K103348 · Kub Technologies, Inc. · MQB · Oct 20, 2011 · Radiology
Device Facts
Record ID
K103348
Device Name
DIGIVIEW 250
Applicant
Kub Technologies, Inc.
Product Code
MQB · Radiology
Decision Date
Oct 20, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Intended Use
The DIGIVIEW 250 is indicated for use in generating radiographic images of human anatomy. The device is intended to provide digital x-ray image capture for conventional film/screen radiographic examinations and replace radiographic film/screen systems in all general purpose diagnostic procedures. The device is not intended for mammography applications.
Device Story
Kubtec DIGIVIEW 250 is a CMOS-based solid-state X-ray imager; intercepts X-ray photons passing through anatomy; converts photons into electrical signals; converts signals into digital values. System includes DIGICOM software for image display, analysis (live or captured), storage, and transmission to PACS. Used in clinical settings for general diagnostic radiography; replaces film/screen systems. Provides high-resolution digital images; enables faster readout (750ms) compared to traditional systems. Healthcare providers use output for diagnostic assessment; facilitates clinical decision-making through digital image review.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
CMOS detector; GdOS (gadolinium oxysulfide) scintillator; 192 x 246 mm imaging area; 2000 x 2560 pixel resolution; 96 micron pixel pitch; 14-bit depth; Ethernet interface; 100-240V power supply; DICOM 3.0 compliant software; molded plastic housing with handle. Performance standard: 21 CFR 1020.30.
Indications for Use
Indicated for generating radiographic images of human anatomy for general purpose diagnostic procedures. Contraindicated for mammography applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K183245 — Carestream DRX-1 System with DRX Plus 2530 Detectors · Carestream Health, Inc. · Feb 8, 2019
K220098 — Digital Radiography CXDI-Pro, Digital Radiography D1 · Canon, Inc. · Feb 4, 2022
K130464 — CARESTREAM DRX-1 SYSTEM WITH DRX 2530C DETECTOR · Carestream Health, Inc. · Jun 7, 2013
K222661 — Digital Radiography CXDI-CS01 · Canon, Inc. · Sep 27, 2022
K222855 — Digital Radiography CXDI-Pro, Digital Radiography D1 · Canon, Inc. · Oct 17, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
1103348
OCT 2 0 2011
# 510(k) Summary
#### Prepared: August 2, 2011
#### Submitter:
Company Name: KUB Technologies, Inc.
Company Address: 270 Rowe Avenue, Unit E
Milford, CT 06460
| Contact Person: | Mr. Vikram Butani |
|-----------------|-------------------|
|-----------------|-------------------|
(203) 364-8544 Phone Number:
- Fax Number: (203) 255-7494
#### Proposed Device:
| Reason For 510(k): New Model | |
|------------------------------|-------------------------------|
| Trade Name: | Kubtec |
| Model Name: | Digiview 250 ® |
| Classification Name: | MQB, Solid State X-ray Imager |
| FDA 510(k) #: | K103348 |
#### Predicate Device:
| Trade Name: | Canon |
|----------------------|-------------------------------|
| Model Name: | CXDI-60G |
| Classification Name: | MQB, Solid State X-ray Imager |
| FDA 510(k) #: | K081648 |
# Description Of Device:
The Kubtec Digital Radiography DIGIVIEW 250 ® is a CMOS based solid state x-ray imager which has a 192 x 246 mm imaging area. The Digital Radiography DIGIVIEW 250® imager intercepts X-ray photons after they pass through anatomy and surrounding air and converts the X-ray photons into electrical signals. These resultant electric signals are converted into digital values which are transmitted for remote viewing. The DIGIVIEW 250 ® system features a DICOM 3.0 compliant software, DIGICOM. The DIGICOM software enables the display and analysis of x-ray images; either live (real
{1}------------------------------------------------
time) or previously captured and the storage and transmission of these images to PACS systems.
### Intended Use:
The DIGIVIEW 250 is indicated for use in generating radiographic images of human anatomy. The device is intended to provide digital x-ray image capture for conventional film/screen radiographic examinations and replace radiographic film/screen systems in all general purpose diagnostic procedures. The device is not intended for mammography applications.
## Comparison Table:
Kubtec DIGIVIEW 250 ® (Proposed) To Canon CXDI-60G (Predicate):
| Characteristic: | Proposed Device | Predicate Device |
|---------------------------|------------------------------------------------------------|------------------------------------------------------------|
| Model: | Kubtec DIGIVIEW 250 ® | Canon CXDI-60G |
| Energy Source: | 100-240 V Power Box(6.6 VAC<br>out) | 100-240V "Power Box" |
| Digital Resolution | 2000 x 2560 pixels (5.12 million)<br>5.2 lp/mm | 1464 x 1776 pixels (2.6 million)<br>3.1 lp/mm |
| Pixel Pitch; | 96 microns | 160 microns |
| Bit depth: | 14 bit | 14 bit |
| DICOM: | Dicom compatible | Dicom compatible |
| Image readout: | 750 milliseconds | approx. 3 seconds |
| Interface: | Ethernet | Ethernet |
| Scintillator: | GdOS, gadolinium oxysulfide | GdOS, gadolinium oxysulfide |
| Dynamic range: | 78 dB | 80 dB |
| Operating<br>temperature: | 0 -50 Degrees C | 5 -35 Degrees C |
| Humidity: | 0 - 80% R.H. | 30 -75% R.H. |
| Method of Control: | Software Driven | Software Driven |
| Performance<br>Standard: | 21 CFR 1020.30 | 21 CFR 1020.30 |
| Classification<br>name: | Solid State X-ray imager | Solid State X-ray imager |
| Detector Type: | CMOS | Amorphous Silicon |
| Housing: | Molded plastic with rounded<br>corners and built-in handle | Molded plastic with rounded<br>corners and built-in handle |
{2}------------------------------------------------
## Comparison:
The imaging area of the Kubtec DIGIVIEW 250 @ is 192 x 246 mm as compared to the 230 x 280mm area of the Canon CXDI-60G.
The Kubtec DIGIVIEW 250 ® has a housing measuring 355 x 285 x 24 mm compared to the Canon CXDI-60G housing at 344 x 380 x 22.5 mm and has a comparable shape with integral handle molded into the case.
Both the proposed and predicate device utilize GdOS (gadolinium oxysulfide) scintillators for the fluorescing screen.
The Kubtec DIGIVIEW 250 @ utilizes CMOS as its detector. Canon CXDI-60G uses amorphous silicon detectors. CMOS (complementary metal-oxide semiconductor) technology offers higher sensitivity and signal to noise performance with faster speed and better resolution.
The image readout time on the Kubtec DIGIVIEW 250 ® is 750 milliseconds as compared to the Canon CXDI-60G time of 3 seconds.
The resolution of the Kubtec DIGIVIEW 250 ® is 2,000 x 2,560 pixels (5.12 million) as compared to the predicate Canon CXDI60G of 1464 x 1776 pixels (2.6 million).
The pixel pitch of the Kubtec DIGIVIEW 250 @ is 96 microns compared to 160 microns on the Canon predicate device. The bit depth on both devices is 14 bits.
Both the proposed device, Kubtec DIGIVIEW 250 ® and the predicate device, Canon CXDI 60G are DICOM compatible, software driven, and utilize an Ethemet interface.
## Conclusion:
The performance data, non-clinical testing, and design process demonstrate that the Kubtec DIGIVIEW 250 ® is as safe and effective as the Canon CXDI-60G, and has no new indications for use, thus rendering it substantially equivalent to the predicate device.
Based on the information submitted, similarity to the predicate device (Canon Digital Radiography CXDI-60G), and the results of our design control activities and non-clinical testing, it is the opinion of Kubtec that the Kubtec Digital Radiography DIGIVIEW 250 ® described in this submission is substantially equivalent to the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo includes the department's emblem on the left, followed by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, sans-serif font. The emblem features a stylized human figure within a circular border, representing the department's focus on health and human well-being.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Mr. Vikram Butani President KUB Technologies 270 Rowe Avenue, Unit E MILFORD CT 06460
AUG 2 3 2013
Re: K103348
Trade/Device Name: Kubtec DIGIVIEW 250 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: September 30, 2011 Received: October 3, 2011
Dear Mr. Butani:
This letter corrects our substantially equivalent letter of October 20, 2011.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, dicross provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
medical device-related adverse events) (21 CFR 803); and good manufacturing practice medical device-lefatce adverse vromb) (21 cr or ots ) regulation (21 CFR Part 820). This letter requirements as set form in the quality of stam ( ( ) sogad in your Section 510(k) premarket whil anow you to begin marketing your act 10 m veence of your device to a legally marketed nontication. The PDF imanig of cassion for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFF Parts 801 and II you desire specific advice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to under the MDK regulation (2) "It it 605), product go of only go the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the T ou may outlu of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Kubtec. The logo features a geometric design to the left of the word "kubtec" in a stylized font. Below the word "kubtec" is the tagline "GOING BEYOND THE SURFACE" in smaller, sans-serif font. The overall design is simple and professional.
# Indications for Use
510(k) Number (if known): 长103348
Device Name: Kubtec DIGIVIEW 250
Indications For Use:
The Digiview 250 is indicated for use in generating radiographic images of human anatomy. The device is intended to provide digital x-ray image capture for conventional film/screen radiographic examinations and replace radiographic film/screen systems in all general purpose diagnostic procedures. The device is not intended for mammography applications.
Prescription Use X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, The Office of In Vitro Diagnostics
Mary S. Pisto
(Division Sign-Off)
Office of In 510K
Page 1 of 1
270 Rowe Ave., Milford, CT 06460. Tel: (203) 364-8544; Fax: (203) 255-7494
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.