K243316 · Polarean, Inc. · MOS · Nov 19, 2024 · Radiology
Device Facts
Record ID
K243316
Device Name
XENOVIEW 3.0T Chest Coil
Applicant
Polarean, Inc.
Product Code
MOS · Radiology
Decision Date
Nov 19, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Polarean XENOVIEW 3.0T Chest Coil is to be used in conjunction with compatible 3.0T Magnetic Resonance Imaging (MRI) scanners and approved xenon Xe 129 hyperpolarized for oral inhalation for evaluation of lung ventilation in adults and pediatric patients aged 12 years and older.
Device Story
Flexible, single-channel, transmit-receive RF coil; tuned to 123xe frequency (35.33 MHz) for 3.0T MRI systems. Patient wears coil while inhaling hyperpolarized 129Xe gas; coil captures MR signals to image regional lung ventilation distribution. Constructed of flexible circuit board, antenna elements, and electronic components; enclosed in non-conductive, water-rated, flame-rated materials with padding for patient comfort. Fixed matching/tuning; no operator adjustment required. Used in clinical MRI settings; operated by MRI technicians/radiologists. Output is MR image data used by clinicians to assess lung ventilation; aids in clinical decision-making regarding pulmonary function.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including NEMA MS 6-2008 (SNR/uniformity), NEMA MS 8-2016 (SAR), NEMA MS 14-2019 (RF heating), electrical safety (IEC 60601-1, IEC 60601-2-33), electromagnetic compatibility (IEC 60601-1-2), and biocompatibility (ISO 10993-5, ISO 10993-10).
Technological Characteristics
Flexible, single-channel, transmit-receive RF coil; 35.33 MHz operating frequency; quadrature antenna configuration (saddle coil pairs). Materials: foam, fabric, flexible circuit board. Fixed tuning/matching. Passive decoupling for 1H. Standards: NEMA MS 6, MS 8, MS 14; IEC 60601-1, 60601-2-33, 60601-1-2; ISO 10993-1, 5, 10. Non-networked, hardware-based device.
Indications for Use
Indicated for evaluation of lung ventilation in adults and pediatric patients aged 12 years and older using hyperpolarized xenon Xe 129 gas with compatible 3.0T MRI scanners.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the agency's name, "U.S. Food & Drug Administration."
November 19, 2024
Polarean, Inc. % Rita King CEO MethodSense, Inc. 1 Copley Pkwy. Ste. 130 Morrisville, NC 27560
Re: K243316
Trade/Device Name: XENOVIEW 3.0T Chest Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: October 18, 2024 Received: October 22, 2024
### Dear Neil Wadehra:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{3}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
D. K.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K243316
Device Name XENOVIEW 3.0T Chest Coil
#### Indications for Use (Describe)
The Polarean XENOVIEW 3.0T Chest Coil is to be used in conjunction with compatible 3.0T Magnetic Resonance Imaging (MRI) scanners and approved xenon Xe 129 hyperpolarized for oral inhalation of lung ventilation in adults and pediatric patients aged 12 years and older.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> X Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
## 510(k) Summary
### Polarean, Inc.
This 510(k) Summary is in conformance with 21CFR 807.92
| Submitter: | Polarean, Inc.<br>PO Box 14805<br>Research Triangle Park, NC 27709-4805 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------|
| Primary Contact: | Rita King, CEO<br>MethodSense, Inc.<br>Email: ritaking@methodsense.com<br>Phone: (919) 313-3961<br>Fax: (919) 313-3979 |
| Company Contact: | Neil Wadehra<br>Title: VP of Operations<br>Email: nwadehra@polarean.com<br>Phone: (919) 206-7900 |
| Date Prepared: | October 17, 2024 |
| Trade Name: | XENOVIEW 3.0T Chest Coil |
| Common Name: | Coil, Magnetic Resonance |
| Classification: | Class II |
| Regulation Number: | 21 CFR 892.1000 |
| Classification Panel: | Radiology |
| Product Code: | MOS |
| Prior Submissions: | K231647 |
## Predicate Device:
| Trade Name | Polarean XENOVIEW 3.0T Chest Coil |
|---------------------------|------------------------------------------------------|
| 510(k) Submitter / Holder | Polarean |
| 510(k) Number | K231647 |
| Requlation Number | 21 CFR 892.1000 Magnetic resonance diagnostic device |
| Classification | Class II |
| Classification Panel | Radiology |
| Product Code | MOS |
The predicate device has not been subject to a design-related recall.
{6}------------------------------------------------
#### Device Description
The Polarean XENOVIEW 3.0T Chest Coil (hereafter Chest Coil) is a flexible, single channel, transmit-receive (T/R) RF coil tuned to 123xe frequency on a 3.0T MRI magnetic field in order to image 129Xe nuclei while the patient is positioned inside a compatible multi-nuclear-capable MRI scanner. The Chest Coil is intended to be worn by a patient who inhales hyperpolarized 100%e gas (XENOVIEW) to obtain an MR image of the regional distribution of hyperpolarized 199xe in the lungs.
The coil is constructed of a durable, flexible circuit board material within which the antenna elements and all electronic components are contained. These components are electrically isolated from the rest of the coil packaging by being enclosed within suitable non-conductive. water-rated, and flame-rated materials. A layer of padding is located on either side of the coil circuitry to provide patient comfort and protection against potential heating generated by circuitry components. The RF coil is a "fixed matching and tuning device" (i.e. not tunable by the operator), thereby eliminating the need to tune and match it for every patient.
#### Indications for Use
The Polarean XENOVIEW 3.0T Chest Coil is to be used in conjunction with compatible 3.0T Magnetic Resonance Imaging (MRI) scanners and approved xenon Xe 129 hyperpolarized for oral inhalation for evaluation of lung ventilation in adults and pediatric patients aged 12 years and older.
#### Substantial Equivalence
The Polarean XENOVIEW 3.0T Chest Coil is substantially equivalent to its predicate device, Polarean XENOVIEW 3.0T Chest Coil (K231647). This submission adds compatibility of the Chest Coil with General Electric Healthcare (GEHC) 3T MR750 and Premier MRI scanners. The safety and performance testing that was performed in the previous submission (K231647) was repeated to demonstrate that the Chest Coil is as safe and effective when used with GEHC 3T MR750 and Premier MRI scanners as when used with Philips 3T MRI scanners.
The table below provides a detailed comparison of Polarean XENOVIEW 3.0T Chest Coil to the predicate device (K231647).
{7}------------------------------------------------
# Detailed Comparison of the Subject and Predicate Devices
| Characteristic | Subject Device<br>Polarean XENOVIEW 3.0T Chest<br>Coil (Models 44315-02 and<br>44315-04) | Primary Predicate Device<br>Polarean XENOVIEW 3.0T Chest<br>Coil (Model 44315-03) (K231647) | Comparison |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indications<br>for Use | The Polarean XENOVIEW 3.0T<br>Chest Coil is to be used in<br>conjunction with compatible 3.0T<br>Magnetic Resonance Imaging<br>(MRI) scanners and approved<br>xenon Xe 129 hyperpolarized for<br>oral inhalation for evaluation of<br>lung ventilation in adults and<br>pediatric patients aged 12 years<br>and older. | The Polarean XENOVIEW 3.0T<br>Chest Coil is to be used in<br>conjunction with compatible 3.0T<br>Magnetic Resonance Imaging<br>(MRI) scanners and approved<br>xenon Xe 129 hyperpolarized for<br>oral inhalation for evaluation of<br>lung ventilation in adults and<br>pediatric patients aged 12 years<br>and older. | Identical |
| Anatomical<br>Area | Chest | Chest | Identical |
| Characteristic | Subject Device<br>Polarean XENOVIEW 3.0T Chest<br>Coil (Models 44315-02 and<br>44315-04) | Primary Predicate Device<br>Polarean XENOVIEW 3.0T Chest<br>Coil (Model 44315-03) (K231647) | Comparison |
| Compatible<br>MRI Systems | GEHC MR750 (44315-02) and<br>Premier (44315-04) 3T | Philips 3T | Equivalent. Subject devices are<br>compatible with GEHC MR750<br>and Premier 3T MRI scanners and<br>the predicate device is compatible<br>with Philips 3T MRI scanners.<br>Both the subject devices and<br>predicate device are compatible<br>with 3T MRI scanners. Both<br>devices use a connector that is<br>compatible with the respective<br>MRI scanner. The addition of the<br>new configurations with<br>compatibility with GEHC MR750<br>and Premier 3T MRI scanners<br>does not affect the intended use of<br>the device and safety and<br>effectiveness of the device has<br>been confirmed with testing. |
| Mode of<br>Operation | Transmit / Receive | Transmit / Receive | Identical |
| Flexible / Rigid | Flexible | Flexible | Identical |
| Nucleus | 129Xe (Multinuclear Channel) | 129Xe (Multinuclear Channel) | Identical |
| Frequency of<br>Operation | 35.33 MHz | 35.33 MHz | Identical |
| Characteristic | Subject Device<br>Polarean XENOVIEW 3.0T Chest<br>Coil (Models 44315-02 and<br>44315-04) | Primary Predicate Device<br>Polarean XENOVIEW 3.0T Chest<br>Coil (Model 44315-03) (K231647) | Comparison |
| Antenna<br>Configuration | Quadrature (co-rotating saddle<br>coil pairs) | Quadrature (co-rotating saddle<br>coil pairs) | Identical |
| Tuning /<br>Impedance<br>Matching | Fixed tuning and matching.<br>Factory set. | Fixed tuning and matching.<br>Factory set. | Identical |
| Method of<br>Decoupling | Passive decoupling for ¹H. | Passive decoupling for ¹H. | Identical |
| Materials | Foam and fabric. | Foam and fabric. | Identical |
| # of receive<br>channels | 1 | 1 | Identical |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
### Non-Clinical Testing
The XENOVIEW 3.0T Chest Coil was verified and validated in accordance with documented Verification & Validation plans and protocols to ensure conformance with established performance criteria. See below for the type of tests performed. Polarean has completed the followina testina:
#### Performance – Bench
Bench testing was repeated by Polarean for the new model of the Chest Coil with compatible MRI scanners to confirm the safety and performance of various components of the Chest Coil in accordance with the following standards:
- NEMA MS 6-2008 (R2014), Determination of Signal-to-Noise Ratio and Image . Uniformity for Sinqle-Channel, Non-Volume Coils in Diagnostic Magnetic Resonance lmaqinq
- . NEMA MS 8-2016, Characterization of the Specific Absorption Rate (SAR) for Magnetic Resonance Imaging Systems
- . NEMA MS 14-2019, Characterization of Radiofrequency (RF) Coil Heating in Magnetic Resonance Imaging Systems
#### Electrical Safety and Electromagnetic Compatibility
Electrical Safety and Electromagnetic Compatibility testing was repeated for the new model of the Chest Coil to confirm the safety and performance of various components of the Chest Coil in accordance with the following standards:
- Basic Safety and Essential Performance (IEC 60601-1:2005/(R)2012 and A1:2012) .
- Basic Safety and Essential Performance of MR Equipment (IEC 60601-2-33:2015) .
- . Electromagnetic Compatibility (IEC 60601-1-2:2020)
#### Biocompatibility
Biocompatibility testing was performed per ISO 10993-1:2018 to confirm the safety and performance of various components of the Chest Coil in accordance with the following standards:
- . Cytotoxicity (ISO 10993-5:2009)
- . Sensitization (ISO 10993-10:2010)
- Irritation (ISO 10993-10:2010)
### Summary Clinical Testing
No clinical tests were required to demonstrate substantial equivalence.
#### Conclusion
The conclusions drawn from the nonclinical testing demonstrate that the subject device, XENOVIEW 3.0T Chest Coil is as safe, as effective, and performs as well as or better than the legally marketed predicate, XENOVIEW 3.0T Chest Coil (K231647).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.