K213709 · Ge Medical Systems, LLC · OUO · Jan 20, 2022 · Radiology
Device Facts
Record ID
K213709
Device Name
SIGNA PET/MR
Applicant
Ge Medical Systems, LLC
Product Code
OUO · Radiology
Decision Date
Jan 20, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K213709 · Jan 20, 2022
SIGNA PET/MR
Ge Medical Systems, LLC
Retrospective clinical patient images
A retrospective reader study was conducted using clinical patient cases to evaluate the performance of the MotionFree Brain software feature compared to images without the feature.
External reader evaluation study for MotionFree Brain; Retrospective blind reader study
Clinical patients (implied by use of clinical cases)
Images reconstructed without MotionFree Brain
Reader preference and assessment of image quality
Indications for Use
The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders. MR is intended to produce transverse, sagittal, coronal and oblique cross-sectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This system may also be used for imaging during interventional procedures when performed with MR compatible devices, such as MR safe biopsy needles. PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The combined system utilizes the MR for radiation correction maps for PET studies. The system provides inherent anatomical reference for the fused PET and MR images due to precisely aligned MR and PET image coordinate systems.
Device Story
SIGNA PET/MR combines 3.0T MR and PET scanners for simultaneous, isocentric, whole-body imaging; used in oncology, neurology, and cardiology. System includes superconducting magnet, gradient coils, RF coils, and PET detectors integrated into MR bore. MotionFree Brain software feature corrects patient head motion using PET data without external tracking devices or additional MR/PET scans. Operates via acquisition/processing workstations; images conform to DICOM standards. Healthcare professionals use fused PET/MR images for disease detection, localization, and diagnosis. Benefits include improved image quality through motion correction, aiding clinical decision-making by providing precisely aligned anatomical and metabolic data.
Clinical Evidence
Retrospective blind reader evaluation study; board-certified readers assessed cases reconstructed with and without MotionFree Brain. Readers blinded to feature status and case history. Results demonstrated device is as safe and effective as predicate; no new safety/effectiveness questions raised.
Technological Characteristics
3.0T superconducting magnet; gradient coil; transmit/receive whole-body RF coil; PET detectors integrated into MR bore; Time of Flight (ToF) coincidence detection; DICOM-compliant software; MotionFree Brain software feature for rigid-body motion correction.
Indications for Use
Indicated for use by trained healthcare professionals to aid in detection, localization, and diagnosis of diseases and disorders (e.g., cardiovascular, neurological, cancer) in humans. Utilizes MR and PET imaging for metabolic and physiologic evaluation.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{1}------------------------------------------------
January 20, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Medical Systems, LLC % Mr. Brian Zielski Regulatory Affairs Leader 3200 N. Grandview Blvd. WAUKESHA WI 53188
Re: K213709
Trade/Device Name: SIGNA PET/MR Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: OUO Dated: November 22, 2021 Received: November 24, 2021
Dear Mr. Zielski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{2}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Julie Sullivan Associate Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
### K213709
Device Name SIGNA PET/MR
#### Indications for Use (Describe)
The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders. MR is intended to produce transverse, sagittal, coronal and oblique cross-sectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This system may also be used for imaging during interventional procedures when performed with MR compatible devices, such as MR safe biopsy needles.
PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
The combined system utilizes the MR for radiation correction maps for PET studies. The system provides inherent anatomical reference for the fused PET and MR images due to precisely aligned MR and PET image coordinate systems.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three water droplet-like shapes around the circle. The logo is simple and recognizable.
# 510(k) Summary
K213709
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | November 22, 2021 |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Medical Systems, LLC<br>3200 N. Grandview Blvd.<br>Waukesha, WI 53188 |
| Primary Contact: | Brian R. Zielski<br>Regulatory Affairs Leader<br>Phone: 262-227-3596<br>Email: brian.zielski@ge.com |
| Secondary Contact: | Andrew Menden<br>Senior Regulatory Affairs Manager<br>Phone: 262-308-5719<br>Email: andrew.menden@ge.com |
| Device Trade Name: | SIGNA PET/MR |
| Common / Usual Name: | Magnetic Resonance Diagnostic Device / Positron<br>Emission Tomography (PET) System |
| Classification Name: | Emission Computed Tomography System 21 CFR<br>892.1200 |
| Product Code: | OUO |
| Predicate Device(s): | SIGNA PET/MR (K163619) |
| Device Description: | The GE SIGNA PET/MR system is a combined Magnetic<br>Resonance Diagnostic Device (MRDD) and Positron<br>Emission Tomography (PET) scanner. The system is<br>designed for whole body oncology, neurology and<br>cardiology examinations. The SIGNA PET/MR system<br>provides simultaneous acquisition of high resolution<br>metabolic and anatomic information from the two major<br>components of each system (MR and PET). Additional<br>components of the system include: a detachable patient<br>table and both the acquisition and processing<br>workstations with associated software.<br><br>The SIGNA PET/MR includes a 3.0T superconducting<br>magnet, gradient coil and a transmit/receive whole body<br>radiofrequency coil. The system includes patient<br>adaptable RF shimming capabilities. The SIGNA PET |
| | detectors are integrated into the MR bore. This allows for<br>simultaneous, precisely aligned whole body MR and PET<br>acquisitions. The PET subsystem supports Time of Flight<br>(ToF) coincidence detection. The SIGNA PET/MR<br>software is used for patient management, data<br>management, scan control, image reconstruction and<br>image archival and evaluation. All images conform to<br>DICOM imaging format requirements. |
| | The modifications to this system include the MotionFree<br>Brain software feature, which allows users the flexibility to<br>correct patient head motion using the acquired PET data<br>from the exam, without the need of external tracking<br>devices, additional MR data, or other motion tracking data<br>schemes. |
| Indications for Use: | The SIGNA PET/MR system combines magnetic<br>resonance diagnostic devices (MRDD) and Positron<br>Emission Tomography (PET) scanners that provide<br>registration and fusion of high resolution physiologic and<br>anatomic information, acquired simultaneously and<br>isocentrically. The combined system maintains<br>independent functionality of the MR and PET devices,<br>allowing for single modality MR and / or PET imaging. |
| | These systems are intended to be utilized by<br>appropriately trained health care professionals to aid in<br>the detection, localization, and diagnosis of diseases and<br>disorders. MR is intended to produce transverse, sagittal,<br>coronal and oblique cross-sectional MR images,<br>spectroscopic images and/or spectra, and displays the<br>internal structure and/or function of the human body.<br>Other physical parameters derived from the images<br>and/or spectra may also be produced. Depending on the<br>region of interest, approved contrast agents may be used,<br>as described in their labeling. This system may also be<br>used for imaging during interventional procedures when<br>performed with MR compatible devices, such as MR safe<br>biopsy needles. |
| | PET images and measures the distribution of PET<br>radiopharmaceuticals in humans to aid the physician in<br>determining various metabolic (molecular) and<br>physiologic functions within the human body for<br>evaluation of diseases and disorders such as, but not<br>limited to, cardiovascular disease, neurological disorders<br>and cancer. |
| | The combined system utilizes the MR for radiation-free<br>attenuation correction maps for PET studies. The system |
| | provides inherent anatomical reference for the fused PET<br>and MR images due to precisely aligned MR and PET<br>image coordinate systems. |
| Comparison of<br>Technological<br>Characteristics: | The SIGNA PET/MR with the proposed software feature<br>employs the same fundamental technology as the<br>predicate device. |
| | The SIGNA PET/MR has been modified to include the<br>MotionFree Brain feature. The user interface provides<br>operators of the system with an option for enabling this<br>feature. MotionFree Brain derives head motion<br>information from the PET data; there is no need for<br>external tracking devices, additional MR data, or<br>additional PET data collection. The feature measures<br>and incorporates the rigid-body motion information into<br>the PET reconstruction, correcting the position of each<br>coincidence event to account for patient head motion. |
| | These technological differences do not raise any different<br>questions regarding safety and effectiveness. Both<br>devices must allow for an effective method to setup an<br>appropriate scan prescription. The performance data<br>described in this submission include results of both bench<br>testing and clinical testing that show the performance of<br>the SIGNA PET/MR compared to the predicate device. |
| Summary of Nonclinical<br>Testing: | The following quality assurance measures were applied<br>to the development of the device:<br>• Risk Analysis<br>• Requirements Reviews<br>• Design Reviews<br>• Verification (functional testing for subsystem and<br>system integration)<br>• Validation (simulated use testing) |
| | Non-clinical tests have been summarized in the<br>verification and validation testing. The testing was<br>completed with passing results per the pass/fail criteria<br>defined in the test cases. This supports<br>substantial equivalence to its predicate because the<br>software feature was developed under quality assurance<br>Design Controls. Performance testing on phantoms as<br>part of non-clinical testing demonstrated MotionFree<br>Brain achieved performance claims for Quantitation,<br>Temporal Resolution, and Spatial Accuracy. |
| | |
| Summary of Clinical<br>Testing: | An external reader evaluation study was performed for<br>MotionFree Brain. The retrospective blind study involved<br>board-certified readers who were asked to evaluate<br>randomly labeled cases that were reconstructed with and<br>without MotionFree Brain. Verification documents, validation documents, and test<br>reports have been provided for more details. Testing<br>demonstrated that the device is as safe, as effective, and<br>performs as well as or better than the predicate device.<br>No new questions of safety and effectiveness were raised<br>during nonclinical testing. |
| | The readers were blinded to feature use (e.g. whether<br>feature was enabled or disabled), report case history, as<br>well as to the assessments made by the other readers.<br>The readers were asked to complete an assessment,<br>including additional commentary, and report their<br>preference on the pair of image series presented. |
| | Clinical testing confirms that MotionFree Brain can be<br>used safely and effectively in a clinical setting. |
| Conclusion Drawn from<br>Performance Testing: | The SIGNA PET/MR with the modified software feature<br>has the same intended use as the predicate. This 510(k)<br>submission includes information on the technological<br>characteristics of the proposed software feature, as well<br>as performance data demonstrating that the feature is as<br>safe and effective as the predicate, and does not raise<br>different questions of safety and effectiveness. |
| | In conclusion, GE Healthcare considers the SIGNA<br>PET/MR to be at least as safe and effective, and its<br>performance is substantially equivalent to the predicate<br>device. |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. Three water droplets or wave-like shapes surround the circle, adding a dynamic element to the design. The logo is simple, recognizable, and associated with the multinational conglomerate.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three white water droplet shapes surrounding the circle. The logo is simple and recognizable, and it is often used to represent the company's brand.
{7}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are three water droplet-like shapes surrounding the circle, one at the top and one on each side. The logo is simple and recognizable, representing the company's brand identity.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.