K011608 · Medical Advances, Inc. · MOS · Aug 14, 2001 · Radiology
Device Facts
Record ID
K011608
Device Name
MODEL 235GE-64; MULTI PURPOSE FLEX ARRAY COIL
Applicant
Medical Advances, Inc.
Product Code
MOS · Radiology
Decision Date
Aug 14, 2001
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Magnetic resonance imaging (MRI) and magnetic resonance angiography (MRA) of the Head: Temporal lobes (hippocampus), internal auditory canals, orbits and anterior optic pathways, cranial nerves, intracranial magnetic resonance angiography (MRA). Upper Neck: Skull base, Cranio-cervical junction, cervical carotid artery. Upper Extremities: Shoulder, acromioclavicular (AC) joint, elbow, peripheral nerves. Lower Extremities: Knee, ankle and Achilles tendon, foot. Pediatric applications.
Device Story
Model 235GE-64 Multi Purpose Flex Array Coil is a specialized accessory for GE Signa 1.5T MRI systems. Device functions as a phased array receiver coil to capture magnetic resonance signals from specific anatomical regions: head, neck, and extremities. Coil is placed on or around the patient; it detects RF signals generated during the MRI scan sequence. Signals are transmitted to the host MRI system for image reconstruction. Used in clinical radiology settings by trained MRI technologists or radiologists. Output is processed into 2D/3D images, including proton density, T1/T2 weighted, time-of-flight, and phase contrast sequences. Clinicians use these images to visualize anatomy and pathology, aiding in diagnostic decision-making. Device provides high-resolution imaging capabilities for various joints and nerves, potentially improving diagnostic accuracy for musculoskeletal and neurological conditions.
Clinical Evidence
No clinical data. Substantial equivalence is based on bench testing and comparison of safety and imaging performance parameters against the predicate system configuration.
Technological Characteristics
Phased array receiver coil designed for GE Signa 1.5T MRI systems. Materials include flame-resistant CPAI-84 material. Complies with UL 2601-1 (Medical Electrical Equipment safety) and UL 94 (flammability). Operates as a passive signal receiver within the MRI environment. No internal energy source; connectivity is via proprietary interface to the GE Signa system.
Indications for Use
Indicated for MRI and MRA imaging of the head, upper neck, upper extremities, and lower extremities in patients, including pediatric populations, requiring diagnostic imaging of specified anatomical regions.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Submission Summary (Full Text)
{0}------------------------------------------------
# AUG 1 4 2001
## 510(k) Summary of Safety and Effectiveness
K011608
Device Name
Applicability
Reason for 510(k)
Classification Name
Device Classification Panel
Device Classification Number
Product Code
Common Name
Proprietary Name
Establishment Registration Number
Address of MFG Facility
Model 235GE-64: Multi Purpose Flex Array Coil
Compatible with GE Signa 1.5T MRI systems with Phased Array option
New indication for use
Magnetic Resonance Diagnostic Device
Radiology
892.1000
90MOS
Magnetic Resonance Specialty
Model 235GE-64: Multi Purpose Flex Array Coil
2183683
Device manufactured for IGC-Medical Advances Inc. and UltraImage Corporation by:
SeaMED Corporation 21621 30th Avenue N.E. Bothell, WA 98021 Telephone: 425.482.1300 Facsimile: 425.482.1401 Establishment Registration Number: 3031699
Contact: Ms. Marcia Page, VP, Quality and Regulatory Affairs
{1}------------------------------------------------
#### Points of Contact
Classification
Intended Uses
Diagnostic Uses
Anatomic Regions
Standards
Performance Standards
Voluntary Safety Standards
IGC-Medical Advances Inc. Michael J. Leigh Director of Operations 414.258.3808 Ext. 206
UltraImage Corporation, a Pathway Medical Technologies Company Louise C. Myers Senior Director, Regulatory Affairs 425.497.0372
Class II
2D, 3D imaging, proton density, T1 and T2 weighted imaging. 2D, 3D time of flight, phase contrast imaging.
Head: Temporal lobes (hippocampus), internal auditory canals, orbits and anterior optic pathways, cranial nerves, intracranial magnetic resonance angiography (MRA) Upper Neck: Skull base, Cranio-cervical junction, cervical carotid artery Upper Extremities: Shoulder, acromioclavicular (AC) joint, elbow, peripheral nerves Lower Extremities: Knee, ankle and Achilles tendon, foot Pediatric applications.
None Established under Section 514
- UL 2601-1 Medical Electrical Equipment, Part 1: General Requirements for Safety
- Tests for Flammability of Plastic UL 94 Materials
General Safety Requirements for IEC 601-1 Medical Electrical Equipment
- Specification for Flame Resistant CPAI-84 Material Used in Camping Tentage
{2}------------------------------------------------
#### Overview
The Radiology Devices Panel considered potential concerns regarding the safe and effective The Nadlorgy Devices Funer combiness when they recommended reclassification operation of Magnetic 1080. After reclassification, the FDA's Center for Devices and to Class if on July 21, 1984) released a draft guidance document for the content and review of Magnetic Resonance Diagnostic Device premarket notification submissions that offered Magical resollance Diagnoute De to considerable technological advances in MRDDs, CDRH Clarineation of these concerns. Due it on November 14, 1998. The following is a summary of Issued an updated gurdained within this premarket notification that addresses these concerns:
The GE 1.5T Signa MRI system operated with the Medical Advances Multi Purpose Flex Array The OL 1.91 bigma MICL of equivalent to the same system operated with the legally marketed predicate devices listed in section 4.0, within the Class II definition of Magnetic Resonance Diagnostic Device with respect to the safety parameter action levels:
#### Safety Parameters
| Maximum Static Magnetic Field: | No change |
|-----------------------------------------|-----------|
| Rate of Magnetic Field Strength Change: | No change |
| RF Power Deposition: | No change |
| Acoustic Noise Levels: | No change |
| Biocompatibility: | No change |
#### Imaging Performance Parameters
| Specification Volume: | No change |
|-----------------------------------|-----------|
| Signal-to-Noise Ratio: | No change |
| Image Uniformity: | No change |
| Geometric Distortion: | No change |
| Slice Thickness and Gap: | No change |
| High Contrast Spatial Resolution: | No change |
{3}------------------------------------------------
#### General Safety and Effectiveness Concerns
The device contains instructions for use. It includes indications for use, precautions, cautions, contraindications, warnings and quality assurance testing. This information assures safe and effective use of the device.
### Substantial Equivalence Summary
The GE 1.5T Signa MRI system operated with the Medical Advances Multi Purpose Flex Array Coil addressed in this PMN, has the same intended use and technological characteristics as the same system operated with the identified legally marketed predicate devices. The use of these coils does not affect the GE Signa system safety parameter specifications.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 4 2001
Mr. Michael Leigh Director of Operations IGC Medical Advances, Inc. 10437 Innovation Dr. MILWAUKEE WI 53226
Re: K011608
Model 235GE-64: Multi Purpose Flex Array Coil (MRI specialty coil) Dated: May 24, 2001 Received: May 25, 2001-Regulatory Class: II 21 CFR 892.1000/Procode: 90 MOS
#### Dear Mr. Leigh:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Nancy C. hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure (s)
{5}------------------------------------------------
Page l of l
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: Model 235 Series: Medical Advances Multi Purpose Flex Array Coil
Indications for Use:
Magnetic resonance imaging (MRI) and magnetic resonance angiography (MRA) of the Head: Magical lobes (hippocampus), internal auditory canals, orbits and anterior optic pathways, I chiporal loocs (inppooaliffalo), interesonance angiography (MRA). Upper Neck: Skull base, Cranio-cervical junction, cervical carotid artery. Upper Extremities: Shoulder, Craino-cavicular (AC) joint, elbow, peripheral nerves. Lower Extremities: Knee, ankle and Achilles tendon, foot. Pediatric applications.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K011608 |
|---------------|---------|
|---------------|---------|
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.