32HL512D This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography. 31HN713D, 32HQ713D This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Device Story
Medical monitors for PACS/radiology; display high-resolution color/grayscale images. Used by trained medical practitioners in clinical settings for image review, analysis, and diagnosis. Input via DisplayPort/HDMI/DVI; device processes signals for visual output on TFT LCD screens. Includes calibration tools (PerfectLum/LG Calibration Studio Medical) to ensure grayscale-to-luminance conformance, luminance uniformity, and stability. Output assists clinicians in diagnostic decision-making; benefits include accurate visualization of radiological data.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against FDA guidance 'Display Devices for Diagnostic Radiology' (Sept 2022), including grayscale-to-luminance function, luminance uniformity/Mura, stability, spatial noise, and veiling glare.
Technological Characteristics
TFT LCD medical monitors. Resolution: 3,840 x 2,160 (32HL512D/32HQ713D) or 4,200 x 2,800 (31HN713D). Connectivity: DisplayPort, HDMI, DVI. Calibration via PerfectLum or LG Calibration Studio Medical. Software validated per IEC 62304. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2.
Indications for Use
Indicated for trained medical practitioners to display, review, analyze, and diagnose radiological images. 32HL512D excludes mammography. 31HN713D and 32HQ713D include full-field digital mammography and digital breast tomosynthesis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
LG Electronics Inc. % Daseul An RA Specialist LG Electronics 222, LG-ro, Jinwi-myeon Pyeongtaek-si, Gyeonggi-do 17709 KOREA
April 27, 2023
# Re: K230845
Trade/Device Name: 32HL512D, 31HN713D, 32HO713D Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: PGY Dated: March 24, 2023 Received: March 28, 2023
Dear Daseul An:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-
combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230845
Device Name 32HL512D, 31HN713D, 32HQ713D
Indications for Use (Describe) 32HL512D
This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
31HN713D, 32HQ713D
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
Type of Use (Select one or both, as applicable)
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|------------------------------------------------------------------------------------|
| <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text "K230845" at the top. Below the text is the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the words "LG Electronics" in gray.
# 510(k) Summary
[As Required by 21 CFR 807.92]
## 1. Date Prepared [21 CFR 807.92(a)(a)]
March 24, 2023
# 2. Submitter's Information [21 CFR 807.92(a)(1)]
| ● | Name of Sponsor: | LG Electronics Inc. |
|---|-----------------------|----------------------------------------------------------------------------------|
| - | Address: | 222, LG-ro, Jinwi-myeon, Pyeongtaek-si,<br>Gyeonggi-do, 17709, Republic of Korea |
| ● | Name of Manufacturer: | LG Electronics Inc. |
| - | Address: | 168, Suchul-daero, Gumi-si, Gyeongsangbuk-do, 39368<br>Republic of Korea |
| ● | Contact Name: | Daseul An / Regulatory Affairs Specialist |
| - | Telephone No.: | +82-10-8914-0116 |
| - | Email Address: | daseul.an@lge.com |
## 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
- Trade Name: 32HL512D, 31HN713D, 32HQ713D ●
- Common Name: .
Medical Monitor
- Classification: ●
| Classification Name | Medical image management and processing system |
|-----------------------|------------------------------------------------|
| Classification Number | 21 CFR 892.2050 |
| Product Code | PGY |
| Device Class | II |
| Review Panel | Radiology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the text "LG Electronics" in gray. The text is in a sans-serif font and is slightly smaller than the circle logo.
## 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow;
#### Predicate Device for 32HL512D
| • | 510(k) Number: | K192925 |
|---|----------------------|------------------------------------------------|
| • | Applicant: | LG Electronics |
| • | Classification Name: | Medical image management and processing system |
| • | Trade Name: | 32HL512D |
#### Predicate Device for 31HN713D
| • 510(k) Number: | K201777 |
|------------------------|------------------------------------------------|
| • Applicant: | LG Electronics |
| • Classification Name: | Medical image management and processing system |
| • Trade Name: | 31HN713D |
#### Predicate Device for 32HQ713D
| 510(k) Number: | K223423 |
|----------------------|------------------------------------------------|
| Applicant: | LG Electronics |
| Classification Name: | Medical image management and processing system |
| Trade Name: | 32H0713D |
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
The Medical monitor is intended to provide high resolution color and grayscale medical imaging for PACS and Radiology system. This Medical Monitor is intended to be used by trained medical practitioners for displaying, reviewing, and analysis of medical images.
#### 6, Indications for use [21 CFR 807,92(a)(5)]
#### - 32HL512D
This Medical Monitor is indicated for use in displaying radiological images for review, analysis, and diagnosis by trained medical practitioners. The display is not intended for mammography.
#### - 31HN713D, 32HQ713D
This Medical Monitor is indicated for use in displaying radiological images (including full-field digital mammoqraphy and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical practitioners.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the words "LG Electronics" in gray. The logo is simple and modern, and it is easily recognizable.
#### 7. Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
The comparison table shows the technical characteristics of the subject device are substantially equivalent to the predicate device. There are no significant differences between the subject devices and the primary predicate devices that would adversely affect the use of the product. The main change is the addition of the available calibration tool which is validated according to IEC 62304.
| | Proposed Device | Predicate Device | Equivalence |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| K Number | TBD | K192925 | - |
| Manufacturer | LG Electronics Inc. | LG Electronics Inc. | - |
| Model Name | 32HL512D | 32HL512D | - |
| Classification<br>Name | Medical image management and<br>processing system | Medical image management and<br>processing system | Same |
| Classification<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Indications for Use | This Medical Monitor is indicated<br>for use in displaying radiological<br>images for review, analysis, and<br>diagnosis by trained medical<br>practitioners. The display is not<br>intended for mammography. | This Medical Monitor is indicated<br>for use in displaying radiological<br>images for review, analysis, and<br>diagnosis by trained medical<br>practitioners. The display is not<br>intended for mammography. | Same |
| Power<br>Consumption | MAX. 65W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W | MAX. 65W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W | Same |
| Screen size | 718.2 x 414.3 mm | 718.2 x 414.3 mm | Same |
| LCD Screen | TFT LCD | TFT LCD | Same |
| Pixel Pitch | 0.18159 x 0.18159 mm | 0.18159 x 0.18159 mm | Same |
| Resolution | 3,840 x 2,160 pixels | 3,840 x 2,160 pixels | Same |
| Horizontal<br>Frequency | 30 kHz to 135 kHz | 30 kHz to 135 kHz | Same |
| Vertical Frequency | 56 Hz to 61 Hz | 56 Hz to 61 Hz | Same |
| Input video signals | DisplayPort x 2<br>HDMI x 1 | DisplayPort x 2<br>HDMI x 1 | Same |
| Calibration Tool | PerfectLum / LG Calibration Studio<br>Medical | PerfectLum | Modified |
Table 1. Comparison of Proposed Device to Primary Predicate Device (32HL512D)
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized white "L" and "." inside, followed by the text "LG Electronics" in gray. The text is in a sans-serif font and is aligned to the right of the circular logo.
| | Proposed Device | Predicate Device | Equivalence |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| K Number | TBD | K201777 | - |
| Manufacturer | LG Electronics Inc. | LG Electronics Inc. | - |
| Model Name | 31HN713D | 31HN713D | - |
| Classification<br>Name | Medical image management and<br>processing system | Medical image management and<br>processing system | Same |
| Classification<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Indications for Use | This Medical Monitor is indicated<br>for use in displaying radiological<br>images for review, analysis, and<br>diagnosis by trained medical<br>practitioners. The display is not<br>intended for mammography. | This Medical Monitor is indicated<br>for use in displaying radiological<br>images for review, analysis, and<br>diagnosis by trained medical<br>practitioners. The display is not<br>intended for mammography. | Same |
| Power<br>Consumption | MAX. 150W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W | MAX. 150W<br>Sleep Mode ≤ 0.5W<br>Off Mode ≤ 0.3W | Same |
| Screen size | 676.9 x 459.7 mm | 676.9 x 459.7 mm | Same |
| LCD Screen | TFT LCD | TFT LCD | Same |
| Pixel Pitch | 0.1554 x 0.1554 mm | 0.1554 x 0.1554 mm | Same |
| Resolution | 4,200 x 2,800 pixels | 4,200 x 2,800 pixels | Same |
| Horizontal<br>Frequency | 30 kHz to 175 kHz | 30 kHz to 175 kHz | Same |
| Vertical Frequency | 56 Hz to 61 Hz | 56 Hz to 61 Hz | Same |
| Input video signals | DisplayPort x 2<br>HDMI x 1 | DisplayPort x 2<br>HDMI x 1 | Same |
| Calibration Tool | PerfectLum / LG Calibration Studio<br>Medical | PerfectLum | Modified |
# Table 2. Comparison of Proposed Device to Primary Predicate Device (31HN713D)
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized white "L" and "." inside, followed by the text "LG Electronics" in gray. The text is in a sans-serif font and is aligned to the right of the circular logo.
| | Proposed Device | Predicate Device | Equivalence |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| K Number | TBD | K223423 | - |
| Manufacturer | LG Electronics Inc. | LG Electronics Inc. | - |
| Model Name | 32HQ713D | 32HQ713D | - |
| Classification<br>Name | Medical image management and<br>processing system | Medical image management and<br>processing system | Same |
| Classification<br>Number | 21 CFR 892.2050 | 21 CFR 892.2050 | Same |
| Indications for Use | This Medical Monitor is indicated<br>for use in displaying radiological<br>images for review, analysis, and<br>diagnosis by trained medical<br>practitioners. The display is not<br>intended for mammography. | This Medical Monitor is indicated<br>for use in displaying radiological<br>images for review, analysis, and<br>diagnosis by trained medical<br>practitioners. The display is not<br>intended for mammography. | Same |
| Power<br>Consumption | MAX. 120W<br>Off Mode ≤ 0.3W | MAX. 120W<br>Off Mode ≤ 0.3W | Same |
| Screen size | 730.7 x 425.2 mm | 730.7 x 425.2 mm | Same |
| LCD Screen | TFT LCD | TFT LCD | Same |
| Pixel Pitch | 0.18159 x 0.18159 mm | 0.18159 x 0.18159 mm | Same |
| Resolution | 3,840 x 2,160 pixels | 3,840 x 2,160 pixels | Same |
| Horizontal<br>Frequency | 30 kHz to 135kHz | 30 kHz to 135kHz | Same |
| Vertical Frequency | 56 Hz to 61 Hz | 56 Hz to 61 Hz | Same |
| Input video signals | DVI IN x 1,<br>DP IN x 1<br>DP OUT x 1 | DVI IN x 1,<br>DP IN x 1<br>DP OUT x 1 | Same |
| Calibration Tool | PerfectLum / LG Calibration Studio<br>Medical | PerfectLum | Modified |
# Table 2. Comparison of Proposed Device to Primary Predicate Device (32HQ713D)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized white face inside, next to the text "LG Electronics" in gray. The text is in a sans-serif font and is aligned horizontally with the circle.
#### 8. Non-Clinical Test summary
- 1) Electrical Safety and Electromagnetic Compatibility
The test results demonstrated that the proposed device complies with the following standards:
- IEC 60601-1:2005/AMD2:2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromaqnetic disturbances -Requirements and tests
- 2) Software Validation
The 32HL512D, 31HN713D and 32HQ713D contain a MODERATE level of concern software, The software was designed and developed according to a software development process and was verified and validated. There have been firmware updates since FDA 510(K) clearance.
The LG Calibration Studio Medical and PerfectLum are moderate level of concern software programs. The software programs were verified and validated according to IEC 62304.
- The software information is provided in accordance with FDA quidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005.
#### 3) Bench Test – Performance Test Report
The performance items suggested in the FDA quidance "Display Devices for Diagnostic Radiology" were tested on the 32HL512D, 31HN713D and 32HQ713D using PerfectLum and LG Calibration Studio Medical.
- -32HL512D
| Measurements | Test Result |
|---------------------------------------------------------------------------------------------------------|-------------|
| a. Conformance to a grayscale-to-luminance function | Pass |
| Measurements | Test Result |
| a. Conformance to a grayscale-to-luminance function | Pass |
| b. Luminance uniformity or Mura test | Pass |
| c. Stability of luminance and chromaticity response with<br>temperature and time of operation (on-time) | Pass |
| d. Spatial noise | Pass |
| e. Veiling glare or small-spot contrast | Pass |
- 31HN713D
32HQ713D
| | Measurements | Test Result |
|----|------------------------------------------------------------------------------------------------------|-------------|
| a. | Conformance to a grayscale-to-luminance function | Pass |
| b. | Luminance uniformity or Mura test | Pass |
| c. | Stability of luminance and chromaticity response with<br>temperature and time of operation (on-time) | Pass |
| d. | Spatial noise | Pass |
| e. | Veiling glare or small-spot contrast | Pass |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the LG Electronics logo. The logo consists of a red circle with a stylized "L" and "G" inside, followed by the words "LG Electronics" in gray. The "L" and "G" are designed to resemble a winking face, with the "L" forming the nose and the "G" forming the face.
All display characteristics of the 32HL512D, 31HN713D and 32HQ713D have met the predefined criteria. Therefore, the performance of 32HL512D, 31HN713D and 32HQ713D were verified through the performance test.
Display Devices for Diagnostic Radiology – Guidance for Industry and Food and Drug Administration Staff, issued on September 28, 2022
#### Clinical Test Summary:
No clinical studies were considered necessary and performed.
## 9. Conclusion [21 CFR 807.92(b)(3)]
The Medical Monitor 32HL512D, 31HN713D and 32HQ713D are found to be substantially equivalent in safety and effectiveness to the predicate devices based on the information provided in this premarket notification.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.