The Sectra MicroDose Mammography L30 is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The Sectra MicroDose Mammography L30 is intended to be used in the same clinical applications as traditional film/screen systems.
Device Story
Full-field digital mammography system; utilizes photon counting technology with crystalline silicon strip detectors; x-ray tube with tungsten target and aluminum filtration; multi-slit scanning technique to reduce scattered radiation. System includes gantry, acquisition workstation, monitor, keyboard, and mouse. Operates in manual, automatic, or SmartAEC exposure modes; SmartAEC continuously adjusts exposure based on transmission at leading detector edge. Used in clinical settings by trained personnel to acquire, process, and transfer digital mammographic images to PACS for diagnostic review. Enables high detection efficiency and noise rejection; provides dose-efficient imaging for breast cancer screening and diagnosis.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated via non-clinical testing including sensitometric response, spatial resolution, noise analysis, DQE, dynamic range, AEC performance, phantom testing, and radiation dose. Image attribute evaluation conducted per FDA Class II Special Controls Guidance Document (Nov 2010) confirmed acceptable image quality for clinical use.
Technological Characteristics
Full-field digital mammography system; photon counting technology using crystalline silicon strip detectors; x-ray tube with tungsten target and aluminum filtration; multi-slit collimator. Connectivity to PACS. Testing performed per IEC standards.
Indications for Use
Indicated for screening and diagnosis of breast cancer in patients requiring mammographic imaging, intended for use in clinical applications equivalent to traditional film/screen systems.
Regulatory Classification
Identification
A full-field digital mammography system is a device intended to produce planar digital x-ray images of the entire breast. This generic type of device may include digital mammography acquisition software, full-field digital image receptor, acquisition workstation, automatic exposure control, image processing and reconstruction programs, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Full-Field Digital Mammography System.”*See* § 892.1(e) for the availability of this guidance document.
Predicate Devices
FUJI'S COMPUTED RADIOGRAPHY MAMMOGRAPHY SUITE (FCRMS) (P050014)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "SECTRA" in large, bold, black letters. The letters are slightly distressed, giving them a textured appearance. The font is sans-serif and the letters are evenly spaced.
Image /page/0/Picture/1 description: The image contains a sequence of characters that appear to be handwritten. The sequence starts with the letter 'K', followed by the number '1' repeated twice, then two zeros '00', the letter 'd', and ends with the number '5'. The characters are written in a dark ink, and the handwriting style is somewhat cursive, particularly in the formation of the 'd' and '5'.
## 510(k) Summary (21 CFR 807.92)
APR 2 8 2011
#### A. SUBMITTER INFORMATION
| Submitter's name | Sectra Imtec AB |
|--------------------------------------|-----------------------------------------------------------------------------|
| Address | Teknikringen 20<br>SE-583, 30 Linköping<br>Sweden |
| Establishment registration<br>number | 9615992 |
| Contact person | Alexander Asklöv |
| Telephone | +46-13-23 52 00 |
| FAX | +46-13-21 21 85 |
| Title | Chief Quality Officer of product division<br>mammography at Sectra Imtec AB |
| Date of the summary preparation | 2010-12-30 |
#### B. DEVICE IDENTIFICATION
Device trade name Device common name Classification name Classification product code Device class Regulation code
Sectra MicroDose Mammography L30 Full-Field Digital Mammography X-ray System Full-Field Digital Mammography X-ray System MUE ll
21 CFR 892.1710 ﮐ
:
#### C. DEVICE DESCRIPTION
The Sectra MicroDose Mammography L30 is a type of full-field digital mammography system comprised of an image acquisition system, a gantry and an acquisition workstation computer equipped with a keyboard, a keypad, a mouse, and a monitor. The image acquisition system includes a digital detector of photon counting technology, x-ray tube (with tungsten target and aluminum filtration), high voltage generator, compression paddle(s), and multi-slit collimator. The acquisition workstation is the user interface for preparing and initiating image acquisition, image processing, and image transfer to the desired destination (e.g. PACS) for diagnosis and archiving.
The Sectra MicroDose Mammography 130 detector is based on photon counting technology and consists of a large number of crystalline silicon strip detectors. The technology enables high detection efficiency of photons and efficient rejection of electronic noise. The Sectra MicroDose Mammography L30 uses a multi-slit scanning technique that prevents image degradation caused by scattered radiation by removing photons scattered in the breast and not directed towards the detector. These factors combine into a dose efficient system.
The Sectra MicroDose Mammography L30 provides three exposure modes; manual, automatic (parameters predefined based on compressed breast thickness), and SmartAEC. SmartAEC
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "SECTRA" in large, bold, black letters. The letters have a slightly distressed or textured appearance, as if they were stamped or printed with a worn-out ink. The word is presented in a straightforward, horizontal arrangement, with each letter clearly legible against the white background. The overall impression is one of a strong, industrial-style typeface.
continuously adjusts the exposure based on measured transmission from the leading detector edge.
#### D. INDICATIONS FOR USE
The Sectra MicroDose Mammography L30 is indicated for generating mammographic images that can be used for screening and diagnosis of breast cancer. The Sectra MicroDose Mammography L30 is intended to be used in the same clinical applications as traditional film/screen systems.
#### E. PREDICATE DEVICES
| | Predicate device #1 | Predicate device #2 |
|------------------------|----------------------------------------------------------|----------------------------------------------------|
| Trade Name | FUJI'S COMPUTED RADIOGRAPHY<br>MAMMOGRAPHY SUITE (FCRMS) | SENOGRAPHE 2000D |
| Classification<br>Name | Full Field Digital, System, X-ray,<br>Mammographic | Full Field Digital, System, X-ray,<br>Mammographic |
| Generic Name | Fuji CR-IR348CL Image Console With<br>FCRMS Software | Digital Mammographic X-ray System |
| PMA Applicant | FUJIFILM MEDICAL SYSTEM U.S.A.,<br>INC. | GE HEALTHCARE |
| PMA Number | P050014 | P990066 |
| Product Code | MUE | MUE |
| Advisory<br>Committee | Radiology | Radiology |
#### ப TECHNICAL CHARACTERISTICS
The Sectra MicroDose Mammography L30 System does not employ exactly the same type of detector as the predicate devices and the readout methods are different, however the fundamental process of digital full field mammographic image creation remains the same for all three devices; x-ray photons are captured in the detector and ultimately converted to an electrical signal, forming a digital image.
Performance data from non-clinical testing of the Sectra MicroDose Mammography L30 covering Sensitometric response, Spatial resolution, Noise analysis, Signal-to-noise-ratio transfer – DQE, Dynamic range, Automatic exposure control performance, Phantom testing, Patient radiation dose, and Image erasure, fading and repeated exposure was compared with data from the PMA Summary of Safety and effectiveness of the predicate devices. This comparison showed that the Sectra MicroDose Mammography L30 device performed as well as or better than the predicate devices in all relevant areas. The testing was performed in accordance with generally accepted test methods, e.g. using IEC standards, published factors for dose calculations etc.
An image attribute evaluation was conducted in accordance with the Guidance for Industry and FDA Staff, Class II Special Controls Guidance Document: Full-Field Digital Mammography System [issued: November 5, 2010] which concluded that the images were of sufficiently acceptable quality for clinical mammographic usage.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SECTRA" in a bold, sans-serif font. The letters are filled with a textured pattern, giving them a distressed or grunge appearance. The overall impression is of a strong, industrial, or perhaps even vintage design.
## G. conclusion
The Sectra MicroDose Mammography L30 described in this substantially equivalent to the predicate devices in respects of indication for use and image quality. The proposed and predicate devices utilize similar technology and materials, comparable safety and effectiveness features, and are similar in design and construction. The information submitted in this application shows that none of the technical differences between the systems raises new questions of safety and effectiveness. All collected performance data demonstrate that the devices are substantially equivalent. Our conclusion is that Sectra MicroDose Mammography L30 is as safe and effective as the legally marketed predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circle around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Alexander Asklöv Chief Quality Officer Sectra Imtec AB - Product Division Mammography Teknikrigen 20 Linköping SE-58330 SWEDEN
APR 2 8 2011
Re: K110025
Trade/Device Name: Sectra MicroDose Mammography L30 Regulation Number: 21 CFR 892.1715 Regulation Name: Mammographic x-ray system Regulatory Class: II Product Code: MUE Dated: April 15, 2011 Received: April 19, 2011
Dear Mr. Asklöv:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{5}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket wification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 11 you don't openice as not in Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Mary Patel
Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:
Sectra MicroDose Mammography L30
Indications for Use:
The Sectra MicroDose Mammography L30 is indicated for The Sectra MicroDoso Mic images that can be used for screening and diagnosis of breast cancer.
and Glaghosis of Steast Mammography L30 is intended to be used in the same clinical applications as traditional film/screen systems.
Prescription Use __ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mury S. Potts
(Division Sign-Off)
..............................................................................................................................................................................
(Division Sign Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
610K. K1100.25
Page _1_ of _1_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.