This product is intended for use in clinical radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Device Story
RadiForce RX570/RX570-AR are 5MP (2,048 x 2,560) color LCD monitors using IPS technology; designed for clinical radiological image review, analysis, and diagnosis. Input signals received via DVI-D or DisplayPort; internal lookup tables store factory-calibrated display modes (including DICOM GSDF) to ensure tone curve compliance regardless of display controller. RX570 features anti-glare treatment; RX570-AR features anti-reflection coating. Used in clinical settings by radiologists/physicians. RadiCS application software provides quality control, calibration, and asset management. Monitors support daisy-chaining and various OSD-controlled adjustments (brightness, contrast, gamma, etc.). Benefits include consistent, high-resolution diagnostic image visualization.
Clinical Evidence
Bench testing only. No clinical or animal data. Performance verified against FDA guidance 'Display Devices for Diagnostic Radiology' (2022) and AAPM TG18 guidelines. Metrics included MTF, luminance, DICOM GSDF conformance, angular dependency, non-uniformity, noise power spectrum, and reflection characteristics.
Technological Characteristics
54.1 cm (21.3 inch) TFT Color LCD (IPS) panel; 5MP resolution (2,048 x 2,560); 0.165 mm pixel pitch; LED backlight. Connectivity: DVI-D (dual link) and DisplayPort. Integrated optical sensor for calibration. RadiCS software for QC/calibration. Calibration stored in internal lookup tables. Anti-glare (RX570) or anti-reflection (RX570-AR) surface treatments.
Indications for Use
Indicated for trained medical practitioners reviewing, analyzing, and diagnosing clinical radiological images, including full-field digital mammography and digital breast tomosynthesis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
EIZO Corporation Hiroaki Hashimoto Senior Manager of Regulatory Compliance and Safety Department 153 Shimokashiwano Hakusan, Ishikawa 924-8566 Japan
Re: K243221
November 27, 2024
Trade/Device Name: RadiForce RX570; RadiForce RX570-AR Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: PGY Dated: October 7, 2024 Received: October 7, 2024
Dear Hiroaki Hashimoto:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory
{3}------------------------------------------------
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb Assistant Director Imaging Software Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
Submission Number (if known)
#### K243221
Device Name
RadiForce RX570; RadiForce RX570-AR
Indications for Use (Describe)
This product is intended for use in clinical radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for EIZO. The logo consists of a stylized square made up of smaller squares in a checkerboard pattern, with a jagged edge on the top and a stepped edge on the bottom. To the right of the square is the word "EIZO" in bold, sans-serif font, with a registered trademark symbol next to the "O".
| EIZO Corporation, 153 Shimokashiwano, Hakusan, Ishikawa<br>924-8566 Japan | Name<br>Department | Hiroaki Hashimoto<br>Regulatory Compliance and<br>Safety |
|---------------------------------------------------------------------------------|--------------------|----------------------------------------------------------|
| U.S. Food and Drug Administration<br>Center for Devices and Radiological Health | Telephone | +81 (76) 274-2468 |
| | E-Mail | hiroaki.hashimoto@eizo.com |
### 510(k) Summary
### 1. Submitter
EIZO Corporation 153 Shimokashiwano, Hakusan, Ishikawa 924-8566 Japan Phone: +81 (76) 274-2468 Contact Person: Hiroaki Hashimoto Date of Prepared: October 7th, 2024
#### 2. Device
- . Name of Device: RadiForce RX570; RadiForce RX570-AR
- Common or Usual Name: 21.3 inch (54.1 cm) Color LCD Monitor ●
- . Classification Name: Medical Image Management and Processing System (21 CFR 892.2050)
- Regulatory Class: . II
- PGY ● Product Code:
## 3. Predicate Device
| • Name of Device: | RadiForce RX560, RX560-AR (K172738) |
|-------------------------|---------------------------------------------------------------------|
| • Common or Usual Name: | 54.1 cm (21.3 inch) class Color LCD Monitor |
| • Classification Name: | Medical Image Management and Processing System<br>(21 CFR 892.2050) |
| • Regulatory Class: | II |
| • Product Code: | PGY |
| • Name of Device: | RadiForce GX540 (K151883) |
| • Common or Usual Name: | 5MP Monochrome LCD Monitor |
| • Classification Name: | Medical Image Management and Processing System<br>(21 CFR 892.2050) |
| • Regulatory Class: | II |
{6}------------------------------------------------
## 4. Device Description
RadiForce RX570 is a color LCD monitor for viewing medical images including those of mammography. The color panel employs in-plane switching (IPS) technology allowing wide viewing angles and the matrix size (or resolution) is 2,048 x 2,560 pixels (5MP) with a pixel pitch of 0.165 mm.
Since factory calibrated display modes, each of which is characterized by a specific tone curve (including DICOM GSDF), a specific luminance range and a specific color temperature, are stored in lookup tables within the monitor, the tone curve is e.g. DICOM compliant regardless of the display controller used.
There are two model variations, RX570 and RX570-AR. The difference of the two variations is the surface treatment of the display screens; the surface treatment of the RX570 is Anti-Glare (AG) treatment and that of the RX570-AR is Anti-Reflection (AR) coating.
Two RX570 monitors mounted on a single stand configuration is available identified by with "MD" like RX570-MD and RX570-AR-MD.
RadiCS is application software to be installed in each workstation offering worry-free quality control of the diagnostic monitors including the RadiForce RX570 based on the QC standards and guidelines and is capable of quantitative tests and visual tests defined by them. The RadiCS and its subset, RadiCS LE, are included in this 510(k) submission as an accessory to the RadiForce RX570.
RadiCS is Basic Documentation level and that it's being used unchanged from the predicate software. RadiCS supports the functions of the monitor RadiForce RX570 and it's not a medical imaging software.
## 5. Indications for use
This product is intended for use in clinical radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
{7}------------------------------------------------
## 6. Comparison of Technological Characteristics with the predicate device
The comparison table below enumerates information derived from the product brochure and measured values of the each device and different technological characteristics are discussed in it:
| Attributes | Proposed Device:<br>RadiForce RX570 | Predicate Device:<br>RadiForce RX560 |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | This product is intended for use<br>in clinical radiological images<br>(including full-field digital<br>mammography and digital<br>breast tomosynthesis) for<br>review, analysis, and diagnosis<br>by trained medical<br>practitioners. | This product is intended to be<br>used in displaying and viewing<br>digital images, including<br>standard and multiframe digital<br>mammography, for review,<br>analysis and diagnosis by<br>trained medical practitioners. It<br>is specially designed for breast<br>tomosynthesis applications. |
| Display Technology | TFT Color<br>LCD Panel (IPS) | TFT Color<br>LCD Panel (IPS) |
| Screen size | 54.1cm / 21.3"<br>Aspect ratio: 4 : 5 | 54.1cm / 21.3"<br>Aspect ratio: 4 : 5 |
| Backlight type | LED | LED |
| Frame rate and refresh rate | | |
| Digital Scanning Frequency<br>(H / V) | 31 - 135 kHz / 23 - 61 Hz<br>Frame synchronous mode: 23.5<br>- 25.5 Hz, 47 - 51 Hz | 31 - 135 kHz / 23 - 61 Hz<br>Frame synchronous mode: 23.5<br>- 25.5 Hz, 47 - 51 Hz |
| Pixel pitch | | |
| Pixel pitch | 0.165 mm x 0.165 mm | 0.165 mm x 0.165 mm |
| Subpixel pattern | RGB Vertical Stripe | RGB Vertical Stripe |
| Pixel aperture ratio | 0.627 | 0.656 |
| Subpixel driving (spatial and temporal dithering) | | |
| Independent Subpixel drive | N/A | N/A |
| Dithering | 3bit FRC | 3bit FRC |
| Display Interface | | |
| Input video signals | DVI-D (dual link) x 1,<br>DisplayPort x 2 | DVI-D (dual link) x 1,<br>DisplayPort x 1 |
| Output video signals | DisplayPort x 1 (daisy chain) | DisplayPort x 1 (daisy chain) |
| Video bandwidth | DVI : 25-290MHz<br>DisplayPort : 25-290MHz | DVI: 290MHz<br>DisplayPort: 290MHz |
| User controls | | |
| OSD | CAL Switch / FineContrast /<br>Color Mode | CAL Switch / FineContrast /<br>Color Mode |
| | Mode Skip | Mode Skip |
| | Hybrid Gamma PXL | Hybrid Gamma PXL |
| | Brightness | Brightness |
| | Contrast | Contrast |
| | Temperature | Temperature |
| | Gamma | Gamma |
| | Hue | Hue |
| | Saturation | Saturation |
| | Gain | Gain |
| | SelfQC | SelfQC |
| | Sharpness Recovery | Picture Expansion |
| | DP Daisy Chain | Sharpness Recovery |
| | PinP | DP Daisy Chain |
| | Orientation(H/W Pivot) | Orientation (H/W Pivot) |
| | Power Save | Power Save |
| | DP Power Save | DP Power Save |
| | Language | EcoView Sense |
| | OSD Information | Language |
| | Indicator | OSD Information |
| | Input Selection | Indicator |
| | Key Lock | Input Selection |
| | LEA<br>(Life Expectancy Analyzer) | Logo |
| | USB Selection for KVM<br>Switch | Key Lock |
| | Grayscale Warning | LEA<br>(Life Expectancy Analyzer) |
| | | DDC for Medical |
| | | Grayscale Warning |
| RadiCS | Complete QC Tests | Complete QC Tests |
| | Calibration | Calibration |
| | Task Setting | Task Setting |
| | History Management | History Management |
| | Asset Management | Asset Management |
| Ambient light sensing | | |
| Ambient light sensor | Photo Diode<br>Position: In the upper right bezel of the screen<br>Software tool: RadiCS | Photo Diode<br>Position: In the upper right bezel of the screen<br>Software tool: RadiCS |
| Touch-screen technology | | |
| | N/A | N/A |
| Luminance calibration tools | | |
| | Integrated optical sensor<br>External optical sensor<br>Calibration software:<br>RadiCS | Integrated optical sensor<br>External optical sensor<br>Calibration software:<br>RadiCS |
| Quality-control procedures | | |
| | Software: RadiCS<br>AAPM On-line Report No.<br>03:2005 compliant | Software: RadiCS<br>AAPM On-line Report No.<br>03:2005 compliant |
| Software/Firmware | | |
| | N/A | N/A |
{8}------------------------------------------------
{9}------------------------------------------------
It is clear that the technological characteristics differences discussed above do not affect the safety and the effectiveness of the RX570.
{10}------------------------------------------------
# 7. Performance Testing
The bench tests below were performed on the RadiForce RX570 following the instructions in "Guidance for Industry and FDA Staff: Display Devices for Diagnostic Radiology" issued on September 28, 2022:
- Measurement of spatial resolution expressed as modulation transfer function (MTF) ●
- The maximum number allowed for each type of pixel defects/faults ●
- . Visual check of presence or absence of miscellaneous artifacts on the display screen as specified in TG18 guideline
- Measurement of temporal response
- Measurement of Luminance ●
- Verification of the conformance to DICOM GSDF as specified in Assessment of . Display Performance for Medical Imaging Systems by AAPM Task Group 18 (TG18 guideline)
- . Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directions
- . Measurement of the luminance non-uniformity characteristics of the display screen as specified in TG18 guideline
- Measurement of the chromaticity non-uniformity characteristics of the display screen ● as specified in TG18 guideline
- . Performance data on luminance stability
- Measurement of noise expressed as noise power spectrum (NPS) .
- Measurement of display reflections including specular, diffuse and haze components
- Measurement of small-spot contrast ratio ●
- Measurement of Color tracking and Gray tracking
The test results showed that the RadiForce RX570 has display characteristics equivalent to those of the predicate device, RadiForce RX560 or RadiForce GX540.
Besides, the display characteristics of the RadiForce RX570 meet the pre-defined criteria when criteria are set.
No animal or clinical testing was performed on the RadiForce RX570.
{11}------------------------------------------------
## 8. Conclusion
The RadiForce RX570 was determined to be substantially equivalent to the predicate device due to the following reasons:
- The stated intended use is substantially the same as that of the predicate device. .
- It was confirmed that the technological characteristics differences from those of the . predicate device do not affect the safety or the effectiveness.
- . The bench tests demonstrated that the display characteristics are equivalent to those of the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.