K170840 · Siemens Medical Solutions USA, Inc. · LNH · Oct 12, 2017 · Radiology
Device Facts
Record ID
K170840
Device Name
MAGNETOM Terra
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 12, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The MAGNETOM Terra system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images and that displays the internal structure and/or function of the head or extremities. Other physical parameters derived from the images may also be produced. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis. The device is intended for patients > 30 kg/66 lbs.
Device Story
7T superconducting MRI system; produces transverse, sagittal, coronal, oblique cross-sectional images; displays internal structure/function of head or extremities. Uses 1Tx32Rx Head Coil and 1Tx28Rx Knee Coil. Operated by healthcare professionals in clinical settings. System utilizes magnetic field, gradients, RF transmitter/receiver, and software to map proton spin distribution. Includes SAR control software based on adult human simulations. Output interpreted by physicians to assist in diagnosis. Benefits include high-field imaging capabilities for improved image quality.
Clinical Evidence
Clinical study of 35 individuals conducted to determine nerve stimulation thresholds for gradient system output (PNS). Bench testing included acoustic noise measurements, surface heating tests for local coils, and image quality assessments across all clinical pulse sequences. Two U.S. board-certified radiologists reviewed image pairs comparing subject and predicate devices.
Technological Characteristics
7T actively shielded superconducting magnet; 60 cm bore; 80mT/m gradient system; 297.18 MHz RF frequency. Local Tx/Rx coils only (no body resonator). Software: syngo MR E11K. Standards: IEC 60601-2-33, IEC 62304, ISO 14971, NEMA MS 4-2010. SAR management via FDTD computational modeling.
Indications for Use
Indicated for patients > 30 kg/66 lbs requiring MR imaging of head or extremities for diagnostic purposes.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MAGNETOM Trio A Tim System with syngo MR B19A (K123938)
{0}------------------------------------------------
October 12, 2017
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue and "ADMINISTRATION" in a smaller font size.
Siemens Medical Solutions USA, Inc. Milind Dhamankar Senior Clinical Affairs Specialist 40 Liberty Boulevard, Mail Code 65-1A Malvern, Pennsylvania 19355
Re: k170840
Trade/Device Name: MAGNETOM Terra Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LNI, MOS Dated: September 5, 2017 Received: April 5, 2017
Dear Milind Dhamankar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
## Page 2 - Milind Dhamankar
k170840
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Section 4: Indications for Use Statement
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Indications for Use | |
| 510(k) Number (if known) | k170840 |
|--------------------------|----------------|
| Device Name | MAGNETOM Terra |
Indications for Use (Describe)
The MAGNETOM Terra system is indicated for use as a magnetic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images and that displays the internal structure and/or function of the head or extremities. Other physical parameters derived from the images may also be produced. These images and the physical parameters derived from the interpreted by a trained physician vield information that may assist in diagnosis.
The device is intended for patients > 30 kg/66 lbs.
| Type of Use (Select one or both, as applicable) | <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented.
## 510(k) Summary: MAGNETOM Terra
| Company: | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard, 65-1A<br>Malvern, PA 19355 |
|----------------|-----------------------------------------------------------------------------------------|
| Date Prepared: | March 20, 2017 |
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
#### 1. General Information
Importer/Distributor: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, Mail Code 65-1A Malvern, PA 19355, USA Establishment Registration Number: 2240869
Manufacturing Site: Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Establishment Registration Number: 3002808157
## Contract Manufaturers:
Quality Electrodynamics for 1Tx28Rx Knee Coil 7T Clinic coil 6655 Beta Drive, Suite 100 Mayfield Village, OH 44143, USA Establishment Registration Number: 3007350713
Nova Medical, Inc. for 1Tx32Rx Head Coil 7T Clinic coil 150 West Street Suite 201 Wilmington, MA 01887, USA
{4}------------------------------------------------
# SIEMENS
#### 2. Contact Person:
Milind Dhamankar, M.D. Senior Clinical Affairs Specialist Siemens Medical Solutions USA, Inc. Phone: (610) 448-6467 E-mail: mailto:milind.dhamankar@siemens-healthineers.com
#### 3. Device Name and Classification:
| Common / Usual Name | 7T Magnetic Resonance Imaging system |
|-------------------------|-------------------------------------------------|
| Trade Name | MAGNETOM Terra |
| Classification Name | Magnetic Resonance Diagnostic Device<br>(MRDD), |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR § 892.1000 |
| Device Class: | II |
| Primary Product Code: | LNH |
| Secondary Product Code: | LNI, MOS |
#### 4. Intended Use
The indications for use of the subject device fall within the intended use of the predicate device. The intended use for the MAGNETOM Terra is as follows:
The MAGNETOM Terra system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images and that displays the internal structure and/or function of the head or extremities. Other physical parameters derived from the images may also be produced. These images and the physical parameters derived from the images when interpreted by a trained physician yield information that may assist in diagnosis.
The device is intended for patients > 30 kg/66 lbs.
{5}------------------------------------------------
#### 5. Predicate and Reference Device Information:
| Predicate Device | FDA Clearance<br>Number | Product<br>Code | Manufacturer |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|------------------|-----------------------------------------------|
| MAGNETOM Trio A Tim<br>System with <i>syngo</i> MR B19A<br>Please note the following parts of the<br>MAGNETOM Trio A Tim System are the<br>predicate device parts related to:<br>• System Architecture<br>• RF transmit/receive system | K123938 cleared<br>February 12,<br>2013 | LNH, LNI,<br>MOS | Siemens AG /<br>Siemens<br>Healthcare<br>GmbH |
| Reference Devices | FDA Clearance<br>Number | Product Code | Manufacturer |
| MAGNETOM Verio with<br><i>syngo</i> MR D13A<br>Please note the following parts of the<br>MAGNETOM Verio are the reference<br>device parts related to:<br>• SAR Management and Control<br>System<br>• Gradient Power Amplifier<br>• Computer Systems | K121434 cleared<br>November 5,<br>2012 | LNH, LNI,<br>MOS | Siemens AG /<br>Siemens<br>Healthcare<br>GmbH |
| MAGNETOM Prisma with<br><i>syngo</i> MR E11C<br>Please note the following parts of the<br>MAGNETOM Prisma are the reference<br>device parts related to:<br>• Software <i>syngo</i> MR E11K<br>• Patient table<br>• Magnet<br>• Gradient Coil<br>• Cover<br>• Physiological Measurement Devices<br>• Patient communication devices | K153343, cleared<br>April 15, 2016 | LNH, LNI,<br>MOS | Siemens AG /<br>Siemens<br>Healthcare<br>GmbH |
### 6. Device Description:
MAGNETOM Terra is a 60 cm bore Magnetic Resonance Imaging system with an actively shielded 7T superconducting magnet. With the interplay of the magnetic field, gradients, radio frequency (RF) transmitter and receiver coil and software this magnetic resonance scanner produces transverse, sagittal, coronal and oblique cross sectional images that represent the spatial distribution of protons with spin. The MAGNETOM Terra uses two local coils 1Tx32Rx Head Coil 7T Clinic and 1Tx28Rx Knee Coil 7T Clinic for head and knee imaging.
{6}------------------------------------------------
## Device Modifications
The subject device, MAGNETOM Terra, has the same fundamental technological characteristics as the predicate device, MAGNETOM Trio A Tim System with syngo MR B19A, (K123938 cleared February 12, 2013). Components such as: magnet, gradient coil, RF Subsystem, software and local Tx/Rx coils have been modified for 7T. Specific details are listed in the Predicate and Reference Device Information table (# 5) above.
The modifications that are new to the subject device are not cleared on other systems and are discussed in detail within the submission. The list below describes the new features and components followed by the modified components.
## Hardware
- New Magnet: The new magnet, OR 107, is based on the 트 already cleared magnet design OR 65 of the reference device MAGNETOM Prisma. The magnet electronics developed for the MAGNETOM Prisma were modified for the subject device magnet. Unlike OR 65, the OR 107 magnet does not support the mounting of the RF components on the magnet. The subject device has a separate RF cabinet instead. OR 107 has two cold heads due to the higher volume of helium in the maqnet.
- I New Gradient Coil: The mechanical dimensions and design of the gradient coil (GC) of the reference device, MAGNETOM Prisma were modified to create the subject device GC to accommodate the wider magnet bore diameter and the additional 3rd order shims. The subject device Gradient svstem 80/200 (80mT/m @@ slewrate 200T/ms simultaneous) has a new Gradient Coil and the Gradient Power Amplifier used in the reference device, MAGNETOM Verio has been modified.
- I New Coils: Two new Tx/Rx local coils for head and knee examinations designed for 7T
- -1Tx32Rx Head Coil 7T Clinic from Nova Medical, Inc. .
- . -1Tx28Rx Knee Coil 7T Clinic from Quality Electrodynamics LLC
- The subject device does not have an integrated body ■ resonator. Only local transmit (Tx) and receive (Rx) coils are used for RF excitation.
- 트 Modified RF Subsystem 1ch Tx: The RF transmit/receive system of the subject device is based on Siemens' Tim technology of the predicate device MAGNETOM Trio A Tim System. The RF Subsystem of the subject device was
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The text is horizontally oriented and centered in the image.
modified to accommodate the 7T System frequency of 297.18 MHz. The RF amplifier, receive electronics and the new MaRS computer are located in separate cabinets.
- 트 Modified patient table: The patient table of the reference device, MAGNETOM Prisma was modified. The horizontal movement speed of the 7T subject device patient table is limited to 35 mm/s compared to 200 mm/s of the 3T reference device to maintain the maximum dB/dt value of 3 T/s defined in IEC 60601-2-33 Ed3, 2010. The subject device patient table has no vertical movement. Coil plugs and patient infrastructure (vacuum, headphone, squeeze ball) are moved to a separate Coil Interface Box. The subject device includes enhancements to the software that controls Specific Absorption Rate (SAR) based on simulations for adult humans only. Therefore the subject device patient table fits the needs for patients from 30 kg up to 200 kg.
## Software
The subject device software, syngo MR E11K, was developed based on the commercially available MAGNETOM software line E11, latest cleared with syngo MR E11C for the reference device MAGNETOM Prisma (K153343, cleared April 15, 2016).
- । There are no new applications available with syngo MR E11K in comparison to syngo MR E11C.
- । Pulse sequence types were adapted according to the different conditions caused by the 7T system frequency of 297.18 MHz. No new 7T specific clinical pulse sequence types were developed.
- One new non-clinical pulse sequence type for B0 field mapping has been added. This sequence is not used for clinical imaging.
- । The subject device includes enhancements to the software that controls Specific Absorption Rate (SAR) based on simulations for adult humans only. Therefore the MAGNETOM Terra is intended for patients > 30 kg/66 lbs. This weight limitation is not there for the 3T predicate device.
The subject device, MAGNETOM Terra is substantially equivalent to the predicate device, MAGNETOM Trio A Tim System with syngo MR B19A.
#### 7. Technological Characteristics
The subject device, MAGNETOM Terra, has the same fundamental technological characteristics as the predicate device, MAGNETOM Trio A Tim System with syngo MR B19A, (K123938 cleared February 12, 2013).
{8}------------------------------------------------
7T specific components such as: magnet, gradient coil, RF Subsystem, software and local Tx/Rx coils have been modified as described in Device Description (#6) above. The subject device includes enhancements to the software that controls Specific Absorption Rate (SAR) based on simulations for adult humans only. Therefore the MAGNETOM Terra is intended for patients > 30 kg/66 lbs. This weight limitation is not there for the 3T predicate device.
The subject device is substantially equivalent to the predicate device with regard to the operational environment, programming language, operating system and performance. MAGNETOM Terra conforms to the standard for software medical devices (IEC 62304:2006) and IEC as well as NEMA standards.
#### 8. Nonclinical Tests
The following performance testing was conducted on the subject device
- Sample clinical images were acquired for all available clinical pulse . sequences and local coils,
- . Image quality assessments of all clinical pulse sequences types, were completed during system test,
- . Acoustic noise measurements according to NEMA standard were performed.
- . Performance Tests according to IEC 62464-1 as well as surface heating test for the local coils were completed,
- Software verification and validation testing was completed in . accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate device to which it has been compared.
#### 9. Clinical Tests
A clinical study of 35 individuals was conducted to determine the nerve stimulation thresholds used to limit the gradient system output. The observed parameters were used to set the PNS (Peripheral Nerve Stimulation) threshold level which is required in IEC 60601-2-33.
No animal testing has been performed on this device.
In addition to providing clinical sample images for all clinical sequences and coils of the subject device, reports from two U.S. board-certified radiologists have been provided after the radiologists reviewed image pairs comparing the subject and the predicate device Their comments on any observed artifacts and concerns have also been included.
{9}------------------------------------------------
#### 10. Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards in a risk analysis beginning early in the design phase and continuing throughout the development of the product. These risks are controlled via measures realized in software development, SW testing and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the operators are healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
The MAGNETOM Terra conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document as stated in the following table.
| Recognitio<br>n Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>date | Standards<br>Development<br>Organization |
|------------------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|------------------------------------------|
| 19-4 | General | Medical electrical equipment - part 1:<br>general requirements for basic safety<br>and essential performance | ES60601-<br>1:2005/(R)<br>2012 and<br>A1:2012 | AAMI / ANSI |
| 19-8 | General | Medical electrical equipment - Part 1-<br>2: General requirements for basic<br>safety and essential performance -<br>collateral standard: electromagnetic<br>disturbances - requirements and<br>tests. | 60601-1-2<br>Edition 4.0<br>2014-02 | IEC |
| 12-295 | Radiology | Medical electrical equipment - Part 2-<br>33: Particular requirements for the<br>basic safety and essential<br>performance of magnetic resonance<br>equipment for medical diagnosis | 60601-2-33<br>Ed. 3.2:2015 | IEC |
| 5-40 | General | Medical devices - Application of risk<br>management to medical devices | 14971 Second<br>edition 2007-<br>03-01 | ISO |
| 5-96 | General | Medical devices - Application of<br>usability engineering to medical<br>devices | 62366-1:2015 | AAMI ANSI<br>IEC |
| 13-79 | Software | Medical device software - Software<br>life cycle processes | 62304 Edition<br>1.1 2015-06 | IEC |
{10}------------------------------------------------
| Recognitio<br>n Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>date | Standards<br>Development<br>Organization |
|------------------------|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|------------------------------------------|
| 12-232 | Radiology | Acoustic Noise Measurement<br>Procedure for Diagnosing Magnetic<br>Resonance Imaging Devices | MS 4-2010 | NEMA |
| 12-300 | Radiology | Digital Imaging and Communications<br>in Medicine (DICOM) Set 03/16/2012<br>Radiology | PS 3.1 - 3.20<br>(2016) | NEMA |
| 2-156 | Biocom-<br>patibility | Biological evaluation of medical<br>devices -- part 1: evaluation and<br>testing within a risk management<br>process. (Biocompatibility) | 10993-<br>1:2009/(R)<br>2013 | AAMI<br>ANSI<br>ISO |
The local transmit coils available with the MAGNETOM Terra are supervised as local transmit coils. The estimation of local (10g average) SAR is based on results of computational modeling on finite-difference time domain (FDTD) algorithm, using eight Human models of the Virtual Population (Christ et al, PhysMedBiol 2010, Gosselin et al, PhysMedBiol 2014) and the MIDA Model (lacono et al, Plos One 2015). The models include male and female models, ages ranging from 6 years to 84 years, weights ranging from 18.6 kg to 119 kg. The FDTD modeling used a mesh size of 2 mm, which has been determined to accurately evaluate the local SAR in the different structures of the human head and reproduce the constructive/destructive interference between anatomical structures.
This local SAR limit effectively also limits whole body SAR, partial body SAR and head SAR to a value that is lower than required by IEC 60601-2-33 for volume coils.
#### 11. Substantial Equivalence
The subject device, MAGNETOM Terra, has the same fundamental technological characteristics as the predicate device, MAGNETOM Trio A Tim System with syngo MR B19A, (K123938 cleared February 12, 2013). 7T specific components such as: magnet, gradient coil, RF Subsystem, software and local Tx/Rx coils have been modified as described in Device Description (#6) above. The subject device is substantially equivalent to the predicate device with regard to the operational environment. programming language, operating system and performance.
The subject device software, syngo MR E11K, was developed based on the commercially available MAGNETOM software line E11, latest cleared with syngo MR E11C for the reference device MAGNETOM Prisma (K153343, cleared April 15, 2016).
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the word "SIEMENS" in all capital letters. The text is a teal color and is centered in the image. The font is sans-serif and the letters are bold.
- । There are no new applications available with syngo MR E11K in comparison to syngo MR E11C.
- Pulse sequence types were adapted according to the different conditions caused by the 7T system frequency of 297.18 MHz. No new 7T specific clinical pulse sequence types were developed.
- One new non-clinical pulse sequence type for B0 field mapping has been added. This sequence is not used for clinical imaging.
- The subject device includes enhancements to the software that controls Specific Absorption Rate (SAR) based on simulations for adult humans only. Therefore the MAGNETOM Terra is intended for patients > 30 kg/66 lbs. This weight limitation is not there for the 3T predicate device.
#### 12. Conclusion as to Substantial Equivalence
The MAGNETOM Terra has the same fundamental technological characteristics as the predicate device MAGNETOM Trio A Tim System with syngo MR B19A (K123938).
7T specific components such as: magnet, gradient coil RF Subsystem, software and local Tx/Rx coils have been modified and this gives the subject device greater capabilities in terms of image quality in the available body regions. All of the modifications do not introduce any new issues of safety or effectiveness. The modifications to the MAGNETOM Terra do not change the fundamental scientific technology of the Magnetic Resonance Diagnostic device and the indications for use of the subject device fall within the intended use of the predicate device.
The subject device includes enhancements to the software that controls Specific Absorption Rate (SAR) based on simulations for adult humans only. Therefore the MAGNETOM Terra is intended for patients > 30 kg/66 lbs. This weight limitation is not there for the 3T predicate device.
believes that the subject device, MAGNETOM Terra is Siemens substantially equivalent to the predicate device, MAGNETOM Trio A Tim System with syngo MR B19A.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.