The VersaVision is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.
Device Story
Cabinet x-ray system for high-detail imaging of surgically excised medical specimens. Input: harvested tissue specimens. Operation: x-ray tube (15-50 kV, tungsten target) generates images with up to 8X geometric magnification via <15 micron focal spot. Output: high-resolution film or DICOM-compliant digital images. Used in clinical settings (OR or nearby) by medical staff to verify correct tissue excision during biopsy. Enables faster case completion; reduces patient time under examination; potentially limits patient recalls. Integrated software manages image acquisition, storage, printing, and modality worklist.
Clinical Evidence
Bench testing only. Device tested for compliance with 21 CFR 1020.40 (Cabinet X-ray Systems), 21 CFR 1040.10/1040.11 (Laser performance standards), European EMC Directive, and IEC 61010 (Safety, 3rd edition).
Indicated for patients undergoing biopsy procedures requiring rapid verification of excised tissue from various anatomical regions.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Faxitron BioVision Plus Specimen Radiography System (K153583)
Bioptics BioVision Specimen Radiography System (K091558)
Faxitron PathVision Specimen Radiography System (K122428)
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 18, 2017
Faxitron Bioptics LLC % Mr. Douglas Wiegman VP Engineering 3440 East Britannia Drive, Suite 150 TUCSON AZ 85706
Re: K170786 Trade/Device Name: VersaVision Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MWP Dated: June 22, 2017 Received: June 23, 2017
Dear Mr. Wiegman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
**Indications for Use**
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| <i>See PRA Statement below.</i> |
| 510(k) Number (if known) | K170786 |
|--------------------------|-------------|
| Device Name | VersaVision |
Indications for Use (Describe)
The VersaVision is a Cabinet x-ray system that is used to provide film and/or digital x-ray images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.
| Type of Use (Select one or both, as applicable) | | | | | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. | | | | |
| | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | | |
| burden, to: | The burden time for this collection of information is estimated to average 79 hours per response,<br>estimate or any other aspect of this information collection, including suggestions for reducing this<br>Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff(@fda.hhs.gov | | including the time to review instructions, search existing data sources, gather and maintain the data<br>needed and complete and review the collection of information. Send comments regarding this burden | | |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number." | | | | |
| FORM FDA 3881 (8/14) | | Page 1 of 1 | | PSC Publishing Services (301) 443 | |
6740
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "faxitron" in a sans-serif font. The letters are a teal color, except for the "x", which is orange and has a pattern of small circles emanating from the top left corner. The word is written in all lowercase letters.
# Premarket Notification 510 (k) Summary As required by 21 CFR 807.92
Device name - as required by 807 . 92 (a)(2):
Trade Name: VersaVision
| Common/Classification Name: | Specimen x-ray System/Cabinet X-ray System |
|-----------------------------|--------------------------------------------|
|-----------------------------|--------------------------------------------|
Classification Regulation: 21 CFR 892.1680
Device Class: Class II
Product Code (Precode): MWP
Radiologic Devices Panel
Company Name: Faxitron Bioptics, LLC
Company Address : 3440 East Britannia Drive ,Suite150 Tucson, Arizona 85706
Contact:
Panel:
VP Engineering
Preparation Date: July 13, 2017
LEGALLY MARKETED PREDICATE DEVICES -as required by 21 CFR 807.92(a)(3)
The Faxitron VersaVision Digital Specimen Radiography System is substantially equivalent to the following predicate devices:
- A. Faxitron BioVision Plus Specimen Radiography System (K153583)
Douglas C. Wiegman ,
- B. Bioptics BioVision Specimen Radiography System (K091558)
- C. Faxitron PathVision Specimen Radiography System (K122428)
{4}------------------------------------------------
DEVICE DESCRIPTION - as required by 21 CFR 807.92(a)(4)
The Faxitron VersaVision Specimen Radiography System is a Cabinet X-ray System specifically designed to provide high detail radiographic imaging of surgically excised medical specimens. The exceptionally high magnification capability (up to 8X) from the <15 micron focal spot with optimized cabinet geometry and the superior contrast available from the low kV capability provides enhanced film and/or digital imaging performance. This device supports radiographic film sizes up to 15 x 24 cm and can be configured to acquire high resolution, DICOM compliant, digital x-ray images up to 15 x 24 cm in size through the use an integrated detector and Faxitron Vision Specimen Radiography software.
DEVICE CLAIMS - as required by 807.92(a)
The Faxitron VersaVision Specimen Radiography System is a fully shielded Cabinet X- ray System that has been designed to comply with 21 CFR 1020.40. The system allows up to 8 times geometric magnification of excised specimens with minimal geometric distortion through the use of a focal spot size that is less than 15 microns. The x-ray coverage of the device allows the use of radiographic film sizes up to 15 x 24 cm. The device can also be configured to provide high resolution, DICOM compliant, digital images through the use an integrated digital camera that is up to 15 x 24 cm in size, and Faxitron Vision Specimen Radiography software. The Faxitron Software supports the DICOM Store, Print and Modality Worklist services.
| DEVICE TECHNICAL SPECIFICATIONS - as required by 807.92(a)(4) |
|-----------------------------------------------------------------|
| Cabinet Specifications: |
| Energy Range: 15-50 kV Constant Potential |
| Tube Current: 1.0 mA max, iso-watt limited to 11.5 watts |
| Focal Spot Size: < 15 microns, 50 micron optional |
| X-ray beam divergence : 40 deg Min. |
| Optional divergence 23 deg. Min. |
| Target material: Tungsten |
| Beryllium Window Filtration: 0.010" |
| X-ray Coverage: 30 x 30 cm |
| Power: 100-240 VAC, ±10%, 50/60 Hz, 150 VA |
| Dimensions: |
| External: 21" W x 24" D x 30" H |
| Internal: 12" W x 12" D x 14" H |
| Weight: 170 lbs. |
| |
| Digital System Specifications: |
| Active image Area: 15 x 24 cm |
| Options: 6 x 15 cm, 12 x 15 cm, 10 x 15 cm |
| Typical Spatial resolution: 10-20 lp/mm depending on detector |
| DICOM 3.0 compliant software includes Store, Print and Modality |
| Work list |
| |
| Network ready workstation and monitor included |
{5}------------------------------------------------
INTENDEDUSE = as required by 807.92(a)(5)
Indications for Use: The VersaVision is a Cabinet x-ray system that is used to provide film and/or digital x-rav images of harvested specimens from various anatomical regions in order to provide rapid verification that the correct tissue has been excised during the biopsy procedure. Doing the verification directly in the same room or nearby enables cases to be completed faster, thus limiting the time the patient needs to be under examination. Specimen radiography can potentially limit the number of patient recalls.
#### LEVEL OF CONCERN - as required by recent FDA guidance
Faxitron has determined that the submitted device has a "moderate" software Level of Concern and has provided a documented record as part of this submission.
#### TECHNOLOGICAL CHARACTERISTICS SUMMARY- as required by 807.92(a)(6)
The VersaVision Specimen Radiography System has the same indications for use, general configuration, and principles of operation as the predicate devices listed above. The technological characteristics of the VersaVision Specimen Radiography System have been compared to the predicate devices sited and is covered in detail in the Substantial Equivalence section of this submission.
### NONCLINICAL PERFORMANCE DATA TESTING AND REVIEW- as required by 807.92(b)(1)
The VersaVision Specimen Radiography System is a Cabinet X-rav System and has been designed and tested to comply with the performance standards set forth in 21 CFR 1020.40 Cabinet X-ray Systems and Laser performance standards 21 CFR1040.10 and 21 CFR 1040.11.Testing and performance data perfaining to these standards has been included as part of the submission. This device has also been successfully tested to the European EMC Directive and Safety testing to IEC 61010 third edition.
Substantial Equivalence Summary
The VersaVision Specimen Radiography System has the same indications for use as the predicate devices sited. The technical characteristics of the VersaVision are very similar to the predicate devices. The x-ray tube and detector technologies used in the Versal/ision are the same as that used in the predicate devices. One major difference that we believe allows the VersaVision Radiography System to perform better than the predicate devices sited is the size of the focal spot (<15 microns), greater size of the digital detector and the optimized cabinet geometry which allows greater x-ray coverage and higher geometric magnification of excised specimens with minimal geometric distortion.
## CONCLUSIONS- as required 807.92(b)(3)
We conclude that the documentation and testing included in this submission indicates that the Versa Vision Specimen Radiography System is safe and effective and substantially equivalent to the predicate devices sited.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.