Digital Expert Access with Remote Scanning

K232346 · GE Healthcare · LNH · Oct 27, 2023 · Radiology

Device Facts

Record IDK232346
Device NameDigital Expert Access with Remote Scanning
ApplicantGE Healthcare
Product CodeLNH · Radiology
Decision DateOct 27, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Digital Expert Access with Remote Scanning is intended as a remote collaboration tool to view and review MR images, to remotely control MR Imaging Devices and to initiate MRI scans remotely.

Device Story

Digital Expert Access with Remote Scanning is a remote collaboration tool enabling real-time interaction between an onsite technologist and a remote expert. The system facilitates remote guidance, procedure assessment, and scanning parameter management. It allows the remote expert to view scanner screens, provide live annotations, and initiate MRI scans on compatible GE HealthCare MRI systems. The device requires an onsite technologist to be present throughout the procedure for patient support. It does not acquire images or perform post-processing; all imaging functions remain with the host MRI system. The system is used in clinical environments where GE HealthCare MRI systems are installed. It benefits patients by providing access to expert support for technologists, potentially improving scan quality and workflow efficiency. The onsite user maintains control by granting access and validating the connection via session password.

Clinical Evidence

No clinical data. Bench testing only. Verification and validation testing were conducted by integrating the device with a subset of GE HealthCare MRI systems to confirm the remote expert's ability to perform remote scanning and ensure proper implementation of design requirements.

Technological Characteristics

Remote collaboration software platform; PC/tablet-based interface; supports touch, keyboard, and touchpad input. Connectivity via hospital network. Operates as an add-on to GE HealthCare MRI systems. Designed under 21 CFR 820 and ISO 13485 standards. No direct patient interface.

Indications for Use

Indicated for use by trained radiologic technologists and experts to facilitate remote collaboration, guidance, and scan initiation on GE HealthCare MRI systems. Requires an onsite technologist to be present for patient support during scanning. Not for diagnostic use or patient safety-related management.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ October 27, 2023 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. GE Healthcare % Allena Holzworth Regulatory Affairs Leader 500 W. Monroe Street Chicago, Illinois 60661 Re: K232346 Trade/Device Name: Digital Expert Access with Remote Scanning Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LLZ Dated: August 4, 2023 Received: August 4, 2023 Dear Allena Holzworth: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). {2}------------------------------------------------ Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Samul for Daniel Krainak, Ph.D. Assistant Director Magnetic Resonance and Nuclear Medicine Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K232346 Device Name Digital Expert Access with Remote Scanning #### Indications for Use (Describe) Digital Expert Access with Remote Scan assistance solution which allows remote control of an MR Imaging Device including the ability to initiate a scan remotely. This access provides real time communication mechanisms between the remote and onsite users to facilitate the acquisition occurring on the device. Access must be granted by the system. Images reviewed remotely are not for diagnostic use. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has a wavy, water-like design around the letters, giving it a dynamic and fluid appearance. # 510(k) Summary This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.92: | Date: | August 4, 2023 | |------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | GE HealthCare<br>500 W. Monroe Street<br>Chicago, IL 60661 | | Primary Contact: | Allena Holzworth<br>Regulatory Affairs Leader<br>GE HealthCare<br>Phone: +1 (414) 208-5162<br>Email: Allena.Holzworth@ge.com | | Secondary Contact: | Christopher Paulik<br>Senior Manager Regulatory Affairs<br>GE HealthCare<br>Phone: +1 (262) 548-2010<br>Email: Christopher.A.Paulik@ge.com | | Subject Device Name: | Digital Expert Access with Remote Scanning | | Device Classification: | Class II | | Regulation Number: | 21 CFR 892.1000 Magnetic Resonance Diagnostic Device<br>21 CFR 892.2050 Medical image management and processing system | | Product Code: | LNH, LLZ | ### Predicate Device Information | Predicate Device 1 | | |--------------------|------------------------------------------------------| | Device Name: | SIGNA Hero | | Manufacturer: | GE Medical Systems, LLC | | 510(k) Number: | K213668 | | Regulation Number: | 21 CFR 892.1000 Magnetic Resonance Diagnostic Device | | Product Code: | LNH | | Predicate Device 2 | | |--------------------|-------------------------------| | Device Name: | Customer Remote Console (CRC) | | Manufacturer: | GE Medical Systems, LLC | | 510(k) Number: | K150193 | {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized font, enclosed within a circular border. There are small, curved lines emanating from the circle, giving the impression of movement or energy. The logo is colored in a light blue hue. | Regulation Number:<br>Product Code: | 21 CFR 892.2050 Medical image management and processing system<br>LLZ | |-------------------------------------|-----------------------------------------------------------------------| | Predicate Device 3 | | | Device Name: | MAGNETOM Systems with syngo Expert-i option | | Manufacturer: | Siemens Medical Solutions, INC. | | 510(k) Number: | K052423 | | Regulation Number: | 21 CFR 892.1000 Magnetic Resonance Diagnostic Device | | Product Code: | LNH | ### Device Description Digital Expert Access is a variation of the Customer Remote Console cleared under K150193. It is a remote scan assistance solution designed to address the skill variability in technologists and their need for ondemand support by allowing them to interact directly with a remote expert connected to the hospital network. By enabling the collaboration between an Onsite Technologist and Remote Expert, Digital Expert Access helps the onsite technologist to seek guidance and real time support on scanning related queries including but not limited to training, procedure assessment, and scanning parameter management. Digital Expert Access with Remote Scanning introduces a feature that enables the Remote Expert to initiate a scan and make changes in real time during the scanning session. This remote scan feature is only available when Digital Expert Access is connected to a compatible GE HealthCare MRI system. Digital Expert Access with Remote Scanning enables the following capabilities for the Onsite Technologist and the Remote Expert: - 1. Collaborative session between an Onsite Technologist and Remote Expert - 2. Real-time scanner screen share and live annotation - ന് Remote console access and control - 4. Remote Scan Initiation Digital Expert Access with Remote Scanning is not intended for diagnostic use or patient safety-related management. This solution is not intended to be used by individuals who are not properly trained in the operation of GE HealthCare Medical Imaging systems. Digital Expert Access with Remote Scanning does not directly interface with any patients and requires the Onsite Technologist to be continuously present throughout the scanning procedure. Digital Expert Access with Remote Scanning does not acquire any MRI images, nor does it do any post image processing. All image acquisition and image processing is conducted by the GE HealthCare MRI system. ### Intended Use Digital Expert Access with Remote Scanning is intended as a remote collaboration tool to view and review MR images, to remotely control MR Imaging Devices and to initiate MRI scans remotely. #### Indications for Use {6}------------------------------------------------ # GE HealthCare 510(k) Premarket Notification Submission Image /page/5/Picture/1 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in white inside. The letters are stylized and connected, and there are small white swirls around the edge of the circle. The logo is simple and recognizable, and it is often used to represent the company. Digital Expert Access with Remote Scanning is a remote scan assistance solution which allows remote control of an MR Imaging Device including the ability to initiate a scan remotely. This access provides real time communication mechanisms between the remote and onsite users to facilitate the acquisition occurring on the device. Access must be granted by the onsite user operating the system. Images reviewed remotely are not for diagnostic use. ### Technology The proposed device, Digital Expert Access with Remote Scanning, employs similar fundamental scientific technology as its predicate devices. ### Comparisons Digital Expert Access with Remote scanning is substantially equivalent to the predicate devices, SIGNA Hero (K213668), Customer Remote Console (CRC) (K150193) and MAGNETOM Systems with syngo Experti option (K052423). The proposed device is intended to be a remote scan assistance solution, allowing the onsite technologist to connect to a remote expert to view/review images, provide real time guidance and initiate an MRI scan remotely. The table below summarizes the substantive feature/technological similarities and differences between the predicate devices and the proposed device: | Items | Predicate 1:<br>SIGNA Hero<br>(K213668) | Predicate 2:<br>Customer Remote<br>Console (CRC)<br>K150193 | Predicate 3:<br>MAGNETOM Systems<br>with syngo Expert-i<br>option (K052433) | Proposed:<br>Digital Expert Access<br>with Remote Scanning | |------------------------------------------------------|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | User Interface<br>Functions | PC-based | PC-based | PC-based | Tablet-based/PC-based | | Operator<br>Interface | Keyboard, mouse | Keyboard, mouse | Keyboard, mouse | Touch LCD, Keyboard,<br>touchpad | | Remote Viewer<br>interactive<br>control | N/A | User profile<br>(preferences) is<br>associated with an<br>Administrator, User<br>with View role, or<br>User with Guidance<br>role. | The Expert-I status<br>dialog is used to<br>determine if<br>the remote user<br>should have Full<br>Access or View Only<br>access to the syngo<br>Workplace | The onsite user initiates<br>a request through DEA<br>which initiates an<br>acceptance screen on<br>the user interface. Once<br>the request is<br>acknowledged on the<br>scanner, the onsite user<br>must validate the<br>connection by entering<br>the session password<br>for the remote session. | | Imaging System<br>Modality for<br>Remote<br>Scanning | N/A | N/A | MR | MR | | Intended Users | Radiologist, Expert<br>Technologist | Radiologist, Expert<br>Technologist | Radiologist, Expert<br>Technologist | Radiologist, Expert<br>Technologist | Table 5-1: Comparison Table for Digital Expert Access with Remote Scanning and its Predicate Devices {7}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. There are three curved lines, resembling water droplets or stylized commas, surrounding the circle. The logo is simple, recognizable, and has been used by GE for many years. # GE HealthCare 510(k) Premarket Notification Submission | Scan Initiation | Scan initiated by<br>onsite technologist | Scan initiated by<br>onsite technologist | Remote user can<br>initiate scan on MR<br>scanners remotely | Remote user can initiate<br>scan on MR scanners<br>remotely | |--------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | Patient/Scanner<br>Support during<br>Acquisition | Onsite technologist<br>oversees patient<br>support during<br>scanning acquisition | N/A – CRC is not<br>intended for scan<br>acquisition | Onsite technologist<br>oversees patient<br>support during<br>scanning acquisition | Onsite technologist<br>oversees patient<br>support during scanning<br>acquisition | ## Clinical and Non-Clinical Tests ## Summary of Non-Clinical Tests Digital Expert Access with Remote Scanning has successfully completed the design control testing per GE HealthCare's quality system. It was designed under the Quality System Regulations of 21CFR 820 and ISO 13485. No new questions of safety and effectiveness and no unexpected test results were observed. The following quality assurance measures have been applied to the development of the system: - . Requirement Definition - Risk Analysis and Control - Technical Design Reviews - Formal Design Reviews - Software Development Lifecycle - Safety Testing (Verification) ● - Performance Testing (Verification, Validation) - Software Release Digital Expert Access with Remote Scanning verification was conducted to demonstrate proper implementation of its design requirements. The testing and results did not raise any new issues of safety and effectiveness. In addition, verification and validation testing was conducted on the bench by integrating Digital Expert Access with Remote Scanning on a subset of GE HealthCare MRI systems to ensure a Remote Expert was able to perform remote scanning on GE HealthCare MRI systems. Design verification and validation testing confirmed that the safety and effectiveness of the device has not been affected. The test plans and results have been executed with acceptable results. ## Summary of Clinical Testing The subject of this premarket submission, Digital Expert Access with Remote Scanning did not require clinical studies to support substantial equivalence to enable the remote scanning feature on GE HealthCare MRI systems. ## Substantial Equivalence Conclusion The introduction of Digital Expert Access with Remote Scanning does not result in any new potential safety risks and uses the same technology as the predicate devices, including the initiation of an MRI scan remotely. After analyzing design verification and validation testing on the bench it is the conclusion of GE {8}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo of General Electric (GE). The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the letters, giving it a classic and recognizable appearance. # GE HealthCare # 510(k) Premarket Notification Submission HealthCare that Digital Expert Access with Remote Scanning to be as safe, as effective, and performance is substantially equivalent to the predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...