DIGITAL RADIOGRAPHY CXDI-60G

K081648 · Canon, Inc. · MQB · Jun 24, 2008 · Radiology

Device Facts

Record IDK081648
Device NameDIGITAL RADIOGRAPHY CXDI-60G
ApplicantCanon, Inc.
Product CodeMQB · Radiology
Decision DateJun 24, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1680
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

DIGITAL RADIOGRAPHY CXDI-60G provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

Device Story

The Canon CXDI-60G is a solid-state X-ray imager. It intercepts X-ray photons using a scintillator, which emits visible light to illuminate an array of photodetectors, generating electrical signals that are converted into digital values. The device is a component requiring a separate control PC with designated system software to perform image capture and DICOM transfer. It is used in clinical settings for general diagnostic radiography. The output is a digital image transmitted to a remote viewing site, allowing healthcare providers to perform diagnostic evaluations. The device benefits patients by providing digital radiography capabilities, replacing traditional film/screen systems.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Solid-state X-ray imager; amorphous silicon photodetector array; scintillator-based X-ray to light conversion; 23x28cm imaging area; includes removable fixed grid for scatter reduction; requires external control PC for operation; DICOM connectivity.

Indications for Use

Indicated for digital image capture in general purpose diagnostic radiographic procedures, replacing conventional film/screen systems. Not indicated for mammography.

Regulatory Classification

Identification

A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K081648 ### Section 10: Summary JUN 2 4 2008 # 510(k) Summary Prepared: ## April 30, 2008 #### Submitter: Company Name: Canon USA, Inc. (U.S. agent for Canon Inc.) Company Address: One Canon Plaza Lake Success, NY 11042 Contact Person: Ms. Sheila Driscoll Phone Number: (516) 328-5602 Fax Number: (516) 328-5169 #### Proposed Device: Reason For 510(k): Manufacturer: Trade Name: Model Name: Classification Name: FDA 510(k) #: New Model Canon Inc. Canon CXDI-60G MQB, Solid State X-ray Imager To be assigned #### Predicate Device: Manufacturer: Trade Name: Model Name: Classification Name: FDA 510(k) #: Canon Inc. Canon CXDI-50G 90MQB, Solid State X-ray Imager K031447 Description Of Device: The Canon digital radiography CXDI-60G is a solid state xrray imagers. The CXDI-60G intercepts x-ray photons and the scintillator of the CXDI-60G emits visible spectrum photons that illuminate an array of photodetectors that create an electrical signals. After the electrical signals are generated, it is converted to digital value. The resultant output signal can be transmitted to remote viewing site The Canon digital radiography CXDI-60C is different from CXDI-50G in the following respect - The imaging area of CXDI.60G is changed from 35x43cm to 23x28cm. The principle of the CXDI-60G is the same as the CXDI-50G, with some modifications of its housing in size and shape. The sensor of the CXDI-60G has the same characteristics as the CXDI-50G The CXDI-60G itself is a component without a control PC. Using a general purpose computer with appropriate specifications and the designated system software installed in it, as a control PC, the CXDI-60G achieves performance stated herein(such as image capturing, DICOM transfer and etc.) {1}------------------------------------------------ # Section 10: Summarv DIGITAL RADIOGRAPHY CXDI•60G provides digital image capture for Intended Use: conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. # Descriptive Comparison: The predicate devices is Digital Radiography CXDI-50G cleared under Document Number K031447 on May 21. The CXDI60C's amorphous silicon array specifications (including image size, pixel pitch, number of pixels), imaging principle and intended use are the same as those of CXDI-50G. However, the differences in the design are as follows: ● The imaging area of CXDI-60G is changed from 35x43cm to 23x28cm. A removable, fixed grid is used for all of the CXDI-60G and the CXDI-50G. Those grids are instated inside the sensor housing and used for eliminating the scatter X-ray in exposures that use films. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services. The logo includes the department's name in bold, sans-serif font. To the left of the text is a symbol that is part of the department's logo. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Canon, Inc. % Mr. Jeff D. Rongero Senior Project Manager Underwriters Laboratories, Inc. 12 Laboratory Drive RESEARCH TRIANGLE PARK NC 27709 AUG 2 3 2013 Re: K081648 Trade/Device Name: CXDI-60G Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: June 10, 2008 Received: June 12, 2008 Dear Mr. Rongero: This letter corrects our substantially equivalent letter of June 24, 2008. We have reviewed your Section 510(k) premarket notification of intent to maket the indication indication indication indication indication indication indication indication ind we nave reviewed your becaon 910(t) presidentially equivalent (for the indications felerenced above and nave uctemined the actoredicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Americans, or to commerce pro- 1970, the chaoninen with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). alle Cosment Act (Act) that to hot required to the general controls provisions of the Act. The 1 ou may, deretore, market me act include requirements for annual registration, listing of general condois provided of wactice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such If your device is classified (see as regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean Flease be advised that I DA 3 issumed on a bucce complies with other requirements of the Act that I DA has made a colorimiation administered by other Federal agencies. You must of any rederal statutes and regulations and limited to: registration and listing (21 comply with an the Act 3 requirements, art.uang, or medical device reporting (reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing pactice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820). This letter requirences as bet for in and quality of evice as described in your Section 510(k) premarket witi anow you to ogin finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFF Pars 801 and 11 you desire specific advice of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket 5 150. Those 1200, process 2019.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to under the MDA regulation (27 Of NY at 000), proma Problem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Tou may obtain other general merchanel and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications Statement 5.10(K)Number(if known): . Device Name: CXDI-60G Indications for Use: DIGITAL RADIOGRAPHY CXDI-60G provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications. Prescription Use (Part 21 CFR 801 Subpart D) ાર Over-The-Counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHERT PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation(ODE) Heilert (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number K081648 **Page** 1 **of** 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...