The Encompass™ 15 Channel Head Coil is intended to be used in conjunction with a Magnetic Resorance Scanner for the MR examination of the human brain just before, during, and at the end of brain surgery. The Encompass™ 15 Channel Head Coil can also be used as a standard diagnostic examinations and fMRI (Functional Magnetic Resonance Imaging). When used with magnetic resonance imaging systems, it is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Device Story
Encompass 15 Channel Head Coil is a rigid, receive-only phased-array coil for 3T MRI systems; co-developed by Qfix and NORAS. Input: RF signals from human head during MRI scan. Transformation: Coil captures signals for image reconstruction by host MRI system software. Output: Diagnostic MR images (transverse, sagittal, coronal, oblique). Used in clinical/surgical settings by trained physicians/technicians. Integrates with Encompass SRS Immobilization System to reduce motion and improve SNR. Benefits: Enables high-quality imaging for surgical guidance and diagnostic evaluation; provides improved homogeneity and patient comfort features (viewing window, mirror accommodations).
Clinical Evidence
No clinical tests conducted. Substantial equivalence supported by non-clinical bench testing, risk analysis, and provided clinical images.
Technological Characteristics
15-channel phased-array receive-only head coil; rigid coil frame; 3T MRI compatible (e.g., Siemens MAGNETOM Skyra). Materials differ from predicate. Connectivity managed by host MRI system software. Compliance with IEC 60601-1 series and ISO 14971:2007. Optimized for use with Encompass SRS Immobilization System.
Indications for Use
Indicated for patients requiring MR examination of the human brain, including pre-operative, intra-operative, and post-operative surgical assessment, standard diagnostic imaging, and fMRI. Used for producing transverse, sagittal, coronal, and oblique images of internal head structures to assist physician diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
September 9, 2019
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Qfix % Alexandra Low Smythe Senior Regulatory Affairs Manager and Intellectual Property Specialist 440 Church Road AVONDALE PA 19311
Re: K190668
Trade/Device Name: Encompass™ 15 Channel Head Coil, 3T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: MOS Dated: July 23, 2019 Received: July 24, 2019
Dear Alexandra Low Smythe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K190668
Device Name Encompass™ 15 Channel Head Coil, 3T
#### Indications for Use (Describe)
The Encompass™ 15 Channel Head Coil is intended to be used in conjunction with a Magnetic Resorance Scanner for the MR examination of the human brain just before, during, and at the end of brain surgery. The Encompass™ 15 Channel Head Coil can also be used as a standard diagnostic examinations and fMRI (Functional Magnetic Resonance Imaging). When used with magnetic resonance imaging systems, it is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures of the head. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
Type of Use (Select one or both, as applicable)
X Prescription Use {Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Q-Fix, which is a red letter Q with the word FIX in black letters inside of it. Below the logo, the text "510k Summary" is written in a large, bold font. The text is black and is aligned to the left.
| I.GENERAL INFORMATION | |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment: | WFR/Aquaplast Corporation/Anholt Technologies Inc., Dba Qfix<br>440 Church Road<br>Avondale, PA 19311 USA |
| Date Prepared: | November 14, 2018 |
| Manufacturer: | Qfix<br>440 Church Road<br>Avondale, PA 19311 USA<br>Registration Number: 2247992 |
| Contact Person: | Alexandra Low Smythe<br>Senior Regulatory Affairs Specialist and Intellectual Property Specialist<br>Qfix<br>440 Church Road<br>Avondale, PA 19311 USA<br>Phone: 610 268-0585 Ext 736<br>Fax: 610-268-0588 |
| Device Name: | Encompass™ 15 Channel Head Coil, 3T |
| Trade Name: | Encompass™ 15 Channel Head Coil, 3T |
| Common Name: | 15 Channel Head Coil, Encompass SRS Coil |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Regulation number: | 21 CFR § 892.1000 |
| Device Class: | II |
| Product Code: | MOS |
#### II.SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING SUBSTANTIAL EQUIVALENCE
#### Indications for Use
The Encompass™ 15 Channel Head Coil is intended to be used in conjunction with a Magnetic Resonance Scanner for the MR examination of the human brain just before, during, and at the end of brain surgery. The Encompass™ 15 Channel Head Coil can also be used as a standard diagnostic head coil for diagnostic examinations and fMRI (Functional Magnetic Resonance Imaging). When used with magnetic resonance imaging systems, it is indicated for use as a diagnostic imaging device to produce
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Q-Fix. The logo is a large, red letter Q with the word "FIX" in black letters inside the Q. The dot of the "i" in "FIX" is a black circle. There is a registered trademark symbol to the right of the "X".
transverse, sagittal, coronal and oblique images of the internal structures of the head. When interpreted by a trained physician, these images provide information that can be useful in determining diagnosis.
### Device Description
Qfix and NORAS MRI products GmbH have jointly developed a head coil designed to enable MRI for diagnosis and to inform additional clinical actions, the Encompass™ 15 Channel Head Coil, 3T. The Encompass 15 Channel Head Coil is a diagnostic imaging device for use in 3T environments for obtaining diagnostic MR images to be used to inform procedures which use MR imaging to guide interventional procedures and other treatments. The Encompass 15 Channel Head Coil has been specifically designed to interface with the Encompass™ SRS Immobilization System (cleared with K152321 December 4, 2015).
The Encompass™ 15 Channel Head Coil is a ridged receive only coil featuring a 7 Channel Top Coil, an 8 Channel Bottom Coil, and an optional mirror holder. Imaging is performed with a 15-Channel "phased array" co-developed by Qfix and NORAS and manufactured by NORAS. The coils are mounted in a rigid Coil Frame and connection to the MRI is managed by the applicable MRI software. The Encompass™ 15 Channel MRI Head Coil has been designed for use with Siemens Healthineers MAGNETOM 3T systems e.g. MAGNETOM Skyra.
Encompass™ 15 Channel Head Coil has been optimized for use with the Encompass™ SRS Immobilization System. The use of the Encompass™ 15 Channel Head Coil together with the Encompass™ SRS Immobilization System allows immobilization of the patient while obtaining diagnostic MR images to be used to inform procedures which use MR imaging to guide interventional procedures and other treatments.
#### Predicate Information
Qfix believes that the subject device, Encompass™ 15 Channel Head Coil is substantially equivalent to the following head coil within the meaning of the Safe Medical Devices Act of 1990.
| Predicate Device Name | FDA Clearance<br>Number and Date | Product<br>code | Manufacturer |
|-----------------------|-----------------------------------|-----------------|----------------------------|
| NORAS OR Head Coil 3T | K091546, cleared<br>June 24, 2009 | MOS | NORAS MRI<br>products GmbH |
The Encompass™ 15 Channel Head Coil and the NORAS OR Head Coil 3T are both intended to be used in surgical settings for diagnostic evaluation of the brain and obtaining diagnostic MR images to be used to inform procedures which use MR imaging to guide interventional procedures and other treatments. To date, this predicate device has not been subject to a design-related recall per information that is publicly accessible in the FDA recall database.
# K190668
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for QFix. The logo consists of a large, red letter "Q" with the word "Fix" in black letters inside the "Q". The "Q" is stylized with a break in the upper right quadrant, and the word "Fix" is positioned to fill the space created by the break. There is a registered trademark symbol to the right of the word "Fix".
## Comparison to Predicate Device
The Encompass™ 15 Channel Head Coil bears many similarities to its predicate, the NORAS OR Head Coil. The Encompass™ 15 Channel Head Coil has a similar Intended Use and purpose as the predicate device. Both the subject device and the predicate device can provide diagnostic information about the head to inform interventional procedures.
The subject device offers the following improvements over the predicate.
- The housing of the subject device is made of different materials than the predicate device.
- -The subject device has 15 channels while the predicate device has 8.
- The subiect device is intended to be used independently or in conjunction with the Encompass SRS Immobilization System. The use of the subject device with immobilization devices in general increases SNR due to motion reduction for visualization of fine details. The predicate device is not compatible with immobilization devices.
- -The subject device offers different adjustments to allow for being as close as possible to patient without disturbing anatomy for improved SNR.
- The subject device offers improved homogeneity across the entire head. -
- The subiect device features additional considerations for patient comfort. including a patient viewing window and accomodations for a mirror in two orientations.
#### Performance Standards and Testing
Testing and analysis have been conducted to show that the verification, and safety requirements have been met per the FDA as established standards. The subject devices have been tested in accordance with FDA recognized NEMA standards for the measurement of performance and safety parameters. Risk Management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Ofix and NORAS adhere to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. This device was tested on 3T Siemens Healthineers MANGETOM MRI scanners and the performance of this device can be considered safe when used with available MAGNETOM 3T systems e.g. MAGNETOM Skyra.
Non-clinical bench testing and customer validation was conducted to support the intended use and to confirm that technological differences do not raise any new issues of safety or effectiveness over the predicate.
#### Clinical Tests
No clinical tests were conducted to support the subject device and the substantial equivalence argument, however clinical images were provided to support the Encompass™ 15 Channel Head Coil.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for QFix. The logo is a red letter "Q" with the word "Fix" inside of it. The "Q" is large and bold, and the word "Fix" is in black. The "X" in "Fix" is stylized to look like it is underlining the word. There is a registered trademark symbol to the right of the "X".
### Safety and Effectiveness
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in development, testing and product labeling. To minimize risks, Ofix & NORAS adhere to recognized and established industry practices and standards to minimize safety and performance risks. Furthermore, the operators and end users of the device are healthcare professionals familiar with and responsible for procedures for which the subject device is intended to be used.
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
#### Substantial Equivalence Conclusions
The subject device, Encompass™ 15 Channel Head Coil, is optimized for use with the Encompass SRS Immobilization System for obtaining diagnostic MR images to be used to inform procedures which use MR imaging to guide interventional procedures and other treatments. While the predicate device, NORAS OR Head Coil 3T, is optimized for its usefulness in interventional surgery applications, both devices are head coils and the fundamental attributes of the subject device and the predicate device are the same.
The conclusions from the non-clinical data suggest that the subject device has the same fundamental technological characteristics with respect to the predicate device and exhibits an equivalent safety and performance profile as that of the predicate device.
Therefore, Qfix believes the Encompass™ 15 Channel Head Coil does not raise new questions of safety or effectiveness and, therefore, is substantially equivalent to the marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.