K120629 · Fujifilm Medical System U.S.A., Inc. · MQB · Aug 28, 2012 · Radiology
Device Facts
Record ID
K120629
Device Name
FDR GO FLEX
Applicant
Fujifilm Medical System U.S.A., Inc.
Product Code
MQB · Radiology
Decision Date
Aug 28, 2012
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The FDR Go Flex is intended to capture for display radiographic images of human anatomy. It is intended for use in general radiographic applications wherever conventional film/screen or CR systems may be used. The FDR Go Flex is not intended for mammography, fluoroscopy, tomography, and angiography applications.
Device Story
FDR Go Flex (DR-ID700) is a portable digital flat panel detector (FPD) system for acquiring and digitizing X-ray exposures. It functions as a replacement for conventional film/screen or CR cassettes in general radiography. The system utilizes indirect X-ray conversion via GOS and CsI scintillators and Fujifilm's Irradiated Side Sampling design. It features enhanced portability through smaller size and battery-powered, fully wireless operation between detectors and a Utility Box. The system includes automatic X-ray detection, eliminating the need for direct physical connection to the X-ray generator. Used in clinical environments where radiographic imaging is required; operated by radiology staff. Output consists of digital radiographic images displayed for clinical review. The device benefits patients by providing a portable, wireless digital imaging solution that integrates into existing radiographic workflows.
Clinical Evidence
Bench testing only. The device successfully completed international IEC testing requirements and followed the FDA Draft Guidance Document for Radio-Frequency Wireless Technology in Medical Devices.
Technological Characteristics
Portable digital flat panel detector; indirect X-ray conversion using GOS and CsI scintillators; Irradiated Side Sampling design; battery-powered; wireless communication between detectors and Utility Box; automatic X-ray detection; Class II device.
Indications for Use
Indicated for capturing radiographic images of human anatomy in general radiographic applications. Not for use in mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Wireless/Wired FDR D-EVO (DR-ID600 w/DR-ID601SE) Flat Panel Detector System (K103596)
Wireless/Wired FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE) (K111548)
{1}------------------------------------------------
K120629
Image /page/0/Picture/1 description: The image shows the logo for Fujifilm Medical Systems, USA. The logo is in black and white and features the word "FUJIFILM" in large, bold letters. Below the main logo, the text "Fujifilm Medical Systems, USA" is written in a smaller font.
AUG 28 2012
## 510(k) Summary FDR Go Flex (DR-ID700)
## Flat Panel Detector System
### Date: February 22, 2012
## Submitter's Information:
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, CT, 06902, USA
### Contact Person:
| Name: | Peter Altman |
|------------|-----------------------|
| Title: | Regulatory Consultant |
| Telephone: | (203) 602-3576 |
| Facsimile: | (203) 363-3950 |
## Identification of the Proposed Device:
| Proprietary/Trade Name: | FDR Go Flex (DR-ID700) |
|-------------------------|------------------------------------------------------|
| Classification Name: | Solid State X-ray Imager (Flat Panel/Digital Imager) |
| Regulations Number: | 21 CFR 892.1650 |
| Product Codes: | 90 MQB |
| Device Class: | Class II |
| Review Panel: | Radiology |
| Common Name: | Flat Panel Digital Detector |
#### I. INDICATIONS FOR USE
The FDR Go Flex is intended to capture for display radiographic images of human anatomy. It is intended for use in general radiographic applications wherever conventional film/screen or CR systems may be used. The FDR Go Flex is not intended for mammography, fluoroscopy, tomography, and angiography applications.
#### II. DEVICE DESCRIPTION
The proposed device, the FDR Go Flex (DR-ID700), is a modification of the Wireless/Wired FDR D-EVO (DR-ID600 w/DR-ID601SE) Flat Panel Detector System, K103596, cleared on 3/29/2011 (submitted as a Special 510(k)) and the Wireless/Wired FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE), K111548, cleared on 8/30/2011. The FDR D-EVO Flat Panel Detector System (DR-ID600) cleared on 3/29/2011 was a modification of the FDR D-EVO (DR-ID600) Flat Panel Detector, K100762, cleared on 7/15/2010. The FDR D-
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Fujifilm Medical Systems, USA logo. The logo is in black and white, with the word "FUJIFILM" in large, bold letters. Below the word "FUJIFILM" is the text "Fujifilm Medical Systems, USA" in a smaller font. The logo is simple and clean, and it is likely used to represent the company's brand.
EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE cleared on 8/30/2011 was a modification of the system cleared on 3/29/2011.
The FDR Go Flex is a portable digital FPD system that acquires and digitizes x-ray exposures from standard radiographic systems and is designed to be used in any environment that would typically use a radiographic cassette. However, the FDR Go Flex is more portable than its predecessors given its smaller size and ability for all components to operate using battery power in a fully wireless configuration (no cable connections are required between the FDR Go Flex components, although it will remain possible to physically connect components if the user so desires). Note: Only wireless communication is supported between detectors and the Utility Box.
The indirect X-ray conversion method using GOS and CsI Scintillators, and FUJIFILM's unique Irradiated Side Sampling design, delivering high image quality, remained unchanged in the proposed device.
Since the detectors (DR-ID601SE, DR-ID602SE, and DR-ID611SE) are capable of automatically detecting X-ray, direct connection with the X-ray system is not needed.
#### SUMMARY OF STUDIES III.
The FDR Go Flex flat panel detector system successfully completed international IEC testing requirements. In addition, the FDA Draft Guidance Document, Radio-Frequency Wireless Technology in Medical Devices, issued on January 3, 2007 was followed for the wireless feature.
#### IV. SUBSTANTIAL EQUIVALENCE
The proposed device, the FDR Go Flex (DR-ID700), is a modification of the Wireless/Wired FDR D-EVO (DR-ID600 w/DR-ID601SE) Flat Panel Detector System. K103596, cleared on 3/29/2011 (submitted as a Special 510(k)) and the Wireless/Wired FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE), K111548, cleared on 8/30/2011. The FDR D-EVO Flat Panel Detector System (DR-ID600) cleared on 3/29/2011 was a modification of the FDR D-EVO (DR-ID600) Flat Panel Detector, K100762, cleared on 7/15/2010. The FDR D-EVO Flat Panel Detector System (DR-ID600 w/DR-ID611SE cleared on 8/30/2011 was a modification of the system cleared on 3/29/2011. All detectors (DR-ID601SE, and DR-ID611SE) used with the FDR Go Flex are capable of automatically detecting X-ray, a feature cleared in the latest predicate device.
#### V. CONCLUSION
The FDR Go Flex (DR-ID700) flat panel detector system is substantially equivalent to the cleared predicate devices and conforms to applicable medical device safety standards.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wing shapes, representing the department's mission to protect the health of all Americans. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which is arranged in a circular fashion around the logo.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
AUG 28 2012
Mr. Peter Altman Regulatory Consultant FUJIFILM Medical Systems USA, Inc. 419 West Avenue STAMFORD CT 06902
Re: K120629
Trade/Device Name: FDR Go Flex (DR-ID700) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: July 30, 2012 Received: August 1, 2012
Dear Mr. Altman:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to commerce profit it May 26, 1776, the encordance with the provisions of the Federal Food, DINA and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmelle Act (Act) market the device, subject to the general controls provisions of the Act. The 1 ou may, therefore, market the device, sabject to the enemal registration, listing of general controls provisions of the Act merelate is and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such alin title 21 II your device is Classified (see above) into enas ffecting your device can be found in Title 21, additional contors. Existing major regulations and the 895. In addition, FDA may publish further Code of Peacharts concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA 3 issumes of a succession is a suith other requirements of the Act that FDA has made a decemmation analy other Federal agencies. You must of any Federal Statutes and regulations administer by not limited to: registration and issting of comply with an the Act Stequirements, therauling, carner called in the reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours, Janine M. Mor
Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K |20 €29
Device Name: FDR Go Flex (DR-ID700)
Indications for Use:
The FDR Go Flex is intended to capture for display radiographic images of human anatomy. It is intended for use in general radiographic applications wherever conventional film/screen or CR systems may be used. The FDR Go Flex is not intended for mammography, fluoroscopy, tomography, and angiography applications.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
***(Division Sign-Off)***
Division of Radiological Devices
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.