SOFTWARE SYNGO MR D13A FOR THE MAGNETOM SYSTEMS AERA/SKYRA/AVANTO/VERIO

K121434 · Siemens Medi Cal Solutions, Inc. · LNH · Nov 5, 2012 · Radiology

Device Facts

Record IDK121434
Device NameSOFTWARE SYNGO MR D13A FOR THE MAGNETOM SYSTEMS AERA/SKYRA/AVANTO/VERIO
ApplicantSiemens Medi Cal Solutions, Inc.
Product CodeLNH · Radiology
Decision DateNov 5, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2

Indications for Use

The MAGNETOM systems Aera/Skyra/Avanto/Verio with software syngo MR D13A are indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross-sectional images, spectroscopic images and/or spectra, and that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. The MAGNETOM systems may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles.

Device Story

Software syngo MR D13A is an upgrade for Siemens MAGNETOM Aera, Skyra, Avanto, and Verio MRI scanners. It introduces new imaging sequences, Dot Engine improvements, and specialized coils for body, neurological, abdominal, orthopedic, and cardiac applications. The system acquires MR signals transformed into cross-sectional images, spectra, and derived physical parameters. Operated by trained clinicians in clinical settings, the output is interpreted by physicians to assist in diagnosis or guide interventional procedures. Benefits include enhanced imaging capabilities, improved workflow via Dot Engines, and expanded multi-nuclear spectroscopy options. The software integrates with existing scanner hardware, including new head, knee, breast, and endo-array coils, to provide high-resolution diagnostic data.

Clinical Evidence

No clinical tests conducted. Substantial equivalence supported by bench testing and phantom validation of new sequences and hardware. Clinical images provided to demonstrate performance of new coils and software features.

Technological Characteristics

MR diagnostic software for 1.5T/3T systems. Features include new imaging sequences, Dot Engine workflow automation, and multi-nuclear spectroscopy. Hardware includes specialized RF coils (TxRx CP Head, 15ch Knee, 16ch Breast, Endo Array). Conforms to IEC 60601-1, ISO 14971:2007, and NEMA standards. Connectivity includes networked scanner capabilities.

Indications for Use

Indicated for use as MR diagnostic devices for patients requiring imaging of head, body, or extremities; produces cross-sectional images, spectroscopic images, and spectra for diagnostic assistance; supports interventional procedures using MR-compatible devices.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ September 28, 2012 121434 ### 510(k) Summary This 510(k) summary is being submitted in accordance with 21 CFR § 807.92. ### General Information | Establishment | Siemens Medical Solutions USA, Inc.<br>51 Valley Stream Parkway<br>Mail Code D02<br>Malvern, PA 19355, USA<br>Registration Number 2240869 | NOV | 5 | 2012 | |----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|---|------| | Manufacturer | Siemens AG<br>Henkestrasse 127<br>D-91052 Erlangen, Germany<br>Registration Number 8010024 | | | | | | SIEMENS SHENZHEN MAGNETIC RESONANCE LTD.<br>Siemens MRI Center<br>Hi-Tech Industrial park (middle)<br>Gaoxin C. Ave., 2nd<br>Shenzhen 518057, P.R. CHINA<br>Registration Number 3004754211 | | | | | Contact Person | Ms. Michelle L. Byrne<br>Regulatory Affairs Technical Specialist<br>Siemens Medical Solutions USA, Inc.<br>51 Valley Stream Parkway<br>Mail Code D02<br>Malvern, PA 19355, USA<br>Phone: (610) 448-4293<br>Fax: (610) 448-1787<br>E-mail: Michelle.L.Byrne@siemens.com | | | | | Device Name | Software syngo MR D13A for the<br>MAGNETOM systems Aera/Skyra/Avanto/Verio | | | | | CFR Code | 21 CFR § 892.1000 | | | | | Classification | Class II | | | | | Product Codes | LNH, LNI, MOS | | | | | | Classification NameMagnetic Resonance Diagnostic Device, MR<br>Spectroscopy, MR Coils | | | | {1}------------------------------------------------ ### Information Supporting Substantial Equivalence ### DEVICE DESCRIPTION The software syngo MR D13A is the latest software version for four Siemens MR scanners: MAGNETOM Aera (1.5T), Skyra (3T), Avanto (1.5T), and Verio (3T). New scanners will be manufactured with this software version; existing scanners can be upgraded to this software version. The new software version includes new software sequences, applications, coils and other hardware for the four MAGNETOM scanners. While some new features (hardware and software) are only available for certain scanners (of the four listed), the basic syngo MR D13A software will run on each of the four MAGNETOM systems. ### Summary of New Features with syngo MR D13A: #### Software - . New sequences for all four systems for body, neurological, abdominal, orthopedic, and cardiac imaging, and for spectroscopy - New or modified applications/software for all four systems: . - Dot Engine improvements o - Image processing and visualization improvements o - o Mapping improvements - Improved iPAT o - Networked Scanner improvements 0 - Dot Engines now available for MAGNETOM Avanto & Verio (previously for . Aera & Skyra with software syngo MR D11) - New Dot Engines for MAGNETOM Aera & Skyra . - Multi Nuclear Option now available for MAGNETOM Skyra (previously for . Avanto & Verio with software syngo MR B17) #### Hardware - New coils for Aera & Skyra only: . - o TxRx CP Head Coil - o TxRx 15ch Knee Coil - o 4 Channel Special Purpose Coil - New coils for Avanto & Verio only: . - Sentinelle Vanguard Breast 16ch Coil o - o Sentinelle Endo Array coil - . Modified Magnet, Skyra only {2}------------------------------------------------ #### INTENDED USE The MAGNETOM systems Aera/Skyra/Avanto/Verio with software syngo MR D13A are indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross-sectional images, spectroscopic images and/or spectra, and that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. The MAGNETOM systems may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles. #### NONCLINICAL TESTS The following performance testing was conducted on the subject device: - The coils were tested for SNR, image uniformity, and heating. . - The magnet was tested for SNR, geometric distortion, image uniformity . and slice profile/thickness/spacing. - Dedicated phantom testing was conducted on the MP2RAGE sequence. . - The performance parameters of MNO Spectroscopy were phantom-. tested. - syngo ASL 3D was tested using a dedicated phantom. . - MR Elastography was tested using a dedicated phantom. . - T1 mapping with B1 correction was phantom-tested. . - All software features were verified and validated. . The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate devices to which it has been compared. #### CLINICAL TESTS There were not any clinical tests conducted to support the subject device and the substantial equivalence argument, however clinical images were provided to support the new coils as well as the new and modified software features of the subject device. #### SUBSTANTIAL EQUIVALENCE Software syngo MR D13A for the MAGNETOM systems Aera/Skyra/Avanto/Verio is substantially equivalent to the following predicate devices: {3}------------------------------------------------ September 28, 2012 | | Predicate Device Name | Manufacturer | 510(k)<br>Number | FDA Clearance<br>Date | |---|-----------------------------------------------------------|-------------------------------------|------------------|-----------------------| | 1 | MAGNETOM Aera &<br>MAGNETOM Skyra with<br>syngo® MR D11 | Siemens<br>Medical<br>Solutions | K111242 | November 23, 2011 | | 2 | MAGNETOM Avanto &<br>MAGNETOM Verio with<br>syngo® MR.B17 | Siemens<br>Medical<br>Solutions | K082427 | November 7, 2008 | | 3 | MR-Touch | GE Medical<br>Systems LLC | K083421 | July 24, 2009 | | 4 | 3D ASL | GE Medical<br>Systems LLC. | K092925 | January 6, 2010 | | 5 | TxRx 1.5T & 3T 15ch<br>Knee Coils | Quality<br>Electrodynamics<br>(QED) | K082636 | September 25,<br>2008 | | 6 | TxRx 1.5T & 3T CP Head<br>Coils | Quality<br>Electrodynamics<br>(QED) | K091114 | May 1, 2009 | | 7 | Sentinelle Endo Coil | Sentinelle<br>Medical Inc | K103274 | August 18, 2011 | | 8 | Sentinelle 16ch Breast<br>Coil | Sentinelle<br>Medical Inc | K112112 | August 25, 2011 | #### SAFETY AND EFFECTIVENESS The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device. Risk Management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions USA, Inc. and Siemens AG adhere to recognized and established industry standards, such as the IEC 60601-1 series, to minimize electrical and mechanical hazards. The MAGNETOM systems Aera/Skyra/Avanto/Verio with software syngo MR D13A conform to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document. ### SUBSTANTIAL EQUIVALENCE CONCLUSION There are no changes to the Indications for Use for the subject device, compared to that of the predicate MAGNETOM scanners with software syngo MR D11 and syngo MR B17. The differences between the subject device and the predicate {4}------------------------------------------------ September 28, 2012 devices, which include the aforementioned new sequences, applications, coils, and hardware, give the systems greater capabilities than the predicate devices, but have the same technological characteristics as the predicate devices, are similar to the functionalities of the predicate devices, and do not introduce any new issues of safety or effectiveness. Therefore, Siemens believes that the subject device, software version syngo MR D13A for MAGNETOM systems Aera, Skyra, Avanto, and Verio, is substantially equivalent to the predicate devices listed above. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus features a staff with a snake winding around it, and the overall design is simple and bold. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Ms. Michelle L. Byrne Regulatory Affairs Technical Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway, D-02 MALVERN PA 19355 Re: K121434 Trade/Device Name: Software syngo MR D13A for the MAGNETOM systems Aera/Skyra/Avanto/Verio Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH, LNI, and MOS Dated: October 10, 2012 Received: October 11, 2012 #### Dear Ms. Byrne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {6}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industrv/support/index.html. Sincerely Yours, Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ ### Indications for Use Section 4 Statement 510(k) Number (if known) Software syngo MR D13A for the Device Names: MAGNETOM systems Aera/Skyra/Avanto/Verio #### Indications for Use The MAGNETOM systems described above are indicated for use as magnetic resonance diagnostic devices (MRDD) that produce transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that display the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra, when interpreted by a trained physician, yield information that may assist in diagnosis. The MAGNETOM systems described above may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room display and MR-safe biopsy needles. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OVID) Aland D. Oken sion Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K121434 Page 1 of 1 Software syngo MR D13A for the MAGNETOM systems Aera/Skyra/Avanto/Verio Page 17 of 7769
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...