrated within the Rapid System. This performance validation testing demonstrated that the Rapid ASPECTS module provides accurate representation of key processing parameters under a range of clinically relevant parameters and perturbations associated with the intended use of the software. Software performance, validation and verification testing demonstrated that the Rapid ASPECTS module met all design requirements and specifications. Design Verification and Validation according to 21 CFR 820.30 passed. {9}------------------------------------------------ 510(k) Summary The Rapid ASPECTS Standalone Performance and Clinical Validation Protocol (VCP) passed. No deviations were found during the execution of the testing. Standalone Performance: The percent agreement of Rapid ASPECTS to the reference at the ASPECTS region level and at the scan level is 82.8%. Both are comparable (overlapping CI) to the pairwise agreement between any two of the three experts. Clinical Validation Reader Improvement: The primary reader improvement endpoint is to demonstrate that reader scoring of the 10 ASPECT regions is more closely aligned with the reference standard when read in conjunction with Rapid ASPECTS than without Rapid ASPECTS. We find that the fixed effect of the Rapid assist increases the percent agreement on average, by about 0.02. That is, agreement increases from 82% without assistance to 84% with assistance excluding the expert, the average agreement increases from 80.4% without assistance to 83.3% with assistance. Supplemental Confounder/Mimic Sensitivity Assessment: Additionally, supplemental data was evaluated to further assess confounders which may appear in case processing. The supplemental data included: Abscess (3), Dural AVF (4), Hydrocephalus (4), Hypertensive Encephalopathy (2), Isodense SDH (4), Multiple Sclerosis (3) and Traumatic Brain Injury (3). Five readers were used to review the confounder supplemental cases for 115 reads. Three of the 115 reads (2.6%) changed based on Rapid results, showing minimal effect of confounders/mimics on ASPECTS performance. #### Demographic & Clinical Subcategories: Demographics are provided for the 102 scans used in the reader improvement study and 88 used for the standalone performance. Case counts are included for reference: | Overview of Case Counts | | | | |------------------------------------------------|-----|------------------------|-------------------------------| | Case Category | N | Include in Stand Alone | Include in Reader Improvement | | Suspected Stroke | 88 | Yes | Yes | | Stroke Mimic<br>(non-ischemic stroke etiology) | 14 | No | Yes | | Technical Exclusion due to contrast | 3 | No | No | | Total | 105 | 88 | 102 | Age: Age ranged from 23 to 99 with a median of 66 years of age. Image /page/8/Figure/10 description: The image shows a histogram, boxplot, and summary statistics for the variable 'Age'. The histogram displays the distribution of ages, while the boxplot provides a visual representation of the quartiles and outliers. The summary statistics include the mean (64.865979), standard deviation (16.358375), median (66), minimum (23), and maximum (99), along with other quantile values and the number of valid (97) and missing (8) observations. {10}------------------------------------------------ ### 510(k) Summary ### Gender: | Gender | N | Column % | |--------|-----|----------| | F | 53 | 51.96% | | M | 49 | 48.04% | | All | 102 | | ### LVO Status: | Stroke LVO | N | Column % | |------------|-----|----------| | no | 37 | 36.27% | | unknown | 14 | 13.73% | | yes | 51 | 50.00% | | All | 102 | 100.00% | ### US/OUS Status | US/OUS | N | Column % | |--------|-----|----------| | OUS | 21 | 20.59% | | US | 81 | 79.41% | | All | 102 | 100.00% | ### Last Known Well - | Last Known Well (6 Hour Splits) | N | Column % | |---------------------------------|----|----------| | <6 | 27 | 31.03% | | ≥6 - <16 | 21 | 24.14% | | ≥16 - 24 | 5 | 5.75% | | UNK | 34 | 39.08% | | All | 87 | 100.00% | #### Data Distribution: Demographics are provided for the 102 scans used in the reader improvement study and 88 used for the standalone performance. The data was distributed from GE (23), Siemens (28), Cannon/Toshiba (22) and Philips (29). The data included 14 confounders/mimics in the primary data set with non-stroke etiology. Standalone Performance cases included US (75) vs OUS (13) for 85% US. Additionally, supplemental data was evaluated to further assess sensitivity of confounders which may appear in case processing. The supplemental data included: Abscess (3), Dural AVF (4), Hydrocephalus (4), Hypertensive Encephalopathy (2), Isodense SDH (4), Multiple Sclerosis (3) and Traumatic Brain Injury (3). #### Prescriptive Statement: Caution: Federal law restricts this device to sale by or on the order of a physician. {11}------------------------------------------------ ### Safety & Effectiveness: Rapid ASPECTS has been designed, verified, and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2019 (risk management). ### Substantial Equivalence: Rapid ASPECTS (K200760) under regulation 21 C.F.R. §892.2060 CADx is the predicate device for Rapid ASPECTS v3. The disease and clinical use are identical and substantially equivalent. There are slight differences between the two versions which do not provide new risks. The Rapid ASPECTS assists the radiologists (clinicians) in the assessment and characterization of morphological features of their respective focus areas using imaging data. The software registers, segments, and analyzes regions of interest (ROI) to provide computer analytics. The identified/extracted features are then synthesized by artificial intelligence algorithms into clinical reference scores. Rapid ASPECTS provides localization information showing which ASPECTS regions are affected. The features are compared in the following table, as well as the Risk Benefit Analysis in the next paragraph is comparative to the predicate: | Substantial Equivalence Table | | | |----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison<br>Feature | Rapid ASPECTS (K200760) | Rapid ASPECTS v3 | | Indications for Use | Rapid ASPECTS is a computer-<br>aided diagnosis (CADx) software<br>device used to assist the clinician in<br>the assessment and characterization<br>of brain tissue abnormalities using<br>CT image data. The Software<br>automatically registers images and<br>segments and analyzes ASPECTS<br>Regions of Interest (ROIs). Rapid<br>ASPECTS extracts image data for<br>the ROI(s) to provide analysis and<br>computer analytics based on<br>morphological characteristics. The<br>imaging features are then<br>synthesized by an artificial<br>intelligence algorithm into a single<br>ASPECT (Alberta Stroke Program<br>Early CT) Score.<br><br>Rapid ASPECTS is indicated for<br>evaluation of patients presenting for<br>diagnostic imaging workup with<br>known MCA or ICA occlusion, for<br>evaluation of extent of disease. | Rapid ASPECTS is a computer-aided<br>diagnosis (CADx) software device used<br>to assist the clinician in the assessment<br>and characterization of brain tissue<br>abnormalities using CT image data.<br>The Software automatically registers<br>images and segments and analyzes<br>ASPECTS Regions of Interest (ROIs).<br>Rapid ASPECTS extracts image data<br>for the ROI(s) to provide analysis and<br>computer analytics based on<br>morphological characteristics. The<br>imaging features are then synthesized<br>by an artificial intelligence algorithm<br>into a single ASPECT (Alberta Stroke<br>Program Early CT) Score.<br><br>Rapid ASPECTS is indicated for<br>evaluation of adult patients presenting<br>for diagnostic imaging workup, for<br>evaluation of extent of disease. Extent<br>of disease refers to the number of<br>ASPECTS regions affected which is<br>reflected in the total score. This device | | | Extent of disease refers to the<br>number of ASPECTS regions<br>affected which is reflected in the<br>total score. This device provides<br>information that may be useful in the<br>characterization of early ischemic<br>brain tissue injury during image<br>interpretation (within 6 hours).<br>Rapid ASPECTS provides a<br>comparative analysis to the<br>ASPECTS standard of care<br>radiologist assessment using the<br>ASPECTS atlas definitions and atlas<br>display including highlighted ROIs<br>and numerical scoring. | provides information that may be useful<br>in the characterization of early ischemic<br>brain tissue injury for ischemic stroke<br>patient (typically < 24 hours since last<br>known well) during image interpretation<br>following the standard of care. Rapid<br>ASPECTS provides a comparative<br>analysis to the ASPECTS standard of<br>care radiologist assessment using the<br>ASPECTS atlas definitions and atlas<br>display including highlighted ROIs and<br>numerical scoring. Rapid ASPECTS<br>presents the original and annotated<br>images for concurrent reads. | | | Limitations:<br>1. Rapid ASPECTS is not intended<br>for primary interpretation of CT<br>images; it is used to assist physician<br>evaluation.<br>2. Rapid ASPECTS has been<br>validated in patients with known<br>MCA or ICA Occlusion prior to<br>ASPECT scoring.<br>3. Use of the Rapid ASPECTS<br>Module in clinical settings other than<br>early brain ischemia (within 6 hours)<br>caused by known ICA or MCA<br>occlusions has not been tested.<br>4. Rapid ASPECTS has been<br>validated and is intended to be used<br>on GE Lightspeed VCT Scanners. | Limitations:<br>1. The ASPECTS score should be only<br>used for ischemic stroke patients<br>following the standard of care.<br>2. Rapid ASPECTS is an adjunct tool<br>and is not intended to replace a<br>clinicians review of the original<br>imaging or their clinical judgement.<br>3. Physicians should not use the CAD<br>generated output as the primary<br>interpretation without their<br>concurrence. | | | Contraindications/Exclusions/Cautio<br>ns:<br>• Patient Motion: excessive motion<br>leading to artifacts that make the<br>scan technically inadequate.<br>• Hemorrhagic Transformation,<br>Hematoma<br>• Very thin or no Ventricles | Contraindications/Exclusions/Cautions:<br>• Patient Motion: excessive motion<br>leading to artifacts that make the<br>scan technically inadequate.<br>• Hemorrhagic Transformation,<br>Hematoma<br>• Very thin or no Ventricles | | Clinical<br>Application/Anatomical Region | Stroke/Head | Stroke/Head | | Standard of Care<br>Representation | ASPECT Scoring | ASPECT Scoring | | Imaging | CT | CT | | Technical<br>Implementation | AI/Random Forest | AI/Random Forest | | Image Overlay | ASPECTS Atlas ROIs, highlighted by algorithms | ASPECTS Atlas ROIs, highlighted by algorithms | | Primary User(s) | Neuroradiologist/Clinician | Neuroradiologist/Clinician | | Alteration of<br>original image<br>data base | No | No | | Alters Standard<br>of Care<br>Workflow | In parallel to | In parallel to | | Analysis<br>Window | ≤ 6 hours | ≤ 24 hours | {12}------------------------------------------------ {13}------------------------------------------------ # 510(k) Summary ### Risk Benefit Analysis: | Risk Benefit Summary | | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Summary of Benefits: | This device provides a systematic, automated analysis of<br>NCCT scans of the head to provide a standardized, automated<br>ASPECT score for Stroke workup. The clinical reader study,<br>which included 1 expert neuroradiologist and 5 non-expert<br>typical readers demonstrated a statistically significant<br>improvement in the accuracy of the 6 readers' scores when<br>scoring was performed in conjunction with the Rapid<br>ASPECTS output. In a subgroup analysis, the benefit of the<br>software was most substantial among the non-neuroradiologist<br>expert reader which typically evaluates CT scans in<br>community hospitals and primary stroke centers. These non-<br>expert readers also evaluate CT scans in comprehensive<br>centers, particularly in the acute setting, when expert<br>neuroradiologists are not immediately available. The software<br>allows the non-expert physician to perform at the expert-like<br>level. Use of the system did not appear to have any significant<br>impact (either positive or negative) on the score of the expert<br>neuroradiologist included in the test reader group. Overall,<br>this system should provide a more consistent and timely<br>benefit of standardized reads regardless of physician and<br>center specialty. | | Summary of the Risks | There are minimal potential risks associated with the use of<br>the device.<br><br>Incorrect scoring which may result in false positive<br>results and results to incorrect patient management with<br>possible adverse effects such as, unnecessary additional | | Risk Benefit Summary | | | | medical imaging and/or unnecessary additional<br>diagnostic workup. | | | Incorrect scoring which may result in false negative<br>results may lead to complications, including incorrect<br>diagnosis and delay in disease management. | | | The device could be misused to analyze images from an<br>unintended patient population or on images acquired<br>with incompatible imaging hardware or incompatible<br>image acquisition parameters, leading to inappropriate<br>diagnostic information being displayed to the user. | | | Device failure could lead to the absence of results, delay of<br>results or incorrect results, which could likewise lead to<br>inaccurate patient assessment. | | | However, based on the performance data and the<br>application of mitigating measures (general controls and<br>special controls established for this device type), use of<br>the device is unlikely to decrease diagnostic<br>performance of the user and possible misuse of the<br>device does not present additional risks compared with<br>misuse of other types of radiological image processing<br>devices. | | Summary of other Factors | The study was enriched to cover the range of ASPECT scores<br>and scanner manufacturers and the readers in practice may not<br>experience a significant improvement in determining<br>ASPECTS. | | Conclusions: Do the<br>probable benefits outweigh<br>the probable risks. | Yes. The probable benefits outweigh the probable risks, given<br>the combination of required general controls and the special<br>controls established for this device. The Special Controls<br>will sufficiently assist in managing risks associated with<br>incorrect brain tissue characterization determining<br>ASPECT scoring, application of the device results to the<br>wrong patient population, analysis of incompatible images,<br>and/or device failure by ensuring proper performance and<br>use of the device. | | | By providing a systematic, automated analysis of NCCT scans<br>of the head to provide a standardized, automated ASPECT<br>score for Stroke workup. The Rapid ASPECTS analytics<br>calculates morphological characteristics of brain tissue using<br>the historical training data and providing results which the<br>attending physician may evaluate and modify based on other<br>presenting conditions of the patient. In addition to the Rapid. | | Risk Benefit Summary | | | ASPECTS clinical module, other clinical information is easily accessible within the Rapid System framework such as CTA and CTP to inform the clinical decision-making process. | | | The clinical reader study demonstrates a statistically significant improvement of ASPECTS reads among a diverse sample of 6 typical readers representing multiple specialties, years of practice, and practice settings. | | | Overall, this system should provide a more consistent and timely benefit of standardized reads regardless of physician and center specialty. Therefore, given the available information concerning the benefits, risks, and supporting data; the probable benefits outweigh the probable risks, given the combination of required general controls and special controls established for this device. |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...