LEO III TX/RX QUADRATURE KNEE COIL; ALPHA III TX/RX QUADRATURE WRIST COIL
K023982 · Usa Instruments, Inc. · MOS · Feb 21, 2003 · Radiology
Device Facts
Record ID
K023982
Device Name
LEO III TX/RX QUADRATURE KNEE COIL; ALPHA III TX/RX QUADRATURE WRIST COIL
Applicant
Usa Instruments, Inc.
Product Code
MOS · Radiology
Decision Date
Feb 21, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Leo III Tx/Rx Quadrature Knee Coil is a transmit-receive quadrature RF coil, used for obtaining diagnostic images of the knee and surrounding anatomy in Magnetic Resonance Imaging Systems. The Leo III Tx/Rx Quadrature Knee Coil is designed for use with the Magnetom Trio 3.0T MRI system manufactured by Siemens Medical Solutions, Inc. The indications for use are the same as for standard MR Imaging. The Alpha III Tx/Rx Quadrature Wrist Coil is a transmit-receive quadrature RF coil, used for obtaining diagnostic images of the wrist and hand in Magnetic Resonance Imaging Systems. The Alpha III T/R Quadrature Wrist Coil is designed for use with the Magnetom Trio 3.0T MRI system manufactured by Siemens Medical Solutions, Inc. The indications for use are the same as for standard MR Imaging.
Device Story
Leo III and Alpha III are multi-element quadrature transmit-receive RF coils for 3.0T MRI systems; designed for knee/surrounding anatomy and wrist/hand imaging respectively. Coils function as passive accessories to the MRI scanner; they receive RF signals from hydrogen nuclei and transmit RF pulses. Enclosed in rigid, fire-rated plastic housings with open windows for patient comfort/positioning. Safety features include PIN diodes for transmit/receive channel isolation, active RF blowing fuses to prevent resonant loops, and non-conductive housing materials. Power deposition is managed by the host MRI system's SAR algorithm. Used in clinical imaging environments by trained MRI technologists/radiologists; output is diagnostic-quality MR images interpreted by physicians to aid clinical diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness supported by comparison of technical specifications, materials, and design features to legally marketed predicate devices.
Technological Characteristics
Multi-element quadrature transmit-receive RF coils. Materials: Flame retardant polyurethane, polycarbonate, Kydex Acrylic/PVC, Delrin Acetal. Leo III: 12 z-inductor design. Alpha III: 2-element saddle design. Connectivity: Direct interface to Siemens Magnetom Trio 3.0T MRI system. Safety: PIN diode switching, active RF blowing fuses, non-conductive housing. SAR limited by host system algorithm.
Indications for Use
Indicated for diagnostic MRI imaging of the knee/surrounding anatomy (Leo III) or wrist/hand (Alpha III) in patients undergoing standard MR imaging procedures on a Siemens Magnetom Trio 3.0T system.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{1}------------------------------------------------
# SUMMARY OF SAFETY AND EFFECTIVENESS
# K023982
| 1. Device Name: | Magnetic Resonance Imaging Accessory |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Proprietary Name: | Leo III Tx/Rx Quadrature Knee Coil |
| 3. Classification: | Class II |
| 4. Establishment Registration #: | 1529041 |
| 5. Manufacture Facility Location: | USA Instruments, Inc.,<br>1515 Danner Drive<br>Aurora, Ohio 44202, USA<br>Telephone: 330-562-1000; Fax: 330-562-1422. |
| 6. Performance Standard: | No applicable performance standards have been<br>issued under Section 514 of the Food, Drug and<br>Cosmetic Act. |
| 7. Intended Use: | The Leo III Tx/Rx Quadrature Knee Coil is a<br>transmit-receive quadrature RF coil, used for<br>obtaining diagnostic images of the knee and<br>surrounding anatomy in Magnetic Resonance<br>Imaging Systems. The Leo III Tx/Rx Quadrature<br>Knee Coil is designed for use with the Magnetom<br>Trio 3.0T MRI system manufactured by Siemens<br>Medical Solutions, Inc. The indications for use are<br>the same as for standard MR Imaging. |
| 8. Device Description: | The Leo III Tx/Rx Quadrature Knee Coil is a multi-<br>element quadrature transmit-receive coil. The coil<br>has a split top design with a latching device and a<br>large open viewing window on the top. The open,<br>patient friendly design eases patient handling and<br>positioning and maximizes patient comfort. The coil<br>elements and accessory electronics are enclosed in<br>a rigid plastic housing, which is fire rated and has a<br>high impact and tensile strength. |
Please turn over
{2}------------------------------------------------
## 9. Safety and Effectiveness
| Leo III Tx/Rx Quadrature Knee Coil<br>Product Features | Comparison to predicate device or other 510(k)<br>cleared products |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: Imaging of the knee and<br>surrounding anatomy. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)<br>-Similar to the Leo 7000 Knee Coil manufactured<br>by USA Instruments, Inc. (K971246) |
| Indications for Use: Identical to routine<br>MRI imaging. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)<br>-Similar to the Leo 7000 Knee Coil manufactured<br>by USA Instruments, Inc. (K971246) |
| Coil Enclosure Material:<br>Flame Retardant Polyurethane<br>Flame Retardant Polycarbonate<br>Kydex Acyrlic/PVC<br>Delrin Acetal | - Similar to the Hi-Res 9000 Phased Array Carotid<br>Coil manufactured by USA Instruments, Inc.<br>(K001210) |
| Coil Design: 12 z-inductor transmit<br>receive quadrature coil. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Decoupling: This is a transmit and<br>receive coil. There is no decoupling, the<br>Transmit / Receive Switch is actively<br>switched during transmit and receive<br>modes. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Prevention of RF Burns: The coil's<br>transmit /receive switch uses pin diodes<br>to isolate the receive channel from the<br>transmit channel; coil elements and<br>circuitry are enclosed in a non-<br>conductive housing. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Radio Frequency Absorption: Power<br>deposition during imaging is limited by<br>SAR algorithm | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Formation of Resonant Loop: Active<br>diodes and fast RF blowing fuses<br>isolate the coil elements from RF fields;<br>length of cable and stiffness does not<br>permit looping | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
{3}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS
| 1. Device Name: | Magnetic Resonance Imaging Accessory |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Proprietary Name: | Alpha III Tx/Rx Quadrature Wrist Coil |
| 3. Classification: | Class II |
| 4. Establishment Registration #: | 1529041 |
| 5. Manufacture Facility Location: | USA Instruments, Inc.<br>1515 Danner Drive<br>Aurora, Ohio 44202, USA<br>Telephone: 330-562-1000; Fax: 330-562-1422. |
| 6. Performance Standard: | No applicable performance standards have been<br>issued under Section 514 of the Food, Drug and<br>Cosmetic Act. |
| 7. Intended Use: | The Alpha III Tx/Rx Quadrature Wrist Coil is a<br>transmit-receive quadrature RF coil, used for<br>obtaining diagnostic images of the wrist and hand in<br>Magnetic Resonance Imaging Systems. The Alpha<br>III T/R Quadrature Wrist Coil is designed for use<br>with the Magnetom Trio 3.0T MRI system<br>manufactured by Siemens Medical Solutions, Inc.<br>The indications for use are the same as for standard<br>MR Imaging. |
| 8. Device Description: | The Alpha III T/R Quadrature Wrist Coil is a multi-<br>element quadrature transmit-receive coil. The coil<br>has a rigid enclosure with windows on top to<br>facilitate patient handling. The open, patient friendly<br>design eases patient handling and positioning and<br>maximizes patient comfort. The coil elements and<br>accessory electronics are enclosed in a rigid plastic<br>housing, which is fire rated and has a high impact<br>and tensile strength. |
Please turn over
{4}------------------------------------------------
### 9. Safety and Effectiveness
| Alpha III Tx/Rx Quadrature Wrist Coil<br>Product Features | Comparison to predicate device or other 510(k)<br>cleared products |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: Imaging of the wrist and<br>hand | -Similar to the Alpha 7000 Wrist Coil manufactured<br>by USA Instruments, Inc. (K972205) |
| Indications for Use: Identical to routine<br>MRI imaging. | -Similar to the Alpha 7000 Wrist Coil manufactured<br>by USA Instruments, Inc. (K972205)<br>-Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Coil Enclosure Material:<br>Flame Retardant Polyurethane<br>Flame Retardant Polycarbonate<br>Kydex Acrylic/PVC<br>Delrin Acetal | - Similar to the Hi-Res 9000 Phased Array Carotid<br>Coil manufactured by USA Instruments, Inc.<br>(K001210) |
| Coil Design: 2 element saddle transmit<br>receive quadrature coil.<br>Decoupling: This is a transmit and<br>receive coil. There is no decoupling, the<br>Transmit / Receive Switch is actively<br>switched during transmit and receive<br>modes. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179)<br>- Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Prevention of RF Burns: The coil's<br>transmit /receive switch uses pin diodes<br>to isolate the receive channel from the<br>transmit channel; coil elements and<br>circuitry are enclosed in a non-<br>conductive housing. | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Radio Frequency Absorption: Power<br>deposition during imaging is limited by<br>SAR algorithm | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
| Formation of Resonant Loop: Active<br>diodes and fast RF blowing fuses<br>isolate the coil elements from RF fields;<br>length of cable and stiffness does not<br>permit looping | - Similar to the Allegra 3.0Tesla TxRx Head coil<br>manufactured by USA Instruments, Inc. (K002179) |
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Christie Shumaker Manager OA and Regulatory USA Instruments, Inc. 1515 Danner Drive AURORA OH 44202
FEB 2 1 2003
Re: K023982 Trade/Device Name: Leo III Tx/Rx Quadrature Knee Coil and Alpha III Tx/Rx Quadrature Wrist Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: 90 MOS Dated: November 27, 2002 Received: December 2, 2002
Dear Ms. Shumaker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:
| 8xx.1xxx | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page 1 of 1
510(k) Number (if known):__ <○23982
Device Name: Leo III Tx/Rx Quadrature Knee Coil Alpha III Tx/Rx Quadrature Wrist Coil
Indications for Use: The Leo III Tx/Rx Quadrature Knee Coil is designed to provide Magnetic Resonance Images of the knee and surrounding anatomy. The Alpha III Tx/Rx Quadrature Wrist Coil is designed to provide Magnetic Resonance Images of the wrist and hand. The Leo III Tx/Rx Quadrature Knee Coil and the Alpha III Tx/Rx Quadrature Wrist Coil are designed for use with the Siemens Magnetom Trio 3.0T scanner manufactured by Siemens Medical Solutions.
> Anatomic Regions: Leo III Tx/Rx Quadrature Knee Coil: Knee and surrounding Anatomy Alpha IIII Tx/Rx Quadrature Wrist Coil: Wrist and Hand Nuclei Excited: Hydrogen
The indications for use are the same as for standard imaging:
The 3.0T MRI system is indicated for use as an NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR signal, (2) depend upon NMR parameters (proton density, spin lattice relaxation time T1, spin-spin relaxation time T2) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images vield information that can be useful in the determination of a diagnosis.
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|---------------------------------------------------------------|---------------|
| Prescription Use (Per 21 CFR 801.109) | <span></span> |
| | OR |
| Over-The-Counter Use (Optional Format 1-2-96) | <span></span> |
| (Division Sign-Off) | <span></span> |
| Division of Reproductive, Abdominal, and Radiological Devices | <span></span> |
| 510(k) Number | K023982 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.