The device was evaluated in a clinical environment over four weeks with more than 30 patients to confirm usability and image quality for its intended use.
The Cios Fusion is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastrointestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, critical care and emergency room procedures. The patient population may include pediatric patients.
Device Story
Mobile fluoroscopic X-ray system; consists of mobile acquisition unit (C-arm, X-ray tube, generator, flat panel detector) and monitor cart (TFT displays, image processing, storage). Used in OR, clinic, and emergency settings by healthcare professionals. Inputs: X-ray radiation through patient anatomy. Transforms inputs via solid-state detector and digital image processing system. Outputs: 2D fluoroscopic images displayed on monitors. Supports DICOM networking, 2D navigation, and injector interfaces. Benefits: provides real-time visualization for interventional procedures; aids clinical decision-making through high-quality imaging; supports dose optimization via Siemens CARE program.
Clinical Evidence
Clinical evaluation conducted over four weeks with >30 patients undergoing Ortho/Trauma procedures (extremities, shoulder, hip/pelvis, spine). Objective: confirm usability and image quality. Results indicated better, more detailed image quality compared to secondary predicate (Arcadis Varic).
Technological Characteristics
Mobile C-arm fluoroscope; solid-state detector (20x20cm or 30x30cm); stationary anode X-ray tube (0.6mm focal spot); 40-110 kV range; 2.3 kW max power. Connectivity: DICOM (wired/wireless), 2D navigation, lithotripsy, and injector interfaces. Software: moderate level of concern. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-28, 60601-2-43, 60601-2-54, ISO 14971, IEC 62304, IEC 62366.
Indications for Use
Indicated for X-ray imaging of anatomical structures in patients, including pediatric patients, during interventional fluoroscopic, gastrointestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, critical care, and emergency room procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 7, 2016
Siemens Medical Solutions USA, Inc. % Ms. Patricia Jones Regulatory Affairs Specialist 40 Liberty Boulevard 65-1A MALVERN PA 19355
Re: K153244
Trade/Device Name: Cios Fusion Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO Dated: January 27, 2016 Received: January 28, 2016
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ools
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K153244
Device Name
Cios Fusion
### Indications for Use (Describe)
The Cios Fusion is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastrointestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, critical care and emergency room procedures. The patient population may include pediatric patients.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary: Cios Fusion
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355
#### Date Prepared: January 27, 2016
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### 1. General Information:
### Importer / Distributor:
Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869
## Manufacturing Site:
Siemens AG / Siemens Healthcare GmbH Roentgenstrasse 19 - 21 95478 Kemnath, Germany Establishment Registration Number: 3002466018
## Legal Manufacturer:
SIEMENS AG Wittelsbacherplatz 2 80333 Muenchen, Germany
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Requlatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 448-6474 Fax: (610) 640-4481 Email: patricia.d.iones@siemens.com
#### 3. Subject Device Name and Classification: Trade Name: Cios Fusion Image-Intensified Fluoroscopic X-Ray Device: System,
{5}------------------------------------------------
| Regulation Medical Specialty: | Radiology |
|-------------------------------|--------------------|
| Review Panel: | Radiology |
| Product Codes: | OWB, OXO |
| Secondary Product Code: | JAA |
| Submission Type: | Traditional 510(k) |
| Regulation Number: | 892.1650 |
| Device Class | 2 |
- 4. Legally Marketed Predicate Devices:
| Primary Predicate: | |
|-------------------------------|------------------------------------------|
| Trade Name: | Cios Alpha |
| 510(k) Number: | K132094 |
| Device: | Interventional Fluoroscopic X-Ray system |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | OWB |
| Secondary Product Code: | OXO |
| Submission Type: | Traditional 510(k) |
| Regulation Number: | 892.1650 |
| Device Class | 2 |
| Recall Information: | Z-1958-2015 |
| | |
| Secondary Predicate: | |
| Trade Name: | Siremobil C-06 (renamed: Arcadis Varic) |
| 510(k) Number: | K040066 |
| Device: | Image-intensified Fluoroscopic X-ray |
| | system, Mobile |
| Regulation Medical Specialty: | Radiology |
| Review Panel: | Radiology |
| Product Code: | OXO |
| Secondary Product Code: | JAA |
| Submission Type: | Traditional 510(k) |
| Regulation Number: | 892.1650 |
| Device Class: | 2 |
| Recall Information: | Z-0118-06; Z-0123-06; Z-0438/9-06 |
| | |
#### 5. Device Description:
The Siemens Cios Fusion mobile fluoroscopy system is an image intensified fluoroscopic X-ray imaging system consisting of two mobile units: a mobile acquisition unit and a monitor cart as the image display station. The mobile acquisition unit is comprised of the X-ray control, the C-arm which supports the single-tank high-frequency generator/X-ray tube assembly, the flat panel detector and user controls. The monitor cart connects to the acquisition unit by a cable. It integrates the TFT flat panel displays, Digital Image Processing System, user controls and image storage devices (DVD, USB).
{6}------------------------------------------------
Interfaces are provided for optional devices such as external monitors, thermal printers, MP3 players, connections for 2D navigation systems, lithotripters or injectors, as well as wired and wireless DICOM network interfaces.
#### 6. Indications for Use:
The Cios Fusion is a mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical applications. Clinical applications may include but are not limited to: interventional fluoroscopic, gastro-intestinal, endoscopic, urologic, pain management, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedures. The patient population may include pediatric patients.
#### 7. Substantial Equivalence:
The new device Cios Fusion is within the same classification requlation for the same indications for use as the Primary predicate Cios Alpha (K132094, cleared 3/11/2014). Other comparable properties of the subject device to the Primary predicate include X-Ray technology, Mechanical design, the Detector technology and Image processing. Similarly, the Indications for Use, X-Ray technology/Output power, and Mechanical design are compared to the Secondary predicate Siremobil C-06 (Arcadis Varic) (K040066, cleared 2/12/2004). Documentation is provided to support a claim of substantial equivalence to these Siemens' predicate devices.
### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The Siemens Cios Fusion design is based on the Siemens Cios Alpha (K132094) including the system control, flat panel detector technology and image processing functions. The mechanical design and X-ray technology/Output power is based on the Siemens Siremobil C 06 (ARCADIS Varic) (K040066).
The subject device has the same fundamental scientific technologies as the predicate devices. The technological comparison table below demonstrates the comparability of the technological characteristics of the new device and the currently cleared predicate devices. The modifications do not affect the intended use of the device.
The table below compares the main performance data of the subject device with the predicate devices to substantiate equivalence of the subject device and predicates.
{7}------------------------------------------------
| Features /<br>Technology | Subject Device<br>Cios Fusion<br>Siemens | Primary Predicate<br>Cios Alpha<br>K132094<br>Siemens | Secondary Predicate<br>Siremobil C06 -<br>ARCADIS Varic<br>K040066<br>Siemens |
|-----------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Mobile fluoroscopic C-<br>arm | Yes | Yes | Yes |
| X-ray tube | Stationary Anode<br>0.6 mm focal spot | Rotating Anode<br>0.3/0.5 mm focal spot | Stationary Anode<br>0.6 mm focal spot |
| Tube housing assembly<br>with high frequency<br>generator | Yes, same design | Yes | Yes, same design |
| kV Range | 40 kV to 110 kV | 40 kV to 125 kV | 40 kV to 110 kV |
| Max power output | 2.3 kW | 12 kW<br>25 kW (optional) | 2.3 kW |
| Pulsed fluoroscopy | 3 mA to 25 mA | 3 mA to 119 mA<br>(12kW) | up to 23 mA |
| Beam limiting device | Yes, rectangular | Yes, rectangular | Yes, Iris collimator |
| X-Ray detector | Solid State Detector | Solid State Detector | Image Intensifier with<br>Optics and TV System<br>(CCD technology) |
| Detector active field size | 20 cm × 20 cm<br>30 cm × 30 cm<br>(optional) | 20 cm × 20 cm<br>30 cm × 30 cm<br>(optional) | ø 23 cm |
| Optional integrated dose<br>measurement device | Yes | Yes | Yes |
| Matrix size | 1024 × 1024<br>(20 cm × 20 cm<br>detector)<br>1536 × 1536<br>(30 cm × 30 cm<br>detector) | 1024 × 1024<br>(20 cm × 20 cm<br>detector)<br>1536 × 1536<br>(30 cm × 30 cm<br>detector) | 1024 × 1024 |
| Displays | 19" TFT Flat Screen<br>Color Display Panels | 19" TFT Flat Screen<br>Display Panels, B/W or<br>Color | 19" TFT Flat Screen<br>Display Panels, B/W or<br>Color |
| 2D Image post<br>processing | Yes | Yes | Yes |
| Dose optimization /<br>Siemens CARE program | Yes | Yes | Yes |
| User Interface | Yes, touch panels | Yes, touch panels | Yes, membrane<br>keyboards |
| NaviLink 2D integrated<br>navigation interface | Yes | No | Yes |
| Lithotripsy Interface | Yes | No | Yes |
| Injector interface | Yes | Yes | No |
| Features /<br>Technology | Subject Device<br>Cios Fusion<br>Siemens | Primary Predicate<br>Cios Alpha<br>K132094<br>Siemens | Secondary Predicate<br>Siremobil C06 -<br>ARCADIS Varic<br>K040066<br>Siemens |
| DICOM functionality | Yes | Yes | Yes |
{8}------------------------------------------------
#### 9. Testing to Voluntary Consensus Standards
The Siemens Cios Fusion has been tested to meet the requirements for conformity to multiple industry standards. Performance testing confirmed, that the Siemens Cios Fusion complies with the following 21 CFR Federal Performance Standards:
1020.30 Diagnostic X-Ray Systems and their major components
1020.31 Radiographic Equipment
1020.32 Fluoroscopic equipment
1040.10 Laser products and with
Also, the following relevant voluntary FDA Recognized Consensus Standards as listed in the table below:
| Recog-<br>nition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>Date | Organization |
|----------------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|--------------|
| 19-4 | General II<br>(ES/EMC) | Medical electrical equipment - Part<br>1:General requirements for basic safety<br>and essential performance | 60601-1:2005 | AAMI<br>ANSI |
| 19-1 | General II<br>(ES/EMC) | Medical electrical equipment - Part 1-2:<br>General requirements for basic safety<br>and essential performance - Collateral<br>standard: Electromagnetic compatibility<br>- Requirements and tests | 60601-1- 2:<br>2007 | IEC |
| 12-210 | Radiology | Medical electrical equipment - Part 1-3:<br>General requirements for basic safety<br>and essential performance - Collateral<br>Standard: Radiation protection in<br>diagnostic X-ray equipment | 60601-1-3:<br>2008 | IEC |
| 12-204 | Radiology | Medical electrical equipment - Part 2-<br>28: Particular requirements for the basic<br>safety and essential performance of X-<br>ray tube assemblies for medical<br>diagnosis | 60601-2-28:<br>2010 | IEC |
| 12-202 | Radiology | Medical electrical equipment - Part 2-<br>43: Particular requirements for the<br>safety and essential performance of X-<br>ray equipment for interventional<br>procedures | 60601-2-43:<br>2010 | IEC |
| 12-274 | Radiology | Medical electrical equipment - Part 2-<br>54: Particular requirements for the basic<br>safety and essential performance of X-<br>ray equipment for radiography and<br>radioscopy | 60601-2-54:<br>2009 | IEC |
{9}------------------------------------------------
| Recog-<br>nition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>Date | Organization |
|----------------------------|--------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------|------------------|
| 5-40 | General I<br>(QS/RM) | Medical devices - Application of risk<br>management to medical devices | 14971:2007 | ISO |
| 12-238 | Radiology | Digital Imaging and Communications in<br>Medicine (DICOM) Set | PS 3.1 - 3.20<br>(2011) | NEMA |
| 12-273 | Radiology | Safety of laser products - Part 1:<br>Equipment classification, and<br>requirements | 60825-1: 2007 | IEC |
| 13-8 | Software/<br>Informatics | Medical device software - Software life<br>cycle processes | 62304:2006 | IEC |
| 5-67 | General I<br>(QS/RM) | Medical devices - Application of<br>usability engineering to medical devices | 62366:2007 | AAMI ANSI<br>IEC |
| 12-229 | Radiology | Medical electrical equipment - Radiation<br>dose documentation - Part1: Equipment<br>for radiography and radioscopy | PAS 61910-1:<br>2007 | IEC |
#### 10. General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk Management is ensured via a Risk Analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification of requirements and validation of the clinical workflow to ensure that the product meets its intended uses. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice and relevant international standards. The following performance data were provided in support of the substantial equivalence determination:
#### 11. Bench Testing:
Bench testing in the form of Unit, Subsystem and System Integration testing was performed to evaluate the performance and functionality of the new features, hardware, and software updates. All testable requirements in the Engineering Requirements Specifications keys. Subsystem Requirements Specifications keys, and the Risk Management Hazard keys have been successfully verified and traced in accordance with the Siemens product development (lifecycle) process. The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria as stated in the test plans.
#### 12. Non-Clinical Testing:
The detectors were evaluated according to the FDA guidance document "Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices"..
Validation activities have been performed at the System test level on production prototype devices by appropriately trained and knowledgeable
{10}------------------------------------------------
test personnel. System level validation and regression testing has been performed successfully, demonstrating that the device meets the acceptance criteria as noted in the system test plans. The verification and validation activities within the meaning of the Quality Regulation (21 CFR 820.30) confirmed design requirements are fulfilled, system functions as intended and the Cios Fusion performs as designed and does not raise new questions regarding safety and effectiveness. Therefore, when compared to the predicate devices the Cios Fusion supports a determination of substantial equivalence to the predicate devices.
#### 13. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical and mechanical hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images.
#### 14. Clinical Testing:
The Cios Fusion has been tested in a clinical environment with the objective to confirm that usability and image quality are adequate for its intended use. In this test Cios Fusion has been evaluated for about four weeks. More than 30 patients have been examined with the Cios Fusion in the course of Ortho/Trauma procedures. These procedures included examinations of the upper and lower extremities, shoulder, hip/pelvis, cervical spine and lumbar spine.
The Cios Fusion was found to provide a better, more detailed image quality as compared to the Secondary predicate device Arcadis Varic.
#### 15. Electrical Safety and Electromagnetic Compatibility (EMC):
Electrical safety and EMC testing were conducted on the Cios Fusion, consisting of the acquisition unit (C-arm system) and the image processing and display station. The system complies with the IEC 60601-1, IEC 60601-2-43 and IEC 60601-2-54 standards for safety and the IEC 60601-1-2 standard for EMC.
#### 16. Software Verification and Validation Testing:
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for
{11}------------------------------------------------
Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, since a malfunction or latent design flaw in the Software Device could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
#### 17. Conclusion as to Substantial Equivalence:
The Cios Fusion has the same indications for use, operating environment, and mechanical design as the Primary and Secondary predicates. Siemens concludes via the documentation provided in this 510(k) submission that the Cios Fusion does not introduce any new potential safety risks and is substantially equivalent to, and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.