K082781 · Ge Healthcare Surgery · OWB · May 1, 2009 · Radiology
Device Facts
Record ID
K082781
Device Name
OEC 9900 ELITE
Applicant
Ge Healthcare Surgery
Product Code
OWB · Radiology
Decision Date
May 1, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The OEC® 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.
Device Story
Mobile fluoroscopic imaging system; consists of C-arm mainframe and mobile workstation. Mainframe houses high-voltage generator, x-ray control, x-ray tube, and image intensifier; supports linear/rotational motion for patient positioning. Workstation supports image display monitors, processing, and recording. Used in OR, clinic, or emergency settings by physicians/technicians to provide real-time fluoroscopic and spot-film images. Output viewed by clinicians to guide diagnostic, surgical, and interventional procedures; facilitates visualization of anatomy during complex interventions. Interfaces provided for optional peripheral devices like printers.
Clinical Evidence
Bench and animal performance testing conducted. Results demonstrate device is safe, effective, and performs equivalently to predicate devices.
Technological Characteristics
Mobile fluoroscopic x-ray system; consists of C-arm mainframe (high voltage generator, x-ray tube, image intensifier) and mobile workstation (monitors, image processing/recording). Supports linear and rotational motion. Interfaces for peripheral printers.
Indications for Use
Indicated for patients undergoing diagnostic, surgical, and interventional procedures requiring fluoroscopic and spot-film imaging, including cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care, and emergency room applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K082781
11 - 1 - 1 - 1 -
## Section 5 510(k) Summary
.
:
The following information is being submitted in accordance with the requirements of 21 CFR 807.92.
| Company Name: | GE Healthcare Surgery |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 384 Wright Brothers Drive<br>Salt Lake City, UT 84116 |
| Contact Person: | Maria C. Frame<br>Vice President Quality Assurance and Regulatory Affairs<br>GE Healthcare Surgery<br>Phone: (801) 517 6440 Fax: (801) 517 6566 |
| Preparation Date: | September 19, 2008 |
| Device (Trade Name): | OEC® 9900 Elite |
| Common/Usual Name: | Mobile Fluoroscopic Imaging System |
| Classification Names: | 21 CFR 892.1650 and 892.1720 (either) Image-intensified fluoroscopic x-ray system and<br>Mobile x-ray system. Product Code: 90.JAA and 90IZL. |
| Predicate Device: | K041932 OEC Olympus Fluoroscopic Imaging System,<br>K073543 OEC 9900 Elite |
| Device Description: | The OEC® 9900 Elite mobile fluoroscopy system is an image intensified fluoroscopic<br>system consisting of two mobile units: a Mainframe (C-Arm) and a Workstation. The<br>Mainframe (C-Arm) is comprised of a high voltage generator, x-ray control, and a "C"<br>shaped apparatus, which supports an X-ray tube and an image intensifier. The<br>Mainframe is designed to perform linear and rotational motions that allow the user to<br>position the x-ray imaging components at various angles and distances with respect to<br>the patient. The Workstation is a mobile platform that supports image display<br>monitors, image processing and recording devices. |
| Intended Use: | Interfaces are provided for optional peripheral devices such as thermal or instant film<br>printers.<br>The OEC® 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic<br>and spot-film images of the patient during diagnostic, surgical and interventional<br>procedures. Examples of clinical application may include cholangiography, endoscopy,<br>urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room<br>procedures. The system may be used for other imaging applications at the physician's<br>discretion. |
| Technology:<br>Determination of<br>Substantial Equivalence: | The OEC® 9900 Elite employs the same fundamental scientific technology as the<br>predicate device(s).<br>The demonstration of substantial equivalence is based on a comparison of features to<br>the predicate device(s) and an assessment of non-clinical performance data.<br>Information is included with this 510(k) submission that supports this determination. |
| Conclusion: | Performance testing included in Section 18 Performance Testing – Bench and Section<br>19 Performance Testing - Animal of this 510(k) demonstrates that the OEC® 9900 Elite<br>is safe, effective and performs in an equivalent manner to the predicate device(s) and<br>in accordance with its labeling. |
..
:
.
.
.
2
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
## Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 .
MAY - 1 2009
Ms. Maria Frame Official Correspondent GE Healthcare 384 Wright Brothers Drive SALT LAKE CITY UT 84116
Re: K082781
Trade/Device Name: OEC® 9900 Elite Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: January 9, 2009 Received: January 12, 2009
## Dear Ms. Frame:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
21 CFR 876.xxx 21 CFR 884.xxx 21 CFR 892.xxx Other
(Gastroenterology/Renal/Urology) (Obstetrics/Gynecology) (Radiology)
(240) 276-0115 (240) 276-0115 (240) 276-0120 (240) 276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
[signature]
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
」
{3}------------------------------------------------
Section 4 Indications For Use
Page 1 of 1
510(k) Number (if known):
OEC® 9900 Elite Device Name:
Indications for Use:
The OEC® 9900 Elite mobile fluoroscopy system is designed to provide fluoroscopic and spot-film images of the patient during diagnostic, surgical and interventional procedures. Examples of clinical application may include cholangiography, endoscopy, urolagic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The system may be used for other imaging applications at the physician's discretion.
Prescription Use_ > (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (Per 21 CFR 801 Subpart C)
1
[PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED]
Concurrence of CDRH, Office of Device Evaluation (ODE)
Iryen M. Whang
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.