K131766 · Apelem-Dms Group · JAA · Dec 12, 2013 · Radiology
Device Facts
Record ID
K131766
Device Name
PLATINUM DRF IMAGING SYSTEM
Applicant
Apelem-Dms Group
Product Code
JAA · Radiology
Decision Date
Dec 12, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Platinum dRF Imaging System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, diagnostic fluoroscopy, conventional linear tomography, angiography and pediatric examinations. The Platinum dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from but not limited to gastrointestinal examinations, cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract. The units may also be used in lymphography, endoscopy, myelography, venography, pediatrics, arthrography, digital angiography and digital subtraction angiography (DSA). The Platinum dRF may be used for outpatient and emergency treatment, as well as for mobile transport (wheelchair and bed) examinations. The Platinum dRF is not indicated for use in interventional radiology.
Device Story
Platinum dRF Imaging System is a remote-controlled radiologic table platform integrating a flat panel detector, X-ray generator, collimator, and imaging software. Device converts X-ray patterns into visible images for diagnostic visualization. Operated by clinicians in clinical settings (outpatient, emergency, mobile transport). System supports general R/F, tomography, and angiography. Clinicians use the control console to manage table positioning, X-ray acquisition, and image viewing. Output facilitates diagnostic decision-making across various anatomical structures. Benefits include versatile patient positioning and dynamic acquisition capabilities for diverse diagnostic needs.
Clinical Evidence
No clinical data. Bench testing only. Compliance with 21 CFR 1020.30, 1020.31, and 1020.32, and electrical/EMC testing per IEC 60601 series standards.
Technological Characteristics
Motorized remote-controlled table with flat panel detector (Trixell Pixium RF 4343). Features include AEC interface (4 ionization chambers), motorized rotation (±180°), and variable table tilting (±90°). Connectivity via DICOM. Software integration for system control. Complies with IEC 60601-1, 60601-1-3, 60601-1-6, 60601-2-54, 62366, and NEMA XR7/PS 3.1-3.20.
Indications for Use
Indicated for patients requiring radiographic and fluoroscopic diagnostic imaging, including pediatric, gastrointestinal, cranial, skeletal, thoracic, lung, and urogenital examinations. Applications include lymphography, endoscopy, myelography, venography, arthrography, and digital subtraction angiography. Not indicated for interventional radiology.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
# DEC 1 2 2013
K131766
Page 1 of 4
### Traditional 510(k) Summary as required by 21 CFR 807.92(a)
| A) Submitted by: | APELEM- DMS GROUP |
|------------------|--------------------------------------------|
| | Parc Scientifique Georges Besse |
| | 175 Allée Von Neumann. 30035 Nîmes cedex 1 |
| | France |
| | Phone : +33 (0)4 67 50 49 00 |
| | Fax : +33 (0)4 67 50 40 09 |
Sharyn Orton. Ph.D. Contact: MEDIcept. Inc. 200 Homer Ave. Ashland. MA 01721
| B) Date Prepared: | October 24. 2013 |
|-------------------------------|-----------------------------------------------|
| C) Common Name: | System, X-Ray Fluoroscopic. Image-Intensified |
| Proprietary Name: | Platinum dRF Imaging System |
| Device Regulations and Class: | 21 CFR 892.1650. Class II |
| Product Code: | JAA |
| Classification Panel: | Radiology |
| C) Predicates: | K062623 Siemens AXIOM Luminos dRF |
D) Device Description:
The Apelem-DMS Platinum dRF Imaging System (*Platinum*) is not a stand-alone device, but functions as a platform for FDA cleared or registered components (i.e. generator, panel detector, detector collimator. X-ray tube and software imaging packages), that are installed with a Apelem-DMS manufactured radiological examination table, control panel with system controller software. and electrical panel.
The Platinum dRF remote controlled table is a radiologic table equipped with a flat panel electronic detector. This table is used to perform general digital radiological, fluoroscopy and peripheral angiography. This device allows for treatment on the whole body, using all angles. It allows the user dynamic acquisition for the whole body, to target the zones to analyze, and to be able to track contrast media.
{1}------------------------------------------------
E) Intended Use/Indication for Use: The Platinum dRF Imaging Svstem is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, diagnostic fluoroscopy. conventional linear tomography, angiography and pediatric examinations.
The Platinum dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from but not limited to gastrointestinal examinations, cranial, skeletal, thoracic and lung exposures as well as examination of the urogenital tract. The units may also be used in lymphography, endoscopy, myelography, venography, pediatrics, arthrography, digital angiography and digital subtraction angiography (DSA).
The Platinum dRF may be used for outpatient and emergency treatment, as well as for mobile transport (wheelchair and bed) examinations.
The Platinum dRF is not indicated for use in interventional radiology.
F) Comparison to Predicate Device(s):
| | Platinum dRF Imaging System | Siemens AXIOM Luminos dRF<br>K062623 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Product code | JAA | JAA, OWB. OXO |
| Components | | |
| Flat detector | Trixell Pixium RF 4343<br>(K080859) | Trixell Pixium RF 4343 |
| Collimator | RALCO R225 ACS<br>(K091517) | Self manufactured |
| Type | Motorized, remote controlled, with<br>± 45° manual rotation according to<br>a vertical axis. automatic lighting<br>on, including positioning camera | Motorized up to a maximum of ±<br>45°. detent at 0° |
| Field coverage | min 4X4 and max 48X48 at<br>100 cm source to image distance | min/max not reported |
| Generator | CMT Indico 100 Series | Self manufactured |
| Power | 50-80 kW | 65 kW or 80 kW (optional) |
| AEC Interface | 4 ionization chambers max (solid<br>state ionization 3 fields) | Not available |
| X-ray tube | Varian 600 kHU tubes RAD92 and<br>G292: 400 kHU tube RAD60 | Self manufactured |
| Use with X-ray<br>ceiling tube | Yes | Yes |
| X-ray Grid | Smit Röntgen | Self manufactured |
| Tube motorized<br>rotation | +180°/-180° | +180/-90° |
{2}------------------------------------------------
| | Platinum dRF Imaging System | Siemens AXIOM Luminos dRF<br>K062623 |
|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| Oblique | 45°/-45° motorized adjustment of | 40°/-40°: motorized adjustment of |
| projections | the height of the rotation axis from | the height of the rotation axis from |
| | 10 to 450 mm above the table (to | 10 to 300 mm above the table (to |
| | correct object parallax) | correct object parallax) |
| | The angle is displayed on both the | The angle is displayed on both the |
| | control console and on the table's | control console and on the table's |
| | control panel | control panel |
| Distance between | 6.2 cm | 7 cm |
| table-top and | | |
| detector<br>Anti-diffusion | Oscillating: ratio 10:1, 12:1 / | Fixed : ratio 15:1. 80 lignes/cm, f0 |
| grid | 401:cm - 60 1:cm / Multifocus 140 | = 125 cm, easy ejection from the |
| | cm for exams in the range of 110 to | front of the table |
| | 180 cm : easy ejection from the | |
| | front of the table | |
| Movement of the | Max 158 cm (remotely), | Max 113 cm (remotely). |
| AB selector | adjustments by increments up to 6 | adjustments by increments up to 7 |
| | cm/sec | cm/sec : PERISTEPPING step to |
| | | step movements of 25 cm each |
| Compression | 0-160 N with a compression | 5-155N |
| cone | indication above 3 kg | |
| Imaging Software | CMT Thales Duet | Flurospot COMPACT digital |
| | (K103028) | imaging system |
| Image resolution | 3.4 lp/mm | 3.4 lp/mm |
| Speed | 12 f/s (RAD) | 8 f/s (RAD) |
| | 18 f/sec for fluoroscopy (large | 15 f/s for fluoroscopy |
| | field) | |
| Dose Reduction | Yes | Yes |
| DICOM | Yes | Yes |
| DAP measuring | 0.1 - 99 999 999 µGy.m2 | Not available |
| range | | |
| Pulse frequency | 15, 7.5 and 3 pulsed per sec | 15, 10, 7.5 and 3 pulsed per sec |
| | Platinum dRF Imaging System | Siemens AXIOM Luminos dRF<br>K062623 |
| Table<br>specifications | | |
| Table Tilting | +90°/-90° | +90°/-45° or +90°/-90° (optional) |
| | Variable tilting up to 10°/sec | Two speeds about 3°/sec and<br>6°/sec. |
| | | Time for complete tilt +90° to 0° :<br>17 sec |
| | Automatic stop in the 0° position is<br>possible / angle indication on the<br>remote control panel as well as the<br>panel on the front of the table. | Automatic stop in the 0° position<br>is possible / angle indication on<br>the remote control panel as well as<br>the panel on the front of the table. |
| Table top height | 50 cm to 130 cm continuously<br>adjustable, about 4 cm/sec | 50 cm to 100 cm continuously<br>adjustable. about 4 cm/sec |
| | 48 cm - 128 cm with the base plate | 48 cm - 98 cm with the base plate<br>not encrusted in the floor |
| Table top | 240 x 85 cm / radio transparent<br>field: 283 x 65 cm / surface<br>inalterable / sliding rail system | 210 X 80 cm / radio transparent<br>field : 193 x 53.5 cm / surface<br>inalterable / sliding rail system |
| | Equivalent attenuation value :<br><0.70 mm (+/- 0.1 Al à 100 kV/3.7<br>mm Al HVL | Equivalent attenuation value : 0.65<br>mm (+/- 0.1 Al à 100 kV/3.7 mm<br>AI HVL |
| Movement | Laterally; longitudinally<br>5 way movement: 8 way movement<br>including Z-motion and tilt motion | Laterally; longitudinally<br>8-way movement |
| Weight tolerance | 265 kg with no restrictions | 230 kg |
| Longitudinal<br>movement of the<br>table top<br>(optional) | 150 cm (±75 cm) | 160cm (±80 cm) |
| Transversal<br>movement of the<br>table top | 35 cm (±17.5 cm) speed around 6<br>cm/sec | 35 cm (±17.5 cm) speed around<br>4.5 cm/sec |
| Patient coverage | 201 cm with 2 way<br>283 cm with 4 way | Not available |
| Focal distance | 110 - 180 cm | 115 - 150 cm |
| Measuring<br>detector/focal<br>point | 50 cm | Not available |
{3}------------------------------------------------
{4}------------------------------------------------
#### Substantial Equivalence Discussion
The Platinum dRF Imaging System has a similar intended use, the same technology, and uses identical or the same general components as the predicate device. Components and/or component specifications are nearly identical between the Platinum and predicate device. Compared to the predicate device, the Platinum has larger table specifications, larger rotation ability, and greater movement of the AB selector. These differences are designed to allow for larger, taller patients to be examined and for the examinations themselves to conduct. This ability has no impact on the imaging functionality of the device, and does not raise new or different issues of safety or effectiveness. All of the features of the Platinum dRF Imaging System are substantially equivalent to the similar features of the predicate device.
#### Performance
The Platinum is not a stand-alone device, but functions as a platform for FDA cleared or registered components. The Platinum is manufactured with FDA cleared/registered commercially available imaging components. Data as required in the appropriate sections of 21 CFR 1020 have been generated by the appropriate component manufacturer, and included in their manuals. Data generated by Apelem-DMS is limited to that required in CFR 1020.31(e)(1). No additional special imaging performance testing was conducted.
Electrical/EMC testing was successfully conducted per the standards listed below.
### Software
The system integration and interconnection software was developed. tested and documented in compliance with FDA Guidance "General Principles of Software Validation: Final Guidance for Industry and FDA Staff". January 2002.
### Conformity to Standards and 21 CFR 1020.30
The Platinum complies with:
- IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) Medical electrical equipment -. Part 1: General requirements for basic safety and essential performance .
- IEC 60601-1-3:2008 Medical electrical equipment Part 1-3: General requirements for . basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- IEC 60601-1-6 :2010 Medical electrical equipment Part 1-6: General requirements for . basic safety and essential performance - Collateral standard: Usability
- IEC 62366: 2007 Medical devices -- Application of usability engineering to medical . devices
- IEC 60601-2-54:2009 Medical electrical equipment Part 2-54: Particular requirements . for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
{5}------------------------------------------------
K131766
Page 6 of 6
- 1EC 60601-1-2: 2007 Medical electrical equipment Part 1-2: General requirements for . basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- NEMA XR7-1995 (R2000) High Voltage X-ray Cable Assemblies and Receptacles (for . the Claymount X-ray Cables)
- NEMA PS 3.1-3.20 (2011) Digital Imaging and Communications in Medicine � (DICOM) Set
- 21 CFR 1020.30. 1020.31 and 1020.32 as applicable .
### Conclusion
The Apelem-DMS Platinum dRF Imaging System features are substantially equivalent to the similar features of the Siemens AXIOM Luminos predicate device.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an abstract human figure embracing a globe or sphere.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 12, 2013
Apelem-DMS Group % Sharyn Orton, Ph.D. Senior Consultant MEDIcept, Inc. 200 Homer Avenue ASHLAND MA 01721
Re: K131766
Trade/Device Name: Platinum dRF Imaging System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: October 25, 2013 Received: October 30, 2013
Dear Dr. Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{7}------------------------------------------------
Page 2 - Dr. Sharyn Orton
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/McdicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.
Sincerely yours.
Smh.7)
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K 131766
## Device Name
Platinum dRF Imaging System
#### Indications for Use (Describe)
The Plainum dRF Imaging System is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic sudies. Using a digital flat detector. it cange of applications including general R/F, diagnostic fluoroscopy, conventional linear tomography, angiography and pediatric examinations.
The Plainum dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image. The system has medical applications ranging from but not limited to gastrointesinal examinations, cranial, shoracic and lung exposures as well as examination of the units may also be used in lymphography, endoscopy, myelography, venography, pediatrics, arthrography, digital angiography and digital subtraction angiography (DSA).
The Platinum dRF may be used for outpatient and emergency treatment, as well as for mobile transport (wheelchair and bed) examinations.
The Platinum dRF is not indicated for use in interventional radiology.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (9/13)
PSU Publishing Services ( 1/2) : 443-6741
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### 'DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.'
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.