D2RS & D2RS9090 Digital Dynamic Remote System is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic, and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
Device Story
Direct digital dynamic remote-controlled fluoroscopy and radiography system; utilizes X-ray source and Pixium RF 4343 FL (Model 4) flat-panel detector (FPD) to capture still and moving images; system includes remotely controlled tilting/elevating table (D2RS9090 variant allows -90 to +90° tilt and 368-1455 cm elevation). Operated by clinicians in patient environments; receptor panel converts X-rays into high-resolution digital signals. Output displayed on console for real-time diagnostic and interventional guidance. Facilitates workflow efficiency and patient throughput; replaces traditional image intensifier technology.
Clinical Evidence
Bench testing only. No clinical data required as the added digital receptor panel was previously cleared (reference device). System performance validated through bench testing, software validation, and compliance with IEC 60601-1, 1-2, 1-3, 1-6, 2-28, 2-54, and 62304 standards.
Technological Characteristics
Digital dynamic remote-controlled R&F system. Detector: Pixium RF 4343 FL (Model 4) CsI scintillator, 148 μm pixel pitch, 16-bit AD conversion. Connectivity: Ethernet. Table: D2RS9090 with -90 to +90° tilt and 368-1455 cm elevation. Software: Moderate level of concern. Compliance: IEC 60601-1, 1-2, 1-3, 1-6, 2-28, 2-54; IEC 62304.
Indications for Use
Indicated for human patients requiring fluoroscopic imaging for vascular angiography, cardiology, diagnostic, and interventional procedures. Not for mammography.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
D2RS & D2RS9090 Digital Dynamic Remote System (K213479)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, featuring a stylized depiction of a human figure. To the right of the HHS logo is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 7, 2022
Stephanix Radiological Systems % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct NAPLES FL 34114
Re: K221335
Trade/Device Name: D2RS and D2RS 9090 Digital Dynamic Remote System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: OWB, JAA, IZI Dated: May 6, 2022 Received: May 9, 2022
## Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT 8B: Division of Radiological Imaging and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221335
Device Name
D2RS & D2RS9090 Digital Dynamic Remote System
Indications for Use (Describe)
D2RS & D2RS9090 Digital Dynamic Remote System is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic, and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over The Counter Use (21 CFR 801 Subpart C) |
===
(1)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) SUMMARY K221335
Image /page/3/Picture/1 description: The image shows the logo for Stephanix Medical Imaging Solutions. The logo features a blue square with a stylized "X" inside, followed by the word "STEPHANIX" in large, bold, blue letters. Below the company name, the words "MEDICAL IMAGING SOLUTIONS" are written in smaller, lighter blue letters.
Stephanix Radiological Systems Rue Jean Moulin - ZI du Bayon La Ricamarie - FRANCE Tel : + 33 4 77 47 81 60 Fax : + 33 4 77 37 55 19 D2RS & D2RS9090 Digital Dynamic Remote System
Date Prepared: June 1, 2022 Summary prepared by: Sandie Perret, Quality Manager
- 1. Trade/Device Name: D2RS & D2RS9090 Digital Dynamic Remote System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA, IZI
- 2. Identification of Predicate Device: K213479 Manufacturer: Stephanix Radiological Systems Trade/Device Name: D2RS & D2RS9090 Digital Dynamic Remote System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA, IZI
- 3. Reference Device: K202235 Trade/Device Name: ArtPIX DRF Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA, MQB
- 4. Device Description: As compared to the predicate system, we have added a newly compatible Digital Dynamic X-ray receptor panel, model: Pixium RF 4343 FL (Model 4). This panel has already received FDA clearance (reference device above). The original Canon panels remain available. The D2RS & D2RS9090 are direct digital dynamic remote-controlled fluoroscopy and radiography systems equipped with the latest generation of Canon Flat Panel Detector (FPD). The single FPD can perform both fluoroscopy and radiography and is detachable and portable for direct projections to create a unique and highly versatile 3-in-1 imaging solution. The receptor panel directly converts the X-ray images captured by the sensor into a high-resolution digital images. The instrument is suited for use inside a patient environment. This unit converts the Xrays into digital signals. The unit can acquire still and moving images. The system includes a
{4}------------------------------------------------
remotely controlled tilting/elevating table. Panel information: Based on a flat-panel digital detector with the largest field-of-view in the market, the Pixium RF 4343 FL (Model 4) is designed for easy integration in classical R&F tables. It allows manufacturers to offer radiologists an alldigital real-time system, generating high-quality images for both routine dynamic applications and high-end static ones. Digital technology facilitates the work of X-ray technicians and speeds up the process, allowing medical centers to optimize patient throughput. Universal: a single detector for both Radiography or Fluoroscopy. A new tilting table has been introduced (D2RS9090). This new variant of D2RS consists of a modification of the remote controlled table. Instead of being down the table, this one is now behind. This allows a tilting from -90 to +90° (versus -25 to 90° with previous version) and an elevation from 368 to 1455 cm (versus 640 to 930 cm with previous version). The control console has been updated to a more modern looking design but is functionally identical to the predicate.
- 5. Indications for Use: The D2RS Digital Dynamic Remote System is indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic, and interventional procedures. It is also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures. It is intended to replace fluoroscopic images obtained through image intensifier technology. Not intended for mammography applications.
- 6. Technological Characteristics and Substantial Equivalence Comparison with the predicate shows the technological characteristics of the modified devices are equal to or better than the predicate device. Please see the detailed comparison table below. The units are functionally identical. Only the digital panel has changed. The new panel is the Pixium RF 4343 FL (Model 4). A new tilting table has been introduced (D2RS9090). This new variant of D2RS consists of a modification of the foot of the remote controlled table. Instead of being down the table, this one is now behind. This allows a tilting from -90 to +90° (versus -25 to 90° with previous version) and an elevation from 368 to 1455 cm (versus 640 to 930 cm with previous version).
| | D2RS & D2RS9090 Digital Dynamic Remote System K213479 | D2RS & D2RS9090 Digital Dynamic Remote System |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>Statement | D2RS & D2RS9090 Digital Dynamic Remote<br>System is indicated for use in generating<br>fluoroscopic images of human anatomy for<br>vascular angiography, diagnostic, and<br>interventional procedures. It is also indicated<br>for generating fluoroscopic images of human<br>anatomy for cardiology, diagnostic, and<br>interventional procedures. It is intended to<br>replace fluoroscopic images obtained<br>through image intensifier technology. Not<br>intended for mammography applications. | D2RS & D2RS9090 Digital Dynamic Remote<br>System is indicated for use in generating<br>fluoroscopic images of human anatomy for<br>vascular angiography, diagnostic, and<br>interventional procedures. It is also indicated<br>for generating fluoroscopic images of human<br>anatomy for cardiology, diagnostic, and<br>interventional procedures. It is intended to<br>replace fluoroscopic images obtained through<br>image intensifier technology. Not intended<br>for mammography applications. SAME THE<br>INDICATIONS FOR USE HAS NOT CHANGED |
## Comparison Table
{5}------------------------------------------------
| | D2RS & D2RS9090 Digital Dynamic<br>Remote System K213479 | D2RS & D2RS9090 Digital Dynamic<br>Remote System |
|--------------------|----------------------------------------------------------|--------------------------------------------------|
| Unit Photos | | |
| Control<br>Console | | |
| Digital Panel Comparison Table | | |
|--------------------------------|----------------------------------------------------------|--------------------------------------------------|
| | D2RS & D2RS9090 Digital Dynamic<br>Remote System K213479 | D2RS & D2RS9090 Digital Dynamic Remote<br>System |
| Name | Canon CXDI-RF Wireless B1 | Pixium RF 4343 FL (Model 4) |
| Dimension | 480 x 460 mm | 500 x 490 mm |
| Useful Area | 42 x 43 cm | 43 x 43 cm |
{6}------------------------------------------------
| | D2RS & D2RS9090 Digital Dynamic<br>Remote System K213479 | D2RS & D2RS9090 Digital Dynamic Remote<br>System |
|--------------------------|--------------------------------------------------------------|------------------------------------------------------|
| Photo | Image: D2RS & D2RS9090 Digital Dynamic Remote System K213479 | Image: D2RS & D2RS9090 Digital Dynamic Remote System |
| Scintillator | Csl | Csl |
| Pixel pitch | 160 μm | 148 μm |
| Matrix | 2592 × 2656 pixels | 2874 x 2840 pixels |
| AD conversion | 16 bits | 16 bits |
| DQE | 60% (0.5 lp/mm) | 73 % typ (0 lp/mm) Similar performance |
| MTF | 38% (2 lp/mm) | 35 % typ. (2 lp/mm) Similar performance |
| Fps Max | 30 fps | 30 fps |
| Interface | Ethernet/Wireless WiFi | Ethernet |
| DICOM 3 | YES | YES |
| Operating<br>temperature | 15-40°C | 15-40°C |
- 7. Bench/Performance Testing Data: Systems covering all generator/panel combinations were assembled and tested and found to be operating properly. Software was validated according to the FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (for a moderate level of concern.) Because the system uses Wi-Fi and Ethernet, we observed the recommendations contained in the FDA Guidance Document: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff. For the new digital x-ray panel we provided testing according to the FDA guidance document: Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Drug Administration Staff. In addition, we addressed all the items detailed in the FDA publication: "Pediatric Information for X-ray Imaging Device Premarket Notifications."
The X-Ray Units and digital panels have been tested to be in compliance with the following International Standards:
{7}------------------------------------------------
IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007), AMD1:2012. Medical electrical equipment Part 1: General requirements for basic safety and essential performance
IEC 60601-1-3:2008, AMD1:2013 for use in conjunction with IEC 60601-1:2005, AMD1:2012 Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray Equipment
IEC 60601-1-2:2014 (Edition 4.0) General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - requirements and tests
IEC 60601-1-3:2008+A1:2013 (Edition 2.1) Medical electrical equipment – Part 1-3: General requirements for basic safety and essential performance – Collateral Standard: Radiation protection in diagnostic X-ray equipment
IEC 60601-1-6:2010 + A1:2013 (Edition 3.1) Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
IEC 60601-2-28:2010 (Edition 2.0) Medical electrical equipment – Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
IEC 60601-2-54:2009+A1:2015 (Edition 1.1) Medical electrical equipment – Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy.
IEC 62304:2006 + A1:2016 (Edition 1.1) . Medical device software life-cycle processes The unit complies with the US Performance Standard for radiographic equipment.
- 8. Clinical Evaluation. Because the added compatible digital x-ray receptor panel had previously received FDA clearance (reference device) the clinical studies as required by the FDA Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices Guidance for Industry and Food and Druq Administration Staff were not required.
- 9. Conclusion: Based on our comparison of technological characteristics and our bench and clinical results, our conclusion is that the modified and updated system is as safe and effective as our predicate device and is therefore substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.