K120613 · Ge Healthcare Surgery GE Oec Medical Systems, Inc. · OXO · Apr 6, 2012 · Radiology
Device Facts
Record ID
K120613
Device Name
OEC 9900 ELITE
Applicant
Ge Healthcare Surgery GE Oec Medical Systems, Inc.
Product Code
OXO · Radiology
Decision Date
Apr 6, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The OEC® 9900 Elite Mobile Fluoroscopy System is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, vascular, cardiac, critical care and emergency room procedures.
Device Story
Mobile fluoroscopic imaging system; assists physicians during surgical/interventional procedures. System consists of C-arm gantry and workstation. X-ray generator in C-arm base produces high voltage; X-ray tube emits radiation directed at patient; image intensifier captures transmitted X-rays. Camera converts intensifier output to digital images displayed on workstation monitor. Operator adjusts C-arm rotation/translation, X-ray technique, image size, and orientation to account for anatomical variation. Used in clinical environments (OR, ER, etc.) by physicians and operators. Provides real-time visualization and records of anatomy/pathology to localize surgical regions, guide procedures, and document outcomes. Benefits include improved visualization for orthopedic, vascular, and cardiac interventions.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Mobile fluoroscopic X-ray system. Components: C-arm gantry, workstation, X-ray generator, X-ray tube, image intensifier, camera. Energy source: High voltage X-ray. Connectivity: Workstation-based display/storage. Sterilization: Not applicable (device is capital equipment).
Indications for Use
Indicated for patients undergoing diagnostic, surgical, and interventional procedures requiring fluoroscopic and spot-film imaging, including cholangiography, endoscopy, urologic, orthopedic, vascular, cardiac, critical care, and emergency room applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
## APR - 6 2012
Image /page/0/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circle. The letters and the circle are black, creating a strong contrast. The logo is a well-known symbol associated with the multinational conglomerate corporation.
# GE Healthcare 510(k) Premarket Notification Submission
K120613
#### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
February 24, 2012 Date:
Submitter:
GE Healthcare Surgery 384 Wright Brothers Drive Salt Lake City, UT 84116
Primary Contact Person:
Gerald Buss Director Regulatory Affairs GE Healthcare Surgery Phone: (801) 517-6591 Fax: (801) 517-6566
Secondary Contact Person:
Karen Russell Regulatory Affairs Leader GE Healthcare Surgery Phone: (801) 536-4930 Fax: (801) 517-6566
Device: (Trade Name): Common/Usual Name: Classification Names: Product Code:
OEC® 9900 Elite
Mobile Fluoroscopic Imaging System
21 CFR 892.1650 Image-intensified fluoroscopic x-ray system
90OXO and 90JAA
K082781 OEC® 9900 Elite Predicate_Device(s):
Device Description:
The OEC® 9900 Elite is a system used to assist trained physicians. The system is used to provide X-Ray images while the physician performs a medical procedure. Images from the system help the physician to visualize the patients' anatomy. This visualization helps to localize surgical regions of interest and The images provide real-time visualization and pathology. records of pre-surgical anatomy, in vivo-surgical activity and post surgical outcomes.
The system is composed of two primary physical elements. The first is referred to as the "C-Arm" because of its "C" shaped image gantry; the second referred to as the "Workstation" because this is the primary user interface to the system.
{1}------------------------------------------------
# GE Healthcare 510(k) Premarket Notification Submission
The system is used in different surgical procedures. Orthopedic physicians may use the system to help perform hip replacements and reconstructions of badly fractured bones. Orthopedic physicians may use the system to help perform hip replacements and reconstructions of badly fractured bones. Vascular physicians may use the system to perform blood flow studies.
Cardiologists may use the system to help see if there are blockages in some of the key arteries supplying blood to the heart. The procedures that these physicians perform are broadly referred to as "Clinical Applications". The system is controlled and run in a clinical environment.
The system employs X-Rays as its imaging technology. An X-Ray Generator located in the base of the C-Arm creates high voltage. High voltage is carried to the X-Ray tube across a set of cables. The X-Ray tube emits X-Rays that are directed toward the patient under the control of the operator. The Physician defines the desired view for the specific clinical procedure and directs the operator. The X-Rays pass through the patient and are captured by the image intensifier (II). Image intensifier images are captured by a camera and displayed on the image monitor located on the Workstation. The Physician or system operator view the images as they are displayed and thev may choose to store the images for later review.
In order to perform these procedures different views of the human anatomy are required, so the system is designed with the ability to rotate and translate the C-Arm's image gantry to obtain different viewing angles. In addition since there is variation in thickness and density of the anatomy the system operator has the ability to adjust the X-Ray Generator technique, image size and orientation
urologic, orthopedic, neurologic, vascular, cardiac, critical care
to account for the anatomical differences. Intended Use: The OEC® 9900 Elite Mobile Fluoroscopy System is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy,
Technology:
The modified OEC® 9900 Elite device employs the same fundamental scientific technology as the predicate device.
and emergency room procedures.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circle. There are decorative swirls around the outside of the circle. The logo is black and white.
00004
# GE Healthcare 510(k) Premarket Notification Submission
#### Determination of Substantial Equivalence:
The OEC® 9900 Elite and its application comply with voluntary standards as detailed in Section 9, of this premarket submission. The modifications from the predicate device OEC® 9900 Elite were completed in accordance with GE Healthcare Surgery's quality management system design controls. Engineering testing and standards compliance testing were successfully conducted and did not raise any new safety questions or concerns or identify new risks. Information is included with this 510(k) submission that supports this determination.
Conclusion:
GE Healthcare considers the modified GE OEC® 9900 Elite to be safe, as effective, and performance is substantially equivalent to the predicate device OEC® 9900 Elite (K082781).
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Gerald Buss Director Regulatory Affairs GE Healthcare Surgery 384 Wright Brothers Drive SALT LAKE CITY UT 84116
Re: K120613
Trade/Device Name: GE OEC 9900 Elite Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO and JAA Dated: March 15. 2012 Received: March 16, 2012
Dear Mr. Buss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
APR - 6 2012
{4}------------------------------------------------
#### Page 2
Enclosure
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed needicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{5}------------------------------------------------
GE Healthcare 510(k) Premarket Notification Submission
### INDICATIONS FOR USE
510(k) Number (if known):
Device Name: GE OEC 9900 Elite
Indications for Use:
The OEC® 9900 Elite is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, vascular, cardiac, critical care and emergency room procedures.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Radiological Devices
Office ofi In Vitro Diagnostic Device Evaluation and Safety
K120613
BIOK
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.