Statistically significant increase in nodule detection performance (sensitivity, FPPI, and JAFROC FOM) for all readers.
—
—
Clinical study containing normal and diseased images.
>1 (radiologists)
Indications for Use
The Auto Lung Nodule Detection is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 10 to 30 mm in size. It is designed to aid the physician to review the PA chest radiographs of adults as a second reader and be used as part of S-Station, which is operation software installed on Samsung Digital X-ray Imaging systems. Auto Lung Nodule Detection cannot be used on the patients who have lung lesions other than abnormal nodules.
Device Story
Auto Lung Nodule Detection (ALND) is a CADe software integrated into Samsung S-Station imaging systems. It processes DICOM PA chest radiographs to identify suspected pulmonary nodules (10-30 mm). The device functions as a second reader; it marks regions of high nodule probability on the image after the physician's initial interpretation. The physician reviews these CAD-generated circular marks to confirm or adjust their findings. By providing automated detection, the device aims to improve radiologist sensitivity and diagnostic performance in identifying lung nodules, potentially facilitating earlier detection of lung cancer.
Clinical Evidence
Clinical study evaluated ALND efficacy in assisting radiologists with pulmonary nodule detection on digital chest radiographs. Study used a test dataset of normal and diseased images. Readers marked regions of suspicion and provided confidence scores before and after reviewing ALND results. Performance metrics included sensitivity, false positives per image (FPPI), and JAFROC figure of merit (FOM). Results showed statistically significant improvement in reader performance with ALND assistance.
Technological Characteristics
Software-based CADe; operates on standard PC/Windows platforms. Inputs: DICOM PA chest radiographs. Outputs: Circular markers on images. Technology: Machine learning. Integrated into Samsung S-Station software on Samsung Digital X-ray Imaging systems. Complies with ISO 14971, IEC 62304, and IEC 62366-1 standards.
Indications for Use
Indicated for adult patients undergoing PA chest radiography to assist physicians in identifying and marking suspected pulmonary nodules (10-30 mm). Contraindicated for patients with lung lesions other than abnormal nodules.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 31, 2021
Samsung Electronics Co., Ltd. % Jaesang Noh Senior Professional, Regulatory Affairs 129, Samsung-ro. Yeongtong-gu Suwon-si, Gyeonggi-do 16677 REPUBLIC OF KOREA
Re: K201560
Trade/Device Name: Auto Lung Nodule Detection Regulation Number: 21 CFR 892.2070 Regulation Name: Medical image analyzer Regulatory Class: Class II Product Code: MYN Dated: July 15, 2021 Received: July 20, 2021
## Dear Jaesang Noh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K201560
Device Name Auto Lung Nodule Detection
### Indications for Use (Describe)
The Auto Lung Nodule Detection is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 10 to 30 mm in size. It is designed to aid the physician to review the PA chest radiographs of adults as a second reader and be used as part of S-Station, which is operation software installed on Samsung Digital X-ray Imaging systems. Auto Lung Nodule Detection cannot be used on the patients who have lung lesions other than abnormal nodules.
Type of Use (*Select one or both, as applicable*)
| <div> <span style="text-decoration: overline;"></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span style="text-decoration: overline;"></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle.
510(k) Premarket Notification - Traditional
# Section 5: 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92
- 1. Date: August 26, 2021
### Submitter 2.
- Company Name: SAMSUNG ELECTRONICS Co., Ltd. A.
- B. Address: Address: 129, Samsung-ro, Yeongtong-qu, Suwon-si, Gyeonggi-do, 16677, Republic of Korea
#### 3. Primary Contact Person
- Name: JAESANG NOH A.
- B. Title: Regulatory Affairs, Senior Professional
- ﻥ Phone Number: +82-2-2193-2444
- D. FAX Number: +82-31-200-6401
- ட் E-Mail: jaesang.noh@samsung.com
#### Secondary Contact Person 4.
- Name: Ninad Gujar A.
- B. Title: Vice President, Regulatory Affairs & Quality Control
- Phone Number: 978-564-8503 ﻥ
- D. FAX Number: 978-560-0602
- ட E-Mail: nqujar@samsungneurologica.com
#### Proposed Device 5.
- A. Trade Name: Auto Lung Nodule Detection
- B. Device Name: Auto Lung Nodule Detection
- C. Common Name: Medical image analyzer
- D. Classification Name: Medical image analyzer
- ட Product Code: MYN
- F. Regulation: 21 CFR 892.2070
### Predicate Device 6.
- A. Manufacturer: Riverain Medical Group, LLC
- Trade Name: ClearRead Detect B.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold, sans-serif font. The word is set against a blue, oval-shaped background.
510(k) Premarket Notification - Traditional
- ﻥ Classification: Medical image analyzer
- D. Product Code: MYN
- ட 510(k) Number or PMA Number: P000041
- F. Decision Date: July 12, 2001
| | Predicate Device |
|---------------------|------------------------|
| Trade Name | ClearRead Detect |
| Classification Name | Medical image analyzer |
| Product Code | MYN |
| Regulation | 21 CFR 892.2070 |
| PMA# | P000041 |
| Decision Date | July 12, 2001 |
*The product code MYN has been reclassified from Class II since February 21, 2020.
## 7. Device Description
Auto Lung Nodule Detection is a Computer-Aided Detection (CADe) device that is designed to perform CAD processing in Chest X-ray images for indication of locations for high nodule probability, which has an effective detection sizes from 10 mm to 30 mm.
Auto Lung Nodule Detection receives images acquired with SAMSUNG Digital X-ray Imaging Systems as an input and identifies suspected nodules, and then sends information of suspected nodules to the visualization part of S-Station, which is installed on all kinds of SAMSUNG Digital X-ray Imaging Systems, to generate output images with circular marks. The CAD performed images, are displayed on the screen by S-Station without defeat of original images and used as a second reader only after the initial read is completed.
### Intended Use 8.
The Auto Lung Nodule Detection is computer-aided detection software to identify and mark regions in relation to suspected pulmonary nodules from 10 to 30 mm in size. It is designed to aid the physician to review the PA chest radiographs of adults as a second reader and be used as part of S-Station, which is operation software installed on Samsung Digital X-ray Imaging systems. Auto Lung Nodule Detection cannot be used on the patients who have lung lesions other than abnormal nodules.
### 9. Summary of Technological characteristic of the proposed device compared with the predicate device
Samsung believes that the proposed device is substantially equivalent to the predicate device because some differences in the design and features is considered low risk and do not raise new questions on safety and effectiveness of the proposed device for its intended use based on the non-clinical and clinical testing.
### A. Comparing with Predicate Device
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The word is placed on a blue, oval background.
510(k) Premarket Notification - Traditional
| | | Predicate Device | Proposed Device |
|--------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device Name | ClearRead Detect | Auto Lung Nodule<br>Detection |
| | Manufacture | Riverain Medical Group,<br>LLC | Samsung Electronics |
| | PMA Number | PMA P000041 | - |
| | Indication for use | CLEARREAD DETECT is<br>a computer-aided detection<br>(CAD) system intended to<br>identify and mark regions<br>of interest (ROIs) on<br>digitized frontal chest<br>radiographs. It identifies<br>features associated with<br>solitary pulmonary nodules<br>from 9 to 30 mm in size,<br>which could represent<br>early-stage lung cancer.<br>The device is intended for<br>use as an aid only after the<br>physician has performed<br>an initial interpretation of<br>the radiograph. | The Auto Lung Nodule<br>Detection is computer-<br>aided detection software to<br>identify and mark regions<br>in relation to suspected<br>pulmonary nodules from 10<br>to 30 mm in size. It is<br>designed to aid the<br>physician to review the PA<br>chest radiographs of adults<br>as a second reader and be<br>used as part of S-Station,<br>which is operation software<br>installed on Samsung<br>Digital X-ray Imaging<br>systems. Auto Lung Nodule<br>Detection cannot be used<br>on the patients who have<br>lung lesions other than<br>abnormal nodules. |
| | Intended users | Physician | Physician |
| | Intended body part | Chest | Chest |
| | Imaging modality | X-ray | X-ray |
| | Key feature | Identification of lung<br>nodules | Identification of lung<br>nodules |
| | Technology | heuristic decision rules,<br>artificial neural network, and<br>fuzzy logic | Machine learning |
| | Operating systems | Standard PC/Windows | Standard PC/Windows |
| Input | Image<br>type | DICOM | DICOM |
| | Applicable<br>Protocols | Chest PA/AP | Chest PA |
| Output | Output<br>type | ROI marked on the<br>duplicated input image | Information for ROI to be<br>marked on the duplicated<br>input image |
| | Marker<br>type/size | Circular/Adjustable | Circular/Fixed |
| | Report | The number of findings | The number of nodule<br>markers |
| | | | |
| | Reader workflow | Second reader workflow | Second reader workflow |
## 10. Safety and Effectiveness Information
Software design description, hazard analysis, and labeling information are provided in support of
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white, bold, sans-serif font. The word is set against a blue, oval-shaped background. The oval is tilted slightly upwards from left to right.
## 510(k) Premarket Notification - Traditional
this premarket notification for the proposed device. The device labeling contains instructions for use with cautions to provide for safe and effective use of the device. The results of the hazard analysis with appropriate risk controls indicate the proposed device is of moderate level of concern, as per the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Device" issued on May 11, 2005.
## 11. Software Verification and Validation
#### Non-clinical Testing A.
Non-clinical tests were conducted for Auto Lung Nodule Detection during the device development in accordance with Samsung verification and validation process complied with the FDA Quality System Regulations, ISO 13485 requirements, and the following standards.
- -ISO14971:2007, Medical devices - Application of risk management to medical devices (2nd Ed.)
- IEC62304:2006, Medical devices Software life cycle processes -
- -IEC62366-1:2015, Medical device – Part1: Application of usability engineering to medical devices
Verification and validation activities for Auto Lung Nodule Detection were conducted to provide evidences that the design meets user needs and intended use and application specification. The testing results support that all the software specifications have met the acceptance criteria and the claims of substantial equivalence.
### B. Clinical Performance Testing
Clinical evaluation was performed to validate the clinical efficacy of Samsung Auto Lung Nodule Detection (ALND) in helping radiologists find pulmonary nodules on digital chest radiographs. In this clinical study, nodule detection performances of human readers were measured using a test dataset containing both normal and diseased images. Readers were asked to mark their region of nodule suspicion on the images while also providing confidence scores on each decision they have made. After independent reading, readers were allowed to adjust their confidence scores after reviewing the ALND's detection results. Nodule detection performances before and after ALND were measured via sensitivity, false positives per image (FPPI), and jackknife alternative free response receiver operating characteristic (JAFROC) figure of merit (FOM). The results have demonstrated that all readers' nodule detection performances using the proposed device have increased with statistical significance. Therefore, the proposed device could provide potential assistance for radiologists in the interpretation and detection of pulmonary nodules when used as an assistant tool.
### 12. Conclusions
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the text "SAMSUNG ELECTRONICS Co., Ltd." The text is in a bold, sans-serif font. The text is black and the background is white.
Image /page/7/Picture/1 description: The image shows the Samsung logo. The logo consists of the word "SAMSUNG" in white letters on a blue oval background. The oval is tilted slightly to the right.
510(k) Premarket Notification - Traditional
Results of all conducted testing were found acceptable in supporting the claim of substantial equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.