VIVOSONIC DIANOSTIC ULTRASOUND IMAGING COUPLING MEDIA
K020956 · Microtek Medical, Inc. · MUI · May 16, 2002 · Radiology
Device Facts
Record ID
K020956
Device Name
VIVOSONIC DIANOSTIC ULTRASOUND IMAGING COUPLING MEDIA
Applicant
Microtek Medical, Inc.
Product Code
MUI · Radiology
Decision Date
May 16, 2002
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
VivoSonic™ will be used during invasive medical diagnostic ultrasound imaging to couple sound waves between patient and medical imaging electronics and to lubricate the insertion and passage of imaging devices, such as ultrasound transducers and endoscopes. VivoSonic™ is intended for use in conjunction with transcutaneous ultrasound image guided biopsy and aspiration, intraoperative ultrasound imaging, endocavity ultrasound imaging and opthalmic ultrasound imaging. VivoSonic™ is unit dose packaged, sterilized and intended for use in all diagnostic ultrasound procedures which currently use and ultrasound coupling gel or fluid alone or in combination with a latex, non-latex, polyurethane or polyethylene transducer cover where sterility and/or in vivo biocompatibility are required.
Device Story
VivoSonic™ is a sterile, in vivo biocompatible ultrasound coupling medium and lubricant. It functions by facilitating acoustic coupling between the patient's body and the ultrasound transducer, while simultaneously providing lubrication for the insertion and passage of imaging devices such as endoscopes and transducers. The device is supplied in unit-dose packaging to maintain sterility. It is intended for use by clinicians in various diagnostic ultrasound settings, including invasive procedures like biopsies, aspirations, intraoperative, endocavity, and ophthalmic imaging. By ensuring effective sound wave transmission and reducing friction during device insertion, it supports accurate diagnostic imaging and patient comfort during invasive ultrasound-guided procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on the device's intended use, formulation, and performance characteristics being equivalent to legally marketed predicate devices.
Technological Characteristics
Sterile, in vivo biocompatible ultrasound coupling gel/fluid. Unit dose packaging. Formulated for acoustic coupling and lubrication. Compatible with latex, non-latex, polyurethane, and polyethylene transducer covers.
Indications for Use
Indicated for use during invasive diagnostic ultrasound imaging procedures, including transcutaneous image-guided biopsy/aspiration, intraoperative, endocavity, and ophthalmic imaging. Used to couple sound waves between patient and transducer and to lubricate imaging devices. Intended for procedures requiring sterility and in vivo biocompatibility, including use with transducer covers.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
Aquasonic 100 sterile (Parker Laboratories, Inc.)
Ultra/Phonic sterile (Pharmaceutical Innovations)
Natural Image (Sontotech, Inc.)
Submission Summary (Full Text)
{0}------------------------------------------------
MAY 1 6 2002
KO20956
## Premarket Notification (510k) Summary
Microtek Medical, Inc. 512 Lehmberg Road Columbus, Mississippi 39702
Telephone: 662-327-1863
Contact Person: Thomas Bonner Date compiled: March 15, 2002
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo for Microtek Medical, Inc. The logo features a stylized "M" with a circle above it, and the words "MICROTEK MEDICAL, inc." printed below. To the left of the "M" is a faded image of the words "IT SYSTEM".
#### Premarket Summary
#### Device Name
Proprietary name: VivoSonic™
Common name: In vivo biocompatible, sterile medical diagnostic ultrasound imaging coupling media.
Establishment Registration Number: 1043582 (Microtek Medical, Inc.)
#### Classification: II
#### Statement of Substantial Equivalence
VivoSonic™ is equivalent in ultrasound imaging coupling performance to:
- 1. Aquasonic 100 sterile (Parker Laboratories, Inc.)
- 2. Ultra/Phonic sterile (Pharmaceutical Innovations)
- 3. Natural Image (Sontotech, Inc.)
#### Description of Device
VivoSonic™ is an in vivo biocompatible sterile ultrasound couplant and lubricant that couples or conducts ultrasound between the body and the ultrasound transducer and electronics.
The VivoSonic™ formulation is in vivo biocompatible with tissue and body fluids and is recognized as safe for oral administration (1)
#### Intended Use
VivoSonic™ will be used during invasive medical diagnostic ultrasound imaging to couple sound waves between patient and medical imaging electronics and to lubricate the insertion and passage of imaging devices, such as ultrasound transducers and endoscopes.
VivoSonic™ is intended for use in conjunction with transcutaneous ultrasound image guided biopsy and aspiration, intraoperative ultrasound imaging, endocavity ultrasound imaging and opthalmic ultrasound imaging.
VivoSonic™ is unit dose packaged, sterilized and intended for use in all diagnostic ultrasound procedures which currently use and ultrasound coupling gel or fluid alone or in combination with a latex, non-latex, polyurethane or polyethylene transducer cover where sterility and/or in vivo biocompatibility are required.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The eagle faces to the left. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### MAY 1 6 2002
Mr. Thomas B. Bonner , Jr. Vice President, Regulatory Affairs/Quality Assurance MICROTEK Medical, Inc. 512 Lehmberg Road COLUMBUS MS 39702
Re: K020956
Trade/Device Name: VivoSonic™ Diagnostic Ultrasound Imaging Coupling Media Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic transducer Regulatory Class: II Product Code: 90 MUI Dated: March 15, 2002 Received: March 25, 2002
Dear Mr. Bonner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
and the same of the same of the same of the same of the same of the same of the states of the states of the states of the states of the states of the states of the states of
,
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number
:
| Device Name: | Microtek Medical, Inc. VivoSonic Diagnostic Ultrasound Imaging Coupling Media |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use Statement: | VivoSonic Media is used during invasive medical diagnostic ultrasound imaging to couple sound waves between the patient and the medical imaging electronics, and to lubricate the insertion and passage of imaging devices. |
| | VivoSonic Media is used in conjunction with transcutaneous ultrasound imaging guided biopsy and apsiration, intraoperative ultrasound imaging, endocavity ultrasound imaging, and opthalmic ultrasound imaging. |
| | VivoSonic Media is intended for use in all diagnostic ultrasound procedures which currently use an ultrasound gel or fluid alone or in combination with a latex, non-latex, polyurethane or polyethylene transducer cover where sterility and vivo biocompatibility are required. |
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | (Division Sign-Off) | |
|--------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------|
| | Division of Cardiovascular, Respiratory, And Neurological Devices | |
| | 510(k) Number | |
| Prescription Use (per 21CFR 801.109) | <div> <table style="border: none;"><tr><td>✓ or Over-The-Counter Use</td></tr></table> </div> | ✓ or Over-The-Counter Use |
| ✓ or Over-The-Counter Use | | |
| | (Division Sign-Off) | |
| | Division of Reproductive, Abdominal, and Radiological Devices | |
| | 510(k) Number R020956 | |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.